← Product Code [NKB](/productcode/NKB) · K100623

# AESCULAP IMPLANT SYSTEMS S4 SPINAL SYSTEM (K100623)

_Aesculap Implant Systems · NKB · Jun 9, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100623

## Device Facts

- **Applicant:** Aesculap Implant Systems
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jun 9, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities: 1 ) degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) 2) spondylolisthesis, 3) trauma (i.e., fracture or dislocation), 4) spinal stenosis, 5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), 6) tumor. 7) pseudoarthrosis, and 8) failed previous fusion.

## Device Story

S4 Spinal System comprises polyaxial/monoaxial screws, hooks, rods, and rod-to-rod connectors; all components top-loading/top-tightening. Used by surgeons in clinical settings for spinal stabilization/fixation. Provides mechanical support to spine during fusion process; aids in correcting deformities or stabilizing instabilities. Benefits patient by restoring spinal alignment and facilitating bone fusion.

## Clinical Evidence

Bench testing only. Static and dynamic compression bending testing performed in accordance with ASTM F1717. Results demonstrated performance similar to predicate devices.

## Technological Characteristics

Materials: Titanium and Titanium alloy (ISO 5832/3, ISO 5832/2). Components: Polyaxial/monoaxial screws, hooks, rods, rod-to-rod connectors. Design: Top-loading, top-tightening. Mechanical fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- S4 Spinal System ([K090657](/device/K090657.md))
- S4 Spinal System ([K071945](/device/K071945.md))
- S4 Spinal System ([K062085](/device/K062085.md))
- S4 Spinal System ([K032219](/device/K032219.md))
- Blackstone Pedicle Screw System ([K082797](/device/K082797.md))
- Stryker XIA 3 Spinal System ([K083393](/device/K083393.md))
- Depuy Expedium Spine System ([K090230](/device/K090230.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) SUMMARY (as required by 21 CFR 807.92) B.

Aesculap® Implant Systems S4 Spinal System March 1, 2010

| COMPANY:           | Aesculap®Implant Systems, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311                                                                                                      |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT:           | Lisa M. Boyle<br>610-984-9274 (phone)<br>610-791-6882 (fax)                                                                                                                                                                               |
| TRADE NAME:        | S4                                                                                                                                                                                                                                        |
| COMMON NAME:       | S4 Spinal System                                                                                                                                                                                                                          |
| REGULATION NUMBER: | 888.3070 - Orthosis, Spinal Pedicle Fixation For<br>Degenerative Disc Disease<br>888.3070 - Orthosis, Spinal Pedicle Fixation<br>888.3070 - Orthosis, Spondyloisthesis Spinal Fixation<br>888.3050 - Appliance, Fixation, Spinal Fixation |
| PRODUCT CODE:      | NKB, MNI, MNH, and KWP                                                                                                                                                                                                                    |

#### Orthopedics REVIEW PANEL:

# PURPOSE FOR PREMARKET NOTIFICATION

The S4 Spinal System described in this submission represents an expanded indication for the entire S4 system and to add components to the S4 Spinal System.

## SUBSTANTIAL EQUIVALENCE

Aesculab® Implant Systems, Inc. believes that the S4 Spinal System is substantially equivalent to the S4 Spinal System (K090657/K071945/K062085/K032219), Blackstone Pedicle Screw System (K082797), Stryker XIA 3 Spinal System (K083393), and the Depuy Expedium Spine System (K090230).

### DEVICE DESCRIPTION

The S4 Spinal System consists of polyaxial screws and monoaxial screws of varying diameters and lengths, various hook styles, rods of varying lengths, and fixed and adjustable rod to rod connectors. All implant components are top loading and top tightening. The S4 Spinal System is manufactured from Titanium and Titanium alloy in accordance with ISO 5832/3 and ISO 5832/2.

{1}------------------------------------------------

KI00623 Page 2 of 2

## INDICATIONS FOR USE

The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities:

1 ) degenerative disc disease (defined as discogenic back pain with

degeneration of the disc confirmed by history and radiographic studies) 2) spondvlolisthesis,

3) trauma (i.e., fracture or dislocation),

4) spinal stenosis,

5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),

6) tumor.

7) pseudoarthrosis, and

8) failed previous fusion.

# TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)}

The components of the Aesculap® Implant Systems S4 Spinal System are offered in similar shapes and sizes as the predicate devices. All the components are manufactured from Titanium and Titanium Alloy, which is the same material as the predicate devices.

## PERFORMANCE DATA

Static and dynamic compression bending testing in accordance with ASTMF1717 was conducted on the subject device, and the results were found to be similar to other legally marketed predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN - 9 2010

Aesculap Implant Systems, Inc. % Ms. Lisa Boyle Senior Regulatory Affairs Associate 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K100623

Trade Name: S4 Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: April 28, 2010 Received: April 29, 2010

Dear Ms. Boyle,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Lisa Boyle

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Qurbara Muehm

Mark N. Melker Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

### INDICATIONS FOR USE STATEMENT A.

510(k) Number:

## Device Name: Aesculap® Implant Systems S4 Spinal System

### Indications for Use:

The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities:

1 ) degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) 2) spondylolisthesis,

3) trauma (i.e., fracture or dislocation),

4) spinal stenosis,

5) deformities or curvatures (i.e., scoliosis, and/or lordosis),

6) tumor.

7) pseudoarthrosis, and

8) failed previous fusion

X ______ and/or Over-the-Counter Use __ Prescription Use

(per 21 CFR 801.109

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K100623 510(k) Number_

002

---

**Source:** [https://fda.innolitics.com/device/K100623](https://fda.innolitics.com/device/K100623)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
