MERIT MANIFOLD WITH INTEGRATED TRANSDUCER, MODEL 203LTCHN-R; 503LTCHN-R; 203LTPHN-R,203LTCHWN-R; 503LTCHWN-R; 203L

K100319 · Merit Medical Systems, Inc. · DTL · Apr 28, 2010 · Cardiovascular

Device Facts

Record IDK100319
Device NameMERIT MANIFOLD WITH INTEGRATED TRANSDUCER, MODEL 203LTCHN-R; 503LTCHN-R; 203LTPHN-R,203LTCHWN-R; 503LTCHWN-R; 203L
ApplicantMerit Medical Systems, Inc.
Product CodeDTL · Cardiovascular
Decision DateApr 28, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4290
Device ClassClass 2

Indications for Use

The Merit Manifold with Integrated Transducer is used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices. The device is also used for measurement of blood pressure.

Device Story

The Merit Manifold with Integrated Transducer (M-T) is a sterile, single-use device combining a 3-port fluid manifold with a pressure transducer. It serves as a fluid pathway for cardiovascular diagnostic, surgical, and therapeutic procedures while simultaneously measuring physiological blood pressure. The device is operated by clinicians in cardiovascular settings. It connects to external tubing and catheters; a reusable interface cable transmits the pressure signal from the transducer to a standard pressure monitor. By integrating two previously stand-alone products into one, the device reduces set-up time and simplifies the clinical workflow. The output is displayed on a monitor, allowing healthcare providers to assess real-time hemodynamic status and make informed clinical decisions during procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation studies were conducted per ISO 10993-1, ISO 10993-7, ISO 11135, and ANSI/AAMI BP 22. Additional testing included device function, transducer/compensation line verification, and bond verification. All results met acceptance criteria.

Technological Characteristics

Polycarbonate manifold body; Acetal (Delrin) valves; Polycarbonate transducer package; Urethane (Meth) Acrylate blend (Dymax 20163) UV adhesive. 3-port manifold configuration. Disposable, sterile, single-use. Connects to reusable interface cable for signal transmission. Sterilized via ethylene oxide. Complies with ANSI/AAMI BP 22:1994[R]2006.

Indications for Use

Indicated for use in cardiovascular diagnostic, surgical, and therapeutic applications for interconnecting tubing, catheters, or other devices and for measuring blood pressure in patients requiring such monitoring.

Regulatory Classification

Identification

A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K100319 # 510(k) SUMMARY #### GENERAL INFORMATION Merit Medical Systems, Inc. Submitter Name: 1600 Merit Parkway Address: South Jordan. Utah 84095 Telephone Number: 801-208-4123 (direct) 801-826-4108 (direct) Fax Number: Jeff Carlstrom Contact Person: Date of Preparation: February 2, 2010 DEVICE INFORMATION Trade or Proprietary Name: Merit Manifold with Integrated Transducer Fluid Manifold; Pressure Transducer Common or Usual Name: Product Code: DTL, DRS Classification Name: Disposable fluid manifold (21 CFR 870.4290) Extravascular blood pressure transducer (21 CFR 870.2850) Classification Panel: Cardiovascular # PREDICATE DEVICE(S) - Merit Manifold (K913161) . - Merit Meritrans Disposable Transducer (K920977) . - Navilyst Perceptor Compensator Morse Manifold (K951722) [combined intended use] . #### DEVICE DESCRIPTION The Merit Manifold with Integrated Transducer (referred to as M-7 in this submission) is a disposable pressure transducer bonded to a 3-port manifold. The Merit M-T device is a sterile, precalibrated, single-use device used for both a means of interconnecting tubing, catheters, or other devices, and for physiological pressure measurement. A separate reusable interface cable is used with this system to transmit the signal from the transducer to a pressure monitor. The M-T device combines two stand-alone Merit products into a single device, for ease of use and reduction of set-up time. #### INTENDED USE The Merit Manifold with Integrated Transducer is used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other device is also used for measurement of blood pressure. APR 2 8 2010 {1}------------------------------------------------ # TECHNOLOGICAL COMPARISON Manifolds and transducers are manually operated devices that provide a user-controlled fluid pathway. The manifold body is composed of polycarbonate, with valves made of Acetal (Delrin''). The transducer package, also made of polycarbonate, is bonded to the manifold using Urethane (Meth) Acrylate blend (Dymax 20163) UV adhesive. The technological characteristics of the Merit M-T device are identical to those of the Merit predicate devices in terms of intended use, clinical utility and mode of operation, user population, basic design, materials, performance, and sterilization method. ## NON-CLINICAL PERFORMANCE TESTING Verification and validation studies were conducted in accordance with in-house protocols to mitigate risks identified in the clinical risk assessment conducted by Merit. Performance testing was conducted or evaluated based on the following FDA Guidance and industry standards: - ISO 10993-1: 2003: Biological Evaluation of Medical Devices Part 1: Evaluation and . Testing - FDA Bluebook memorandum G95-1: Required Biocompatibility Training and Toxicology . Profiles for Evaluation of Medical Devices, May, 1, 1995 - ISO 10993-7: 1995: Biological Evaluation of Medical Devices Part 7: Ethylene Oxide . Sterilization residuals - ISO 11135: 1994: Medical Devices Validation and routine control of ethylene oxide . sterilization - ANSI/AAMI BP 22: 1994[R]2006: Blood Pressure Transducers . Additional Testing not covered by these standards: - Device Function Verification � - Transducer and Compensation Line Verification . - Bond Verification . Results of performance testing met the acceptance criteria and demonstrate substantial equivalence to the predicate devices. ## SUMMARY OF SUBSTANTIAL EQUIVALENCE Based on CDRH's substantial equivalence decision tree, the Merit Manifold with Integrated Pressure Transducer is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of an eagle or bird-like figure, with stylized lines representing its wings and body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 APR 2 8 2010 Merit Medical Systems, Inc. c/o Mr. Jeff Carlstrom Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, UT 84095 Re: K100319 > Trade/Device Name: Merit Manifold with Integrated Transducer Regulatory Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary bypass adaptor, stopcocks, manifold, or fitting Regulatory Class: II (two) Product Code: DTL, DRS Dated: February 02, 2010 Received: April 23, 2010 Dear Mr. Carlstrom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Jeff Carlstrom Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Duna R. Vahner A Bram D. Zuckerman, M.D. Director > · Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Section 5.0 # INDICATIONS FOR USE STATEMENT K100319 510(k) Number (if known): Merit Manifold with Integrated Transducer Indications for Use: Device Name: The Merit Manifold with Integrated Transducer is used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices. The device is also used for measurement of blood pressure. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R. Vaches (Division Sign-Off) Division of Cardiovascular Devices KI00319 510(k) Number Traditional 510(k) Premarket Notification Manifold with Integrated Transducer Merit Medical Systems, Inc 0025
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...