VISTAKON (NARAFILCON B) CONTACT LENS VISIBILITY TINTED WITH UV BLOCKER
K100349 · Johnson & Johnson Vision Care, Inc. · LPL · May 21, 2010 · Ophthalmic
Device Facts
Record ID
K100349
Device Name
VISTAKON (NARAFILCON B) CONTACT LENS VISIBILITY TINTED WITH UV BLOCKER
Applicant
Johnson & Johnson Vision Care, Inc.
Product Code
LPL · Ophthalmic
Decision Date
May 21, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The VISTAKON® (narafilcon B) Soft Contact Lens (spherical) is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The VISTAKON® (narafilcon B) Multifocal Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic, phakic or aphakic persons with non-diseased eyes who may have 4.00D of ADD power or less and 0.75D of astigmatism or less. The VISTAKON® (narafilcon B) Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of visual acuity in phakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00D or less of astigmatism. The VISTAKON® (narafilcon B) Multifocal-Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eves who may have 4.00D of ADD power or less and 10.00D of astigmatism or less. VISTAKON® (narafilcon B) Contact Lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye. The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.
Device Story
VISTAKON® (narafilcon B) contact lens; silicone hydrogel material with internal wetting agent; spherical, multifocal, toric, and toric-multifocal designs. Features blue visibility tint and benzotriazole UV-absorbing monomer. Supplied sterile in buffered saline with methyl ether cellulose. Prescribed by eye care professionals for daily wear, single-use; discarded after removal. Provides refractive correction for phakic/aphakic patients; UV blocker reduces transmission of UV radiation to cornea/eye.
Clinical Evidence
One-month clinical study, 48 subjects, 1:1 ratio (test vs. 1-DAY ACUVUE® control). Endpoints: slit lamp findings, symptoms, visual acuity, wear time, adverse reactions, keratometry, and discontinuation rates. Results showed similar performance in vision and health compared to concurrent controls.
Indicated for daily wear, single-use optical correction of myopia, hyperopia, and presbyopia in phakic or aphakic persons with non-diseased eyes. Supports astigmatism correction up to 10.00D and ADD power up to 4.00D depending on lens design.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{1}------------------------------------------------
K100349
MAY 2 ] 2010
# 510(k) Summary
.
| Submitter<br>Information | Company: | Johnson & Johnson Vision Care, Inc.<br>7500 Centurion Parkway<br>Suite 100<br>Jacksonville, FL 32256 | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--|
| | Contact Person: | Catherine Dillon | |
| | Email: | cdillon9@its.jnj.com | |
| | Telephone: | 904-443-3180 | |
| | FAX: | 904-443-1424 | |
| | Date Prepared: | February 8, 2010 | |
| Identification of<br>the Device | Common Name: | Soft Contact Lens | |
| | Device Name: | VISTAKON® (narafilcon B) Contact Lens | |
| | Classification Name: | Soft Hydrophilic Contact Lens, Daily Wear | |
| | Device Classification: | Class II, 21 CFR 886.5925 (b) (1). | |
| Predicate<br>Device(s) | Material<br>VISTAKON® (narafilcon A) Contact Lens – K073485<br>(FDA Group I; low water, nonionic polymer) | | |
| | Indication, Wear Schedule<br>VISTAKON® (etafilcon A) Contact Lenses – K962804<br>(Daily wear, single use) | | |
.
{2}------------------------------------------------
| Description of Device | The VISTAKON® (narafilcon B) Contact Lens Visibility Tinted with UV Blocker is available as a spherical lens, multifocal lens, toric lens, and toric-multifocal lens. The lenses are made of a silicone hydrogel material containing an internal wetting agent. The VISTAKON® (narafilcon B) Contact Lens is tinted blue using Reactive Blue Dye #4 to make the lens more visible for handling. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 1% in the UVB range of 280-315nm and less than 10% in the UVA range of 316-380nm. The VISTAKON® (narafilcon B) Contact Lens is a hemispherical or hemitoric shell. The lens is supplied in a sterile state, packaged in a buffered saline solution with methyl ether cellulose. The composition of the lens is 52% narafilcon B and 48% water by weight when hydrated and stored in the buffered saline solution. |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Continued on next page
.
{3}------------------------------------------------
・ Ind
| Indications for<br>Use | Lens Design | Indication |
|------------------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Spherical | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of refractive ametropia (myopia and<br>hyperopia) in phakic or aphakic persons with non-<br>diseased eyes who may have 1.00D or less of<br>astigmatism. |
| | Multifocal | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of distance and near vision in presbyopic,<br>phakic or aphakic persons with non-diseased eyes who<br>may have 4.00D of ADD power or less and 0.75D of<br>astigmatism or less. |
| | Toric | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of visual acuity in phakic or aphakic persons<br>with non-diseased eyes that are hyperopic or myopic<br>and may have 4.00D of ADD power or less and 10.00D<br>or less of astigmatism. |
| | Multifocal<br>Toric | The VISTAKON® (narafilcon B) Contact Lens is<br>indicated for daily wear single use only for the optical<br>correction of distance and near vision in presbyopic<br>phakic or aphakic persons with non-diseased eyes who<br>may have 4.00D of ADD power or less and 10.00D of<br>astigmatism or less. |
- VISTAKON® (narafilcon B) Contact Lenses contain a UV . Blocker to help protect against transmission of harmful UV . radiation to the cornea and into the eye.
- . The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.
