The Pioneer Poise Anterior Cervical Plate System is intended for anterior cervical fixation for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
Device Story
System consists of titanium alloy plates and screws for anterior cervical spine fixation; includes static and dynamic plate designs (1-4 levels) to facilitate postoperative load sharing; features integrated Nitinol screw-retention mechanism. Used by surgeons in clinical settings to secure vertebral bodies; screws (4.0/4.5mm diameter, 10-20mm length) available in self-tapping or self-drilling configurations. Provides mechanical stabilization of cervical spine segments to promote fusion; benefits patients by addressing instability caused by degenerative disease, trauma, or deformity.
Clinical Evidence
Bench testing only. Mechanical performance characterized via static and fatigue compression bending and static torsion per ASTM F1717, and axial pullout strength per ASTM F543. Additional internal protocol testing assessed screw pull-through and retention characteristics.
Technological Characteristics
Materials: Titanium alloy (ASTM F136) and Nitinol (ASTM F2063). Components: Plates (10-84mm) and screws (4.0/4.5mm diameter, 10-20mm length). Design: Static and dynamic (uni-directional axial movement) plates with integrated screw-retention mechanism. Energy source: None (manual surgical instrumentation).
Indications for Use
Indicated for anterior cervical fixation in patients with degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
Pioneer Anterior Cervical (PAC) Plate System (K083663)
Pioneer SlimFuse Anterior Cervical Plating System (K072703)
{0}------------------------------------------------
-
## 510(k) Summary
| Sponsor: | JUN - 4 2010<br>Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855<br>(906) 225-5602<br>Contact: Emily M. Downs<br>Prepared May 20, 2010 |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Pioneer Poise Anterior Cervical Plate System |
| Classification<br>Name: | The classification of the Pioneer Poise Anterior Cervical Plate System is<br>Class II, as per the Code of Federal Regulations, Title 21, Section<br>888.3060: Spinal intervertebral body fixation orthosis. |
| Product Codes: | The product code is KWQ. The Panel code is 87. |
| Predicate Device: | K083663- Pioneer Anterior Cervical (PAC) Plate System (SE 2/25/09)<br>K072703- Pioneer SlimFuse Anterior Cervical Plating System (SE 1/10/08)<br>K053053- Pioneer Anterior Cervical Plate System (SE11/16/05)<br>K043066- Pioneer Anterior Cervical Plate System (SE 2/3/05) |
| Description: | The Pioneer Poise Anterior Cervical Plate System consists of an assortment<br>of plates and screws. The screws are used to secure the plates to the<br>vertebral bodies of the cervical spine through an anterior approach. The<br>Poise system consists of static and dynamic plates in lengths that range<br>from 10-84mm and include one, two, three and four level designs. The<br>dynamic design provides for uni-directional axial movement for<br>postoperative load sharing between the plate and graft. The plates have an<br>integrated screw retention mechanism. Screws range from 10-20mm in<br>length, 4.0 or 4.5mm in diameter, and are available in self tapping or self<br>drilling. The plates and screws are manufactured from Titanium alloy<br>(ASTM F136) and Nitinol (ASTM F2063). |
| | The system also contains Class 1 manual surgical instruments and cases<br>that are considered exempt from premarket notification. |
| Intended Use: | The Pioneer Poise Anterior Cervical Plate System is intended for anterior<br>cervical fixation for the following indications: degenerative disc disease<br>(DDD) (defined as neck pain of discogenic origin with degeneration of the<br>disc confirmed by history and radiographic studies), spondylolisthesis,<br>trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures<br>(i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed<br>previous fusion. |
{1}------------------------------------------------
The subject device has indications for use identical to those of the predicate Comparison to Predicate Devices: PAC system and employs the same principles of operation. Identical materials are used for both subject and predicate plates and screws. A vailable screw lengths (10-20mm), screw diameters (4.0/4.5mm), overall plate length (10-84mm), and plate types (one, two, three and four level; static and dynamic) are identical for both systems.
The Pioneer Poise Anterior Cervical Plate System plates and screws are Material: manufactured from titanium alloy (Ti6Al4V ELI, according to ASTM The plates incorporate a screw-retention mechanism ("spring"), F136). which is manufactured from Nitinol (ASTM F2063).
Non-Clinical Mechanical testing was presented to characterize construct and component performance, including testing static and fatigue compression bending and Performance Data: static torsion per recognized ASTM F1717, axial pullout screw strength per ASTM F543, and additional testing performed per internal protocols to assess screw pull through plate and screw retention characteristics. The test results of verification testing demonstrate that the mechanical performance of the Pioneer Poise Anterior Cervical Plate System is substantially equivalent to the predicate devices.
Comparisons of device performance data, materials, indications and Performance and design/function to predicate devices were provided in making a SE Determination: determination of substantial equivalence.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or feathers. The bird is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## JUN - 4 2010
Pioneer Surgical Technology % Ms. Emily M. Downs 375 River Park Circle Marquette, Michigan 49855
Re: K100708
Trade/Device Name: Pioneer Poise Anterior Cervical Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: April 13, 2010 Received: April 14, 2010
Dear Ms. Downs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Emily M. Downs
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sincerely yours,
Mark M. Mulkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K100708
-Device-Name:
Indications for Use:
The Pioneer Poise Anterior Cervical Plate System is intended for anterior cervical fixation for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and
Pioneer-Poise Anterior Cervical Plate
Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
failed previous fusion.
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
K100708 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.