← Product Code [NEW](/productcode/NEW) · K100461

# PDM, VILET, VILET QUICK, MONO Q, PLAIN GUT:# 6 CHROMIC GUT (K100461)

_Riverpoint Medical · NEW · Jul 8, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100461

## Device Facts

- **Applicant:** Riverpoint Medical
- **Product Code:** [NEW](/productcode/NEW.md)
- **Decision Date:** Jul 8, 2010
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4840
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

PDO: PDO surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery other than contact with comea and/or sclera. PDO Suture is not intended for use in adult cardiovascular tissue, microsurgery or neural tissue. PGA: PGA surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue. Fast Absorbing PGA (Trade Name: Vilet Quick): Vilet Quick surgical suture is indicated for use in general soft tissue approximation, including use in ophthalmic procedures, where only short-term wound support is required. Vilet Quick is not intended for use in cardiovascular or neurological procedures. PGCL: PGCL surgical suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological procedures.

## Device Story

Absorbable surgical sutures (PDO, PGA, Fast Absorbing PGA, PGCL) used to secure tissues or close wounds; provided with or without medical-grade needles. Used by surgeons in clinical settings. Sutures provide mechanical tissue approximation; absorb over time. Benefit: wound closure with reduced need for removal. Performance verified against USP 32 requirements for diameter, tensile strength, and needle attachment strength.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity (bacterial reverse mutation, chromosomal aberration, bone marrow micronucleus), subchronic toxicity (4-week subcutaneous), and muscle implantation (12-week). Mechanical testing verified USP 32 requirements for diameter, tensile strength, and needle attachment strength.

## Technological Characteristics

Materials: Polydioxanone (PDO), Polyglycolic Acid (PGA), Poly(glycolide-co-caprolactone) (PGCL). Dimensions: USP 32 standard sizes (12/0 through 7). Sterilization: Provided sterile. Biocompatibility: ISO 10993 compliant. Mechanical specs: USP 32.

## Regulatory Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- PDO ([K061037](/device/K061037.md))
- PGA ([K972566](/device/K972566.md))
- Fast Absorbing PGA ([K023710](/device/K023710.md))
- PGCL ([K960653](/device/K960653.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

point

K100461 Page 1/3

# 510(k) Summary

Page 1-23

JUL = 8 2010

### Submitter Information

Submitter's Name:

Address:

825 NE 25" Ave. Portland, OR 97232

Riverpoint Medical

Phone Number:

Contact Person:

(503) 517-8001 or 866 445-4923

(503) 517-8002 Fax Number:

Registration Number:

Douglas Rowley (503) 517-8001

February 03, 2010

3006981798

Date of Preparation:

### Device Names

Trade Names:

PGCL: PGA: Fast Absorbing PGA:

PDO:

PDM Mono Q Vilet Vilet Quick

| Common Names:         | Absorbable Surgical Sutures: PDO, PGA, PGCL |
|-----------------------|---------------------------------------------|
| Classification Names: | All: General and Plastic Surgery Devices    |

### Device Classification

FDA Class:

2 (All varieties listed)

Product Classification:

- Polydioxanone (PDO): 1. 878.4840, Absorbable polydioxanone surgical suture 2. Polyglycolic Acid (PGA):
- 3. Fast Absorbing Polyglycolic Acid (PGA):
- 4. Poly(glycolide-co-caprolactone){PGCL):

878.4493, Absorbable poly(glycolide/l-lactide) surgical suture 878.4493, Absorbable poly(glycolide/l-lactide) surgical suture 878.4493, Absorbable poly(glycolide/l-lactide) surgical suture

510(k) Summary - Absorbable Suture

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a stylized drawing of the letter 'r' in a simple, blocky font. Below the letter, the word 'point' is written in a bold, sans-serif typeface. The letter 'r' has a distinct, almost cartoonish appearance due to its rounded top portion. The overall image has a clean and straightforward design.

K100461 Aa.

Page 1-24

Codes:

- 1. Polydioxanone (PDO): NEW
- 2. Polyglycolic Acid (PGA): GAM
- 3. Fast Absorbing PGA: GAM
- 4. Poly(glycolide-co-caprolactone)(PGCL): GAM

Classification Panel: All: Class II (special controls); General and Plastic Surgery

# Predicate Devices (applicable 510(k) number listed):

| 1. PDO:                | K061037 |
|------------------------|---------|
| 2. PGA:                | K972566 |
| 3. Fast Absorbing PGA: | K023710 |
| 4. PGCL:               | K960653 |

# Special Controls

FDA Guidance "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" was followed during the preparation of this submission.

# Device Description

Riverpoint Medical Absorbable surgical sutures are medical devices used to secure tissues together or create wound closures during a surgical procedure or after an injury. They are composed of the applicable suture material and a standard medical grade suture needle as applicable (sutures can be provided without needles as well).

Available Suture sizes will be standard according to USP 32 requirements (12/0 through 7, depending on suture type).

# Intended Use

PDO: PDO surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery other than contact with comea and/or sclera. PDO Suture is not intended for use in adult cardiovascular tissue, microsurgery or neural tissue.

PGA: PGA surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.

510(k) Summary - Absorbable Suture

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a symbol that looks like the letter 'r' with a curved top. Below the symbol, the word 'point' is written in a bold, sans-serif font. The image quality is somewhat degraded, with visible noise and artifacts, particularly around the edges of the symbol and text.

