DIMENSION VISTA LOCI CA 15-3 FLEX REAGENT CARTRIDGE, DIMENSION VISTA LOCI 7 CALIBRATOR, MODELS: K6456 AND KC605
Applicant
Siemens Healthcare Diagnostics
Product Code
MOI · Immunology
Decision Date
Apr 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K100344 · Apr 18, 2011
DIMENSION VISTA LOCI CA 15-3 FLEX REAGENT CARTRIDGE, DIMENSION VISTA LOCI 7 CALIBRATOR, MODELS: K6456 AND KC605
Siemens Healthcare Diagnostics
Retrospective clinical patient samples; Medical records/clinical history
Retrospective clinical samples were used to evaluate the clinical performance of the CA 15-3 assay by comparing changes in CA 15-3 concentrations against physician-determined disease status (Active/Progressive, Responding, Stable, or No Evidence of Disease).
Retrospective study; Clinical performance; Disease monitoring; Breast cancer
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of serial clinical samples
75 female patients with breast cancer (Stage I-IV); Sample Size: 75 patients (258 sequential pairs)
Not applicable for this study
Concordance of CA 15-3 concentration changes with clinical disease status (progression vs. no progression)
Indications for Use
The LOCI CA 15-3 method is an in vitro diagnostic test for the quantitative measurement of CA 15-3 in human serum and lithium heparin and EDTA plasma on the Dimension Vista® System. When used in conjunction with other clinical and diagnostic procedures, serial testing with the LOCI CA 15-3 assay may be used as an aid in the management of previously treated stage II and III breast cancer patients and for monitoring response to therapy in metastatic breast cancer patients. The LOCI 7 CAL is an in vitro diagnostic product for the calibration of the Cancer Antigen 15-3 (CA 15-3) and Cancer Antigen 19-9 (CA 19-9) methods on the Dimension Vista® System.
Device Story
LOCI CA 15-3 is a homogeneous, sandwich chemiluminescent immunoassay for the Dimension Vista® System. Input: human serum, lithium heparin, or EDTA plasma. Process: sample incubated with biotinylated anti-CA 15-3 antibody and Chemibeads (coated with anti-CA 15-3 monoclonal antibody); Sensibeads (streptavidin-coated with photosensitizer) added to form bead-pair immunocomplexes. Illumination at 680 nm triggers singlet oxygen diffusion from Sensibeads to Chemibeads, inducing chemiluminescence measured at 612 nm. Signal intensity is proportional to CA 15-3 concentration. Used in clinical laboratories by technicians to provide quantitative results to physicians. Output aids in managing stage II/III breast cancer and monitoring metastatic therapy response. LOCI 7 Calibrator is a bovine serum albumin-based product used to calibrate CA 15-3 and CA 19-9 methods on the system.
Clinical Evidence
Retrospective study of 75 female patients (258 serial samples) across stages I-IV. Clinical status (Active/Progressive, Responding, Stable, NED) determined by physician. Using 27.9% RCV, positive percent agreement with progression was 66.7% (95% CI: 56.3%-76.0%); negative percent agreement was 79.0% (95% CI: 71.9%-85.0%). Total concordance with predicate was 90.7%.
Technological Characteristics
Homogeneous sandwich chemiluminescent immunoassay using LOCI® technology. Reagents: two synthetic bead types (Chemibeads with anti-CA 15-3 antibody and chemiluminescent dye; Sensibeads with streptavidin and photosensitizer). Energy source: 680 nm illumination. Detection: 612 nm chemiluminescence. Matrix: bovine serum albumin (calibrator). Connectivity: Dimension Vista® System integration. Form factor: reagent cartridge and liquid calibrator.
Indications for Use
Indicated for quantitative measurement of CA 15-3 in human serum, lithium heparin, and EDTA plasma. Used as an aid in management of previously treated stage II and III breast cancer patients and monitoring therapy response in metastatic breast cancer patients. For professional use on the Dimension Vista® System.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
Predicate Devices
CA 15-3 Assay for the ADVIA Centaur System (K012357)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k100344
B. Purpose for Submission:
New device
C. Measurand:
CA 15-3
D. Type of Test:
Quantitative, Sandwich chemiluminescent immunoassay
E. Applicant:
Siemens Healthcare Diagnostics Inc.
F. Proprietary and Established Names:
Dimension Vista® LOCI CA 15-3 Flex® Reagent cartridge
Dimension Vista® LOC 7 Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR § 866.6010, Tumor-Associated antigen immunological test system
21 CFR § 862.1150, Calibrator
2. Classification:
Class II
3. Product code:
MOI – System, Test, Immunological, Antigen, Tumor
JIX – Calibrator, Multi-analyte Mixture
4. Panel:
Immunology (82)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
The LOCI CA 15-3 method is an in vitro diagnostic test for the quantitative measurement of CA 15-3 in human serum and lithium heparin and EDTA plasma on the Dimension Vista® System. When used in conjunction with other clinical and diagnostic procedures, serial testing with the LOCI CA 15-3 assay may be used as an aid in the management of previously treated stage II and III breast cancer patients and for monitoring response to therapy in metastatic breast cancer patients.
The LOCI 7 CAL is an in vitro diagnostic product for the calibration of the Cancer Antigen 15-3 (CA 15-3) and Cancer Antigen 19-9 (CA 19-9) methods on the Dimension Vista® System.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
Prescription use only
{2}
2
4. Special instrument requirements:
Siemens Dimension Vista® System - device performance was established on the Dimension Vista® 1500 instrument.
