← Product Code [HRS](/productcode/HRS) · K100604

# OTIS -C PLUS (K100604)

_Sciences ET Bio Materiaux (Sbm) · HRS · Jun 14, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100604

## Device Facts

- **Applicant:** Sciences ET Bio Materiaux (Sbm)
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Jun 14, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

OTIS-C Plus is intended to be used in conjunction with bone screws to provide fixation following Proximal Tibial opening wedge osteotomies.

## Device Story

OTIS-C Plus is a stainless steel bone fixation plate and screw system designed for stabilization during high tibial osteotomy (HTO) via a medial approach. The system features an anatomically shaped, thin, short plate with four holes and self-tapping screws (6.5 mm diameter). The locking mechanism allows for graft compression and stable fixation, facilitating early weight-bearing. Used by orthopedic surgeons in a clinical/surgical setting, the device enables mini-invasive surgery. The surgeon selects mono- or bi-cortical fixation based on screw length. The system provides mechanical stability to the osteotomy site, supporting bone healing and patient recovery.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including fretting corrosion, torsional yield strength, ultimate torque, insertion/removal torque, and pull-out strength.

## Technological Characteristics

Stainless steel bone fixation plate and screw system. Plate features 4 holes; 3 mm thickness. Screws are 6.5 mm diameter, self-tapping. Mechanical fixation principle. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- NUMELOCK II SYSTEM ([K041709](/device/K041709.md))
- ARTHREX PUDDU OSTEOTOMY SYSTEM ([K973812](/device/K973812.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K100604#1/6

# 510(k) OTIS-C Plus

# SB

# 510(k) Summary

As required by section 807.92

JUN 1 4 2010

## 1. GENERAL INFORMATION

| Type of 510(k)                        | TRADITIONAL                                                                                                                                                                                                          |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                            | OTIS-C Plus                                                                                                                                                                                                          |
| CFR section                           | 21CFR 888.3030                                                                                                                                                                                                       |
| Classification Name                   | Single/multiple component metallic bone fixation appliances<br>and accessories                                                                                                                                       |
| Device panel                          | ORTHOPEDIC                                                                                                                                                                                                           |
| Product Code                          | HRS: PLATE, FIXATION, BONE<br>HWC : SCREW, FIXATION, BONE                                                                                                                                                            |
| Class                                 | II                                                                                                                                                                                                                   |
| Legally marketed<br>predicate devices | K041709 NUMELOCK II SYSTEM manufactured by<br>HOWMEDICA OSTEONICS CORP (now STRYKER<br>ORTHOPAEDICS) and K973812 ARTHREX PUDDU<br>OSTEOTOMY SYSTEM manufactured by ARTHREX, INC.                                     |
| Submitter                             | SCIENCE FOR BIOMATERIALS<br>Sciences et Bio Matériaux<br>ZI du Monge<br>F 65100 LOURDES - FRANCE<br>Registration Number : 3004549189                                                                                 |
| Contact                               | Denis CLEMENT, CEO<br>Tel : +33 (0)5 62 42 21 01<br>Fax : +33 (0)5 62 42 21 00<br>e-mail : denis.clement@sbm-fr.com<br>Regulatory contact: Idée Consulting (FRANCE)<br>Isabelle DRUBAIX e-mail : idrubaix@nordnet.fr |

.

{1}------------------------------------------------

510(k) OTIS-C Plus

### 2. DEVICE DESCRIPTION

The OTIS-C Plus plate is designed for the stabilization of High Tibial Osteotomy with a medial approach. Anatomically shaped, thin and short, the OTTS-C Plus plate enables mini-invasive surgery. Its locking system provides immediate compression of the graft as well as stable fixation, thus allowing early weight-bearing. The selftapping OTIS-C Plus screws allows easy and reliable one step locking, without counternut, in a simple and concise approach. With its range of twelve screws lengths, fixation can be either mono or bi-cortical, upon the choice of the surgeon.

#### 3. INDICATIONS FOR USE

OTIS-C Plus is intended to be used in conjunction with bone screws to provide fixation following Proximal Tibial opening wedge osteotomies.

#### 4. SUBSTANTIAL EQUIVALENCE

Both OTIS-C Plus system and its predicates are plate and screw systems made of stainless steel. Plates included in all these systems present 4 holes. Both OTIS-C Plus system and its predicate NUMELOCK II SYSTEM include screw 6.5 mm in diameter and plate 3 mm thick. Both plates and screws are supplied in a similar range of length.

Non clinical performance testing including fretting corrosion, and determination of torsional yield strength, ultimate, insertion and removal torque and Pull-out strength demonstrate that OTIS-C Plus system is as safe, as effective, and performs at least as safely and effectively as its predicate devices.

Summary preparation date: revised May 31, 2010

{2}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Science Et Bio Materiaux (SMB) % Mr. Denis Clement Zi du Monge Lourdes France 65100

JUN 1 4 2010

Re: K100604 Trade/Device Name: OTIS-C Plus Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS, HWC Dated: June 01, 2010 Received: June 03, 2010

Dear Mr. Clement:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

### Page 2 - Mr. Denis Clement

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark N Milliken

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) OTIS-C Plus

# INDICATIONS FOR USE

510(k) Number (if known): K100604 Device Name: OTIS-C Plus Indications for Use:

OTIS-C Plus is intended to be used in conjunction with bone screws to provide fixation following Proximal Tibial Opening wedge osteotomies.

Prescription Use |▲ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Smith for nxa

and Restorative Devic

510(k) Number K100604

Page 6 / 25

---

**Source:** [https://fda.innolitics.com/device/K100604](https://fda.innolitics.com/device/K100604)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
