K100624 · Invendo Medical Gbmh · FDF · Feb 23, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K100624
Device Name
INVENDO C20 COLONOSCOPY SYSTEM
Applicant
Invendo Medical Gbmh
Product Code
FDF · Gastroenterology, Urology
Decision Date
Feb 23, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The invendo C20 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopy surgery. The colonoscope component of the invendo C20 Colonoscopy System, the SC20 Colonoscope, is a single use disposable device. The SC20 Colonoscope cannot be reprocessed.
Device Story
System provides visualization and access to adult lower GI tract; consists of single-use disposable SC20 colonoscope and base system. Unlike manual colonoscopes, SC20 uses motorized drive unit and lubricated inverted sleeve technology for advancement/retraction; pressure-controlled fluid-filled bellows for tip deflection. Distal tip contains CMOS camera and LEDs. Operated by physician via handheld controller connected to base unit. Base unit manages insufflation, irrigation, suction, and drive functions. Output displayed for physician to guide navigation, perform biopsies, or polypectomies. Benefits include automated navigation and disposable design eliminating reprocessing needs.
Clinical Evidence
Prospective single-arm feasibility study (n=61) in asymptomatic adults (50-75 years). Primary endpoint: cecal intubation rate. Results: 98.4% (60/61) cecal intubation success (95% CI: 91.2-99.9%). 39 lesions detected in 26 subjects; 32 polyps removed. No device-related adverse events reported. Bench testing confirmed forces exerted during navigation were below perforation thresholds.
Technological Characteristics
Materials: Silicone and polycarbonate (biocompatible per ISO 10993). Sensing/Actuation: CMOS camera, LED illumination, electrohydraulic bellows for deflection. Energy: Electric drive unit, 100-240V power supply. Dimensions: 18mm diameter, 2100mm working length. Connectivity: Handheld controller to base unit. Sterilization: Single-use disposable colonoscope. Software: Base unit controls drive, suction, irrigation, and insufflation.
Indications for Use
Indicated for adult patients requiring visualization and diagnostic/therapeutic access to the lower gastrointestinal tract. Contraindicated in patients with fulminant colitis, acute diverticulitis, severe cardiopulmonary illness, or lack of patient cooperation.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
K100624
Pages 1 of 7
FEB 2 3 2012
# 510(k) Summary invendo medical C20 Colonoscopy System
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
| Owner: | invendo medical GmbH |
|----------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Peterhofstr. 3b<br>86438 Kissing<br>Germany |
| Contact: | Oliver v. Ruepprecht<br>Manager QA/RA<br>Phone : +49 - 8233 - 744 98 16<br>Fax: +49 - 8233 - 744 98 15<br>Mail: oliver.ruepprecht@invendo-medical.de |
{1}------------------------------------------------
## Device Name, 21 CFR807.92(a){2}.
Trade Name: invendo C20 Colonoscopy System; includes SC20 Colonoscope and the C20 Base System (base unit, drive, and handheld) Common Name: Colonoscope Classification Name: Endoscope and accessories, 21CFR 876.1500, product code: FDF
# Predicate Devices, 21 CFR 807.92(a)(3):
| Predicate Device | 510(k) No. |
|-------------------------------------------------------------|------------|
| Olympus CF-Q160AL/I | K001241 |
| Fujinon EN 450P5/20 Double Balloon | K040048 |
| ColonoSight 510B | K032688 |
| NeoGuideEndoscopy System | K052930; |
| | K070622 |
| Avantis 3rd Eye Retroscope | K070330 |
| Softscope CAD 12 | K092156 |
| Medela Dominat 35c/i suction pump | K043544 |
| Medela Basic 30 | K021368 |
| PROTOCO2LTM Colon Insufflator with Performance Improvements | K030854 |
| EndoGatorEndoscopy Irrigation Pump | K060962 |
| Olympus Exera 160A videoprocessor | K051645 |
{2}------------------------------------------------
#### Device Description, 21 CFR 807.92(a)(4):
The invendo C20 Colonoscopy System has many features in common with currently marked colonoscopes. The invendo C20 Colonoscopy System consists of two separate components: a single use disposable colonoscope and a base system that supplies and controls the colonoscope according to user commands via a handheld control unit. The distal tip of the colonoscope is deflectable and equipped with a CMOS camera and LEDs for illumination. A working channel is incorporated for biopsies and polypectomies. The invendo C20 Colonoscopy System is equipped with insufflation, irrigation and suction functions. Deflection, rinsing, insufflation and suction are activated and controlled by the base unit, according to user commands.
