The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood and blood products to a human patient under the direction or supervision of physician or other certified health care professional.
Device Story
Z-800 Infusion System is a pole-mounted volumetric linear peristaltic pump; delivers parenteral fluids, blood, and blood products. System comprises pump unit and approved external IV administration sets. User interface includes 128x64 LCD, membrane keypad, and LED flow rate display. Embedded software executes state machine to monitor infusion and safety alarms; watchdog timer (50ms) triggers fail-safe state upon CPU reset. Pump features all-metal casing (milled aluminum) and tubing guide cavity to prevent free flow. Healthcare providers program infusion parameters via scroll keys; device provides visual/audible alerts for occlusion, air-in-line, and door status. Benefits include precise fluid delivery and integrated safety mechanisms to mitigate data entry errors and free-flow risks.
Clinical Evidence
Bench testing only. Hemolysis testing performed on human blood using production units to assess the impact of the new administration set. Results met acceptance criteria: plasma free hemoglobin < 2 mg/mL; hemoglobin standard curve correlation coefficient > 0.95. Risk management report confirms no incremental risk to patient/user safety.
Technological Characteristics
Volumetric linear peristaltic pump; milled aluminum/sheet metal casing. Power: 100-240V AC or internal 4700mA NiMH battery. Connectivity: Bidirectional serial communication. User interface: 128x64 dot matrix LCD, 14-key membrane keypad. Software: Embedded state machine with watchdog timer. Sterilization: Not specified (device is pump unit).
Indications for Use
Indicated for intravenous infusion of parenteral fluids, blood, and blood products in human patients under the direction or supervision of a physician or certified healthcare professional. No specific age or gender restrictions; no known contraindications.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
K100705
### Appendix G
## Summary of Safety and Efficacy
## 510(K) SUMMARY
Zyno Medical LLC. 9 Tech Circle Natick, MA, 01760
#### Contact Person:
Chaoyoung Lee President Zyno Medical LLC 9 Tech Circle Natick, MA, 01760 Fax: 508-650-2006 Phone: 508-650-2008 Ext 103 Email: cylee@careeverywhere.com
Date Originated: March 5, 2010 Trade Name: Z-800 Infusion System Common Name: Volumetric Infusion Pump Classification Name: Infusion Pump Product code: FRN 880.5725
#### Predicate Devices
-Z-800 Infusion pumps (K071545) -B. Braun Vista basic (formerly known as Infusomat P. K003029) -Sigma 8000 (K950766) -Abbott Acclaim Encore (K991501)
#### Intended Use
The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood and blood products to a human patient under the direction of physician or other certified health care professional.
#### Device Description:
The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids. blood and blood products to a human patient under the direction of physician or other certified health care professional.
The Z-800 Infusion System consists of the Z-800 Infusion Pump and approved external IV administration sets. The scope of this 510K is to qualify CareFusion's blood administration sets
#### Z-800 Premarket Notification - Modified on June 2, 2010 in response to FDA request
JUL -- 8 2010
{1}------------------------------------------------
(K882302) for use with the currently marketed Z-800 infusion pump system so as to extend the intended use of the Z-800 Volumetric Infusion Pumps to include intravenous administration of blood and blood products. The infusion pump remains the same as currently marketed Z-800 infusion pump, as in the predicated device Z-800 infusion system (K0717545).
The Z-800 infusion system contains the following components:
- The infusion delivery mechanism: Volumetric linear peristaltic pumping mechanism assembly. .
- Approved IV administration sets include External IV sets for intravenous parenteral fluid delivery . and for intravenous blood and blood product delivery.
- . User Interface consists of a 128 X 64 dot matrix LCD display; a 14 key membrane keypad; a LED flow rate display; two device status indicator LEDs; power status indicator LEDs (AC & Battery); an audio speaker; an external alarm light accessory; an software alarm handler.
- The embedded software on the main CPU execute an infusion pump state machine, which monitors . the execution of currently programmed infusion as well as a collection of safety related alarm conditions.
- A watchdog timer is set at 50ms interval. Fail to satisfy the watchdog timer will cause both main . CPU and motor control CPU reset, pump device parks at a fail safe state, and an audible alarm to sound.
