CONTIPLEX D CONTINUOUS NERVE BLOCK SET, MODEL 331687, 331686, 331685, 331680
Applicant
B.Braun Medical, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Apr 16, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The B. Braun Contiplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate, by stimulation, peripheral nerves by transferring electrical impulses from a nerve stimulator. The needle is used to inject and facilitate the continuous administration of local anesthetics or analgesics to the targeted nerve bundle for general and orthopedic surgery. The Contiplex D Continuous Nerve Block Set is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management during the pre-operative, perioperative and post-operative periods associated with general and orthopedic surgery. The Contiplex D polyamide catheter may remain indwelling for up to 72 hours. The B. Braun Stimuplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate by stimulation peripheral nerves by transferring electrical impulses from a nerve stimulator. The needle is used to inject a single dose of anesthetic or analgesic fluid to the targeted nerve bundle for general and orthopedic surgery. The B. Braun Stimuplex® D Plus Insulated Echogenic Needle is intended for use in regional anesthesia and pain therapy to target peripheral nerves by transferring electrical impulses from a nerve stimulator and by visualization of an echogenic reflective pattern at the needle tip using an ultrasound imaging device. The needle is used to inject a single dose of anesthetic or analgesic fluid to the targeted nerve bundle for general and orthopedic surgery.
Device Story
Insulated needles for peripheral nerve localization; used in regional anesthesia/pain therapy. Inputs: electrical impulses from external nerve stimulator; ultrasound waves (Stimuplex D Plus only). Operation: dielectric coating/hub connection transfers electrical current to needle tip for nerve stimulation; etched markings on Stimuplex D Plus reflect ultrasound for visualization. Output: needle tip position relative to nerve; injection of anesthetic/analgesic fluid via integrated tubing. Used in OR/clinic by physicians. Contiplex D includes catheter for continuous infusion (up to 72 hours). Benefits: improved nerve targeting accuracy via dual-modality (electrical/ultrasound) or electrical stimulation, facilitating effective anesthesia delivery.
Clinical Evidence
Bench testing only. Biocompatibility testing performed to address material differences. Performance testing verified design inputs against ISO standards (ISO 9626, ISO 7864, ISO 594-2). Comparative ultrasound evaluation performed for Stimuplex D Plus vs. EchoStim.
Technological Characteristics
Insulated stainless steel needles with dielectric coating; hub includes electrical connection and injection tubing. Stimuplex D Plus features etched tip markings for echogenicity. Polyamide open-tip catheters (Contiplex D). Conical 6% (Luer) lock fittings per ISO 594-2. Sterilized for single use.
Indications for Use
Indicated for patients undergoing general or orthopedic surgery requiring regional anesthesia or pain therapy via peripheral nerve block. Used for single-shot (Stimuplex D/D Plus) or continuous (Contiplex D) administration of local anesthetics/analgesics.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
#### 5. 510(k) SUMMARY
April 13, 2010
APR 1 0 2010
KLOOZ41
#### SUBMITTER:
B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500
Contact: Lisa Giaquinto, Sr. Analyst, Regulatory Affairs Phone: (610) 596-2354 Fax: (610) 266-4962 E-mail: lisa.giaquinto@bbraun.com
DEVICE NAME:
Contiplex® D Insulated Needles Stimuplex® D Insulated Needles Stimuplex® D Plus Insulated Echogenic Needles
#### COMMON OR USUAL NAME:
Needles, Conduction, Anesthetic (W/Wo Introducer)
DEVICE CLASSIFICATION:
Class II, Product Code BSP 868.5150
PREDICATE DEVICES:
Contiplex® Continuous Nerve Block Set, B. Braun Medical, Inc., K090995, Class II, CAZ, 868.5140 (containing Contiplex Tuohy and Straight Needle components, Class II, BSP 868.5150)
EchoStim® Facet Tip, Havel, Inc., K063380, Class II, BSP, 868.5150
DESCRIPTION:
The Contiplex® D Insulated Needles are intended for locating by stimulation, peripheral nerves for the administration of anesthetic or analgesic fluid to the targeted nerve. The Contiplex D needle includes an introducer catheter over the needle to facilitate the advancement of an indwelling catheter. The dielectric needle coating and electrical connection within the hub facilitate the transfer of electrical impulses from a nerve stimulator to the tip of the needle. The needle hubs include an aspiration or injection tubing line in order to facilitate the initial injection of anesthetic or analgesic fluid as well as a connecting cable, which can be attached to the cables of any of the B. Braun Nerve Stimulators listed on the product labeling. The Contiplex D needle/catheter assemblies will be available in two lengths, gauges and bevel geometries.
