The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician.
Device Story
Actiwave is a compact, battery-operated, ambulatory physiological signal recorder. It captures physiological waveform voltages via skin-surface gel electrodes and records activity counts via an internal accelerometer (Cardio model). The device logs digital data to internal storage for later transfer to a PC; it performs no onboard data analysis or real-time viewing. Used in clinical or research settings, it is worn by patients during normal daily activities. Clinicians retrieve the stored data from the device to a computer for subsequent review and decision-making. The device provides a portable, unobtrusive method for long-term physiological monitoring, enabling data collection outside of traditional medical environments.
Clinical Evidence
Bench testing only. Performance evaluated against Special Controls for ambulatory electrocardiographs, including input dynamic range, input impedance, gain stability, system noise, multichannel crosstalk, frequency response, timing accuracy, and temporal alignment. Electrical safety and electromagnetic interference testing performed per international standards.
Technological Characteristics
Polycarbonate/ABS case; stainless steel snap or no-touch connectors; 3.0V lithium rechargeable battery. Sensing via gel-type electrodes (voltage) and accelerometer (activity). 1, 2, or 4-channel configurations. Dimensions range from 19x27x9 mm to 37x27x9 mm; weight 5-10g. Data stored digitally for PC transfer. No onboard analysis or display.
Indications for Use
Indicated for patients requiring ambulatory recording of physiological waveforms and activity. Suitable for use in clinical settings or scientific research under clinician supervision; patients may perform normal daily activities while wearing the device.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
SEER Light Compact Digital Holter Recorder (K021470)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Premarket Notification Actiwave™M
#### 5.0 510(K) SUMMARY
## APR 2 9 2010
#### 5.1 Summary information
#### 5.1.1 Submitter's name and address
Howard Smith, Engineering Director CamNtech, UK, Ltd. Upper Pendrill Court Ermine Street North Papworth Everard Cambridge, United Kingdom CB233UY Tel: 011-44-1480-831223 011-44-1480-831733 Fax:
Date summary was prepared: 01/25/2010
#### 5.1.2 Name of device
| Trade Name: | ActiwaveTM |
|----------------------|-----------------------------------------|
| Common Name: | Digital Ambulatory Monitor |
| Classification Name: | Electrocardiograph, Ambulatory (Without |
| Analysis) | |
| Product Regulation: | 21 CFR 870.2800 |
Product Code:
MWJ
#### 5.1.3 Identification of predicate device
SEER Light Compact Digital Holter Recorder, 510(k) Number : K021470
{1}------------------------------------------------
K100266 P2/4
510(k) Premarket Notification Actiwave™M
#### 5.2 Device description
#### 5.2.1.1 Functions of the device
Activave is a compact, ambulatory, battery-operated physiological signal recorder. Actiwave is commonly known as an Ambulatory Electrocardiograph. Because of its small size and battery operation, Actiwave may be used in a clinical or non-clinical setting to record physiological signal waveforms while the patient is unattended. The device obtains its signal from the skin surface through the use of standard gel-type signal electrodes. Actiwave may be worn during daily activities and is fully ambulatory due to its small size and its battery-operated power source. Recorded data may be transferred later to a personal computer for storage and archiving. Actiwave does not provide any data viewing or data analysis functions.
#### 5.2.1.2 Basic scientific concepts
The Actiwave device acquires and logs digital data whose values represent the amplitude-varying physiological waveform signal voltages at progressive time intervals. The scientific concepts and technologies that are used to sense the signals are summarized in Table 5.1.
| Physiological Parameter | Recorder Used | Technology | Value Obtained |
|-------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Physiological waveform | Actiwave1-, 2-, and 4-<br>channel recorders and<br>Actiwave Cardio | Amplifier senses and<br>amplifies voltage signal. | Digital value<br>corresponding to<br>instantaneous signal<br>voltage. |
| Activity | Actiwave Cardio | Signal from accelerometer<br>amplifier is digitized and<br>recorded. | Digital value<br>corresponding to Activity<br>Counts or acceleration. |
# TABLE 5.1: BASIC TECHNOLOGIES USED FOR PHYSIOLOGICAL SIGNAL RECORDING IN ACTIWAVE RECORDERS
{2}------------------------------------------------
510(k) Premarket Notification Actiwave™e™
#### 5.2.1.3 Physical characteristics of Recorder
Pertinent physical characteristics of the Actiwave Recorder are shown in Table 5.2.
