GUIDE CATHETERS AND ACCESSORIES

K101015 · Thomas Medical Products, Inc. · DQY · Sep 23, 2010 · Cardiovascular

Device Facts

Record IDK101015
Device NameGUIDE CATHETERS AND ACCESSORIES
ApplicantThomas Medical Products, Inc.
Product CodeDQY · Cardiovascular
Decision DateSep 23, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

ScoutPro Inner Catheters are indicated for the delivery of contrast media or Biotronik devices into the left ventricular coronary venous system. ScoutPro Inner Catheters are not intended to introduce left ventricular leads through their lumen.

Device Story

ScoutPro Inner Catheters are percutaneous catheters used to access the coronary venous system. The device acts as a conduit for guidewires (0.014" to 0.035") or delivery of contrast media. It features a multi-durometer shaft with metal braid reinforcement and a lubricious coating. The distal end includes a 50° to 90° curve and a platinum/iridium marker ring for fluoroscopic visualization. The proximal end features a female luer-lock hub for connection to syringes or stopcocks. Used in clinical settings by trained cardiologists, the device is intended for short-term application (< 24 hours). It can be used alone or in a telescopic assembly with outer catheters. The device assists clinicians in navigating coronary vasculature to deliver diagnostic or therapeutic agents, potentially improving procedural access to the coronary venous system.

Clinical Evidence

Bench testing only. Testing included simulated use (insertion/removal forces, kink/buckling resistance, curve configuration, joint integrity) and physical/dimensional testing (leak resistance, pull forces, dimensional accuracy). All tests passed.

Technological Characteristics

Percutaneous catheter with multi-durometer shaft, metal braid reinforcement, and medical-grade lubricious coating. Features platinum/iridium marker ring for radiopacity. Distal tip angles 50°-90°. Includes female luer-lock hub. Compatible with 0.014"-0.035" guidewires. Single-use, short-term application.