I. Device Description:
The LOCI CA 15-3 method is a homogeneous, sandwich chemiluminescent immunoassay based on LOCI® technology. The LOCI® reagents include two synthetic bead reagents and a biotinylated anti-CA 15-3 monoclonal antibody (DF3) fragment. The first bead reagent (Chemibeads) is coated with an anti-CA 15-3 monoclonal antibody (115D8) and contains a chemiluminescent dye. The second bead reagent (Sensibeads) is coated with streptavidin and contains a photosensitizer dye.
The LOCI 7 CAL is a multi-analyte liquid, frozen bovine serum albumin, based product containing CA 15-3 from human cell culture. The kit consists of ten vials, two vials per level (A-E), 2.0 mL per vial.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ADVIA Centaur® CA 15-3 assay
Access® BR Monitor Calibrator
2. Predicate 510(k) number(s):
k012357 (ADVIA Centaur® CA 15-3 assay)
k072612 (Access BR Monitor Calibrator)
3. Comparison with predicate:
| Similarities and differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Name | LOCI CA 15-3 Flex® reagent cartridge (k100344) | ADVIA Centaur System CA 15-3® Assay (k012357) |
| Intended Use | The LOCI CA 15-3 method is an in vitro diagnostic test for the quantitative measurement of CA 15-3 in human serum and lithium heparin and EDTA plasma on the Dimension Vista® System. When used in conjunction with other clinical and diagnostic procedures, serial testing with the LOCI CA 15-3 assay may be used as an aid in the management of previously treated stage II and III breast cancer patients and for monitoring response to therapy in metastatic breast cancer patients. | The ADVIA Centaur CA 15-3 assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 15-3 in human serum using the ADVIA Centaur® System. When used in conjunction with other clinical and diagnostic procedures, serial testing with the ADVIA Centaur CA 15-3 assay is useful for monitoring the course of disease and therapy in metastatic breast cancer patients, and for detection of recurrence in previously treated patients with Stage |
{3}
3
| Similarities and differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Name | LOCI CA 15-3 Flex® reagent cartridge (k100344) | ADVIA Centaur System CA 15-3® Assay (k012357) |
| | | 11 with greater than two positive lymph nodes, or Stage III breast cancer patients. This assay is not intended for use on any other system. |
| Assay technology | Chemiluminescent | Same |
| Sample type | Serum, lithium heparin and EDTA plasma | Serum |
| Assay format | Sandwich bead-pair immunoassay based on LOCI® technology. | Sandwich immunoassay using direct chemiluminometric technology |
| Capture and Detection Antibodies | Monoclonal antibodies DF3 and 115D8 | Same |
| Labeled Antibody Formats | DF3 is biotinylated 115D8 is coated on Chemibeads | Antibody-DF3 is acridinium ester labeled 115D8 is fluorescein labeled |
| Measuring Range | 1 - 300 U/mL | 0.5 - 200 U/mL |
| Sample Size | 1 μL | 20 μL |
| Instrument Platform | Siemens Dimension Vista | ADVIA Centaur automated analyzer |
| Similarities and differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | |
| Feature | LOCI 7 Calibrator (k100344) | Access® BR Monitor Calibrator (k072612) | Dimension Vista LOCI 7 CAL (k100375) |
| Calibrator Intended Use | The LOCI 7 CAL is an in vitro diagnostic product for the calibration of Cancer Antigen 15-3 (CA 15-3) and Cancer Antigen 19-9 (CA 19-9) methods on the Dimension Vista® system. | For in vitro diagnostic use for the calibration of the Access BR Monitor (CA 15-3 Antigen) Assay. | The LOCI 7 CAL is an in vitro diagnostic product for the calibration of the Cancer Antigen 19-9 (CA 19-9) method on the Dimension Vista® system. |
| Calibrator Levels | 5 levels Target Concentrations approximately: Level 1 (Cal A): 0 U/mL Level 2 (Cal B): 20 U/mL Level 3 (Cal C): 60 U/mL Level 4 (Cal D): 150 U/mL | 6 Levels Target Concentrations approximately: Level 1: 0 U/mL Level 2: 10 U/mL Level 3: 50 U/mL Level 4: 100 U/mL | None |
{4}
4
| Similarities and differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | |
| Feature | LOCI 7 Calibrator (k100344) | Access® BR Monitor Calibrator (k072612) | Dimension Vista LOCI 7 CAL (k100375) |
| | Level 5 (Cal E): 315 U/mL | Level 5: 500 U/mL
Level 6: 1000 U/mL | |
| Calibrator Preparation | Liquid | Lyophilized | Liquid |
| Calibrator Storage | Store at -15 to -25°C. | Store at 2 to 8°C. | Store at -15 to -25 °C. |
| Calibrator Matrix | Bovine Serum Albumin | Same | Bovine Serum Albumin |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline
CLSI -EP09-A2, Method Comparison and Bias Estimation Using Patient Samples; Approved Guidelines.
CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation
L. Test Principle:
The LOCI CA 15-3 method is a homogeneous, sandwich chemiluminescent immunoassay based on LOCI® technology. The LOCI® reagents include two synthetic bead reagents and a biotinylated anti-CA 15-3 monoclonal antibody (DF3) fragment. The first bead reagent (Chemibeads) is coated with an anti-CA 15-3 monoclonal antibody (115D8) and contains a chemiluminescent dye. The second bead reagent (Sensibeads) is coated with streptavidin and contains a photosensitizer dye. Sample is incubated with biotinylated antibody and Chemibeads to form bead-CA 15-3-biotinylated antibody sandwiches. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex at 680 nm generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 nm and is a direct function of the CA 15-3 concentration in the sample when measured against a calibration curve.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
i) Assay:
Precision testing was conducted by following CLSI EP5-A2 (Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline) over 20 days. During each day of testing, two separate runs, with two test samples (n=80), for each test material were analyzed. The duplicates were run with two separate sample cups. The studies included two
{5}
instruments (Dimension Vista® 1500 model), two flex lots, and two calibrator lots.