- Power supply consists of AC/DC power. Z-800 infusion pump is intended to be used as a pole . mounted pump. It contains a medical device grade switch power supply (input 100-240V 50-60Hz, 0-1.4A. output 15vdc) and a rechargeable 4700mA. 8X1.2 volt Nickel Metal Hydride battery pack. A fully charged new battery will power at least 8hr of continuous infusion (at 125mL/hr). A full battery charging time on a new battery is around 5hr. The battery charging circuitry has a dedicated controlling CPU, which manages the charging current as well as the battery safety monitoring. Redundant temperature sensor is embedded in the battery pack capable of shut off the charging current if battery pack temperature is above the safety threshold.
- . There is 128K flash EPROM memory which stores the embedded software along with 8K NVRAM for user specified configuration as well as preserving infusion parameters between power cycles.
- . Z-800 infusion pump uses linear peristaltic pumping mechanism driven by a step motor through driving belt. The ID/OD, wall thickness and the durometer of the IV set are specified by Zyno to the IV set manufacturers to ensure the accuracy specification of flow rate.
- . The user interface design requirement of the Z-800 pump is focused on simplicity and intuitiveness. The keypad contains 3 groups of keys: the navigation key group; the data entry key group; and the action key group. There is no numeric data entry keypad. All data entry kevs are scroll key. This design is to focus user's attention to the data entry results on the display while entering the infusion parameters, so as to mitigate data entry error caused by key bounce or user hitting the wrong keys,
- . The Z-800 infusion pump has an all metal casing to improve the durability of the pump and the integrity of the pumping/free flow protection mechanism. The pump door and the mating pump front panel are made in milled aluminum. The tubing guide cavity is designed to mitigate tubing loading error. If the tubing is not loaded correctly, the door can not be closed. Even user jam close the door, the mating pump door and the front panel will shut off the tubing to prevent free flow, and the drip sensor accessory will report no flow to alert user of misloaded tubine.
- . There is no known contraindication.
The equivalency matrix (Table 1) compares the Z-800 Infusion pump with the predicate devices.
### Z-800 Premarket Notification – Modified on June 2, 2010 in response to FDA request
{2}------------------------------------------------
## Table 1: Equivalency Matrix
| Parameter | Z-800 | B. Braun Vista basic | Sigma 8000 | Abbott Acclaim Encore |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pump Type | Volumetric Infusion Pump | Volumetric Infusion Pump | Volumetric Infusion Pump | Volumetric Infusion Pump |
| Intended use | Intra-venous | Intra-venous<br>Enteral | Intravenous<br>Epidural | Intravenous<br>Enteral |
| Units of delivery | ml/hr | mg/kg/min<br>mg/kg/hr<br>mg/min<br>mg/hr<br>mcg/kg/min<br>mcg/kg/hr<br>mcg/min<br>mcg/hr<br>g/kg/hr<br>g/hr<br>g/kg/min<br>g/min<br>U/kg/hr<br>U/kg/min<br>U/min<br>U/hr<br>ml/hr<br>ml/min | units/hr<br>units/kg/min<br>units/kg/hr<br>milliunits/min<br>grams/hr<br>mcg/kg/min<br>mcg/kg/hr<br>mcg/min<br>mcg/hr<br>mg/kg/min<br>mg/kg/hr<br>mg/min<br>mg/hr<br>ml/hr | ml/hr<br>mcg/min<br>mcg/kg/min |
| Pumping<br>mechanism | Linear peristaltic pump | Linear peristaltic pump | Linear peristaltic pump | Linear peristaltic pump |
| Free Flow<br>Protection | IV set based free flow<br>protection (Option)<br>Pump based free flow<br>protection YES | IV set based free flow<br>protection (Option)<br>Pump based free flow<br>protection YES | IV set based free flow<br>protection YES.<br>Pump based free flow<br>protection YES | IV set based free flow<br>protection YES.<br>Pump based free flow<br>protection YES |