{4}------------------------------------------------
#### Technological Characteristics
The technological characteristics of the VISTAKON® (narafilcon B) Contact Lenses are compared to the characteristics of the predicate device, VISTAKON® (narafilcon A) Contact Lens, in the following tables.
| | Predicate Device | Subject Device |
|----------------------|---------------------------------------------|---------------------------------------------|
| Product Name | VISTAKON®<br>(narafilcon A)<br>Contact Lens | VISTAKON®<br>(narafilcon B)<br>Contact Lens |
| Material USAN Name | narafilcon A | narafilcon B |
| 510(k) Number | K073485 | TBD - This submission |
| FDA Category (Group) | Group I | Group I |
| Manufacturing Method | Molded | Molded |
| Sterilization | Moist Heat | Moist Heat |
| Packaging | Blister | Blister |
| Visibility Tint | Blue | Blue |
| Parameter Comparison | | | | |
|-------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------|----------|---------|
| | Predicate Device<br>VISTAKON®<br>(narafilcon A) Contact<br>Lens-K073485 | Subject Device<br>VISTAKON®<br>(narafilcon B)<br>Contact Lens | | |
| | Measured | Labeled | Measured | Labeled |
| Water Content, % | 47 | 46 | 48 | 48 |
| Refractive Index @ 20°C | 1.40 | 1.41 | 1.41 | 1.41 |
| Dk, edge corrected | 96 | 100 | 52 | 55 |
| Base Curve, mm | 8.53 | 8.5 | 8.49 | 8.5 |
| Diameter, mm | 14.26 | 14.2 | 14.19 | 14.2 |
| Power, D | -0.92 | -1.00 | -0.98 | -1.00 |
Polarographic Method, Dk units: 10 -11(cm+/sec)(ml O2/ml x mmHg)
{5}------------------------------------------------
| Non-clinical<br>Testing | A series of in-vitro and in-vivo preclinical toxicology and<br>biocompatibility tests were performed to assess the safety and<br>effectiveness of the contact lens. All tests were conducted in<br>accordance with the GLP regulation (21 CFR Part 56) or according to<br>valid scientific protocols. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The results of the non-clinical testing/evaluation demonstrate that: |
| | the lens material and/or extracts are non-toxic, non-irritating and<br>non-sensitizing under the experimental conditions; and the lens physical and material properties are consistent with<br>currently marketed lenses. |
| Clinical Testing | A one-month clinical study was completed to evaluate the safety and<br>efficacy of the VISTAKON® (narafilcon B) Contact Lens for single.<br>use daily wear only. |
| | The study evaluated 48 subjects with a 1:1 ratio between the test lens<br>and the control lens (1·DAY ACUVUE® Brand Contact Lenses). The<br>primary endpoints were slit lamp findings, symptoms, problems and<br>complaints, visual acuity and average wear time. Additional |
parameters measured included adverse reactions, keratometry changes, reasons for discontinuation, and the number and reasons for unscheduled lens replacements.
The clinical evaluation demonstrated similar overall performance in the clinically relevant areas of vision and health as compared to concurrent controls when used under daily wear single use conditions.
{6}------------------------------------------------
#### Conclusions Drawn from Studies
| Validity of<br>Scientific Data | A contract laboratory under Good Laboratory<br>Practice Regulations conducted toxicology studies.<br>Microbiology, chemistry, shelf-life stability, and<br>leachability studies were conducted by VISTAKON®<br>laboratories and followed scientific protocols. The<br>data were determined to be scientifically valid under<br>21 CFR 860.7. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence | Information presented in this Premarket Notification<br>establishes that the VISTAKON® (narafilcon B)<br>Contact Lens is as safe and effective as the predicate<br>device when used in accordance with the labeled<br>directions for use and for the requested indication. |
| Risk and<br>Benefits | The risks of the subject device are the same as those<br>normally attributed to the wearing of soft<br>(hydrophilic) contact lenses on a daily wear single<br>use basis. The benefits to the patient are the same as<br>those for other soft (hydrophilic) contact lenses. |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle or bird-like figure with its wings spread, positioned above a wavy line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002
Johnson & Johnson Vision Care, Inc. C/O Catherine C. Dillon Regulatory Affairs 7500 Centurion Parkway, Suite 100 Jacksonville, FL 32256
Re: K100349
Trade/Device Name: VISTAKON® (narafilcon B) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: May 4, 2010 Received: May 5, 2010
Dear Ms. Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
MAY 2 1 2010
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 – Ms. Catherine C. Dillon
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 70 http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/7/Picture/8 description: The image shows a handwritten word that appears to be "for". The script is cursive and the letters are connected. The "f" has a prominent loop extending below the baseline, and the "o" and "r" are joined together at the top.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known):
Device Name:
VISTAKON® (narafilcon B) Contact Lens Visibility Tinted, with UV Blocker
Indications for Use:
The VISTAKON® (narafilcon B) Soft Contact Lens (spherical) is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
100349
The VISTAKON® (narafilcon B) Multifocal Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic, phakic or aphakic persons with non-diseased eyes who may have 4.00D of ADD power or less and 0.75D of astigmatism or less.
The VISTAKON® (narafilcon B) Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of visual acuity in phakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00D or less of astigmatism.
The VISTAKON® (narafilcon B) Multifocal-Toric Soft Contact Lens is indicated for daily wear single use only for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eves who may have 4.00D of ADD power or less and 10.00D of astigmatism or less.
VISTAKON® (narafilcon B) Contact Lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye.
The Eye Care Professional should prescribe the lenses for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use
(21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Kaur Warburton
vision Sign-Off Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K100349
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.