K100980 · Hologic, Inc. · NPQ · Jun 1, 2011 · Hematology
Device Facts
Record ID
K100980
Device Name
INVADER FACTOR V
Applicant
Hologic, Inc.
Product Code
NPQ · Hematology
Decision Date
Jun 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7280
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Invader® Factor V test is an in vitro diagnostic test intended for the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene in isolated genomic DNA obtained from whole blood potassium EDTA samples from patients with suspected thrombophilia.
Device Story
Invader® Factor V test detects Factor V G1691A mutation in isolated genomic DNA from whole blood (potassium EDTA). Process: PCR amplification of target sequence; Invader Plus® chemistry for signal generation; Cleavase® enzyme-mediated cleavage of allele-specific primary probes and FRET cassettes. Input: purified genomic DNA. Output: genotype call (WT, HET, MUT) generated by software. Used in clinical laboratory settings; operated by laboratory technicians. Workflow: DNA extraction; PCR amplification on third-party thermal cycler; fluorescence measurement on multi-well fluorometer; data analysis via Invader® Call Reporter™ and Invader® Factor V software. Output assists clinicians in diagnosing thrombophilia risk.
Clinical Evidence
Bench testing only. Method comparison study (n=352) against bi-directional sequencing showed 100% agreement. Reproducibility studies (n=540) across three sites showed 99.81% final agreement. Lot-to-lot reproducibility (n=60) showed 100% agreement. Analytical sensitivity established at 5-80 ng/µL input gDNA. Interfering substances (heparin, cholesterol, bilirubin, hemoglobin, K2EDTA, ethanol) showed no impact on performance. Stability confirmed for 6 months real-time and 15 freeze-thaw cycles.
Technological Characteristics
In vitro diagnostic genotyping system. Uses PCR amplification and Invader Plus® chemistry (Cleavase® enzyme). Signal detection via FRET cassettes (FAM and RED fluorophores). Requires third-party thermal cycler and multi-well fluorometer. Software-based genotype calling. Compatible with standard DNA extraction methods. Isothermal reaction at ~63°C.
Indications for Use
Indicated for patients with suspected thrombophilia to detect and genotype the Factor V G1691A mutation. For prescription use only.
Regulatory Classification
Identification
Factor V Leiden mutation detection system is a device that allows the detection and genotyping of a single point mutation of the human Factor V gene, referred to as Factor V Leiden mutation, from DNA isolated from human whole peripheral blood. The system consists of different reagents and instruments which includes polymerase chain reaction (PCR) primers, hybridization matrices, thermal cyclers, imagers, and software packages. The detection system is an aid to diagnosis in the evaluation of patients with suspected thrombophilia.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Factor V Leiden DNA Mutation Detection Systems.” (See § 864.1(d) for the availability of this guidance document.)
Predicate Devices
Roche Factor V Leiden Kit (k033607)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k100980
B. Purpose for Submission:
New device
C. Measurand:
Factor V Leiden
D. Type of Test:
Qualitative genotyping test for single nucleotide polymorphism detection
E. Applicant:
Hologic Inc.
F. Proprietary and Established Names:
Invader® Factor V
G. Regulatory Information:
1. Regulation section:
21 CFR §864.7280 Factor V Leiden DNA mutation detection systems
2. Classification:
Class II
3. Product code:
NPQ: Test, Factor V Leiden Mutations, Genomic DNA PCR
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The Invader® Factor V test is an *in vitro* diagnostic test intended for the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene in isolated genomic DNA obtained from whole blood potassium EDTA samples from patients with suspected thrombophilia.