| Administration<br>sets | Approved standard PVC<br>gravity tubing | B. Braun basic pump<br>administration sets. | Approved standard PVC<br>gravity tubing | Approved standard PVC<br>gravity tubing |
| Power source | AC: 100-250V 50-60 Hz<br>DC: Internal Nickel<br>Metal Hydride | AC: 100-120V 50-60 Hz<br>DC: Internal NiCd | AC:<br>105-135V 50/60 Hz<br>220-240V 50/60 Hz<br>DC: Sealed lead acid | AC: 100-130V 47-63 Hz<br>DC: Internal<br>rechargeable 8V battery |
| Battery Life | 8 hours at 125 ml/hr | 3 hours at 125 ml/hr, | 5 hours | 8 hours at 125ml/hr |
| Display | Program controlled dot<br>matrix LCD | Custom LCD | LED | Custom LCD |
| Case construction | Milled Aluminum &<br>Sheet Metal Enclosure | Sheet Metal and<br>Injection Molded<br>Plastic Enclosure | High impact Injection<br>Molded Plastic<br>Enclosure | High impact Injection<br>Molded Plastic<br>Enclosure |
| Size (inch) | 8.6H x 5.7W x 5.3D | 9.4Hx 5.5W x 7.8D | 9.5Hx 6.5Wx 6.75D | 8.25Hx7Wx5.74D |
| Weight | 5.7 lbs | 6.8 lbs. | 10 lbs | 7 lbs |
| Volume and Rate<br>Accuracy | +/- 5% | +/- 5% | +/- 5% | +/- 5% |
| Occlusion<br>Pressure Accuracy | Low: 4 psi<br>Medium 16 psi<br>High: 30 psi<br>adjustable | Low: 8 psi, unspecified<br>accuracy<br>High 17 psi,<br>unspecified accuracy | Low 2 psi<br>High 15 psi<br>adjustable | Low 6 psi<br>Medium 10pis<br>High 20 psi |
| Parameter | Z-800 | B. Braun Vista basic | Sigma 8000 | Abbott Acclaim Encore |
| Rate Range | 1 ml/hr to 999 ml/hr in<br>1 ml/hr increment | 0.1 to 800 ml/hr | 0.1 to 99.9 ml/hr (in 0.1<br>ml/hr increments)<br>100 to 999 ml/hr (in 1<br>ml/hr increments) | 1.0 to 99.9 ml/hr (in<br>0.1ml/hr increments)<br>100 to 999 ml/hr (in 1<br>ml/hr increments) |
| Warnings / Status | Low Battery<br>Near End<br>Infusion Complete<br>Pump unattended<br>KVO | Low Battery<br>Near End<br>Infusion Complete<br>Pump Stopped<br>Bag Empty (with Drip<br>Sensor)<br>KVO | Low battery<br>Improper IV set loading<br>Infusion Complete<br>Close Roller Clamp<br>PM Due<br>KVO | Battery<br>Check Settings<br>VTBI complete<br>Stopped (unattended)<br>No Slide Clamp<br>KVO |
| Alarms | Occlusion<br>Battery Empty<br>Air-In-Line<br>Door Open<br>System Error<br>No Drip (with drip<br>sensor)<br>Drip sensor connection<br>(with drip sensor) | Occlusion<br>Plug In Now<br>Air-In-Line<br>Door Open<br>Unit Malfunction | Occlusion<br>Plug in<br>Air-In-Line<br>Pump Malfunctions | Occlusion<br>Battery Off<br>Air-In-Line<br>Check Door<br>Check Slide Clamp |
| Inputs | Continuous Mode<br>Primary Rate<br>Primary VTBI<br>Secondary Rate<br>Secondary VTBI<br>Volume/Time<br>Primary VTBI<br>Primary Time<br>Secondary VTBI<br>Secondary Time<br>TPN<br>Total VTBI<br>Total Time<br>Ramp Up Time<br>Ramp Down Time<br>10 Step Sequence<br>Nth Rate<br>Nth VTBI<br>(N=1 to 10) | Continuous Mode<br>Primary Rate<br>Primary VTBD<br>Secondary Rate<br>Secondary VTBD<br>Volume/Time<br>Primary VTBD<br>Primary TIME<br>Secondary VTBD<br>Secondary TIME<br>Rate & Time<br>Primary Rate<br>Primary Time<br>Secondary Rate<br>Secondary Time<br>TPN<br>Total Bag Volume<br>Total Time<br>Ramp Up time<br>Taper Down Time<br>10 Step Sequence<br>Nth Rate<br>Nth VTBD<br>(N=1 to 10)<br>Dose Calc Mode<br>Concentration Value<br>Concentration Unit<br>Body Weight<br>Drug Dose amount<br>Drug Dose Unit<br>Diluent Volume<br>Delay A Mode<br>Delay Time Amount | Continuous Mode<br>Primary Rate<br>Primary VTBI<br>Piggyback Rate<br>Piggyback VTBI<br>Volume/Time<br>Primary VTBD<br>Primary TIME<br>Piggyback VTBD<br>Piggyback TIME<br>Rate & Time<br>Primary Rate<br>Primary Time<br>Piggyback Rate<br>Piggyback Time<br>TPN<br>Volume Limit<br>Main Rate<br>Ramp Time<br>Taper Time<br>21 Step Sequence<br>Nth Rate<br>Nth VTBI<br>(N=1 to 21)<br>Dose Rate Mode<br>Drug Dose amount<br>Drug Dose Unit<br>Body Weight<br>Diluent Volume<br>Volume Limit<br>Delay Start Mode<br>Delay Time Amount<br>KVO option | Continuous Mode<br>Primary Rate<br>Primary VTBI<br>Secondary Rate<br>Secondary VTBI<br>Volume/Time<br>Primary VTBI<br>Primary TIME<br>Secondary VTBI<br>Secondary TIME<br>TPN<br>Total VTBI<br>Total TIME<br>Taper up TIME<br>Taper down TIME<br>Dose Calc for<br>MICROGRAM delivery<br>Patient Weight<br>Drug amount<br>Diluent Volume<br>Dose Rate |