The test samples consisted of three levels of Bio-Rad Liquichek™ Tumor Marker control (Level-1, 2 and 3), three serum and two plasma pools. One of the serum samples was in the lower portion of the normal range, one approximately at the medical decision level and one at the upper end of the assay range. Two of the serum pools (spiked pool-B and high-spiked pool-E) and one of the plasma pools (Lithium heparin pool-C) were spiked with CA 15-3 antigen.
The range of samples tested encompassed the analytical measuring range. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of EP5-A2. This experiment allowed a determination of within-run precision (repeatability), within-day, between days, between runs (within lab) precision for each site and a determination of between site precision.
Multi-Instrument/ Multi-Flex® lot study; Between-Site plus Between-Lot
| Material | Mean U/mL | Standard Deviation (%CV) |
| --- | --- | --- |
| Control Level 1 | 23.6 | 1.81 (7.6) |
| Control Level 2 | 65.0 | 4.37 (6.7) |
| Control Level 3 | 158.1 | 8.29 (5.2) |
| Serum Pool A | 13.6 | 1.46 (10.8) |
| Serum Pool B | 35.5 | 2.80 (7.9) |
| Plasma Pool C (Li-Hep) | 185.5 | 11.18 (6.0) |
| Plasma Pool D (EDTA) | 16.96 | 1.88 (11.1) |
| Serum Pool E | 279.05 | 6.29 (2.3) |
ii) Multicenter and multi-calibrator lot study:
The studies were done at 3 sites in accordance with CLSI EP05-A2. The personnel at all three sites are laboratory personnel and were trained in the operation of the Dimension Vista® system. These sites and personnel are representative of the expected final users as the Dimension Vista® system is designed to be used in clinical laboratories by laboratory personnel who have been trained in the operation of the Dimension Vista® system. Study was performed over 20 days utilizing 3 calibrator lots, 1 flex lot, 3 instruments (Dimension Vista® 1500 model instruments) and 2 replicates (n=360); the duplicates were run with two separate sample cups. The three different sites provides side-by-side comparison of three different instruments using identical lots of the devices; calibrators and reagents. The test samples consisted of three levels of Bio-Rad Liquichek™ Tumor Marker control (Level-1, 2 and 3), three serum and two plasma pools. One of the serum
{6}
samples was in the lower portion of the normal range, one approximately at the medical decision level and one at the upper end of the assay range. Two of the serum pools (spiked pool-B and high-spiked pool-E) and one of the plasma pools (Lithium heparin pool-C) were spiked with CA 15-3 antigen
The range of samples tested encompassed the analytical measuring range. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of EP5-A2. This experiment allowed a determination of within-run precision (repeatability), within-day, between days, between runs (within lab) precision for each site and a determination of between site precision.
Single site precision study:
| Material | Mean U/mL | Standard Deviation (%CV) | | | |
| --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Run | Between-Day | Within-Lab |
| Control Level 1 | 22.3 | 0.48 (2.1) | 0.54 (2.4) | 0.00 (0.0) | 0.72 (3.2) |
| Control Level 2 | 61.9 | 1.03 (1.7) | 1.16 (1.9) | 1.00 (1.6) | 1.85 (3.0) |
| Control Level 3 | 152.2 | 2.13 (1.4) | 2.90 (1.9) | 0.76 (0.5) | 3.68 (2.4) |
| Serum Pool A | 12.6 | 0.28 (2.2) | 0.30 (2.4) | 0.00 (0.0) | 0.41 (3.3) |
| Serum Pool B | 33.5 | 0.62 (1.8) | 0.82 (2.4) | 0.15 (0.5) | 1.04 (3.1) |
| Plasma Pool C (Li-Hep) | 177.6 | 3.41 (1.9) | 2.17 (1.2) | 3.48 (2.0) | 5.33 (3.0) |
| Plasma Pool D (EDTA) | 15.6 | 0.33 (2.1) | 0.21 (1.3) | 0.17 (1.1) | 0.43 (2.7) |
| Serum Pool E | 274.6 | 7.1 (2.6) | 4.1 (1.5) | 5.2 (1.9) | 9.7 (3.5) |
Multi-Instrument/multi-calibrator lot study:
| Material | Mean U/mL | Standard Deviation (%CV) | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Repeatability | Between-Instrument | Between-Calibrator | Between-Day | Total |
| Control Level 1 | 24.2 | 0.5 (2.1) | 0.15 (0.6) | 0.10 (0.4) | 0.61 (2.5) | 0.81 (3.4) |
| Control Level 2 | 66 | 1.18 (1.8) | 0.20 (0.3) | 0.51 (0.8) | 1.68 (2.5) | 2.13 (3.2) |
| Control Level 3 | 160 | 3.01 (1.9) | 0.00 (0.0) | 1.42 (0.9) | 4.22 (2.6) | 5.37 (3.4) |
| Serum Pool | 13.7 | 0.38 (2.8) | 0.23 (1.6) | 0.02 (0.2) | 0.39 (2.8) | 0.59 (4.3) |
| Serum Pool | 35.7 | 0.70 (1.9) | 0.14 (0.4) | 0.19 (0.5) | 1.02 (2.9) | 1.26 (3.5) |
| Plasma Pool (Li-Hep) | 184.9 | 3.44 (1.9) | 0.00 (0.0) | 1.54 (0.8) | 4.97 (2.7) | 6.24 (3.4) |
| Plasma Pool (EDTA) | 17.5 | 0.51 (2.9) | 0.04 (0.2) | 0.05 (0.3) | 0.47 (2.7) | 0.70 (4.0) |
| Serum Pool | 276.5 | 5.55 (2.0) | 1.49 (0.5) | 1.98 (0.7) | 6.55 (2.4) | 8.93 (3.2) |
| Data was collected for twenty days, one run per day, two replicates per run, on three separate instruments, using three different calibrator lots on each instrument | | | | | | |
{7}
7
b. Linearity/assay reportable range:
i) The linear range was determined according to CLSI EP06-A. Based on the results of linearity range testing and that from the Limit of Detection Study, the analytical measuring range of 1 – 300 U/mL was established in three matrices: serum, Lithium Heparin plasma, and EDTA plasma.