K100461 Age 3/3

Fast Absorbing PGA (Trade Name: Vilet Quick): Vilet Quick surgical suture is indicated for use in general soft tissue approximation, including use in ophthalmic procedures, where only short-term wound. support is required. Vilet Quick is not intended for use in cardiovascular or neurological procedures.

PGCL: PGCL surgical suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological procedures.

### Safety and Effectiveness

Each variety of Riverpoint Medical absorbable sutures has been designed and manufactured to be substantially equivalent to the predicate devices listed for safety and effectiveness. Materials used were selected based on known biocompatibility (per ISO 10993) and established histories of use in the medical device industry for implantable devices, and are identical or substantially equivalent to the materials used in the predicate devices listed. Biocompatibility testing performed on each variety of absorbable sutures within this submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Genotoxicity - Bacterial Reverse Mutation and Chromosomal Aberration, Bone Marrow Micronucleus, Subchronic Toxicity (4-week, following subcutaneous implantation), Muscle Implantation (12-week).

Riverpoint Medical absorbable sutures have been designed to meet the requirements for diameter, tensile strength, and needle attachment strength as specified within USP 32. Testing is performed on each lot of product to verify that USP requirements have been met prior to release.

#### 510(k) Summary - Absorbable Suture

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Riverpoint Medical % Mr. Doug Rowley RA/QA Manager 825 NE 25" Avenue Portland, Oregon 97232

JUL 8 2010

Re: K100461

Trade/Device Name: Polyglycolic Acid (PGA) Absorbable Surgical Suture, PGA Absorbable Surgical Suture (Fast Absorbing), Poly(glycolide co caprolactone) PGCL) Absorbable Surgical Suture, Polydioxanone (PDO) Absorbable Suture

Regulation Number: 21 CFR 878.4493

Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: II Product Code: GAM, NEW Dated: June 7, 2010 Received: July 2, 2010

Dear Mr. Rowley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Doug Rowley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a symbol that resembles the letter 'p' with a circular loop at the top. Below the symbol, the word 'point' is written in lowercase letters. The symbol and the word are likely related, possibly indicating a specific point or location.

# Indications for Use Statement

| 510(k) Number: | Unknown at this time                               |
|----------------|----------------------------------------------------|
| Device Name:   | Polyglycolic Acid (PGA) Absorbable Surgical Suture |
| Trade Name:    | Vilet                                              |

## Indications for Use:

PGA surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue. PGA suture is provided sterile as a single use device.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krone for MKM
(Division Sign-Off)

Division of Surgical, Orthopedic. and Restorative Devices

510(k) Number K100461

Page 1 of 1

510(k) Indications for Use Statement - Absorbable Suture

Page 1-19

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a drawing of the letter 'P' with the word 'point' written below it. The 'P' is drawn with a thick line and has a circular shape at the top. The word 'point' is written in a simple, sans-serif font. There are some blurry artifacts around the image.

# Indications for Use Statement

| 510(k) Number: | Unknown at this time                            |
|----------------|-------------------------------------------------|
| Device Name:   | PGA Absorbable Surgical Suture (Fast Absorbing) |
| Trade Name:    | Vilet Quick                                     |

## Indications for Use:

Vilet Quick surgical suture is indicated for use in general soft tissue approximation, including use in ophthalmic procedures, where only short-term wound support is required. Vilet Quick is not intended for use in cardiovascular or neurological procedures. Vilet Quick suture is provided sterile as a single use device.

Prescription Use ਮੁ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krane for MKM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100461

Page 1 of 1 510(k) Indications for Use Statement - Absorbable Suture Page 1-20

{7}------------------------------------------------

K100461 RESPONSE

Image /page/7/Picture/1 description: The image shows a stylized drawing of the letter 'p' with the word 'point' written below it. The letter 'p' is drawn with a thick outline, and the circular part of the 'p' is a complete circle. The word 'point' is written in a simple, sans-serif font.

Page 1-21

# Indications for Use Statement

| 510(k) Number: | Unknown at this time |
|----------------|----------------------|
|----------------|----------------------|

Poly(glycolide co-caprolactone) (PGCL) Absorbable Surgical Suture Device Name:

Trade Name: Mono Q

Indications for Use:

PGCL surgical suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological procedures. PGCL suture is provided sterile as a single use device.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Keane for MXM
(Division Sign-Off)

Division of Surgical, Orthoped and Restorative Devices

510(k) Number K100461

Page 1 of 1

510(k) Indications for Use Statement - Absorbable Suture

{8}------------------------------------------------

K100461 RESPONSE

Image /page/8/Picture/1 description: The image shows a stylized letter "P" above the word "point". The "P" is drawn with a thick line and has a circular top. The word "point" is written in a simple, sans-serif font and is positioned directly below the letter.

Page 1-18

# Indications for Use Statement

| 510(k) Number: | Unknown at this time                           |
|----------------|------------------------------------------------|
| Device Name:   | Polydioxanone (PDO) Absorbable Surgical Suture |
| Trade Name:    | PDM                                            |

## Indications for Use:

PDO surgical suture is indicated for use in general soft tissue approximation and/or ligation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery (other than contact with cornea and/or sclera). PDO Suture is not intended for use in adult cardiovascular tissue, microsurgery or neural tissue. PDO suture is provided sterile as a single use device.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BECOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Device Evaluation (SDE)
David Krane for MXM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100461

510(k) Indications for Use Statement - Absorbable Suture

---

**Source:** [https://fda.innolitics.com/device/K100461](https://fda.innolitics.com/device/K100461)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
