← Product Code [FEQ](/productcode/FEQ) · K100899

# OLYMPUS ENDOSCOPIC FLUSHING PUMP, MODEL OPF-2 (K100899)

_Keymed (Medical & Industrial Equipment), Ltd. · FEQ · Sep 13, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100899

## Device Facts

- **Applicant:** Keymed (Medical & Industrial Equipment), Ltd.
- **Product Code:** [FEQ](/productcode/FEQ.md)
- **Decision Date:** Sep 13, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The OLYMPUS FLUSHING PUMP (OFP-2) and its accessories are intended for use during endoscopic procedures to wash blood, feces and other organic matter from the gastric and colonic mucosa site being visualized, diagnosed and to aid in the filling areas of the gastrointestinal tract with water in order to aid in examinations preformed with trans-endoscopic ultrasound probes.

## Device Story

Motorized peristaltic pump transfers water from container to endoscope; used in medical facilities by trained physicians. Input: water from container; Output: water flow through endoscope instrument or auxiliary channel. User controls flow speed via front-panel buttons (nine settings) or footswitch/endoscope button. Digital controller manages pump head rotation speed; LEDs indicate selected speed. Benefits: clears visualization site of blood/feces; aids ultrasound probe examinations. System includes water container, sterile single-use/single-day tubes, and instrument channel adapter.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-5 (L929 MEM elution) and ISO 10993-10 (Intracutaneous Injection, Kligman Maximization). Electrical safety testing per IEC 60601-1, EN 60601-1, UL 60601-1, CAN/CSA-C22.2 No. 601.1-M90, IEC 60601-1-1, and IEC 60601-2-18. Steam sterilization validation performed for water container.

## Technological Characteristics

Peristaltic pump; digital controller; nine-step adjustable flow speed; LED speed indicators; footswitch/endoscope button operation. Materials: water container, tubing, instrument channel adapter (MAJ-1606). Connectivity: Luer-compatible fittings. Standards: ISO 14971 (risk management), ISO 10993 (biocompatibility), IEC 60601-1 series (electrical safety).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Olympus Endoscopic Flushing Pump, OFP-1 ([K000948](/device/K000948.md))
- Olympus AFU-100 ([K073207](/device/K073207.md))
- Olympus UWS1 Water Supply Unit ([K951994](/device/K951994.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are black and appear to have a slightly distressed or textured effect, as if they have been stamped or printed with some wear. The background is plain and white, which makes the word stand out.

OLYMPUS KEYMED KeyMed House, Stock Road, Southend-on-Sea, Essex, SS2 5QH, UK

Your Vision, Our Future

Telephone: +44 (0)1702 616333 Facsimile: +44 (0)1702 465677

K 100899-

SEP 1 3 2010

# 510 (K) SUMMARY OLYMPUS ENDOSCOPIC FLUSHING PUMP OFP-2

Contact Name: Company: KeyMed (Medical & Industrial Equipment) Ltd Address: KeyMed House Stock Road Southend-on-Sea Essex SS2 5QH UK Telephone: +44 (0)1702 616 333 Fax: +44 (0)1702 465 677 E-mail:

| Owner/Operator #:             | 8020728 |
|-------------------------------|---------|
| Establishment Registration #: | 9611174 |

Date of Preparation:

22 March 2010

| Name of Device:      | Olympus Endoscopic Flushing Pump (OFP-2)                                                     |
|----------------------|----------------------------------------------------------------------------------------------|
| Common Name:         | Endoscopic Flushing or Lavage Pump                                                           |
| Classification Name: | Pump, Air, Non-Manual, For Endoscope                                                         |
| Product Code:        | FEQ (21 CFR 876.1500)                                                                        |
|                      | (as per classification of predicate device)                                                  |
| Basis of submission: | This submission claims substantial equivalence to the<br>following legally marketed devices: |
|                      | Olympus Endoscopic Flushing Pump, OFP-1<br>(K000948)                                         |

- Olympus AFU-100 (K073207) .
- Olympus UWS1 Water Supply Unit (K951994) .

Image /page/0/Picture/13 description: The image contains three logos. The first logo is a circle with the word "REGISTERED" around the edge and a symbol in the center. The second logo is a square with a crown and the letters "UKAS" above the number "009". The third logo is partially visible and contains the words "POSITIVE" and "DISABLED".

Image /page/0/Picture/14 description: The image contains two distinct graphic elements. On the left, a portion of text is visible, with the word "PEOPLE" clearly discernible. To the right, there is a logo featuring a stylized letter "Q" with a wing extending from it, topped by a crown. Below the "Q", the text "THE QUEEN'S" is present, although the rest of the text is not clear.

Image /page/0/Picture/15 description: The image shows two logos for The Queen's Awards for Enterprise. Each logo features a stylized letter "e" with a crown above it. The left logo has the text "QUEEN'S AWARDS FOR ENTERPRISE" below the letter, while the right logo has the text "THE QUEEN'S AWARD FOR ENTERPRISE".

Image /page/0/Picture/16 description: The image shows a logo for "The Queen's Awards for Enterprise 2006". The logo features a stylized letter "E" with a wing-like design extending from its left side. Above the "E" is a crown, symbolizing the British monarchy. The text "THE QUEEN'S AWARDS FOR ENTERPRISE 2006" is arranged below the "E" in a stacked format.

Otympus KeyMed is an operating division of KeyMed (Medical & Industrial Equipment) Ltd. Company Registered in England No. 966735

ed in England (No. 900700)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. Below the word is a thick black line. Underneath the line, the phrase "Your Vision, Our Future" is written in a smaller font.