a) Serum
Linearity study of serum was completed over a range that is 20 to 30% wider than the anticipated measuring range for both the full assay range and the low end. The full assay range included serum as diluent, the low-end linearity which extends to 1 U/mL has water as diluent (as no samples were available with CA 15-3 values close to 0). The observed values in each case represent the mean of 3 replicates. Both studies indicate that the assay is linear over the assay range.
The mean concentration value was calculated for each sample. The percent recovery of observed CA 15-3 concentration relative to expected concentration was calculated. The mean recoveries for the three individual samples in the high-end range (serum diluent in serum weighted) shows acceptable linearity up to 366 U/mL using weighted linear regression model assuming constant CV. Weighted linear regression analysis was performed on the observed versus expected CA 15-3 assay results. The regression analysis yielded a slope of 1.005 and a y-intercept of -0.02 U/mL.
For the low end of the assay range, linearity testing was performed on a serum sample with dilutions ranging from 0 – 35 U/mL. All levels meet the acceptance criteria and that the method can be accepted as linear across the tested range; the analysis yielded R²=0.999 with 95% CI for the slope of 0.98 to 0.99 and for the intercept of 0 to 0.
b) Lithium Heparin and EDTA Plasma
Linearity of Lithium Heparin plasma shows acceptable linearity up to 363 U/mL, yielding a slope of 1.00 and a y-intercept of 0.02 U/mL. EDTA plasma demonstrates acceptable linearity up to 360 U/mL using weighted data, yielding a slope of 1.00 and a y-intercept of 0.01 U/mL.
ii) Spiking and Dilution Recovery Studies:
a) Spiking Recovery: Known amounts of CA 15-3 at concentrations between 10.3 and 235.5 U/mL were added to human serum with baseline CA 15-3 values of 4.3 and 4.7 U/mL. The recovered CA 15-3 concentrations on these samples were then measured and the calculated percent recovery ranged from 96.3 to 105.1%.
{8}
8
Value Recovery Following Spiking of Samples
| Serum (4.3 U/mL) | | | |
| --- | --- | --- | --- |
| CA 15-3 Spiked In U/mL | Expected U/mL | Recovered U/mL | % Recovery |
| neat | 4.3 | 4.3 | n/a |
| 10.3 | 14.4 | 14.7 | 102.1 |
| 15.4 | 19.4 | 19.4 | 99.8 |
| 20.6 | 24.5 | 25.2 | 102.9 |
| 30.8 | 34.6 | 35.1 | 101.5 |
| 41.1 | 44.7 | 46.9 | 105.0 |
| 61.7 | 64.9 | 65.0 | 100.2 |
| 82.2 | 85.1 | 89.4 | 105.1 |
| 123.3 | 125.5 | 125.0 | 99.6 |
| 164.4 | 165.8 | 170.3 | 102.7 |
| Serum (4.7 U/mL) | | | |
| --- | --- | --- | --- |
| CA 15-3 Spiked In U/mL | Expected U/mL | Recovered U/mL | % Recovery |
| neat | 4.7 | 4.7 | n/a |
| 33.6 | 37.7 | 36.5 | 96.6 |
| 67.3 | 70.8 | 68.4 | 96.5 |
| 100.8 | 103.8 | 99.9 | 96.3 |
| 134.6 | 136.9 | 134.8 | 98.4 |
| 168.2 | 170.0 | 166.2 | 97.8 |
| 201.9 | 203.1 | 202.1 | 99.5 |
| 235.5 | 236.1 | 236.6 | 100.2 |
b) Dilution recovery: Serum samples were diluted with water according to the dilution levels shown in the table. The % recovery was calculated by dividing the observed result by the expected result and multiplying by 100. The percent recovery ranged from 92 - 102%.
| Sample | Dilution | Expected (U/mL) | Observed (U/mL) | % Recovery |
| --- | --- | --- | --- | --- |
| S1-1 | neat | 245.4 | 245.4 | |
| S1-2 | 1:1.5 | 163.6 | 164.8 | 101% |
| S1-3 | 1:2 | 122.7 | 125 | 102% |
| S1-4 | 1:3 | 81.8 | 81.6 | 100% |
| S1-5 | 1:4 | 61.4 | 62.3 | 102% |
| S1-6 | 1:6 | 40.9 | 41.1 | 100% |
{9}
| Sample | Dilution | Expected (U/mL) | Observed (U/mL) | % Recovery |
| --- | --- | --- | --- | --- |
| S1-7 | 1:8 | 30.7 | 30.9 | 101% |
| S1-8 | 1:12 | 20.5 | 20.5 | 100% |
| S1-9 | 1:16 | 15.3 | 15.2 | 99% |
| S1-10 | 1:24 | 10.2 | 10.2 | 99% |
| | | | | |
| S2-1 | neat | 269.2 | 269.2 | n/a |
| S2-2 | 1:1.14 | 235.5 | 233.4 | 99% |
| S2-3 | 1:1.33 | 201.9 | 197.9 | 98% |
| S2-4 | 1:1.6 | 168.2 | 160.5 | 95% |
| S2-5 | 1:2 | 134.6 | 126.8 | 94% |
| S2-6 | 1:2.67 | 100.9 | 95.8 | 95% |
| S2-7 | 1:4 | 67.3 | 62.1 | 92% |
| S2-8 | 1:8 | 33.6 | 31.1 | 92% |
c) Auto-dilution: A study was completed to compare side-by-side auto-dilution vs. manual-dilution. Samples used were from above assay range of 300 U/mL. The results of the auto-dilution in percent of manual results for the same sample ranged from 97- $102\%$ demonstrating that the results are accurate and precise.
