SMARTLINK GOLD BLOOD GLUCOSE MONITORING SYSTEM

K100398 · Standard Diagnostics, Inc. · NBW · Jun 9, 2010 · Clinical Chemistry

Device Facts

Record IDK100398
Device NameSMARTLINK GOLD BLOOD GLUCOSE MONITORING SYSTEM
ApplicantStandard Diagnostics, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateJun 9, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The SD Check GOLD blood glucose monitoring system is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger, palm, forearm, upper arm, calf and thigh. It is intended for use by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Story

SmartLink™ GOLD Blood Glucose Monitoring System is a portable, battery-operated in vitro diagnostic device for measuring glucose concentration in capillary whole blood. System components include a meter, test strips, lancets, lancing device, and control solutions. User applies blood sample to test strip; meter processes sample and displays glucose concentration in mg/dL. Device stores 500 results with 7, 14, and 30-day averages for normal, pre-meal, and post-meal states. Features include hypo-warning, alarms, and pre/post-meal marking. Used by patients at home or healthcare professionals in clinical settings to monitor diabetes management. Healthcare providers use stored trends to assess glycemic control. Benefits include convenient, rapid self-monitoring of blood glucose levels to aid in diabetes management.

Clinical Evidence

No clinical sensitivity/specificity testing performed. System accuracy evaluated via method comparison with predicate device per EN ISO 15197:2003. User performance study conducted with lay consumers using capillary blood from fingertip, palm, forearm, and upper arm to demonstrate accuracy.

Technological Characteristics

Portable, battery-operated (3V type 2032) glucose meter. Uses electrochemical test strips. Features include 500-result memory, LCD display, and no-coding technology. Complies with EN ISO 15197:2003, IEC 61010-1, and IEC 61010-2 safety standards. Connectivity includes internal memory storage for trend analysis.

