K100837 · Fimreite Software AS · LLZ · Dec 3, 2010 · Radiology
Device Facts
Record ID
K100837
Device Name
PACENTRIC CONNECT, TRAVELER AND WEB
Applicant
Fimreite Software AS
Product Code
LLZ · Radiology
Decision Date
Dec 3, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
PaCentric Web is a software device intended for viewing of images acquired from CT, MR, CR, US and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated, processed, and displayed within the system and or across computer networks at distributed locations. The device is not intended for clinical viewing of mammography images.
Device Story
PaCentric Web is a web-based software application for read-only review of DICOM images and reports. It enables remote access to clinical data via web browsers, requiring a key and password provided by the sender. The device facilitates image display, measurement (distance, area, volume, angle, ratios), window-leveling, brightness control, color mapping, and image manipulation (flipping, zooming). It generates simple printable reports containing patient demographics and clinical data. Used by radiologists, specialists, and other medical professionals in clinical settings to review images acquired at different geographic locations. The system ensures secure, time-limited access to data through encrypted internet communication. By providing remote access to diagnostic images, it supports clinical decision-making and patient care coordination across distributed networks.
Clinical Evidence
No clinical data. Bench testing only. The device underwent risk analysis, requirements reviews, design reviews, module verification, system integration testing, final acceptance validation, and performance testing to ensure compliance with voluntary standards.
Technological Characteristics
Software-only device; operates on commercial standard hardware. Uses DICOM standard for image communication. Connectivity via internet using VeriSign Class 3 Extended Validation SSL CA, HTTP with Privacy TLS 1.0, AES 128-bit encryption, and RSA 1024-bit key exchange. Features include DICOM-compliant image processing tools (window/level, measurements, color maps).
Indications for Use
Indicated for radiologists, cardiologists, surgeons, chiropractors, dentists, and other trained medical professionals to review clinical DICOM images (CT, MR, CR, US) acquired at remote locations. Not indicated for clinical viewing of mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K100Y37 Fimreite SOFTWARE AS". The text "Fimreite" is in a larger, bold font. There is a partial circle to the right of the text. The text appears to be a logo.
510(k): Abbreviated Pre-market Notification Submission
## 5 510(k) Summary
As required by section 807.92{c}
DEC - 3 2010
| 510(k) owner's<br>name: | Fimreite Software AS |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Torggata 26, 3181 Horten, Norway |
| Phone: | (47) 93 21 56 27 |
| Contact person: | Svein Fimreite |
| Date of preparation: | 29 October 2010 |
| Name of Device | PaCentric Web |
| Trade Name: | PaCentric Web |
| Common Name: | Picture Archiving and Communications Systems and Workstation |
| Classification Name: | 21 CFR 892.2050 System, Image Processing, Radiological |
| Predicate Device: | K062488 iQ-System PACS |
| Device Description: | PaCentric Web is an Internet application for read-only review of external image<br>deliveries. |
| | PaCentric Web facilitates the display and/or transfer of clinical DICOM images<br>and reports to a Web browser. PaCentric is ensuring secure and time-limited<br>access to clinical data. The receiver will only have access to the relevant data<br>for a particular delivery. Using PaCentric, patient data can be delivered to a<br>specific recipient anywhere in the world. |
| | PaCentric Web gives a qualified user read-only access to an image delivery.<br>Access is given with a key and password, which must be provided by the<br>Sender. |
| | PaCentric Web features commonly used tools and features found on DICOM<br>workstations: |
| | • Preview image icons from which the user may select images to review in<br>full size |
| | • Measurements: Distance, Area, Volume, Angle and ratios. Measurement<br>calibration based on information provided by the relevant DICOM tag. The<br>measurements are performed by having the user select a caliper and<br>marking end points by clicking with a mouse button. The results are being<br>displayed and updated in a results area on the screen. |
| | • Horizontal and Vertical invert, i.e. flipping the image sideways or upside-<br>down. |
| | • Color maps. A pixel consists of three sets of values from 0 to 255 that<br>together constitute a color or a gray level. Technically, these RGB values<br>are reduced according to desired color hue. A grayscale image therefore<br>will not lose any details in the image for higher or lower values. This<br>function is mostly used for MONOCHROME1 and MONOCHROME2 images. |
| | • Window-leveling and Brightness control. These Attributes is only used for<br>Images with Photometric Interpretation (0028,0004) values of |
| | MONOCHROME1 and MONOCHROME2. They have no meaning for other<br>Images. |