| Sample ID | Dilution Type | dilution (1:10) | | Mean (U/mL) | SD | %CV |
| --- | --- | --- | --- | --- | --- | --- |
| | | run - 1 | run - 2 | | | |
| S31 | Auto | 479.7 | 487.5 | 483.6 | 5.52 | 1.14 |
| S32 | Auto | 2469.0 | 2476.0 | 2472.6 | 4.88 | 0.20 |
| S31 | Manual | 495.0 | 500.0 | 497.5 | 3.54 | 0.71 |
| S32 | Manual | 2499.0 | 2327.0 | 2413.0 | 121.62 | 5.04 |
# iii) High Dose Hook-effect
Hook effect was evaluated using samples prepared by adding commercial CA 15-3 antigen to calibrator matrix. The Dimension Vista® instrument reports an error message "Above Assay range" when the result is above the assay range of 300 U/mL. Users are instructed in the labeling (Limitations of procedure) that samples having this error code should be repeated with dilution. Based on this procedure, the data shows that CA 15-3 method shows no hook effect up to 20,000 U/mL.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# i) Traceability:
There is no recognized standard or reference material for assay calibration. The LOCI 7 calibrator is a liquid, frozen, bovine serum albumin based product containing CA 15-3 from human tissue culture.
{10}
10
ii) Calibrator:
The Dimension Vista® LOCI 7 Calibrator (KC605) is an in-vitro diagnostic product intended to be used to calibrate the Vista® CA 15-3 method. It is a multi-analyte frozen liquid product consists of ten vials, packaged as two vials for each of five levels (A-E), 2.0 mL per vial. The matrix is 6% bovine albumin with buffer and preservatives; level A is zero, while levels B to E contain purified CA 15-3 antigen. Target values for CA 15-3 are 0, 20, 60, 150 and 315 U/mL. The master pool is a frozen liquid five-level material, with composition identical to the calibrator, held in reserve.
a) Value assignment:
An anchor pool of purified CA 15-3 is prepared and a value is assigned by comparing to patient samples assayed by the predicate device. Values are assigned to each lot of calibrator from the master pool using the Dimension Vista® System, through a patient sample comparison to the predicate ADVIA® Centaur CA 15-3 assay.
b) Stability:
Frozen liquid calibrator stability is 12 months from the date of manufacture when stored unopened at -20°C. Thawed liquid calibrator stability is 30 days when stored unopened at 2-8°C. Once the vial stopper is punctured, the assigned calibrator values are stable for 30 days on board the Dimension Vista System. LOCI 7 calibrators are not to be used on board the instrument once the cap is removed.
iii) Kit Stability:
Unopened flexes were stored at 4°C as control condition and on board a Dimension Vista instrument for 15 and 33 days. After 15 days and 33 days, the instrument was calibrated and the recovery of calibrators and QC materials with an on-board stored flex was compared to recovery with a control flex. The losses as a result of storage were within acceptable limits of ±5%, indicating Sealed wells on the instrument are stable for 30 days.
Open vial stability: CA 15-3 flexes were tested for open well stability by using a linear depletion approach. The lot was calibrated on Day 0 and then tested on Day 0, 3, 4, 5, 6 and 7. Two flex well sets were used per sample. All wells were opened on day 0. For each well set, 5 tests were run on Day 0, and 3 on each of the other test days, for a total of 20 tests. Samples were calibrator level A and E, and a serum pool between 30 and 40 U/mL. The testing time for day 7 was at least 2 hours later than on day 0, to ensure the tested interval was wider than the claim of 7 days open-well stability. The losses as a result of storage were within acceptable limits of ±5%, indicating Open wells are stable for 7 days.
d. Detection limit:
CLSI/NCCLS EP 17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation, were followed to determine the lowest concentration of CA 15-3 that can be detected with at least 95% probability; that is, with proportions of false positives (α) less than 5% and false negatives (β) less than 5%; based on 120
{11}
11
determinations, with 5 blank and 5 low level samples. Five CA 15-3 free test samples and five non-zero samples were tested for three days, one run per day, 2 replicates per run, with two reagent (Flex®) lots, on two instruments. Testing was performed using the same lot of calibrator, by the same operator, instrument and operator.
**LOB** – Highest measurement result which has a 95% probability to be observed for a blank sample. It is the 95th percentile of a blank distribution. CA 15-3 negative sera were not available. Therefore, five lots of LOCI-7 calibrator A (0 level calibrator) were tested; the calibrator matrix has a 6% BSA base. The Limit of Blank (LoB) calculated was 0.31 U/mL (claimed 0.3 U/mL).
**LOD** – No serum samples in the concentration range of the expected LoD were available. Five serum samples with concentrations in the range of 3-6 U/mL were diluted between 4-fold and 10-fold with calibrator A (0 level calibrator) to concentrations covering the range where LoD was expected to fall. The Limit of Detection (LoD) for CA 15-3 calculated was 1.0 U/mL.
**LOQ** – Lowest actual amount of analyte that can be reliably detected and at which total error meets lab requirements for accuracy. LOQ was not claimed for the Dimension Vista CA 15-3 assay.
