RAYSTATION VERSION 1.0

K100552 · Raysearch Laboratories AB · MUJ · Mar 12, 2010 · Radiology

Device Facts

Record IDK100552
Device NameRAYSTATION VERSION 1.0
ApplicantRaysearch Laboratories AB
Product CodeMUJ · Radiology
Decision DateMar 12, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Device Story

RayStation is a software-based radiation treatment planning system. Input: patient CT images via DICOM. Workflow: user defines regions of interest (ROIs) manually or semi-automatically; specifies treatment setup, machine, energy, and delivery type (SMLC); defines beam configuration (isocenter, dose grid, fluence grid). System uses a three-dimensional collapsed cone dose engine to optimize treatment parameters and compute clinical dose distribution. Output: deliverable step-and-shoot IMRT plan, 2D/3D dose displays, DVH curves, and plan data. Used in clinical settings by radiation therapy staff. Clinicians review and approve plans; approved data is exported via DICOM to servers for treatment delivery. Benefits: enables precise radiation therapy planning and analysis to support clinical decision-making.

Clinical Evidence

Bench testing only. No clinical data provided. Accuracy of the dose calculation algorithm was validated using test specifications previously applied to the predicate RayAutoplan 1.0. Testing covered a wide variety of field geometries, treatment units, treatment setups, patient positions, and dose grid resolution settings. All tests were completed successfully.

Technological Characteristics

Software-based treatment planning system. Core technology: three-dimensional collapsed cone dose engine. Connectivity: DICOM import/export. Output: IMRT treatment plans, dose distributions, DVH curves. Functionality: manual/semi-automatic contouring, plan optimization, electronic approval.

Indications for Use

Indicated for radiation therapy treatment planning and analysis. Intended for use by clinically qualified radiation therapy staff trained in the system to generate treatment unit set-up parameters and dose distribution estimates for patient treatment.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ | K100 | | | ર | ﺎ ﮨﮯ ﮨﮯ ﮨﮯ | ನಿ | |------|--|--|---|------------|----| |------|--|--|---|------------|----| | Title: | Document ID: | Version: | |-------------------------------------|--------------|----------| | 510(k) Application - RayStation 1.0 | RSL-D-61-04 | 1.1 | ### 510(k) Summary 5. 5.1 510(k) owner > RaySearch Laboratories AB Sveavägen 25, plan 9 111 34 Stockholm Sweden MAR 1 2 2010 Tel: +46 (8) 54506130 Fax: +46 (8) 54506139 #### 5.2 Contact person Anders Murman ### 5.3 Preparation date 12/11/2009 #### 5.4 Trade name ## RayStation RayStation is the product name RaySearch will use introducing this device into the world market. Aliases may be added later, if RayStation is sublicensed to other radiation therapy companies. Throughout the documentation it is our practice to write the trade name and version number together, i.e. "RayStation 1.0". #### 5.5 Common name Radiation treatment planning system #### રે છે. દિવેલી તેમને છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેત-ઉત્પાદની ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખે Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUI) #### 5.7 Predicate devices | RayAutoplan 1.0 | 510(k) number K083264 | |-------------------------|-----------------------| | Oncentra MasterPlan 3.1 | 510(k) number K081281 | #### 5.8 Device description RayStation is a treatment planning system, i.e. a software program for planning and analysis of radiation therapy plans. Typically, a treatment plan is created by importing patient images obtained from a CT scanner, defining regions of interest either manually or semi-automatically, deciding on a treatment setup and objectives, optimizing the treatment parameters, comparing rival plans to find the best compromise, computing the clinical dose distribution, approving the plan and exporting it. The main workflow, creating a step-and-shoot IMRT plan from an imported patient and case, is described below: | | Flow of Events | | | |----|------------------------------------------|----|-----------------------------------------| | | User | | System | | 1. | The user launches RayStation 1.0 | | | | 2. | The user imports a patient and case with | | | | | CT images through DICOM | 3. | The system imports the data and checks | | | | | consistency of in-data | | 4. | The user enters the Structure Definition | | | | | module and creates ROIs using the | | | | | contouring tools | 5. | The system adds the ROIs to the patient | | | | | case | | 6. | The user enters the Treatment | | | | | Specification module and creates a plan | | | {2}------------------------------------------------ | Title: | Document ID: | Version: | |-------------------------------------|--------------|----------| | 510(k) Application - RayStation 1.0 | RSL-D-61-04 | 1.1 | | | and a treatment setup with specified<br>machine, treatment energy and delivery<br>type (SMLC) | | | |-----|--------------------------------------------------------------------------------------------------------|-----|---------------------------------------------------------------------------------------------| | 7. | The user specifies beam configuration<br>including isocenter, dose grid and fluence<br>grid resolution | | | | | | 8. | The system adds the plan and treatment<br>setup to the patient case | | 9. | The user enters the Plan Optimization<br>module and creates an optimization<br>problem | | | | 10. | The user defines the algorithm and<br>segmentation settings and starts the<br>optimization | | | | | | | 11. The system generates a deliverable step-<br>and-shoot plan | | | | | 12. The system displays the plan as | | | | | - 2D and 3D dose and patient displays<br>- DVH curves<br>- Plan data (beams, segments etc.) | | 13. | The user reviews the plan | | | | 14. | The user enters the Plan Evaluation<br>module and evaluates the plan | | | | 15. | The user approves and exports the plan<br>together with dose, structure sets and<br>images | | | | | | 16. | The system exports the plan and patient<br>data to a DICOM server | #### 5.9 Intended use RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system. #### 5.10 Technological characteristics summary The technological characteristics are the same for RayStation 1.0 as for RayAutoplan 1.0 and Oncentra MasterPlan v3.1. All devices produce IMRT treatment plans with corresponding dose distributions computed using a three dimensional collapsed cone dose engine. All devices have a function of electronic approval of treatment plans by trained and authorized staff, and export in DICOM format for commencing treatment or archiving. #### Assessment of non-clinical performance data 5.11 The dose algorithm in both RayStation 1.0 and RayAutoplan 1.0 is the same. This is supported by the dose algorithm accuracy testing, which has used the same test specification 1.0 as was previously used for Ray Autoplan 1.0. The tests include dose calculation for a wide variety of field geometries, treatment units, treatment setups and patient positions, including different dose grid resolution settings. All tests were run successfully for RayStation 1.0. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized lines that resemble a human figure. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 # MAR 1 2 2010 RaySearch Laboratories AB % Mr. Casey Conry Senior Project Engineer Underwriters Laboratories, Inc. 1285 Walt Whitman Rd. MELVILLE NY 11747 Re: K100552 Trade/Device Name: RayStation Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: February 19, 2010 · Received: February 26, 2010 ## Dear Mr. Conry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register, Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Amelia J. Roth Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K100552 Device Name: RayStation Indications for Use: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system. Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) (Division Sign-Off) Division Sign-Off) of Radiological Devices Office of In Vitro Diagnostic Dev 510K K100552 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...