| | When working with DICOM images, the user has the ability to make<br>changes to the displayed images, on a global level, by manipulating the<br>Window Level. Tonal changes to specific areas of an image. Window Center<br>DICOM TAG 0028,1050 and Window Width DICOM TAG 0028,1051 specify<br>a linear conversion from stored pixel values (after any Modality LUT or |
| | Rescale Slope and Intercept specified in the IOD have been applied) to |
| | values to be displayed. Window Center contains the input value that is the<br>center of the window. Window Width contains the width of the window.<br>Note: The terms "window center" and "window width" are not consistently<br>used in practice, nor were they defined in previous versions of the<br>standard. The definitions here are presented for the purpose of defining<br>consistent meanings for identity and threshold transformations while<br>preserving the common practice of using integral values for center and<br>width.<br>• Zoom - Image is downloaded in its original size, but the height and width<br>definition can be adjusted and will therefore function as zoom.<br>• Language selection. Default language is based on current location derived<br>from the IP address of the viewer.<br>• Simple printable report with preview and selectable fields: Display of<br>patient demographics: Sex, age, date of examination, nationality, operator,<br>weight, referring physician, height, performing physician, body surface<br>area (BSA), location, BP, Sender address, Description, Diagnosis,<br>Comments, Images, selectable signature lines<br>• Ability to display up to eight independent image areas at the same time |
| Intended Users: | Intended users PaCentric Web include radiologists, specialists including<br>cardiologists, surgeons, chiropractors, dentists and other trained medical<br>professionals who have a need for reviewing clinical data acquired at some<br>other geographic location. |
| Intended Use: | PaCentric Web is a software device intended for viewing of images acquired<br>from CT, MR, CR, DR, US and other DICOM compliant medical imaging systems<br>when installed on suitable commercial standard hardware.<br>Images and data can be captured, stored, communicated, processed, and<br>displayed within the system and or across computer networks at distributed<br>locations.<br>The device is not intended for clinical viewing of mammography images. |
| Technology: | PaCentric Web use the same functional scientific technology as its predicate<br>devices including a transfer protocol based on Query/Retrieve within the<br>DICOM standard. In addition, PaCentric uses secure transfer protocols with<br>VeriSign Class 3 Extended Validation SSL CA HTTP with Privacy TLS 1.0, AES with<br>128 bit encryption (High); RSA with 1024 bit exchange for internet<br>communication. |
| Test Summary: | PaCentric Web complies with the voluntary standards as detailed in Section 9.<br>The following quality assurance measures were applied to the development:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Testing on unit level (Module verification)<br>• Integration testing (System verification)<br>• Final acceptance testing (Validation)<br>• Performance testing |
| Conclusion: | Fimreite Software AS considers features of PaCentric Web substantially<br>equivalent to those of the predicate device, but believes that PaCentric should<br>be considered a device with Moderate Level of Concern. |
{1}------------------------------------------------
510(k): Abbreviated Pre-market Notification Submission
Image /page/1/Picture/1 description: The image shows the logo for "Fimreite SOFTWARE AS". The word "Fimreite" is in bold, and the words "SOFTWARE AS" are in a smaller font size below the word "Fimreite". To the right of the word "Fimreite" is a partial circle with three dots inside of it.
-
.
・
. .
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Svein Fimreite Managing Director Fimreite Software AS Torggata 26. Horten. 3181 NORWAY
DEC - 3 2010
Re: K100837
Trade/Device Name: PaCentric Web Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 28, 2010 Received: November 9, 2010
Dear Mr. Fimreite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
.
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Dan
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Fimreite SOFTWARE AS. The word "Fimreite" is in bold, black letters, with "SOFTWARE AS" in smaller letters below. To the right of the word "Fimreite" is a partial circle with three dots inside.
510(k): Abbreviated Pre-market Notification Submission
## 4 Indications for Use Statement
510(k) Number: K100837
Device Name: PaCentric Web
Indications for Use:
PaCentric is a software only device intended for viewing of images acquired from CT, MR, CR, US and other DICOM compliant medical imaging systems when installed on suitable commercial hardware. Images and data can be captured, stored, communicated, processed, and displayed within the system and or across computer networks at distributed locations. The device is not intended for clinical viewing of mammography images.
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use NO
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K K160837
Fimreite Software AS Torggata 26, 3181 Horten, Norway
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.