The results are noted to support a claim for LOB of 0.3 U/mL and of LOD of 1.0 U/mL. Therefore, the lowest value for the range of the assay is 1.0 U/mL.
e. Analytical specificity:
Interference testing was performed according to CLSI/NCCLS EP7-A2 to determine the effect of various endogenous and exogenous substances on the Dimension Vista® CA 15-3 assays. For all interferents the percent bias was determined by testing a control sample without the interferent and comparing it to the value obtained from a test sample to which the potential interferent had been added.
i) Endogenous Substance Interference:
Testing was performed at two CA 15-3 concentrations: 30 ± 6 U/mL and 130 ± 25 U/mL using two lots of Flex® reagents. Test samples were prepared by spiking the potential interferent into serum. Bias exceeding 10% was considered interference. Results for hemoglobin, bilirubin and intralipid are summarized below:
| Substance tested | Substance concentration | LOCI CA 15-3 U/mL | Bias % |
| --- | --- | --- | --- |
| Hemoglobin (hemolysate) | Hemoglobin (monomer) 1000 mg/dL [0.62 mmol/L] | 33 | <10 |
| | | 141 | <10 |
| Bilirubin (unconjugated) | 60 mg/dL [1026 μmol/L] | 28 | <10 |
| | | 132 | <10 |
| Bilirubin (conjugated) | 20 mg/dL [342 μmol/L] | 28 | <10 |
| | | 132 | <10 |
| Lipemia (Intralipid®) | 3000 mg/dL [33.9 mmol/L] | 30 | <10 |
| | | 123 | <10 |
{12}
12
Seven additional endogenous substances, including Uric acid, Urea, Creatinine, Cholesterol, IgG and serum proteins, were also tested with bias <10%.
ii) Exogenous Substance Interference:
Fifty two exogenous substances were tested for interference including common over-the-counter drugs and cancer drugs. The substances did not interfere with the LOCI CA 15-3 method when present in serum and plasma (lithium heparin, EDTA). Inaccuracies (biases) due to these substances were less than 10% at CA 15-3 concentrations of 30 ± 6 U/mL and 130±25 U/mL.
iii) HAMA interference:
A number of optimized concentrations of HAMA blockers were employed in the reaction, as well as reagents designed to minimize non specific binding (NSB) interference. Interference from HAMA was evaluated by testing commercially available and in-house sera containing HAMA or heterophilic antibodies. Results from samples to which 10% of a blocker cocktail was added served as controls and were compared to the same samples without blocker cocktail. The blocker cocktail consisted of several commercially available HAMA/ heterophilic blocking substances from multiple vendors. Thirty-six in-house HAMA/NSB sera and 19 commercial HAMA sera were tested by processing two replicates for each sample and control. All samples recovered within ± 10% of their controls.
iv) Method Cross-Reactivity:
Interference/ cross-reactivity of cancer marker antigens in the LOCI CA 15-3 assay was tested by spiking the tumor marker antigens into two sera and comparing the recovery of the spiked sample to control samples spiked with Phosphate Buffered Saline. No cross-reactivity was observed for the tested cancer marker antigens.
| Spiked Antigen | Testing concentration | Serum Sample U/mL | Bias (%) |
| --- | --- | --- | --- |
| CA19-9 | 11,608 U/mL | 28.8 U/mL | -4.0% |
| | | 123.0 U/mL | -2.0% |
| CA125 | 10,466 U/mL | 32.0 U/mL | -3.0% |
| | | 135.3 U/mL | -2.0% |
| AFP | 500 ng/L | 33.3 U/mL | -2.0% |
| | | 140.2 U/mL | 1.0% |
| CEA | 5,000 ng/L | 32.9 U/mL | -1.0% |
| | | 143.6 U/mL | -2.0% |
| PSA | 361 ng/L | 33.8 U/mL | -4.0% |
| | | 140.4 U/mL | 1.0% |
{13}
13
f. Assay cut-off:
For the Dimension Vista® LOCI CA 15-3 assay, the reference change value (RCV) was used to determine if a significant change occurred. The RCV is a percentage change in measured CA 15-3 values when compared to a baseline value. The RCV takes into account the maximum imprecision of the assay within the reportable range, the intra-individual biological variation, and the 95% confidence interval (z value for p<0.05). The RCV was derived by taking into account the published biological variation; within-subject biological variation (6.2%) was obtained from the literature. In determining the RCV, the analytical variation used was 7.9% (which is the reported multi-instrument/ multi-Flex®-lot variability (Multi-Instrument/ Multi-Flex® lot study) at 35.5 U/mL. The RCV for the Dimension Vista® LOCI CA 15-3 method was calculated to be 27.9%.
2. Comparison studies:
a. Method comparison with predicate device:
The predicate device CA 15-3 Assay for the ADVIA Centaur® System was used to demonstrate substantial equivalence to the LOCI CA 15-3 Flex® Reagent Cartridge according to EP9-A2. One serum sample randomly chosen from each of the 75 female patients who were tested for the monitoring of their disease status were combined with 248 excess, de-identified serum samples containing measurable amounts of CA 15-3. Each sample was assayed for the observed values (323 samples) using the Dimension Vista® CA 15-3 (Y) assay and compared with the expected values using the Advia Centaur® CA 15-3 (X).