{1}------------------------------------------------
> The Contiplex D Insulated Needles will be available in a set configuration containing one Contiplex D introducer catheter over insulated nerve stimulating needle, one polyamide open-tip catheter with threading assist guide, and one clamp style catheter connector. After needle removal, the short introducer catheter is used to thread a longer open-tip polyamide catheter into the patient. The polyamide catheter and catheter connector are used to administer the continuous flow of anesthetic or analgesic fluid to the patient for up to 72 hours.
> Stimuplex® D Insulated Needles are individually The packaged insulated needles, intended for the localization and stimulation of peripheral nerve bundles for "single shot" anesthesia. The Stimuplex D needle hubs, like the Contiplex D needles, include an aspiration or injection tubing line in order to facilitate the injection of anesthetic or analgesic fluid. The dielectric needle coating and electrical connection within the hub facilitate the transfer of electrical impulses from a nerve stimulator to the tip of the needle. The Stimuplex D needles can be used with any of the B. Braun Nerve Stimulators listed on the product labeling. The needles will be available in various lengths and gauge sizes with two bevel angles for some sizes.
> The Stimuplex® D Plus Insulated Echogenic Needles are designed to provide the clinician with two modalities to locate peripheral nerves for nerve block procedures; through electrical stimulation and ultrasound imaging. The dielectric coating and electrical connection within the hub facilitate the transfer of electrical impulses from a nerve stimulator to the tip of the needle. The Stimuplex D Plus needles can be used with any of the B. Braun Nerve Stimulators listed on the product labeling. In addition, the needles are designed with etched markings to help reflect an image of the needle when using an ultrasound imaging machine. Once the nerve is located, the Stimuplex D Plus needles include a tubing line in order to permit the 'single shot' injection of local anesthetic fluid to the nerve. The needles will be available in various lengths and gauge sizes.
#### INTENDED USE:
The B. Braun Contiplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate, by stimulation, peripheral nerves by transferring electrical impulses from a nerve stimulator. The needle is used to inject and facilitate the continuous administration of local anesthetics or analgesics to the targeted nerve bundle for general and orthopedic surgery.
The Contiplex D Continuous Nerve Block Set is intended to provide continuous and/or intermittent infusion of local
{2}------------------------------------------------
> anesthetics and analgesics for peripheral plexus anesthesia and pain management during the pre-operative, perioperative and post-operative periods associated with general and orthopedic surgery. The Contiplex D polyamide catheter may remain indwelling for up to 72 hours.
> The B. Braun Stimuplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate by stimulation peripheral nerves by transferring electrical impulses from a nerve stimulator. The needle is used to inject a single dose of anesthetic or analgesic fluid to the targeted nerve bundle for general and orthopedic surgery.
> The B. Braun Stimuplex® D' Plus Insulated Echogenic Needle is intended for use in regional anesthesia and pain therapy to target peripheral nerves by transferring electrical impulses from t a nerve stimulator and by visualization of an echogenic reflective pattern at the needle tip using an ultrasound imaging device. The needle is used to inject a single dose of anesthetic or analgesic fluid to the targeted nerve bundle for general and orthopedic surgery.
