The Aquamantys Double Cone Bipolar Sealer is a single use, sterile, bipolar device, intended to be used in conjunction with the Aquamantys Pump Generator for delivery of radio-frequency (RF) energy and saline for blunt dissection and for haemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, open abdominal, orthopaedic, and thoracic surgery. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Device Story
Aquamantys Double Cone Bipolar Sealer is a single-use, sterile, bipolar electrosurgical device. It functions by delivering radio-frequency (RF) energy and saline irrigation to soft tissue and bone. The device features a dual electrode tip and a handpiece with an integrated on-off button that simultaneously activates RF energy and saline flow. It connects to the Aquamantys Pump Generator via a three-pin electrical connector and a saline fluid delivery line. Used by surgeons in open abdominal, orthopaedic, and thoracic procedures, the device facilitates blunt dissection and hemostatic sealing. By controlling bleeding and enabling tissue separation, the device assists in surgical site management, potentially reducing blood loss and improving operative efficiency.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing and in-vivo porcine studies. Porcine testing evaluated tissue resection depth versus fluence, stationary and moving depth/diameter of effect, and usability (angle of usage, tissue states, generator compatibility, and finger switch operation). Electromagnetic compatibility was verified per IEC 60601-1-2 and IEC 60601-2-2. Visual and dimensional inspections were performed.
Technological Characteristics
Bipolar electrosurgical device; RF energy source; saline irrigation system. Features dual electrode tip, handpiece with integrated activation switch, and fluid delivery line with pump tubing/drip chamber. Connects to external Aquamantys Pump Generator. Single-use, sterile. Complies with IEC 60601-1-2 and IEC 60601-2-2 for electromagnetic compatibility.
Indications for Use
Indicated for patients undergoing open abdominal, orthopaedic, or thoracic surgery requiring blunt dissection, haemostatic sealing, or coagulation of soft tissue and bone. Contraindicated for contraceptive tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Aquamantys Pump Generator System (including AQM Bipolar Sealers 2.3 and 6.0) (K052859)
Submission Summary (Full Text)
{1}------------------------------------------------
K181057
## 510(k) Summary of Safety and Effectiveness
APR - 5 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990, 21 CFR 807.87, 21 CFR 807.92, Format for Traditional and Abbreviated 510(k)s.
- ﺎ . Name of Submitter, Contact Person and Date Summary Prepared:
| Company Name: | Salient Surgical Technologies, Inc. |
|---------------|--------------------------------------------------------------------|
| Address: | 180 International Drive<br>Portsmouth, NH 03801 |
| Phone: | (603) 742-1515 |
| Fax: | (603) 742-5943 |
| Contact: | Tara N. Turney Urling, RAC<br>Senior Regulatory Affairs Specialist |
Date of Preparation: March 29, 2011
2. Device Trade Name and Common Name:
3.
| Trade Name: | Aquamantys Double Cone Bipolar Sealer |
|-----------------------------------------------------------|---------------------------------------------------------------------------|
| Common/Usual Name:<br>Classification Name:<br>Accessories | Electrosurgery Bipolar Sealer<br>Electrosurgical, Cutting & Coagulation & |
| Product Code: | GEI<br>21CFR 878.4400 |
| Device Class: | Class II |
4. Legally Marketed Equivalent Device: TissueLink Solid Cylinder Monopolar Device 510(k) K014260
> Aquamantys Pump Generator System (including AQM Bipolar Sealers 2.3 and 6.0) 510(k) K052859
{2}------------------------------------------------
- న. Performance Standards: None applicable
- Description of the Device: The Aquamantys Double Cone Bipolar Sealer is a 6. disposable, single-use, sterile, bipolar device. The device employs radio-frequency (RF) energy and saline irrigation for hemostatic sealing and for coagulation and blunt dissection. The device is equipped with a dual electrode tip. Saline and electrical lines exit the opposite end of the handpiece from the electrodes. The handpiece is equipped with an on-off button that simultaneously activates both RF and saline flow. A saline fluid delivery line is provided with the device, and includes a section of pump tubing and a drip chamber or spike. The three-pin electrical connector is designed to be plugged into the Aquamantys Pump Generator.
- 7. Intended Use of the Device: The Aquamantys Double Cone Bipolar Sealer is a single use, sterile, bipolar device, intended to be used in conjunction with the Aquamantys Pump Generator for delivery of radio-frequency (RF) energy and saline for blunt dissection and for haemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, open abdominal, orthopaedic, and thoracic surgery.
The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
- 8. Comparison of technological characteristics with Predicate Device: The device is similar to the TissueLink Solid Cylinder Monopolar device in that it is capable of blunt dissection as well as hemostatic sealing and coagulation. Physically, the device employs similar characteristics to the Aquamantys 6.0 Bipolar Sealer including connection to the Aquamantys Pump Generator.
- 9. Pre-Clinical Evaluation: Where applicable, non-clinical testing of the proposed device relied upon to determine substantial equivalence to the predicate were as follows:
- . In-vivo porcine testing, including:
- o Tissue resection showing depth versus fluence
- o Stationary depth of effect on tissue versus time
- o Moving depth and diameter of effect versus time
- o Usability
- · Angle of usage
- . Various tissue usage states
- Compatibility with appropriate generator .
- a Finger switch
- Electromagnetic Compatibility per IEC 60601-1-2 as modified by IEC . 60601-2-2
- Visual and Dimensional .
{3}------------------------------------------------
Testing has confirmed the proposed device's performance for blunt dissection and for haemostatic sealing and coagulation is at least as safe and effective as that of the predicate devices.
- 10. Conclusion: The proposed and predicate devices have similar intended uses, similar design, performance and functional features as well as the same fundamental scientific technology. The proposed device is found to be substantially equivalent to the named predicates. 1
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of the department's name, which is arranged in a circular fashion around the left side of the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Salient Surgical Technologies, Inc. % Regulatory Technology Services LLC Mr. Mark Job 1394 25th Street NW Buffalo, MN 55313
APR - 5 2011
Re: K101057
Trade/Device Name: Aquamantys Double Cone Bipolar Sealer Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 1, 2011 Received: April 4, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Mark Job
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse overse (CDFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1158011, please
the Center for Devices and bodie DA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli ine (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
edia.dps.mn.gov/KoreanServices/ResourcesforYourIndustry/default.htm
Sincerely yours,
Sincerely yours,
for Peter D. Rieman
Mark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K101057
Device Name: Aquamantys Double Cone Bipolar Sealer
Indications for Use: The Aquamantys Double Cone Bipolar Sealer is a single use, sterile, bipolar device, intended to be used in conjunction with the Aquamantys Pump Generator for delivery of radio-frequency (RF) energy and saline for blunt dissection and for haemostatic sealing and coagulation of soft tissue and bone at the operative site. It is intended for, but not limited to, open abdominal, orthopaedic, and thoracic surgery.
The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101057
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.