2. Indication(s) for use:
Same as Intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Multi-well fluorometer capable of:
| Multi-Labeling Measurement Parameters | Measurement 1 | Measurement 2 |
| --- | --- | --- |
| Read mode | Top | Top |
| Excitation wavelength | 485/20 nm | 560/20 nm |
| Emission wavelength | 535/25 nm | 612/10 nm |
| Gain | Manual | Manual |
| Number of flashes | 10 | 10 |
| Integration time | 20 μs | 20 μs |
I. Device Description:
The Invader® Factor V test consists of the following components:
{1}
- Factor V Oligo Mix
Universal Buffer
Universal Enzyme Mix
No DNA Control
- Factor V Wild Type Control
- Factor V Heterozygous Control
- Factor V Mutant Control
- Invader® Call Reporter™ Software - Version 5.3
- Invader® Factor V Software - Version 2.0
# J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s): Roche Factor V Leiden Kit (k033607)
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | An in vitro diagnostic test intended for the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene in isolated genomic DNA obtained from whole blood potassium EDTA samples from patients with suspected thrombophilia. | Same |
| Target Population | Patients with suspected thrombophilia | Same |
| Specimen Type | Purified DNA isolated from human whole peripheral blood | Same |
| Target Amplification Technology | PCR | Same |
| Signal Generation Technology | Fluorescence Resonance Energy Transfer (FRET) | Same |
| Output Data | Assay signal results are interpreted by a software program and are assigned a genotype that is presented to the end user in a report format | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Allele Discrimination | Targeted cleavage of distinct FRET cassette bound to allele-specific primary probe with a | Melting curve analysis of hybridizations to allele-specific FRET probes |
{2}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | unique 5'-flap | |
| Reaction Conditions | 20-μL reaction in multiple plastic microtiter wells | 10-20 μL reaction in glass capillaries loaded in a sample carousel |
| Signal Detection | End-point detection of amplified sequences after PCR amplification | Real-time detection of amplified sequences during PCR cycles |
| Hardware | Non-specified, third-party fluorometer and thermal cycler | LightCycler® Instrument V1.2 including a cycler and a fluorimeter |
| Software Interface | Java-based software installed on a standalone PC capable of converting raw fluorescence data into genotype calls | LightCycler® Software V3.5 |
K. Standard/Guidance Document Referenced (if applicable):
- Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Factor V Leiden DNA Mutation Detection Systems issued on March 16, 2004
- Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued May 11, 2005
- Guidance for Industry and FDA Staff – Format for Traditional and Abbreviated 510(k)s issued on August 12, 2005
L. Test Principle:
The Invader® Factor V test combines target amplification by PCR using a third-party thermal cycler with signal generation by the Invader Plus® chemistry and signal detection using interpretative software and third-party fluorometer for genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene.
During the signal generation phase, a discriminatory Primary Probe transiently hybridizes to the amplified target sequence along with an Invader® oligonucleotide, to form an overlapping structure. The 5'-end of the Primary Probe includes a 5'-flap that does not hybridize to the target DNA. The 3'-nucleotide of the bound Invader® oligonucleotide overlaps the Primary Probe, and does not hybridize to the target DNA. The Cleavase® enzyme recognizes this overlapping structure and cleaves off the unpaired 5'-flap of the Primary Probe, releasing it as a target-specific product. The Primary Probe is designed to have a melting temperature aligned with the Invader® reaction temperature so that under the isothermal reaction conditions (~63°C) the Primary Probes cycle on and off the target DNA. This allows for multiple rounds of Primary Probe cleavage for each DNA target resulting in an accumulation of released 5'-flaps. The released 5'-flap transiently hybridizes with a corresponding fluorescent resonance energy transfer (FRET) cassette forming an overlapping structure that is recognized and the fluorophore is cleaved from the FRET cassette by the Cleavase® enzyme. The 5'-flap is designed to have a melting temperature aligned with the Invader® reaction temperature, so that the 5'-flaps cycle on and off of the corresponding FRET cassettes. This allows for multiple rounds of FRET cassette cleavage for each 5'-flap, and an
{3}
accumulation of released fluorophore. When the FRET cassette is cleaved, a fluorophore and quencher are separated, generating detectable fluorescence signal.
The Invader® Factor V test uses two different discriminatory Primary Probes, one for the mutant allele and one for the wild type allele. Each Primary Probe is assigned a unique 5'-flap, and distinct FRET cassette, with a spectrally distinct fluorophore. The released 5'-flaps will bind only to their respective FRET cassettes to generate a target-specific signal, linking the wild type allele with one fluorophore (Fluorescence 1: RED) and the mutant allele with the second fluorophore (Fluorescence 2: FAM).