| Parameter | Z-800 | B. Braun Vista<br>basic | Sigma 8000 | Abbott Acclaim<br>Encore |
| Serial<br>Communications | Bidirectional | Bidirectional | Bidirectional | None |
## Z-800 Premarket Notification – Modified on June 2, 2010 in response to FDA request
{3}------------------------------------------------
Z-800 Premarket Notification – Modified on June 2, 2010 in response to FDA request
·
36
.
{4}------------------------------------------------
.
-
.
Z-800 Premarket Notification – Modified on June 2, 2010 in response to FDA request
.
·
,
37
:
{5}------------------------------------------------
System validation for the Z-800 Volumetric Infusion Pump was performed according to the Z-800 System Validation Plan in compliance with the FDA guidance "Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions (DRAFT GUIDANCE) issued in April 23, 2010, and Zyno Medical's Quality Development Procedure. Assurance case report has been generated to demonstrate there is no incremental risk to patient and user safety as well as to the Z-800 infusion system efficacy has been introduced by qualifying CareFusion blood IV set (K882302) to be used with Z-800 infusion pump device. Hemolysis was identified as incremental risk. As a result of this risk analysis assessment, flowing blood hemolysis test with human blood was performed using production units which are identical to the currently marketed Z-800 infusion pumps. Per Zyno Medical LLC request, the test was conducted by a certified lab according to recognized Good Laboratory Practices following FDA recognized ASTM standard for assessment of hemolytic properties. All acceptance criteria were met (Each blood sample had a plasma free hemoglobin level that is less than 2 mg/mL; the correlation coefficient for the hemoglobin standard curve is greater than 0.95). Risk management activity has been incorporated into the device design/development process. A risk management report has been compiled to document risk level assessment for the list of FDA identified risks for infusion pump systems in Appendix A of "Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions (DRAFT GUIDANCE) issued in April 23, 2010. Z-800 Validation Summary Report TR800-2010-02 submitted in Appendix E summarized the validation test results, which supports the conclusion which indicates that adding CareFusion blood set (K882302) to be used with Z-800 infusion pump for intravenous delivery of blood and blood products does not introduce any incremental risk to the user/patient safety.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird-like figure with three curved lines representing wings or snakes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Chaoyoung Lee President Zyno Medical LLC 9 Tech Circle Natick, Massachusetts 01760
JUL - 8 2010
Re: K100705
Trade/Device Name: Zyno Z-800 Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Dated: June 8, 2010 Received: June 10, 2010
Dear Mr. Lee:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Ctim U.m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Z-800 Volumetric Infusion System
Indications for Use: The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood and blood products to a human patient under the direction or supervision of physician or other certified health care professional.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rill C. An 7/8/10
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: k/00705
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.