NOVABONE DENTAL PUTTY-BIOACTIVE SYNTHETIC BONE GRAFT
K100671 · Novabone Products, LLC · LYC · Mar 30, 2010 · Dental
Device Facts
| Record ID | K100671 |
| Device Name | NOVABONE DENTAL PUTTY-BIOACTIVE SYNTHETIC BONE GRAFT |
| Applicant | Novabone Products, LLC |
| Product Code | LYC · Dental |
| Decision Date | Mar 30, 2010 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 872.3930 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The intended use of NovaBone Dental Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include: - Periodontal/Infrabony defects - Ridge Augmentation (sinusotomy, osteotomy, cystectomy) - Extraction sites (ridge maintenance/augmentation, implant preparation/ placement) - Sinus lifts - Cystic cavities - Cranio-facial augmentation For larger defects, a mixture of NovaBone Dental Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Device Story
NovaBone Dental Putty is an osteoconductive, synthetic bone void filler composed of calcium-phosphorus-sodium-silicate particulate and a synthetic binder. Supplied as a sterile, pliable, cohesive material in a cartridge, syringe, or tray. Used by dental professionals in oral, dental intraosseous, and craniofacial defect sites. Upon implantation, the binder is absorbed, allowing tissue infiltration between bioactive glass particles. The particles act as a scaffold for new bone growth, eventually being absorbed and replaced by natural bone tissue. The device is used alone or as an extender mixed with autograft or allograft bone. The cartridge packaging format is a product line extension to facilitate easier placement and application.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Calcium-phosphorus-sodium-silicate bioactive glass particulate; synthetic binder; osteoconductive scaffold; supplied sterile; packaged in PET-G tray, syringe, or cartridge; non-structural; biocompatible.
Indications for Use
Indicated for patients requiring bone grafting in oral, dental intraosseous, and craniofacial defects, including periodontal/infrabony defects, ridge augmentation, extraction sites, sinus lifts, and cystic cavities.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- NovaBone Dental Putty – Bioactive Synthetic Graft (K091484)
- NovaBone Dental Putty – Bioactive Synthetic Graft (K063549)
Reference Devices
Related Devices
- K091484 — NOVABONE DENTAL PUTTY - BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Jun 15, 2009
- K063549 — NOVABONE DENTAL PUTTY-BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Feb 12, 2007
- K110368 — NOVABONE PUTTY - BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Apr 14, 2011
- K082672 — NOVABONE PUTTY-BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Dec 9, 2008
- K112428 — NOVABONE DENTAL MORSELS - BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Dec 16, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 3 0 2010
Image /page/0/Picture/1 description: The image shows the logo for NOVABONE, a company that drives innovation in osteobiologics. The logo is in black and white, with the company name in large, bold letters. Below the name is the company's slogan, "DRIVING INNOVATION IN OSTEOBIOLOGICS." The logo is simple and modern, and it conveys the company's focus on innovation and technology.
# Special 510(k): Device Modification Summary NovaBone Dental Putty - Bioactive Synthetic Bone Graft
03/04/10
## 1. Submitter Information:
| Name: | NovaBone Products, LLC |
|------------|----------------------------------------------------------|
| Address: | 13631 Progress Boulevard, Suite 600<br>Alachua, FL 32615 |
| Telephone: | (386) 462-7660 |
| Facsimile: | (386) 418-1636 |
| Contact: | David M. Gaisser |
#### 2. Name of Device:
| Trade Name: | NovaBone Dental Putty - Bioactive Synthetic Bone Graft |
|--------------------|------------------------------------------------------------------------------|
| Common Name: | Osteoconductive Bone Void Filler<br>Synthetic Resorbable Bone Graft Material |
| Regulation Number: | 21 CFR 872.3930 |
| Regulation Name: | Bone Grafting Material, Synthetic |
#### Legally Marketed Predicate Device: 3.
| Predicate #1: | NovaBone Dental Putty – Bioactive Synthetic Graft<br>[K091484] |
|---------------|----------------------------------------------------------------|
| Predicate #2: | NovaBone Dental Putty – Bioactive Synthetic Graft<br>[K063549] |
## Device Description 4.
NovaBone Dental Putty is an osteoconductive, bone void filler device. The device is intended for dental intraosseous, oral, and cranio-/maxillofacial bony defects and is supplied sterile. It is composed of a calcium-phosphorussodium-silicate particulate mixed with a synthetic binder that acts as a temporary binding agent for the particulate. The particulate and binder are provided premixed as a pliable cohesive material. The mixed device is supplied sterile, packaged either in a PET-G tray or in a disposable plastic syringe or cartridge. On implantation, the binder is absorbed to permit tissue infiltration between the Bioglass particles. The particles are slowly absorbed and replaced by new bone tissue during the healing process. The device is intended for dental intraosseous, oral, and cranio-/maxillofacial bony defects. It is supplied sterile.