K 100899

Page 2 of 3

# DEVICE DESCRIPTION

The Olympus Endoscopic Flushing Pump (OFP-2) is a motorised pump designed to transfer water from a bottle, via a tube, to an endoscope.

The OFP-2 uses a peristalic pump head to move the water through the tube. The speed of rotation of the pump head is adjustable by the user, over nine steps, using flow control buttons on the front of the pump. The flow control buttons are connected to a digital controller that governs speed of rotation of the pump head. Nine LEDs on the front of the pump indicate the speed setting that has been selected.

The pump is operated by a footswitch or a button on the endoscope. It has an on/off button and a standby button.

The OFP includes a water container supplied non-sterile to be processed before use. Two water tubes of different lengths and an instrument channel adapter are supplied sterile. The Instrument Channel Adapter (MAJ-1606) and Instrument Channel Water Tube (MAJ-1607) are supplied sterile for single use only. to pump water through the instrument channel of the endoscope. The Auxilliary Channel Water Tube (MAJ-1608) is supplied sterile for single day only use.

#### INDICATIONS FOR USE

The OFP-2 and accessories are intended for use in medical facilities under the direction of a trained physician.

The OFP-2 is intended for use:

- . During endoscopic procedures to wash blood, feces and other organic matter from the site being visualised, diagnosed or treated.
- To aid in filling areas of the gastrointestinal tract with water in order to aid in . examinations performed with trans-endoscopic ultrasound probes

#### SUMMARY OF TESTING

Risk Management of the OFP-2 and its tubing was performed to ISO 14971 and included a device hazard analysis.

Biocompatibility of the Water Container and tubing has been established by subjecting the container, a tube and the Instrument Channel Adapter to the following tests:

- L929 MEM elution test to ISO 10993-5 .
- Intracutaneous Injection Test to ISO 10993-10 .
- Kligman Maximization Test to ISO 10993-10 .

Electrical safety testing of the OFP-2 and its accessories was performed to:

- . IEC 60601-1:1988
- EN 60601-1:1990 ●
- UL 60601-1:2003 .
- CAN/CSA-C22.2 No. 601.1-M90 .

Additional testing was performed to:

- IEC 60601-1-1 .
- IEC 60601-2-18 .

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. A horizontal line is underneath the word. Below the line, the text "Your Vision, Our Future" is written in a smaller font.

i<100899
Page 3 of 3

The effectiveness of steam sterilization on the water container was validated, as was the robustness of the water container when subjected to steam sterilization over the expected useful life of the container.

The performance of the OFP-2 and its accessories was verified and validated.

### EQUIVALENCE TO PREDICATE DEVICES

# Substantial equivalence to the Olympus Flushing Pump (OFP-1)(K000948)

The proposed device, the OFP-2, is substantially equivalent to the OFP-1 in the following ways:

- . Both devices share the same intended use
- . Both devices utilize the same method of operation
- . Both devices utilize the same method of connection - a Luer compatible fitting - to connect the water supply tube to the Auxiliary Channel of the endoscope

#### Substantial equivalence to the Olympus AFU-100 peristaltic pump unit (K073207)

The proposed device, the OFP-2, is substantially equivalent to the Olympus AFU-100 in the following ways:

- Both devices share the same intended use .
- Both devices share the same method of operation. The OFP-2 is supplied with a foot . switch which is optional on the AFU-100.
- Both devices utilize the same method of connection a Luer compatible fitting to . connect the water supply tube to the endoscope. The OFP-2 uses an additional adapter - MAJ1606, included in this application - to connect to the Instrument Channel of some endoscopes.

#### Substantial equivalence to the Olympus UWS-1 water supply unit (K951994)

The proposed device, the OFP-2, is substantially equivalent to the Olymous UWS-1 water supply unit in the following ways:

- . Both devices share the same intended use
- . Both devices share a similar method of operation. The circuit formed by the water container, tube and endoscope is functionally equivalent. The OFP-2 uses a peristaltic pump to move the fluid along the tube while the UWS-1 uses an electromagnetic pump.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

KeyMed (Medical & Industrial Equipment) Ltd % Mr. Ron Warren Principal Consultant, Regulatory Services Medical Device Consultants, Inc. 11440 West Bernardo Drive, Suite 300 SAN DIEGO CA 92127

SEP 1 3 2010

Re: K100899

Trade/Device Name: Olympus Endoscopic Flushing Pump (OPF-2) Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FEQ Dated: March 29, 2010 Received: June 16, 2010

Dear Mr. Warren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Huckel Lemur us

erbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{5}------------------------------------------------

# Indications for Use

# SEP 1 3 2010

510(k) Number (if known):

OLYMPUS ENDOSCOPIC FLUSHING PUMP (OPF-2) Device Name:

| And accessories : MAJ – 1603: Water Container (2L) 3 pack |
|-----------------------------------------------------------|
| MAJ - 1606: Instrument Channel Adapter                    |
| MAJ - 1607: Instrument Water Tube                         |
| MAJ - 1608: Auxiliary Channel Water Tube                  |

Indications For Use:

The OLYMPUS FLUSHING PUMP (OFP-2) and its accessories are intended for use during endoscopic procedures to wash blood, feces and other organic matter from the gastric and colonic mucosa site being visualized, diagnosed and to aid in the filling areas of the gastrointestinal tract with water in order to aid in examinations preformed with trans-endoscopic ultrasound probes.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K100899](https://fda.innolitics.com/device/K100899)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