Unlike other colonoscopes, the invendo C20 Colonoscopy System is not manually pushed or otherwise manually manipulated by the operator, but instead, moves under the operator's direction through use of a handheld controller that controls, among other things, the device's drive unit. The colonoscope uses a lubricated inverted sleeve technology for advancement and retraction. Unlike other colonoscopes, the invendo C20 Colonoscopy System uses pressure controlled bellows for the deflecting function. The endoscope deflects in any direction depending on the pressure in the bellows. The handheld control is connected to the base unit. Depending on the user commands, the base unit activates and controls the functions of the invendo SC20 Colonoscope.
The colonoscope component of the system is a single-use disposable device. The SC20 colonoscope cannot be reprocessed. The main materials with direct contact with the patient are silicone and polycarbonate. The whole colonoscope, including the lubricant, were demonstrated to be biocompatible per ISO 10993.
| Viewing direction | Forward |
|---------------------------------------|--------------------------------|
| Field ofview | 114° |
| Chip | CMOS, 250.000 Pixels |
| Max. Diameter Device | 18mm |
| Working length | 2100mm |
| Total length | 3600mm |
| Bending capacity | 180° in each direction at 37°C |
| Diameter working channel | 3.1mm |
| Max. speed colonoscope via drive unit | 500mm/min |
| Power supply baseunit | 100 - 240V 50/60Hz |
| Max. internal pressure bas eunit | 600kPa |
| Max. internal vacuum base unit | -75kPa |
#### Specification Chart:
{3}------------------------------------------------
#### Intended Use / Indications for Use, 21 CFR 807.92(a){5):
The invendo C20 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopy surgery.
The colonoscope component of the invendo C20 Colonoscopy System, the SC20 Colonoscope, is a single use disposable device. The SC20 Colonoscope cannot be reprocessed.
The device is contraindicated in patients with fulminant colitis; acute diverticulitis; severe cardiopulmonary illness; or where there is a lack of cooperation on the part of the patient.
## Predicate Comparison, 21 CFR 807.92(a)(6):
The invendo C20 Colonoscopy System has the same intended use and indications for use as the predicate devices. The C20 Colonoscopy System has similar technological characteristics, however, the principles of operation with respect to advancing and retracting the endoscope in the colon are different from the predicate devices. The invendo SC20 Colonoscope is not manually pushed or otherwise manually manipulated by the operator, but instead, moves under the operator's control and direction with the help of the drive unit and user selected commands. The colonoscope uses an inverted sleeve technology for advancing and retracting.
Like the predicate devices, the invendo SC20 Colonoscope is a long, flexible endoscope that features a working channel, deflectable tip, and optics at the tip of the device. The working length of the predicate devices is somewhat less since those devices adjust the length and conformation of the colon to the device, this minor difference does not raise new questions of safety and effectiveness.
The invendo SC20 Colonoscope uses an overtube mechanism for the transmission of advanced power like the SoftScope CAD12. In both cases the sleeve systems are disposable.
The deflecting tip of the invendo C20 Colonoscopy System is electrohydraulically controlled via fluid-filled bellows while the deflecting tips of the predicate devices are controlled by Bowden wires. The performance bench and clinical testing demonstrate that these differences do not raise new questions of safety or effectiveness as compared
{4}------------------------------------------------
to the cleared predicate devices and that the device is as safe and effective as its predicates.
The invendo C20 Colonoscopy System is capable of performing rinsing, suction and insufflation like the predicate devices.
The invendo C20 Colonoscopy System uses the same basic technology for visualization and illumination as the predicate devices.
The materials used in the invendo C20 Colonoscopy System are somewhat different from those of the predicate devices, the company has provided complete biocompatibility data demonstrating that the materials used are biocompatible and thus as safe as those used in the predicate devices.
Performance data demonstrate that the device is as safe and effective as the predicate devices. Thus, the invendo C20 Colonoscopy System is substantially equivalent.
Clinical data confirms that the device can successfully navigate the colon and perform polyp/lesion removals.
#### Nonclinical performance data, 21 CFR 807.92(b)(1):
To assess the performance of the invendo C20 Colonoscopy System, invendo conducted bench tests to measure forces encountered during device use, product stability, biocompatibility of device components (according ISO Standard ISO-10993 and FDA Guidance G959-1), and constructive safety (IEC 60601-2-18). In all instances, the SC20 Colonoscope and the invendo C20 base system functioned as intended and met individual test specifications.
Forces: The forces exerted by the colonoscope while moving through small curves with and without deflection were measured. The maximum forces exerted by the deflection, and advancement of the colonoscope were less than the force necessary to cause a perforation as determined by Thomas, K. Wu; Gastrointest. Endosc. 1978 Volume 24.
Biocompatibility: The tested materials met the biocompatibility requirements as per ISO 10993.
Product stability: The invendo SC20 Colonoscope has been demonstrated to be stable over the labeled 6 month shelf life of the product.