The Invader® Factor V software, in combination with Invader® Call Reporter™ software, is a data analysis software that provides a working template for the setup of reaction mixes and sample placement, and following the import of fluorescence data, it determines results and validity for controls and samples.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Repeatability/Reproducibility**: Two operators each from 3 different sites (2 external sites and 1 internal site) performed the testing in duplicate twice daily over 5 non-consecutive days using the same testing materials including a panel of nine (9) whole blood samples, representing all possible genotypes including three (3) wild types (WT), three (3) heterozygous mutants (HET) and three (3) homozygous mutants (MUT). There were 19/540 (3.5%) first pass No Calls due to “Invalid Control” result on a single run. Upon retraining of the operator and retesting of the run, all controls reported “Valid” and all 18 samples were found to be in agreement with sequencing. One sample with Final No Call was called correctly upon re-extraction and re-testing. There were zero Miscalls, and there was no statistical difference due to Site, Operator or Day.
| Site | Operator | Samples Tested | First pass | | | Final | | % Agreement |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Correct Calls | No-Calls | Miscalls | Correct Calls | Miscalls | |
| Site 001 | 1 | 90 | 90 | 0 | 0 | 90 | 0 | 100% |
| | 2 | 90 | 90 | 0 | 0 | 90 | 0 | 100% |
| Site 002 | 1 | 90 | 90 | 0 | 0 | 90 | 0 | 100% |
| | 2 | 90 | 90 | 0 | 0 | 90 | 0 | 100% |
| Site 003 | 1 | 90 | 90 | 0 | 0 | 90 | 0 | 100% |
| | 2 | 90 | 71 | 19 | 0 | 89 | 0 | 98.89% |
| All | All | 540 | 521 | 19 | 0 | 539 | 0 | 99.81% |
**Lot-to-Lot Reproducibility**: A total of five (5) genomic DNA samples (three (3) wild type and two (2) heterozygous) were tested in quadruplicate using three (3) different kit lots of the Invader® Factor V test. The percent agreement between Invader® Factor V test and sequencing was 100% (n=60).
**Genomic DNA Extraction Reproducibility**: Four commonly used extraction methods were used to process thirty (30) human whole blood samples and ten (10) leukocyte depleted whole blood (LDWB) spiked with cell lines with the following genotypes:
{4}
| Sample Type | Number of Samples | | | |
| --- | --- | --- | --- | --- |
| | WT | MUT | HET | Total |
| Human whole blood samples | 20 | 0 | 10 | 30 |
| LDWB spiked with cell lines | 5 | 5 | 0 | 10 |
| Total | 25 | 5 | 10 | 40 |
The 160 extracted samples were assayed in singlet in a single run using a single kit lot of the Invader® Factor V test. With the exception of one sample that was removed from the study due to loss of traceability of the sample identification, all other samples, regardless of extraction methods, were correctly called on the first pass and gave 100% agreement with DNA sequencing.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality Control:
Each test contains positive and negative controls to assure proper functioning of the system. Failure of any controls will be indicated as "Invalid" in the test results section of the report. The genotyping test result will not be reported for any sample for which a positive or negative control failure occurs.
Positive Control: The genotype controls (WT, HET, MUT) ensure reagents were assembled correctly and perform according to the specifications.
Negative Control: The No DNA Control is used by the interpretive software to set the "noise" component of the run for "signal-to-noise" calculations.
Hardware and Software Control: The genotyping test result will not be reported for any sample for which a positive or negative control failure occurs.
Real-Time Stability Study
The ongoing Stability Study tests seven (7) gDNA samples including three (3) WT, two (2) HET and two (2) MUT genotypes using three (3) lots of Invader® Factor V product stored under two recommended conditions: 1) -20°C (Standard Storage of intermediate components) and 2) 4°-8°C (Standard Storage of Genotype-Specific Controls). Functional testing is performed in quadruplicate at each time point and the interim test results have demonstrated 6 months stability for the device.
| Sample/Control | Genotype | T0 Result | | | T3 Result | | | T6 Result | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 |
| Control 1 | WT | WT | WT | WT | WT | WT | WT | WT | WT | WT |
| Control 2 | HET | HET | HET | HET | HET | HET | HET | HET | HET | HET |
| Control 3 | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT |
| gDNA 1 | WT | WT | WT | WT | WT | WT | WT | WT | WT | WT |
| gDNA 2 | WT | WT | WT | WT | WT | WT | WT | WT | WT | WT |
| gDNA 3 | HET | HET | HET | HET | HET | HET | HET | HET | HET | HET |
| gDNA 4 | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT | MUT |
| Percent Agreement | | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
Reagent Freeze-Thaw Stability Study
Product in the final configuration was subject to 15 freeze-thaw cycles prior to the
{5}
final thaw at the time of testing. Functional testing was performed in replicates of 3-8 using genomic DNA samples isolated from cell lines, representing all possible genotypes including two (2) WT, two (2) HET and two (2) MUT. Results showed that all samples were correctly called using Invader® Factor V test reagents that underwent repeated freeze-thaw cycles.