Indications for Use

Indicated for people with diabetes for quantitative glucose measurement in fresh capillary whole blood (finger, palm, forearm, upper arm, calf, thigh) to monitor diabetes control. Not for neonates, diagnosis, or screening.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ and the control control consideration of the JUN - 9 2010 Exhibit #1 ## 510(K) SUMMARY .. . . . . . . . . . This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. .. The assigned 5l0(k) number is:_k100398 ## 1. Submitter's Identification: Standard Diagnostics, Inc. Head Office: 156-68 Hagal-dong, Giheung-gu, Yongin-si, Kyonggi-do, 446-930 KOREA Tel: 82-31-899-9818 FAX: 82-31-899-9755 ## Manufacturing Site: C-4th & 5th Floor Digital Empire Building 980-3 Yeongtong-dong, Yeongtong-gu Suwon-si, Kyonggi-do 443-702 KOREA Tel: 82-31-300-0423 FAX: 82-31-300-0499 Contact: Mr. Sun Young Jeong QA/RA Manager Standard Diagnostics, Inc. Date Summary Prepared: June 7, 2010 . - 2. Name of the Device: SmartLink™ GOLD Blood Glucose Monitoring System - - - - - - - - - - - - - - - - - - - - - - - - {1}------------------------------------------------ #### Common or Usual Name:Blood·Glucose Test System, Over-the-Counter 3. #### Predicate Device Information: 4. SD CHECK GOLD Blood Glucose Monitoring System, K082683 #### 5. Device Description: The SmartLink™ GOLD Blood Glucose System is an Rx/OTC blood glucose system to be used by professional healthcare personnel or diabetics at home to measure the glucose concentration for aiding diabetes management. The SmartLink™ GOLD Blood Glucose System consists of a SmartLink™ GOLD Blood Glucose Meter which measures and displays test results, SmartLink™ GOLD test strip, lancet and lancing device for blood sampling, control solution to check the meter and test strip, check strip to check the meter, 3 V battery (type 2032) for power, carrying case, user guide, packaging inserts for the test strip and control solution and a Quick Guide for helping the user with a self test diary to document test results for monitoring the trend of blood glucose. The system is a battery-operated portable device and stores 500 test results in memory. The user can search the stored results with three presentations of 7, 14 and 30-day averages of test results stored in memory: normal, pre-meal and post-meal state averages. The system can set the beep, hypo warning, date, time, post-meal alarm and alarm. The svstem can also set the pre-meal and postmeal mark. Test results are displayed with mg/dL unit. A check strip allows the meter to check a problem and the control solution allows the meter and test strip to be checked. #### 6. Intended Use: SmartLink™ GOLD Blood Glucose Monitoring System is indicated for monitoring qlucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm or upper arm. SmartLink™ GOLD meter must be used with SmartLink™ GOLD blood glucose test strip and SD Control solutions. The SD control solutions Level M and Level H are for use with SmartLink™ GOLD test system as quality controls to verify the accuracy of blood glucose test results. and and personal programment and province and comments of the comments of . . . . . . . . . . . . . Testing is done outside the body (in vitro diagnostic use). {2}------------------------------------------------ This system is indicated for home (over-the-counter, OTC) by person with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. This system should not be used for the screening or diagnosis of diabetes or for testing neonates. #### 7. Comparison to Predicate Devices: The SmartLink 100 GOLD Blood Glucose System is substantially equivalent to our SD CHECK GOLD Blood Glucose Monitoring System, K082683 with the main difference being no coding (one code). Both the subject and predicate devices are similar in intended use and basic fundamental scientific technology with differences in: - A. SmartLink™ GOLD Meter Appearance, Size, Weight, Color, Codechip, PCB, LCD, MCU - B. SmartLinkTM GOLD Test strip Printed Film, shelf life ### 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Based on our risk analysis evaluation results, and, in accordance with the FDA "Draft Guidance for Industry and FDA Staff - Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems, 10/24/06", outlined performance characteristics, the following testing was conducted to support the modifications found in our subject device: - Software Verification and Validation Testing . - . Precision Evaluation" (Exhibit #11) - . Linearity Testing - . Performance Evaluation at Sea Level and High Altitude - Equipment Temperature & Humidity Exposure Test . - Mechanical Resistance to Vibration-Environmental Testing . - Document TD-8 Testing Temperature Study . - Electromagnetic Compatibility Study (EN ISO 15197:2003) . - Electrical Safety Study (IEC 61010-1 & IEC 61010-2) . {3}------------------------------------------------ None of the testing demonstrated any design characteristics that violated the requirements of the FDA recognized standards or resulted in any safety hazards. It was our conclusion that testing met all relevant standards requirements. #### 9. Discussion of Clinical Tests Performed: 1 Clinical sensitivity and clinical specificity testing is not applicable. A system accuracy evaluation (Method Comparison with Predicate Device) for the SmartLink™ GOLD Blood Glucose Monitoring System was performed according to EN ISO 15197:2003 A user performance study was performed to demonstrate that lay consumers could obtain accurate results using the SmartLink " GOLD Blood Glucose Monitoring System. The study was performed using capillary whole blood from fingertip, palm, forearm, and upper arm sample sites. #### 10. Conclusions: Based on documentation supplied with this submission, conclusions drawn from clinical and bench testing of the subject device demonstrates that the subject device is as safe, as effective, and performs as well as our legally marketed predicate device. 12 un any is {4}------------------------------------------------ Public Health Service Building 66 Image /page/4/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an eagle-like emblem with three stylized lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the emblem. The seal is black and white. Standard Diagnostics, Inc. c/o Ms. Susan Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021-1734 Re: k100398 Trade/Device Name: SmartLink GOLD Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW. CGA Dated: May 7, 2010 Received: May 10, 2010 Dear Ms. Goldstein-Falk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. **JUN** 0 3 2000 Silver Spring, MD 20993 Food & Drug Administration 10903 New Hampshire Avenue {5}------------------------------------------------ ## Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, CA Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement 510(k) Number ... .. .. .. k100398 Device Name SmartLink™ GOLD blood glucose monitoring system ## INDICATIONS FOR USE SmartLink™ GOLD blood glucose monitoring system is indicated for monitoring glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm or upper arm. SmartLink™ GOLD meter must be used with SmartLink™ GOLD blood glucose test strip and SD Control solutions. The SD control solutions Level M and Level H are for use with SmartLink™ GOLD test system as quality controls to verify the accuracy of blood glucose test results. Testing is done outside the body (in vitro diagnostic use). This system is indicated for home (over-the-counter; OTC) by person with diabetes, or in · clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. This system should not be used for the screening or diagnosis of diabetes or for testing neonates. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ X (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH, Office of In Vitro.Diagnostic Device (OIVD) - - - - Signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) R100398
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...