Constructive safety: The system met the general requirements of safety and the particular safety requirements of endoscopic equipment per IEC 60601.
{5}------------------------------------------------
The results of the tests demonstrate that the device is as safe and effective as the predicate devices.
#### Clinical Study Summary, 21 CFR 807.92(b)(2):
Groth et al., "High Cecal Intubation Rates with a New Computer-Assisted Colonoscope: A Feasibility Study," The American Journal of Gastroenterology, vol. 106, pp. 1075-1080, June 2011.
A study was conducted in Germany to measure the rate of successful cecal intubation for the assessment of the safety and effectiveness of the invendo C20 Colonoscopy System. The study was a prospective single arm study of 61 asymptomatic subjects (between 50 and 75 years of age) that were willing to undergo screening colonoscopy and were at average risk for colorectal cancer. Exclusion criteria included (but were not limited to) subjects with:
- history of colorectal neoplasia including familial adenomatous polyposis (FAP) or o hereditary non-polyposis colorectal cancer;
- suspected diagnosis of inflammatory bowel disease, bowel obstruction, acute o diverticulitis, known severe diverticulosis or any known large bowel disease
- GI-related symptoms, complaints or diseases; o
- previous abdominal surgery except for uncomplicated cholecystectomy, o appendectomy or minor pelvic surgery (e.g., hernia repair, oophorectomy).
Investigators underwent "pre-study" training where 2-4 training cases were needed to gain proficiency. The investigators were experienced endoscopists that insufflated the colon with either CO2 or water. The subjects were examined in the left lateral position, with shifting to the supine position in most cases.
Cecal intubation was achieved in 98.4% (60/61) of the subjects (95% Cl: 91.2 - 99.9%). Two device malfunctions and no device-related adverse events were reported. Thirtynine lesions in 26 subjects (42.6%) were detected. Of those lesions, 36 were polyps in 23 subjects (37.7%), ranging in size from 2mm to 18mm (average 4.8mm). The remaining 3 lesions consisted of one abnormally large ileocecal valve which underwent biopsy, one varicose knot (no intervention) and one lipoma (no intervention). Thirty-two polyps (88.9%) were removed by either forceps (22) or snare (10). There were two cases where a second conventional colonoscope was used for the removal of a 12mm flat lesion and for the treatment of post-polypectomy bleeding.
Therapeutic maneuvers other than biopsy and polypectomy were not evaluated because cecal intubation was the primary endpoint. Although not required by the
{6}------------------------------------------------
protocol, investigators performed retroflexions as needed to ensure thoroughness of examination; retroflexion did not pose a problem and was well tolerated.
This study demonstrates that the invendo C20 Colonoscopy System can successfully provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopy surgery.
#### Summary, 21 CFR 807.92(b)(3)
The laboratory studies, which include determinations of exerted forces, product stability, biocompatibility and constructive safety, demonstrate that the invendo C20 Colonoscopy System performs as well as the predicate devices. The clinical data demonstrate that the invendo C20 Colonoscopy System can successfully provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum and ileocecal valve) for endoscopy and endoscopy surgery.
{7}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 2 3 2012
Mr. Oliver V. Ruepprecht Manager OA/RA Invendo Medical GmbH Peterhofstr. 3b KISSING 86438 GERMANY
Re: K100624
Trade/Device Name: Invendo C20 Colonoscopy System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: February 15, 2012 Received: February 15, 2012
Dear Mr. Ruepprecht:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Occinetie fore, market the device, subject to the general controls provisions of the Act. The I ou mayy insteresy size of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your de subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oe as note a determination that your device complies with other requirements of the Act that I Drinas Intact and regulations administered by other Federal agencies. You must or uny with all the Act's requirements, including, but not limited to: registration and listing confirs will and the North of I CFR Part 801); medical device reporting (reporting of medical
{8}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 Jou action ips . fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin K. Twitchell
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use Statement
Indications for Use Statement
510(k) Number (if known): K100624
Device Name: invendo C20 Colonoscopy System
Indications for Use:
The invendo C20 Colonoscopy System is intended to provide visualization and diagnostic / therapeutic access to the adult lower gastrointestinal tract (including but not limited to, the anus, rectum, sigmoid colon, colon, cecum an ileocecal valve) for endoscopy and endoscopic surgery.
The colonoscope component of the invendo C20 Colonoscopy System, the rne colonoscope, is a single use disposable device. The SC20 colonoscope cannot be reprocessed.
Prescription Use × (Part 21 CFR 801 Subpart D) AND Over-The-Counter Use (21 CFR 801 Subpart C) OR
121
00018
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Beniamin K. Kihara
(Division Sign-Off)
ylston Sign-Off sion of Reproductive, Gastro-Renal, and I Device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.