EASY TALK BLOOD GLUCOSE MONITORING SYSTEM MODEL: 6277

K100560 · Biotest Medical Corp. · NBW · May 12, 2010 · Clinical Chemistry

Device Facts

Record IDK100560
Device NameEASY TALK BLOOD GLUCOSE MONITORING SYSTEM MODEL: 6277
ApplicantBiotest Medical Corp.
Product CodeNBW · Clinical Chemistry
Decision DateMay 12, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The SuperCheck 1 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, and forearm. It is intended for use by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

Easy Talk Blood Glucose Monitoring System measures glucose in capillary whole blood samples via test strips. System includes meter, test strips, control solutions, lancing device, and lancets. Meter features bilingual speaking function for audible results, aiding users with low vision. Used by healthcare professionals or patients at home for monitoring diabetes control programs. Forearm testing permitted only during steady-state glucose conditions. Device provides quantitative glucose concentration output. Modifications from predicate include added backlight, changed error message (E7 to E4), and updated physical form factor.

Clinical Evidence

Clinical performance study conducted to validate usability and accuracy for consumer use. Results demonstrated substantial equivalence to predicate device.

Technological Characteristics

Glucose oxidase-based electrochemical test system. Includes meter with backlight and bilingual audio output for low vision users. Capillary blood sampling from finger or forearm. Standalone device.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from finger, palm, and forearm; intended for home use by individuals with diabetes to monitor glycemic control; not for diabetes diagnosis/screening; not for neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kivosto O # MAY 1 2 2010 # 510(k) Summary (Per 21 CFR 807.92) Updated on April 6, 2010 #### 1. Submitter Information | Company name | Biotest Medical Corporation | |----------------|------------------------------------------------------------------------------------| | Contact person | Maggie Chu, President | | Address | No. 3-2, Chien-Kuo Road, TEPZ, Tantzu<br>Taichung 427, Taiwan<br>Republic of China | | Phone | 886-919099431 | | Fax | 886-425326593 | #### 2. Name of Device | Trade Name | Easy Talk Blood Glucose Monitoring System | |-----------------|---------------------------------------------------------------------------------------------------------| | Common Name | Blood Glucose Test System | | Classifications | NBW, Over the Counter Blood Glucose Test, 862.1345<br>CGA, Glucose Oxidase, 862.1345<br>Class II device | #### 3. Predicate Device | Trade name | SuperCheck 1 Blood Glucose Monitoring System (SuperCheck 1)<br>Model 6268 | |---------------|---------------------------------------------------------------------------| | Common name | Blood Glucose Test System | | Submitter | Biotest Medical Corporation | | 510(k) number | K091815 | #### 4. Device Description The Easy Talk Blood Glucose Monitoring System, consists of a blood glucose meter, test strips, control solutions, lancing device, and commercially available lancets. The meter has a bilingual speaking feature that provides audible test results for users with low vision. The forearm may be used as an alternate site for capillary blood. #### 5. Intended Use The Easy Talk Blood Glucose Monitoring System, is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the forearm. It is intended for use by healthcare professionals and people with diabetes mellitus at home and as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis or screening for diabetes mellitus, nor for use with neonates. {1}------------------------------------------------ The alternative site testing (forearm) in this system can only be used during steady-state blood glucose conditions. This system contains a speaking function that provides audible test results for users with low vision. #### 6. Comparison to Predicate Device Modifications to the cleared device include additional backlight feature and change of one error message from E 7 to E 4. The size and shape of the device have also been modified to differentiate from the predicate. The Easy Talk Blood Glucose Monitoring System has the same indented use and fundamental scientific technology as the previous version, the SuperCheck 1 Blood Glucose Monitoring System, Model 6268. #### 7. Performance Studies #### Non-clinical Tests Biotest Medical Corp. has performed the extensive verification, validation and testing activities to establish the performance, functionality and reliability characteristics of the device. The device passed all of the tests based on pre-determined Pass/Fail criteria: # Clinical Tests The clinical performance study was conducted for purpose of validating consumer's use of the Easy Talk Blood Glucose Monitoring System in term of usability and accuracy? The test results showed substantial equivalence. #### 8. Conclusion The conclusion drawn from the non-clinical and clinical tests is that the Easy Talk Blood Glucose Monitoring System is as safe, as effective, and performs as well as the legally marketed predicate device, SuperCheck 1 Blood Glucose Monitoring System, model 6268 (k091815). Therefore, the subject device is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 MAY 1.2 2010 Biotest Medical Corporation c/o Maggie Chu President No. 3-2, Chien-Kuo Road, TEPZ, Tantzu Taichung, China (Taiwan) 427 Re: k100560 Trade/Device Name: Easy Talk Blood Glucose Monitoring System, Model: 6277 Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA Dated: April 12, 2010 Received: April 14, 2010 Dear Ms. Chu We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ ## Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Signature Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): 45100560 Device Name: Easy Talk Blood Glucose Monitoring System Indications for Use: The Easy Talk Blood Glucose Monitoring System, is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the forearm. It is intended for use by healthcare professionals and people with diabetes mellitus at home and as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis or screening for diabetes mellitus, nor for use with neonates. The alternative site testing (forearm) in this system can only be used during steady-state blood glucose conditions. This system contains a speaking function that provides audible test results for users with low vision. Prescription Use _____________________________________________________________________________________________________________________________________________________________ × · AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_14100560 ============================================================================================================================================================== Page 1 of _1_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...