d. Detection limit:
Two (2) genomic DNA samples with different genotypes (i.e., one WT and the other HET) were extracted from whole blood collected in potassium EDTA. They were serially diluted and tested 40 times each at eight (8) different input amounts of (0.5, 5, 20, 40, 80, 200, 400, 800 ng/μL). The lower and upper limits of input gDNA when all samples were called correctly were determined to be 5 ng/μL and 80 ng/μL, respectively.
e. Analytical specificity:
Interfering Substances
Test performance was not affected by addition of the following substances to four (4) whole blood samples of different genotype (3 WT, 1 HET) prior to extraction:
- Heparin (1500 U/dL human whole blood)
- Cholesterol (300 mg/dL human whole blood)
- Bilirubin (10 mg/dL human whole blood)
- Hemoglobin (up to 0.2% in whole blood)
- Potassium EDTA (1.8 mg/mL human whole blood)
- Ethanol-based Wash Buffer (5% in DNA sample)
Interfering Mutations
The effect of interfering mutations on G1691A genotyping by the Invader® Factor V Test was evaluated with five (5) contrived DNA samples, including one WT sample with no secondary polymorphism, one HET sample and three WT samples each with a known secondary polymorphism (i.e., G1689A, A1692C or A1696G). Forty replicates for each of the five different samples were tested, and no effect of secondary polymorphisms was observed on G1691A genotyping by the Invader® Factor V Test.
| Non-Panel Mutation or Polymorphism | Panel Mutation |
| --- | --- |
| 1692A>C | 1691G>A |
| 1689G>A | |
| 1696A>G | |
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison:
A total of 352 gDNA samples extracted from whole blood samples were genotyped using the Invader® Factor V Test and DNA sequencing. All samples gave 100% agreement with DNA sequencing.
Table: Comparison of Invader® Factor V Test and DNA Sequencing Results
| Genotype[1] | First Pass Results | | | | | | | Final Results | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | WT | HET | MUT | No- | Mis- | % | 95% | WT | HET | MUT | No- | Mis- | % |
| 1692A>C | 1692A | 1692A | 1692A | 1692A | 1692A | 100% | 100% | 1692A | 1692A | 1692A | 1692A | 1692A | 100% |
| 1689G>A | 1689G | 1689G | 1689G | 1692A | 1692A | 100% | 100% | 1692A | 1692A | 1692A | 1692A | 1692A | 100% |
{6}
7
| | | | | Call | Call | Agree -ment | LCB[2] | | | | Call | Call | Agree -ment | LCB |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WT | 289 | 0 | 0 | 0 | 0 | 100.00 | 98.97 | 289 | 0 | 0 | 0 | 0 | 100.00 | 98.97 |
| HET | 0 | 45 | 0 | 0 | 0 | 100.00 | 93.56 | 0 | 45 | 0 | 0 | 0 | 100.00 | 93.56 |
| MUT | 0 | 0 | 18 | 0 | 0 | 100.00 | 84.67 | 0 | 0 | 18 | 0 | 0 | 100.00 | 84.67 |
Note:
[1] Genotype as determined with DNA sequencing;
[2] Lower boundary of the 95% confidence interval
b. Matrix comparison:
Not applicable
c. Instrument Equivalency:
Twenty-nine (29) human whole blood samples and ten (10) leukocyte depleted whole blood samples spiked with cell lines were extracted using two (2) commonly used extraction methods. The extracts were tested with the Invader® Factor V test using three (3) commercially available thermal cyclers and the raw fluorescent data acquired on three (3) commercially available fluorometers. Results from the three (3) fluorometers were transferred into the interpretive software and genotype calls compared to bi-directional sequencing.
| Concordance by Instrument | | | |
| --- | --- | --- | --- |
| Fluorometer | Thermal Cycler | | |
| | 1 | 2 | 3 |
| A | 78 of 78 = 100% | 77 of 78 = 98.7% | 78 of 78 = 100% |
| B | 78 of 78 = 100% | 77 of 78 = 98.7% | 78 of 78 = 100% |
| C | 78 of 78 = 100% | 77 of 78 = 98.7% | 78 of 78 = 100% |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The Factor V mutation being genotyped in the Invader® Factor V Test is present at a frequency of 5% in the general population.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.