There were no missing responses or invalid results. Samples above the assay range were re-tested to confirm the results; the repeat tests were not included in any analysis. Using least squares regression, four (4) statistical outliers were identified. These 4 samples and 18 samples that were outside the range of the assay were excluded from the data analysis. Using a Passing Bablok regression, all results were included in the analysis. CA 15-3 values on the Dimension Vista® spanned 4.7 - 4814.1 U/mL across the samples tested; values from 4.7 - 264.4 U/mL (301 samples) were included in the analysis.
| Comparative Method | Slope (95% CI) | Intercept U/mL (95% CI) | Correlation Coefficient | n |
| --- | --- | --- | --- | --- |
| Dimension Vista® CA 15 vs. ADVIA Centaur® CA 15-3 | 0.91 (0.89 to 0.92) | 2.43 (1.53 to 3.33) | 0.988 | 301 |
b. Matrix comparison:
The Dimension Vista Flex Assay for CA 15-3 was evaluated in serum, LiHeparin plasma, and EDTA plasma.
i) Serum samples were evaluated with matched samples of Lithium Heparin plasma. The two matrices were evaluated using 68 samples over the measuring range of the device (linear regression line y=0.98x + 0.4 and R²=1.000).
{14}
ii) Serum samples were evaluated with matched samples of EDTA plasma. The two matrices were evaluated with 82 samples over the measuring range of the device (linear regression line $y = 0.97x + 0.5$ and $R^2 = 0.999$ ).
| Sample comparison | Slope (95% CI) | Intercept U/mL (95% CI) | Correlation Coefficient | n |
| --- | --- | --- | --- | --- |
| Lithium heparin versus serum | 0.98 (0.97 to 0.99) | 0.4 (0 to 0.9) | 1.0 | 68 |
| EDTA versus serum | 0.97 (0.96 to 0.98) | 0.5 (0 to 1.0) | 0.999 | 82 |
# 3. Clinical studies:
Serial samples from 75 female patients were obtained from a supplier of stored retrospective samples to the medical device industry. Samples were selected for age (range 21.8 years old to 75.9 years old), ethnicity and stage of disease (stage 1 through IV). A minimum of 3 serial samples were obtained for each of the 75 patients. Clinical information that details the status of the subject's disease was required for inclusion of samples in the study. Changes in CA 15-3 concentrations and in disease status were analyzed on a per visit basis. Patients were categorized as Active/Progressive, Responding, Stable, or No Evidence of Disease (NED) by the attending physician based on the clinical information (medical imaging, physical examination, and other clinical investigations). All 75 patient sets were analyzed to determine the change in disease status per sequential pair $(n = 258)$ . Table below shows the distribution of results when compared to the disease status
Disease State Frequency using the Dimension Vista® LOCI CA 15-3 Assay
| | Change in Disease State | | | | |
| --- | --- | --- | --- | --- | --- |
| Change in CA 15-3 | Responding n (%) | Stable n (%) | No Evidence of Disease n (%) | Progression n (%) | Total |
| >27.9% increase | 4 (1.5%) | 27 (10.5%) | 3 (1.2%) | 64 (26.8%) | 98 (38%) |
| No significant Change | 10 (3.9%) | 42 (16.3%) | 37 (14.3%) | 17 (6.6%) | 106 (41%) |
| >27.9% decrease | 8 (3.1%) | 28 (10.9%) | 3 (1.2%) | 15 (5.8%) | 54 (21%) |
| Total | 22 (8.5%) | 97 (37.7%) | 43 (16.7%) | 96 (37.2%) | 258 (100%) |
For the Dimension Vista® LOCI CA 15-3 assay, the reference change value (RCV) was used to determine if a significant change occurred. A change of $21\%$ was considered to be significant for the predicate method (AVDIA Centaur® CA 15-3 assay). This value was obtained from the manufacturer's published insert sheet. Per patient visit clinical performance results for the Dimension Vista® LOCI CA 15-3 test and predicate devices are given in following two tables. In this evaluation, disease status was classified as "Progression" and "No Progression" with "No Progression" consisting of responding, stable, and no evidence of disease. Using a cut-off of $>27.9\%$ rise in CA 15-3 value, $66.7\%$ of subject visits ( $95\%$ confidence interval $56.3\% - 76.0\%$ ) had a rise in CA 15-3 value when the patient's disease
{15}
15
status was classified as progression. This value represents the positive percent agreement of significant CA 15-3 rise with a clinical disease status classified as progression. Using a cut-off of a >27.9% rise in CA 15-3 value, 79.0% of subject visits (95% confidence interval 71.9% - 85.0%) had no rise in CA 15-3 value when the patient's disease status was classified as no progression. This value represents the negative percent agreement of CA 15-3 rises with a progressive disease status.
Dimension Vista® LOCI CA 15-3 Value vs. Disease Progression
| | Progression | No-Progression | Total |
| --- | --- | --- | --- |
| >27.9% increase | 64 | 34 | 98 |
| ≤27.9% increase | 32 | 128 | 160 |
| Total | 96 | 162 | 258 |
| | | Estimate | Exact 95% Confidence Interval |
| Total Concordance | | 74.40% | (68.6% - 79.6%) |
| Positive Concordance | | 66.70% | (56.3% - 76.0%) |
| Negative Concordance | | 79.00% | (71.9% - 85.0%) |
The positive and negative agreement for the FLEX CA 15-3 method, when taken together, show similar performance to the predicate method (below)
Predicate CA 15-3 Value vs. Disease Progression
| | Progression | No-Progression | Total |
| --- | --- | --- | --- |
| >21% increase | 67 | 47 | 116 |
| ≤21% increase | 29 | 115 | 142 |
| Total | 96 | 162 | 258 |
| | | Estimate | Exact 95% Confidence Interval |
| Total Concordance | | 70.50% | (64.6% - 76.0%) |
| Positive Concordance | | 69.80% | (59.6% - 78.8%) |
| Negative Concordance | | 71.00% | (63.4% - 77.8%) |
All specimens were analyzed for agreement between the two assays using the Reference Change Value for the Dimension Vista® LOCI CA 15-3 assay and the predicate assay when used according to the manufacturer's instructions; these results are shown in the table below.
