K100631 · Medtronic, Inc. · KFM · Jun 22, 2010 · Cardiovascular
Device Facts
Record ID
K100631
Device Name
AFFIITY CP CENTRIFUGAL BLOOD PUMP
Applicant
Medtronic, Inc.
Product Code
KFM · Cardiovascular
Decision Date
Jun 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AFFINITY™ CP Centrifugal Blood Pump is used to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to 6 hours). It is also indicated for use in extracorporeal support systems (for periods up to 6 hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants). The AFFINITY™ CP Centrifugal Blood Pump is driven by the External Drive Motor or the Emergency Handcrank.
Device Story
Sterile, single-use centrifugal blood pump; moves blood through extracorporeal circuit via centrifugal force. Input: blood entering inlet port; internal cone with impeller blades rotates to accelerate blood toward outlet. Driven by external motor or emergency handcrank via magnetic coupling. Used in OR/clinical settings by perfusionists/surgeons. Output: blood flow through bypass circuit. Benefits: reduced priming volume (40 mL), reduced bearing heat via ceramic pivot system, and lower hemolysis compared to predicate. Device includes pump (AP40), external drive motor (560A), and emergency handcrank (HC150A).
Clinical Evidence
Bench testing only. Validation included maximum pressure generation, flow rates, pivot bearing wear, maximum speed, port strength, magnetic coupling force, biocompatibility, noise generation, and prime volume. Simulated use testing with expert users demonstrated effectiveness. Reduced hemolysis and heat generation compared to predicate.
Technological Characteristics
Centrifugal blood pump; sterile, single-use. Features ceramic pivot system for bearing heat reduction. Magnetic coupling to external motor/handcrank. Priming volume: 40 mL. Non-pyrogenic. No software/algorithm described.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support during cardiopulmonary bypass or other extracorporeal support procedures (e.g., valvuloplasty, mitral valve reoperation, vena cava/aorta surgery, liver transplants) for up to 6 hours. Contraindicated as a cardiotomy suction device.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
K100631
# Summary of Safety and Effectiveness
#### I. Applicant Information:
JUN 2 2 2010
| Date Prepared: | June 21, 2010 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Perfusion Systems |
| Address: | 7611 Northland Drive<br>Minneapolis, MN 55428 |
| Establishment<br>Registration No.: | 2184009 (owner operator number 2112641) |
| Contact Person: | Sue Fidler, Regulatory Affairs Manager<br>Medtronic, Inc.<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112 |
| Telephone Number:<br>Fax Number: | (763) 514-9839<br>(763) 367-8360 |
#### II. Device Information:
.
| Trade Name: | Medtronic AFFINITYTM CP Centrifugal Blood Pump (Model<br>AP40); Medtronic External Drive Motor (Model 560A); Medtronic<br>Emergency Handcrank (Model HC150A) |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | AFFINITYTM Cardiopulmonary Centrifugal Blood Pump; External<br>Drive Motor; Emergency Handcrank |
| Classification Name: | Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type;<br>Cardiovascular bypass pump speed control |
| Classification: | Class III, 21 CFR 870.4360; Class II, 21 CFR 870.4380 |
| Product Code: | KFM; DWA |
| Predicate Devices: | Medtronic Bio-Medicus® Bio-Pump (BPX-80) Centrifugal Pump<br>(K973011); Medtronic External Drive Motor (Model 540T)<br>(K936091); Emergency Handcrank (Model H150) (K926357) |
{2}------------------------------------------------
#### Predicate Device Intended Uses:
The BPX-80 centrifugal blood pump is indicated for use only with the Medtronic Bio-Console pump speed controller to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours).
It is also indicated for use in extracorporeal support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g. valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants).
The External Drive Motor and Emergency Handcrank are intended for use with the Bio-Console System and Bio-Pump Centrifugal Blood Pump. The Bio-Console System is a centrifugal blood pumping system intended for use as an extracorporeal centrifugal blood pumping system during cardiopulmonary bypass procedures.