The Contiplex® D Insulated Needles, Stimuplex® D Insulated Needles and Stimuplex® D Plus Insulated Echogenic Needles have the same intended use and are similar in design to the Contiplex Tuohy and Straight needles included in the Contiplex® Continuous Nerve Block Set (K090995). The proposed devices and predicate device incorporate insulated stimulating needles to locate by stimulation, targeted nerve bundles, in order to perform peripheral nerve block procedures. Biocompatibility testing was completed to address any differences in materials and to verify that no new issues of safety have been introduced with the proposed devices. Additionally, performance testing was completed to verify that all design input requirements of the proposed devices have been met. The following performance standards were considered in evaluating the performance of the proposed devices: ISO 9626:1991/Amd. 1:2001(E) "Stainless steel needle tubing for the manufacture of medical devices." ISO 7864:1993(E) "Sterile hypodermic needles for single use." ISO 594-2:1998 "Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock fittings." The results of biocompatibility and performance testing demonstrate the proposed devices raise no new issues of safety or effectiveness when compared to the predicate device.
The Stimuplex D Plus Insulated Echogenic Needles also have similar indications and technological characteristics as Havel Inc.'s EchoStim® needles. Both the proposed needles and predicate needles are indicated for locating by stimulation,
### SUBSTANTIAL EQUIVALENCE:
{3}------------------------------------------------
> targeted nerve bundles, in order to perform peripheral nerve block procedures. In addition, both the Stimuplex D Plus needles and the EchoStim® needles incorporate etched markings near the needle tip, which help the needles to be visualized under ultrasound. A comparative evaluation of the needles under ultrasound was completed to demonstrate there are no differences between the proposed and predicate device, which raise new issues of safety or effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Lisa Giaquinto Senior Analyst, Regulatory Affairs B. Braun Medical, Incorporated 901 Marcon Boulevard Allentown, Pennsylvania 18109-9341
Re: K100241
Trade/Device Name: Stimuplex D Plus Insulated Echogenic Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: January 21, 2010 Received: January 27, 2010
APR 1 6 2010
Dear Ms. Giaquinto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, nowever, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Ms. Giaquinto
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nh for
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
January 20, 2010
#### INDICATIONS FOR USE STATEMENT
Page of
510(k) Number (if known): _______________
Device Names: Stimuplex® D Plus Insulated Echogenic Needle
#### Indications For Use:
The B. Braun Stimuplex® D Plus Insulated Echogenic Needle is intended for use in regional anesthesia and pain therapy to target peripheral nerves by transferring electrical impulses from a nerve stimulator and for visualization of an echogenic reflective pattern at the needle tip using an ultrasound imaging device. The needle is used to inject a single dose of local anesthetic or analgesic to the targeted nerve bundle for general and orthopedic surgery.
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
18
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital · fection Control, Dental Devices
1100241 J10(k) Number: _
{7}------------------------------------------------
January 20, 2010
#### INDICATIONS FOR USE STATEMENT
Page l
510(k) Number (if known): _
Device Names: Stimuplex® D Insulated Needle
Indications For Use:
The B. Braun Stimuplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate by stimulation peripheral nerves by transferring electrical impulses from a nerve stimulator. The needle is used to inject a single dose of local anesthetic or analgesic to the targeted nerve bundle for general and orthopedic surgery.
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K100241 510(k) Number:
{8}------------------------------------------------
# 4. INDICATIONS FOR USE STATEMENT
| Page | 1 of 1 |
|------|--------|
|------|--------|
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
## Device Names: Contiplex® D Insulated Needles
#### Indications For Use:
The B. Braun Contiplex® D Insulated Needle is intended for use in regional anesthesia and pain therapy to locate, by stimulation, peripheral nerves by transferring electrical impulses from a therapy to tocate, by summation, perspiect and facilitate the continuous administration of local anesthetics or analgesics to the targeted nerve bundle for general and orthopedic surgery.
The Continuous D Continuous Nerve Block Set is intended to provide continuous and/or I he Countines D Continuous Tre re and analgesics for peripheral plexus anesthesia and paint management during the pre-operative, perioperative and post-operative periods associated with general and orthopedic surgery. The Contiplex D polyamide indwelling catheter may remain indwelling for up to 72 hours.
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use _
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulthe
(Division Sign-Üir ; Division of Anestosibiligy Coneral Hospital Infection Control, Dental Bonces
5.10(k) Number: K100241
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.