← Product Code [LLZ](/productcode/LLZ) · K100292

# CAAS QVA 3D, CAAS QCA 3D (K100292)

_Pie Medical Imaging BV · LLZ · Apr 30, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100292

## Device Facts

- **Applicant:** Pie Medical Imaging BV
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Apr 30, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

CAAS QxA 3D is used as follows: - 3D Reconstruction of coronary arteries and peripheral vessels from a set of angiographic X-ray images; - Calculation of dimensions of the vessel corrected for out-of-plane magnification and foreshortening errors; - To determine acquisition parameters for optimal imaging of part of interest on a vessel tree; - To assist in the positioning of implantable devices in the vessel segment of interest. CAAS QxA 3D software is designed for use in clinical practice to support diagnoses and interventional treatment of cardiovascular conditions. The software is used by or under supervision of a cardiologist or radiologist.

## Device Story

CAAS QxA 3D is a software system for 3D reconstruction and analysis of vascular structures. Input consists of multiple 2D angiographic X-ray images from different projections. The system performs automatic contour detection on selected arterial segments in 2D images; these contours are used to generate a 3D reconstruction of the vessel. The software calculates vessel dimensions (area, diameter, length), reconstructs healthy vessel shapes, and performs subsegment analysis. Results are corrected for out-of-plane magnification and foreshortening. The device provides a virtual display of implantable devices to assist in positioning. Used in clinical settings by cardiologists or radiologists to support diagnosis and interventional planning. Benefits include objective, accurate, and reproducible vessel assessment and improved procedural planning.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing and validation of the software's performance in vessel reconstruction and measurement accuracy.

## Technological Characteristics

Software-based system for 3D vascular reconstruction and analysis. Operates on angiographic X-ray image inputs. Features include automatic contour detection, 3D reconstruction, and geometric measurement (diameter, area, length). Corrects for magnification and foreshortening. Designed for clinical use in cardiovascular diagnosis and intervention.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- CAAS ([K052988](/device/K052988.md))
- CAAS QCA 3D ([K063344](/device/K063344.md))
- IC-Pro ([K083745](/device/K083745.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

(510)(63) number: K5100292

510(k) Summary

#### CAAS QxA 3D

[QA376] v.2.0

APR 3 0 2010

K100292

Submitter Name Submitter Address Contact Person: Phone Number Fax Number Email Address Preparation Date

Pie Medical Imaging BV Becanusstraat 13 D; 6216 BX Maastricht, The Netherlands Saskia Lloyd, Quality Assurance Officer +31 43 32 81 328 +31 43 32 81 329 Saskia.lloyd@pie.nl 17 March 2010

Cardiovascular Angiography Analysis System

Class II (21 CFR, part 892.2050, LLZ)

CAAS QxA 3D

CAAS QCA 3D CAAS QVA 3D

Trade Name Common Name

Regulation Class

Predicate Devices

CAAS, cleared for marketing under K052988 CAAS QCA 3D, cleared for marketing under K063344 IC-Pro, cleared under K083745

Device Description

The CAAS QxA 3D covers the QCA 3D and QVA 3D software modules intended to run on the Cardiovascular Angiography Analysis System (CAAS). The CAAS QxA-3D is designed for objective, accurate and reproducible assessment of the vessel geometry from a set of angiographic X-ray images from different projections. The variant QCA 3D is intended for coronaries and the variant QVA 3D is intended for peripheral vessels.

On each of the 2D images an arterial segment is selected resulting in automatic contour detection. The detected 2D arterial contours in each image are used to generate a 3D reconstruction of the arterial segment. A number of analysis results can be calculated.

- Vessel dimensions such as area, diameter and length; 1.
Reconstruction of healthy vessel shape; 2.

User defined subsegments analysis. 3.

Results are corrected for out-of-plane magnification and foreshortening errors. CAAS QxA 3D features a virtual display of an implantable device on the 2D images and assists to plan its position based on the 3D reconstruction.

Intended Use

CAAS QxA 3D is used as follows:

- 3D Reconstruction of coronary arteries and peripheral vessels from a set of 】. angiographic X-ray images;
- Calculation of dimensions of the vessel corrected for out-of-plane 2. magnification and foreshortening errors;
- To determine acquisition parameters for optimal imaging of part of interest on 3. a vessel tree;
- To assist in the positioning of implantable devices in the vessel segment of 4. interest.

CAAS QxA 3D software is designed for use in clinical practice to support diagnoses and interventional treatment of cardiovascular conditions. The software is used by or under supervision of a cardiologist or radiologist.

•

{1}------------------------------------------------

CAAS QxA 3D is developed, tested, validated and produced under the same Performance Data Quality Assurance system applicable to the development and production of products currently marketed by Pie Medical Imaging. Substantial Equivalence The intended use and technological characteristics of CAAS QxA 3D are substantial equivalent to a combination of the intended use and technological characteristics of the predicate devices. The testing reported in this 510(k) establishes that CAAS QxA 3D is substantial Conclusion equivalent to a combination of predicate devices and is safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design with three stylized wing segments. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002

# APR 30 2010

Mr. Saskia Lloyd Submission Correspondent-Quality Assurance Officer Pie Medical Imaging b.v. Becanusstraat 13D, 6216 BX Maastricht THE NETHERLANDS

Re: K100292

Trade/Device Name: CAAS QxA 3D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 30, 2010 Received: April 5, 2010

Dear Mr. Lloyd:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter. will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours, -

Donald J. Trump

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

### Enclosure

{4}------------------------------------------------

## 510(k) Notification - CAAS QxA 3D

#### INDICATION FOR USE STATEMENT

page 1 of 1

K100292 510(k) number (if known): ___

CAAS QxA 3D Device Name:

Indications For Use:

CAAS QxA 3D is used for:

- 3D Reconstruction of coronary arteries and peripheral vessels from a set of angiographic X-ray 1. images;
- Calculation of dimensions of the vessel corrected for out-of-plane magnification and 2. foreshortening errors;
- To determine acquisition parameters for optimal imaging of part of interest on a vessel tree; 3.
- To assist in the positioning of implantable devices in the vessel segment of interest. 4.

The CAAS QxA 3D software is designed for use in clinical practice to support diagnoses and interventional treatment of cardiovascular conditions. The software is used by or under supervision of a cardiologist or radiologist.

Indications for use: CAAS QCA 3D: coronary arteries CAAS QVA 3D: peripheral vessels

Prescription Use OR (Per 21 CFR 801.109)

Over-The-Counter Use_

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

concurrence of CDRH, Office of Device Evaluation (ODE)

Robert Becker

(Division Sign-Off)

Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K100292

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K100292](https://fda.innolitics.com/device/K100292)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
