The Quintex Cervical Plating System is intended for the immobilization and stabilization of the spine as an adjunct to fusion in patients with: - Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), - Spondylolisthesis, - Trauma (i.e. fracture or dislocation), - Spinal Stenosis, - Deformity (i.e., scoliosis, kyphosis, and/or lordosis), - Tumors. - Pseudoarthrosis as a result of failed spine surgery, - Failed previous fusions, - Symptomatic cervical spondylosis - Instability following surgery for the above indications. Levels of anterior cervical intervertebral body screw fixation for this indication are from C2-T1. Warning: This device is not approved or intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
Device Story
Quintex Cervical Plating System consists of dynamic and semi-constrained plates, bone screws, and associated surgical instruments. Used by surgeons via anterior approach to the cervical spine (C2-T1) to provide fixation and stabilization as an adjunct to spinal fusion. Device components are manufactured from titanium/titanium alloy. System provides mechanical support to the vertebral body to facilitate fusion in patients with degenerative, traumatic, or neoplastic conditions. Output is physical stabilization of the spinal segment. Benefits include restoration of spinal stability and support for bone healing.
Clinical Evidence
Bench testing only. Biomechanical testing performed per FDA guidance for spinal systems to demonstrate safety and effectiveness comparable to predicate devices.
Technological Characteristics
Materials: Titanium/Titanium Alloy. Components: Dynamic and semi-constrained plates, screws, and instruments. Energy source: None (mechanical). Connectivity: None. Sterilization: Provided non-sterile. Form factor: Anterior cervical intervertebral body fixation system.
Indications for Use
Indicated for patients requiring spinal immobilization and stabilization as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, deformity (scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, failed previous fusions, symptomatic cervical spondylosis, and post-surgical instability. Levels C2-T1. Contraindicated for posterior element (pedicle) fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
K100243
Page 1 of 1
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
Quintex Cervical Plating System January 26, 2010
SEP - 2 2010
COMPANY: Aesculap Implant Systems, Inc. 3773 Corporate Parkwav Center Valley, PA 18034 Establishment Registration Number: 3005673311
CONTACT: Lisa M. Boyle 800-258-1946 (phone) 610-791-6882 (fax)
TRADE NAME: Quintex Cervical Plating System
COMMON NAME: Anterior Cervical Screw Spinal Fixation System
CLASSIFICATION NAME: Spinal Intervertebral Body Fixation Orthosis (87KWQ)
REGULATION NUMBER: 888.3060
# SUBSTANTIAL EQUIVALENCE
Aesculap Implant Systems (AIS), Inc., believes that the Quintex Cervical Plating System is substantially equivalent to the AIS ABC Cervical Plating Systems (K050813/K040017 /K000486/K974706), AIS Caspar Plate & Screw System (K953720/K936269/K913730) and the AIS Spectrum Cervical Spinal System (K022997/K050804).
# DEVICE DESCRIPTION
The Aesculap Implant Systems (AIS) Quintex Cervical Plating System consists of dynamic and semi-constrained plate, screws and associated instruments. Fixation is provided by bone screws inserted into the vertebral body of the cervical spine using an anterior approach. The AIS 3G Cervical Plating System is manufactured from Titanium/ Titanium Alloy and will be provided non-sterile.
## INDICATIONS FOR USE
The Quintex Cervical Plating System is intended for the treatment of cervical spinal instability resulting from:
- Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with 0 degeneration of the disc confirmed by history and radiographic studies),
- Spondylolisthesis, ●
- Trauma (i.e. fracture or dislocation), 0
- Spinal Stenosis,
- Deformity (i.e., scoliosis, kyphosis, and/or lordosis),
{1}------------------------------------------------
- . Tumors and instability associated with major reconstructive surgery related to tumor excision,
- Pseudoarthrosis as a result of failed spine surgery, .
- . Failed previous fusions.
- Symptomatic cervical spondylosis, .
- . Instability following surgery for the above indications.
Levels of anterior cervical intervertebral body screw fixation for this indication are from C2-T1.
Warning: This device is not approved or intended for screw attachment of fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
# TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The AIS Quintex Cervical Plating System is considered substantially equivalent to other legally marketed predicate systems. Biomechanical testing of the subject device was found to be similar in performance to previously cleared spinal systems with similar indications.
## PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where applicable. In addition, testing per the "Spinal System 510(k)s" was completed where relevant. Testing results demonstrate the AIS Quintex Cervical Plating System is safe and effective comparable to other predicate systems currently on the market.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in the center. The bird is composed of three curved lines, suggesting movement or flight. Encircling the bird is text, which appears to be part of the logo. The text is arranged in a circular fashion, following the contour of the bird and adding to the overall design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room --WO66-G609 Silver Spring, MD 20993-0002
SEP - 2 19 1
Aesculap Implant Systems, LLC % Ms. Lisa M. Boyle Sr. Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K100243
Trade/Device Name: Quintex Cervical Plating System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: August 30, 2010 Received: August 31, 2010
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Lisa M. Boyle
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Barbara Buchen
Mark N. Mell Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
#### INDICATIONS FOR USE STATEMENT A.
### SEP = 2 200
510(k) Number:_
## Device Name: Quintex Cervical Plating System
### Indications for Use:
The Quintex Cervical Plating System is intended for the immobilization and stabilization of the spine as an adjunct to fusion in patients with:
- . Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
- Spondylolisthesis,
- Trauma (i.e. fracture or dislocation), ●
- Spinal Stenosis,
- . Deformity (i.e., scoliosis, kyphosis, and/or lordosis),
- . Tumors.
- Pseudoarthrosis as a result of failed spine surgery,
- Failed previous fusions, .
- Symptomatic cervical spondylosis
- . Instability following surgery for the above indications.
Levels of anterior cervical intervertebral body screw fixation for this indication are from C2-T1.
Warning: This device is not approved or intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
Prescription Use X
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
K100243 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.