#### ડ. Intended Use
The intended use of NovaBone Dental Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include:
{1}------------------------------------------------
- · Periodontal/Infrabony defects
- · Ridge Augmentation (sinusotomy, osteotomy, cystectomy)
- · Extraction sites (ridge maintenance/augmentation, implant preparation/ placement)
- Sinus lifts
- · Cystic cavities
- · Cranio-facial augmentation
For larger defects, a mixture of NovaBone Dental Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
## Technological Characteristics 6.
The technological characteristics of the NovaBone Dental Putty device are similar to those of the predicates. The device and the predicates are designed as osteoconductive space-filling devices to be gently packed into defect sites and used as non-structural scaffolds for the body's natural healing and bone regeneration process. The device indications are unchanged from those of the predicate devices. The device is intended to be used alone, or in combination with autogenous or allograft bone.
The NovaBone Dental Putty device of this submission is identical in composition to the NovaBone Dental Putty predicate devices; no changes to the device formulation or function have been made. The primary component of NovaBone Dental Putty is a calcium phospho-silicate particulate, a synthetic material that is both biocompatible and osteoconductive. The NovaBone Dental Putty includes a synthetic binder as an inert carrier for ease of handling and delivery, forming a premixed cohesive material. The binder is biocompatible and is absorbed after implantation, opening space between the bioactive glass particles for cell infiltration and bone formation. The bioactive glass particulate remains for a longer post-implantation period, acting as a scaffold for bone ingrowth. This particulate is absorbed and replaced by new bone tissue.
The filled-cartridge package format as described in this submission is a product line extension, increasing the device package format availability. The new packaging aids in device placement, permitting a different mode of application distinct from that of the previously cleared predicates packaged in tray-in-pouch [K063549] and syringe-in-pouch [K091484] formats. The cartridge packaging system itself is similar to the Class I devices as defined in premarket applications K882551 (cartridge) and K832662/K915147 (cartridge syringe applicator).
#### 7. Conclusion
The NovaBone Dental Putty device modification subject to this submission is to include a filled-cartridge package format to augment the current package configurations. This device modification does not result in a change in
{2}------------------------------------------------
technological characteristics of the device. NovaBone Putty continues to be safe and effective as a non-structural osteoconductive bone void filler for osseous defects.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. David M. Gaisser Vice President NovaBone Products, LLC 13631 Progress Boulevard, Suite 600 Alachua, Florida 32615
MAR 3 0 2010
Re: K100671
Trade/Device Name: NovaBone Dental Putty-Bioactive Synthetic Bone Graft Regulation Number: 21CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: March 4, 2010 Received: March 9, 2010
Dear Mr. Gaisser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Gaisser '
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital. Infection Control and dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
. Enclosure
{5}------------------------------------------------
Confidential
K100671
# STATEMENT OF INDICATIONS FOR US
510(k) Number (if known):
Device Name: NovaBone Dental Putty - Bioactive Synthetic Bone Graft
Indications For Use:
The intended use of NovaBone Dental Putty is to provide a safe, biocompatible synthetic bone graft material for use in oral, dental intraosseous, and craniofacial defects. It is used alone in a manner comparable to autogenous bone graft chips or allograft bone particulate (Demineralized Freeze Dried Bone) or may be mixed with either (typically 1:1 ratio v/v) as a bone graft extender. Typical uses include:
- · Periodontal/Infrabony defects
- · Ridge Augmentation (sinusotomy, osteotomy, cystectomy)
- (ridge maintenance/augmentation, implant · Extraction sites preparation/ placement)
- Sinus lifts
- · Cystic cavities
- · Cranio-facial augmentation
For larger defects, a mixture of NovaBone Dental Putty with an equal volume of allograft or autograft bone and bone marrow may improve new bone formation.
Prescription Use XX
OR (Per 21 CFR 801.109) Over-The-Counter Use_
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSBetz DDS for Dr. K.P. Mulug
Division Sign-Off) Division of Anesthesiology, General Hospital infection Control. Dental Devices
:10(k) Number: K100671