Device Description: The Medtronic AFFINITY CP Centrifugal Blood Pump is a sterile, single-use centrifugal blood pump. It is a non-invasive, nonpyrogenic device designed to move blood through the extracorporeal circuit by centrifugal force. The pump is the disposable portion of the pumping system and it is electromagnetically coupled to an instrument that monitors and displays the flow and pressure of the blood. Blood enters the inlet port of the pump, where a cone with impeller blades within the pump housing rotates and the blood is gently accelerated towards the outlet of the pump.
> The AFFINITY CP Centrifugal Blood Pump can be driven through magnetic coupling by the following two devices:
- External Drive Motor (Model 560A) a
- . Emergency Handcrank (Model HC150A)
The AFFINITY™ CP Centrifugal Blood Pump is used to pump Intended Use: blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to 6 hours).
> It is also indicated for use in extracorporeal support systems (for periods up to 6 hours) not requiring complete cardiopulmonary bypass (eg, valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants). "
{3}------------------------------------------------
The AFFINITY™ CP Centrifugal Blood Pump is driven by the External Drive Motor or the Emergency Handcrank.
Contraindications: The AFFINITY CP Centrifugal Blood Pump is contraindicated as a cardiotomy suction device. Comparison to Predicate Device: The AFFINITY CP Centrifugal Blood Pump is substantially equivalent to the predicate device, in that its intended use is the same, and the basic design is the same, with a magnetic coupling to an external motor that drives the rotating cones. Both pump drives are controlled with Medtronic controllers. The new pump includes the following improvements: The bearing heat generation is reduced due to a ceramic pivot system. -Priming volume is reduced by 50%, to a volume of 40 mL. Hemolysis is reduced compared to the predicate device. -The External Drive Motor and Emergency Handcrank are substantially equivalent to the predicate devices, in that their intended uses are the same and the basic design is the same. Necessary design changes were made to allow for compatibility with the AFFINITY CP Centrifugal Blood Pump: dimensional changes and rotational direction. Test Data: Verification and validation testing has demonstrated that the AFFINITY CP Centrifugal Blood Pump and its associated drive systems are safe and effective. The blood pump is more efficient, as evidenced by less heat generation, and less damaging to red blood cells, as evidenced by reduced hemolysis. Other validation tests included: maximum pressure generation, flow rates, pivot bearing wear, maximum speed, port strength, magnetic coupling force, biocompatibility, noise generation and prime volume. Validation testing under simulated conditions of use with expert users also demonstrated the effectiveness of the AFFINITY CP Centrifugal Blood Pump. Based on the accumulated technical information, intended use, Summary: laboratory verification tests and performance data provided, the AFFINITY CP Centrifugal Blood Pump and its associated drive systems are substantially equivalent to the currently marketed predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
## JUN 2 2 2010
Medtronic, Inc. c/o Ms. Susan C. Fidler 8200 Coral Sea Street NE Mounds View, MN 55112
Re: K100631
Trade/Device Name: Medtronic AFFINITY™ CP Centrifugal Blood Pump (Model AP40); Medtronic External Drive Motor (Model 560A); Medtronic Emergency Handcrank (Model HC150A)
Regulation Number: 21 CFR 870.4360
Regulation Name: Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type Regulatory Class: III Product Code: KFM, DWA Dated: May 24, 2010 Received: May 25, 2010 /_
Dear Ms. Fidler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of . devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Susan C. Fidler
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Statement of Indications for Use
510(k) Number: K100631
Medtronic AFFINITY™ CP Centrifugal Blood Pump (Model AP40); Medtronic External Drive Motor (Model 560A); Medtronic Emergency Handcrank (Model HC150A)
Indications for use:
The AFFINITY™ CP Centrifugal Blood Pump is used to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to 6 hours).
It is also indicated for use in extracorporeal support systems (for periods up to 6 hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants).
The AFFINITY™ CP Centrifugal Blood Pump is driven by the External Drive Motor or the Emergency Handcrank.
Prescription Use (Part 21 CFR 801 Subpart D) OR
Over-The-Counter-Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
rrence of operative of Device
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K100631
6/22/2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.