Indications for Use

Indicated for delivery of contrast media or Biotronik devices into the left ventricular coronary venous system. Not for introduction of left ventricular leads. For use by trained cardiologists/medical personnel.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K/0/0/5 # 510(k) SUMMARY #### Submitter's Information: A. Name: Address: Title: SEP 2 3 2010 Thomas Medical Products - A GE Healthcare Company 65 Great Valley Parkway Malvern, PA 19355 Telephone Number: 610.651.5000 Facsimile: 610.651.5003 Contact Person: Tim Stoudt Manager, QA / RA Date Submission Prepared: February 26, 2010 #### B. Device Information: | Trade name: | ScoutPro Inner Catheters | |--------------------------|------------------------------------------| | Classification Name(s): | Percutaneous Catheter (21 CFR §870.1250) | | Common or usual name(s): | Guide catheters and accessories | #### C. Legally marketed device to which equivalence is claimed: Guidant Corporation Rapido™ Guiding Catheter (K021455) and Merit Medical Systems, Inc. Impress Radiology Catheter (K053171) #### D. Description of the device: The Thomas Medical Products, Inc. (TMP) ScoutPro Inner Catheters are intended to access the coronary venous system, either alone or in a telescopic assembly with other introducers. The guide catheters serve as a conduit to guide devices, including guidewires, or to deliver contrast medium into specific branches of the coronary venous system. ScoutPro Inner Catheters are not intended to introduce left ventricular leads through their lumen. They have direct contact to the inner heart. ScoutPro Inner Catheters come with a 50° to 90° tip angle. They are designed as single use devices and for short term application (< 24 hours). Only medical doctors and medical personnel, who are well trained in cardiology, should apply these catheters. The ScoutPro Inner Catheters are designed for guiding devices, including guidewires, with an outer diameter smaller than 1.1 mm. In addition, the ScoutPro Inner Catheters can work with outer catheters. The ScoutPro Inner Catheters are compatible with the following devices: - ScoutPro Guidewire @ 0.035" (BIOTRONIK #359805) - - ScoutPro hemostatic valve (BIOTRONIK #345779) - - Various ScoutPro 7F catheter shapes (Extended Hook Right BIOTRONIK#35937, Extended Hook -#350369, Straight #359371, MPEP #359373, BIO2 #359374, Amplatz 6.0 #350235, Hook #350236, Multipurpose Hook #350237, Extended Hook Right L #361531, Extended Hook L 3361530, Straight L #361536, MPEP L #361533, BIO2 L #361529, Amplatz 6.0 L #361527, Hook L #361532, Multipurpose Hook L #361534) - Commercially available Guide Wires (0.014" 0.035") - - Commercially available medical devices with male Luer-Lock (stopcocks, syringes) 1 - -Commercially available fluid x-ray diagnostic contrast media. The ScoutPro Inner Catheters have a shaft design with four (4) gradually decreasing stiffness segmentations from proximal to distal. The shaft is reinforced by a metal braid from the proximal end until approximately 32 mm from the distal end. The shaft is coated by a medical-grade coating that provides enhanced lubricity when advanced through an outer catheter or in the coronary vasculature. {1}------------------------------------------------ K101015 The proximal end is equipped with a hub that has a female luer lock connector for adapting a syringe or a three way stopcock with luer lock connections. , There are differing versions in the curve form, 50° to 90° curve at the distal end, that is used according to the anatomy of the present coronary vasculature. The distal soft tip has a tapered outer diameter and the distal tip further contains a platinum/iridium marker ring for enhanced x-ray visibility. Package contents: - One (1) guide catheter - One (1) 3-way stopcock - One (1) Instructions for use manual. #### E. Intended use of the device: ScoutPro Inner Catheters are indicated for the delivery of contrast media or Biotronik devices into the left ventricular coronary venous system. ScoutPro Inner Catheters are not intended to introduce left ventricular leads through their lumen. - F. Summary of the technological characteristics of the device compared to the predicate devices: | Features | ScoutPro<br>Inner<br>Catheters | Guidant<br>Rapido<br>Guiding<br>Catheter<br>K021455 | Merit<br>Impress<br>Radiology<br>Catheter<br>K053171 | |-----------------------------------------------|--------------------------------|-----------------------------------------------------|------------------------------------------------------| | Hemostasis valve provided | No | No | No | | Compatible with .038" guide wire | Yes | Yes | Yes | | Introducer available in 65cm length | Yes | No | Yes | | Introducer set available in 7F | Yes | No | No | | Multi-durometer catheter shaft | Yes | Yes | Yes | | Teflon inner liner | Yes | Yes | Yes | | Wire reinforcement completely<br>encapsulated | Yes | Yes | Yes | | Radiopaque tip or marker | Yes | Yes | Yes | | Luer lock hub | Yes | Yes | Yes | - Testing: G. | | Test Description | Results | |-----------------------|-------------------------------------------------------------------|---------| | Simulated Use Testing | | | | | ScoutPro Inner Catheter | | | | insertion and removal forces of<br>.014" and .035" guidewires | Pass | | | insertion and removal forces<br>through outer sheath and<br>valve | Pass | | | shaft kink and buckling<br>resistance | Pass | | | tip section kink, corrugation,<br>and peel back resistance | Pass | | | curve configuration after<br>straightening | Pass | | | shaft joint integrity when bend<br>around mandrel | Pass | {2}------------------------------------------------ K/0/0/5 | Test Description | Results | |-----------------------------------------------|---------| | Physical/Dimensional Testing | | | Stopcock | | | leak resistance of stopcock/catheter assembly | Pass | | ScoutPro Inner Catheter | | | pull forces of shaft joints | Pass | | pull force of hub/tube joint | Pass | | tube shaft O.D. | Pass | | tip section O.D. | Pass | | tip I.D. | Pass | | length | Pass | | curve configuration | Pass | ### H. Substantial equivalence rationale: The ScoutPro Inner Catheters substantially equivalent to the currently marketed predicate devices. This assessment is based upon analysis of similar technological characteristics, bench testing, and indications for use. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the upper portion of the circle. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Thomas Medical Products c/o Mr. Tim Stoudt Manager, QA/RA 65 Great Valley Parkway Malvern, PA 19355 SEP 2 3 2010 Re: K101015 ScoutPro Inner Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: DQY Dated: September 8, 2010 Received: September 9, 2010 Dear Mr. Stoudt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Tim Stoudt Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. · Sincerely yours. M Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K1010/5 Page 1 of 1 510(k) Number (if known): ScoutPro Inner Catheters Device Name: SEP. 2 3 2010 Indications For Use: ScoutPro Inner Catheters are indicated for the delivery of contrast media or Biotronik devices into the left ventricular coronary venous system. Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter-Use (Per 21 CFR 801 Subpart C) (Optional Format 11-13-03) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular De nik) Number 510(k) "Indications For Use" Form (Replica of FDA Form)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...