# TABLE 5.2: PHYSICAL CHARACTERISTICS OF ACTIWAVE
### RECORDER
| Parameter | Value | Condition/Note | |
|------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------|
| Size | Actiwave 1-channel: 19x27x9 mm<br>Actiwave 2-channel: 25x27x9 mm<br>Actiwave 4-channel: 37x27x9 mm<br>Actiwave Cardio: 32diam x 10 mm | Outer dimensions | |
| Weight | 1-channel: 5 gm<br>2-channel: 6 gm<br>4-channel: 9 gm<br>Cardio: 10 gm | With no electrodes<br>attached | |
| | Case material | Polycarbonate/ABS | |
| | Attachment type | Stainless steel snap connector or<br>standard no-touch connectors | |
| | Battery type | 3.0 volt lithium rechargeable | Not user replaceable |
| Indicators | None | | |
#### 5.3 Statement of the intended use of the device
The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder having 1, 2, or 4 channels of physiological waveform storage. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the alphanumeric strings "K100266" and "P4/4". The strings are written in a handwritten style with black ink on a white background. The strings appear to be some kind of identifier or code.
510(k) Premarket Notification Actiwave™
### Technological characteristics of the device compared to predicate 5.4 device
Actiwave and the SEER Light Compact Digital Holter Recorder (FDA 510(k) Number: K021470) are each physiological recording systems based upon the concept of an ambulatory, unattended physiological recorder that logs physiological data to the recorder. Each of these devices is a solid-state recorder with electronic amplifiers, data collection means, and with the ability to store data until it is transferred to a personal computer. Actiwave and the SEER Light Compact Digital Holter Recorder are of similar mechanical materials, construction, size, and human interface characteristics. Activave and the SEER Light Compact Digital Holter Recorder are of similar electronic design and their operational characteristics are functionally equivalent.
#### 5.5 Assessment of non-clinical performance data and safety testing
The performance of the Actiwave Recorder has been tested in accordance with a Special Control specific to this device type. The categories of tests applied to this device are entirely contained within the Special Control. These categories are: Input dynamic range, Input Impedance, Gain Stability, System noise, Multichannel crosstalk, Frequency response, Timing accuracy, and Temporal alignment. For all the categories, we have determined that Actiwave performance meets the test criteria in the Special Control. In addition to performance testing, the device has been tested for electrical safety and electromagnetic interference according to internationally recognized standards. The testing according to these standards has raised no issues as to the safety and effectiveness of the present device or the present device compared to the predicate device.
#### 5.6 Conclusion
The results of the tests conducted in accordance with the Special Control demonstrate that the Actiwave is as safe, as effective, and performs as well as the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling a stylized bird or human figure, with flowing lines forming the shape of a wing or hair.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
CamNtech, UK, Ltd. c/o Mr. Howard Smith Engineering Director Upper Pendrill Court, Ermine Street North Papworth Everard, Cambridge, UK CB233UY
APR 2 9 2010
Re: K100266
Device Name: Actiwave™ Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory Electrocardiograph (Without Analysis) Regulatory Class: Class II (Two) Product Code: MWJ Dated: January 25, 2010 Received: January 29, 2010
Dear Mr. Smith:
. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. Howard Smith
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRF/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Brom D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### 1.0 INDICATIONS FOR USE STATEMENT
## Indications for Use
510(k) Number (if known): K100266
Device Name: Actiwave™
Indications For Use:
The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician.
Prescription Use AND/OR -Over-The-Counter Use (Part 21 CFR 801 Subpart D) . (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K100266
Page 4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.