{16}
16
# Dimension Vista® LOCI CA 15-3 Concordance to Comparative Method (on a per visit basis)
| Dimension Vista®
LOCI CA 15-3 | Comparative Method | | Total |
| --- | --- | --- | --- |
| | >21% increase | ≤21% increase | |
| >27.9% increase | 94 | 4 | 110 |
| ≤27.9% increase | 20 | 140 | 148 |
| Total | 114 | 144 | 258 |
| | | Estimate | Exact 95%
Confidence Limits |
| Total Concordance | | 90.70% | (86.5% - 94.0%) |
| Positive Concordance | | 82.50% | (74.2% - 88.9%) |
| Negative Concordance | | 97.20% | (93.0% - 99.2%) |
## 4. Clinical cut-off:
Clinical Cut-off is based upon a 27.9% reference change value (RCV). When the measurement of CA 15-3 varies by >27.9% of the baseline measurement (either positive or negative) the change is considered significant. The RCV was used to ensure that the change in CA 15-3 value is not attributed to assay variation or biological variation. This percent variation represents values within the 95% confidence interval for real alteration in CA 15-3 values above system noise.
## 5. Expected values/Reference range:
The distribution of CA 15-3 values determined in specimens from patients with nonmalignant or malignant disease are shown in the two tables below. The distribution of CA 15-3 values was determined in specimens from healthy individuals (n = 350; 150 males and 200 females) and from patients with nonmalignant or malignant diseases. From apparently healthy adult females (age 18-64 years), 99.0% of 200 samples had CA 15-3 levels of less than or equal to 35 U/mL. Each laboratory should establish its own expected values for CA 15-3 as performed on the Dimension Vista® System. Expected values of CA 15-3 measurements are dependent upon the individual patient's baseline reading for CA 15-3. Decreases from baseline value of 27.9% or more in subsequent repeated measurements are indicative of response to therapy or remission. Increases of 27.9% or more suggest no response to therapy and are weakly correlated with progression of disease. Estimation and empirical distributions of CA 15-3 values in various populations of subjects was performed. The distribution of CA 15-3 values in 350 apparently healthy individuals, 200 patients with nonmalignant and 398 patients with malignant disease is shown in the tables below.
{17}
17
Non-malignant Disease and Healthy Normals: # of Patients (% Patients) with CA 15-3 level U/mL
| Sample Category | n | 0-35 (%) | 35.1-60 (%) | 60.1-120 (%) | 120.1-300 (%) | >300 (%) |
| --- | --- | --- | --- | --- | --- | --- |
| Healthy Normal Females | 200 | 198 (99.0) | 2 (1.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Females < 50 yrs age | 151 | 150 (99.3) | 1 (0.7) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Females ≥ 50 yrs age | 49 | 48 (98.0) | 1 (2.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Healthy Males | 150 | 148 (98.7) | 2 (1.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Benign Diseases | | | | | | |
| Breast | 30 | 27 (90) | 2 (6.7) | 1 (3.3) | 0 (0.0) | 0 (0.0) |
| Ovarian Cyst* | 30 | 29 (96.7) | 1 (3.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Cervix / Uterus | 33 | 33 (100.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| GI Tract | 20 | 20 (100.)) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| UTI | 30 | 28 (93.3) | 2 (6.7) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Pancreas | 27 | 22 (81.5) | 5 (18.5) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Colon | 10 | 7 (70.0) | 3 (30) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Congestive Heart Failure | 20 | 16 (80.0) | 3 (15.0) | 1 (5.0) | 0 (0.0) | 0 (0.0) |
| Malignant Diseases: | | | | | | |
| Breast: stage I | 4 | 3 (75.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 1 (25.0) |
| Breast: stage II | 30 | 22 (73.3) | 7 (23.3) | 0 (0.0) | 0 (0.0) | 1 (3.3) |
| Breast: stage III | 12 | 11 (91.7) | 1 (8.3) | 0 (0.0) | 0 (0.0) | 0 (0) |
| Breast: stage IV | 40 | 14 (35.0) | 8 (20.0) | 3 (7.5) | 8 (20.0) | 7 (17.5) |
| Breast: (stage unknown) | 20 | 16 (80.0) | 1 (5.0) | 0 (0) | 2 (10.0) | 1 (5.0) |
| Ovarian | 54 | 27 (50.0) | 14 (25.9) | 8 (14.8) | 3 (5.6) | 2 (3.7) |
| Cervix / Uterus | 41 | 33 (80.5) | 7 (17.1) | 0 (0.0) | 0 (0.0) | 1 (2.4) |
| Liver* | 23 | 21 (91.3) | 1 (4.3) | 0 (0.0) | 0 (0.0) | 1 (4.3) |
| Colorectal | 41 | 38 (92.7) | 3 (7.3) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Pancreatic | 30 | 23 (76.7) | 6 (20.0) | 1 (3.3) | 0 (0.0) | 0 (0.0) |
| Lung | 30 | 22 (73.3) | 6 (20.0) | 1 (3.3) | 0 (0.0) | 1 (3.3) |
| Lymphoma | 30 | 25 (83.3) | 5 (16.7) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Prostate/Testicle | 28 | 26 (92.9) | 2 (7.1) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Upper GI* | 13 | 12 (92.3) | 1 (7.7) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
| Renal | 2 | 1 (50.0) | 1 (50.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
Liver* = Gall Bladder, Bile Duct
Upper GI* = Oral, Larynx, Esophagus, Stomach
Ovarian Cyst* = Polycystic Ovaries/Ovarian Cyst
{18}
18
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.