SYMBIA.NET, SYMBIA.NET SELECT MODEL 10520828, 10520827
K100619 · Siemens Medical Solutions USA, Inc. · LLZ · May 7, 2010 · Radiology
Device Facts
Record ID
K100619
Device Name
SYMBIA.NET, SYMBIA.NET SELECT MODEL 10520828, 10520827
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
May 7, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. Software: The MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies taken from SPECT, PET, CT and other imaging modalities.
Device Story
Symbia.net is a client/server software solution for SPECT and SPECT-CT systems; enables deployment of MI Applications on standard hardware (desktops, laptops, workstations). Server-side software processes imaging data; client-side software displays rendered images. Clinicians access system via PCs or Macs. System ensures clinical equivalence between server and client image rendering; no degradation in resolution, color, or timing. Used by physicians and healthcare professionals in clinical settings to aid in assessment, quantification, and diagnosis of pathologies; supports radiotherapy planning and interventional procedures. Benefits include flexible remote access to diagnostic imaging tools while maintaining diagnostic quality.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through technical comparison and verification of image rendering consistency between server and client platforms.
Technological Characteristics
Client/server software architecture; compatible with standard desktop/laptop/workstation hardware. Supports SPECT, PET, CT, and other imaging modalities. Software-based display and analysis package. Risk management compliant with ISO 14971.
Indications for Use
Indicated for use by trained healthcare professionals to aid in detection, localization, diagnosis, staging, and restaging of lesions, tumors, disease, and organ function. Applicable to cardiovascular disease, neurological disorders, and cancer. Used for radiotherapy treatment planning and interventional radiology. Software aids in assessment and quantification of pathologies from SPECT, PET, CT, and other imaging modalities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Klooiq
Siemens Symbia.net
Special 510(k) Premarket Notification
Strictly Confidential
# 510(k) Summary
as required by 21 CFR Part 807.87(h)
MAY - 7 2010
| Submitter: | Elaine Chang<br>Regulatory Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number: | (865) 898-4170 |
| Fax Number: | (865) 218-3019 |
| Name / Address of<br>Manufacturer | Siemens Medical Solutions USA, Inc<br>Molecular Imaging<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192<br>USA |
| Date of Submission: | March 3, 2010 |
| Identification of the product | |
| Device Proprietary Name: | Symbia.net |
| Common Name: | Picture Archiving and Communication System |
| Classification Name: | Picture Archiving and Communication System per 21<br>CFR 892.2050<br>Emission Computed Tomography System per 21 CFR<br>892.1200 |
| Product Code: | LLZ and KPS. |
| Classification Panel: | Radiology |
| Device Class: | Class II |
{2}------------------------------------------------
### Marketed Devices to which Equivalence is claimed
| Device | Manufacturer | 510(k) Number |
|---------------------------------------|-------------------------------|---------------|
| Symbia 4.0 (MI<br>Applications 2009A) | Siemens Medical Solutions | K082506 |
| Visage PACSICS | Mercury Computer Systems, Inc | K072205 |
#### Device Description:
Symbia.net introduces client/server functionality allowing the MI Applications product to be deployed on any compatible hardware (includinq desktops, laptops and workstations) that meet minimal hardware reguirements. Symbia.net is a solution for SPECT and SPECT-CT systems.
The software application is provided on a server. Clients running on any personal computer or Mac, meeting minimum requirements, can have access to the system. The rendered images on the client and server side are clinically equivalent with each other. No imaqe quality degradation on any of the imaqes shown in the clients compared to the images shown on the server related to resolution, color and timing.
#### Safety and Effectiveness:
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitiqation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens considers that the proposed device does not introduce new safety concerns, and is substantially the same in indications for use. design, materials, energy sources and technology as the predicate devices. Siemens believes that Symbia.net is substantially equivalent to the predicate devices.
#### Indications for Use:
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Software: The MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies taken from SPECT, PET, CT and other imaging modalities.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
MAY - 7 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Elaine Chang Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. Molecular Imaging Group 2501 North Barrington Road HOFFMAN ESTATES IL 60192 .
Re: K100619
Trade/Device Name: Symbia.net Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 7, 2010 Received: April 8, 2010
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895: In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Donald J. Bohn
Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K1006 19
Device Name: Symbia.net
Indications for Use:
The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Software: The MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies taken from SPECT, PET, CT and other imaging modalities.
Prescription Use × (Part 21 CFR 801 Subpart D) OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of ORH, Office of Device Evaluation (ODE)
Booker
Rbott
K 100619
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.