← Product Code [DQY](/productcode/DQY) · K100184

# FUSTAR STEERABLE INTRODUCERS (K100184)

_Lifetech Scientific (Shenzhen) Co., Ltd. · DQY · Feb 14, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100184

## Device Facts

- **Applicant:** Lifetech Scientific (Shenzhen) Co., Ltd.
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Feb 14, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

FuStarTM Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

## Device Story

FuStar Steerable Introducer functions as a guiding or introducer sheath to facilitate percutaneous access. Device features low-profile design, atraumatic tip, and manual deflection mechanism. Physician operates device to navigate vasculature; visible knob with scale on controller allows estimation of sheath deflecting angle (0 to 180 degrees). Curl options available in 30mm (small) and 50mm (large) radii. Connector options include luer lock or silicone valve. Device provides pathway for various catheters and instruments; assists in precise positioning within cardiac or peripheral vessels. Benefits include reduced risk of trauma during puncture and improved control during interventional procedures.

## Clinical Evidence

Bench testing only. Performance testing included primary dimension, X-ray visibility, bending, fatigue (cycle), conical fitting, hub, connection strength, corrosion resistance, and leakage/aspiration tests. Biocompatibility testing included cytotoxicity, acute systemic toxicity, intracutaneous reactivity, hypersensitivity, hemolysis, complement system, thrombosis, pyrogen, and endotoxicity tests. Sterilization and package integrity were also validated.

## Technological Characteristics

Steerable introducer sheath; deflectable tip (0-180 degrees); manual deflection via controller knob with scale. Available in 5F-10F diameters and 550mm-900mm lengths. Connector types: luer lock or silicone valve. Materials: unspecified medical-grade polymers/metals. Sterilization: validated. Connectivity: none (mechanical device).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- AMPLATZER Torqvue Delivery System ([K072313](/device/K072313.md))
- Enpath Medical Steerable Sheath ([K061119](/device/K061119.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Premarket Notification Section 510(k) Submission
 FuStar Steerable Introducers
 Exhibit #5 510(k) Summary
 Ref No.: A2009-018-075

Image /page/0/Picture/1 description: The image shows the logo for SesrCare Lifetech Scientific. The logo is in black and white, with the word "SesrCare" in a bold, sans-serif font. The letters are connected, giving the logo a sense of unity. The letters "TM" are in superscript to the right of the word "SesrCare". Below the word "SesrCare" is the phrase "Lifetech Scientific" in a smaller, sans-serif font.

1

FEB 1 4 2011

## Exhibit #6 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

## The 510(k) Number is K100184

| Date of Submission: | 02 SEPT 2010                                                                                                                                                                                                                                                                                                                                                             |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:            | Lifetech Scientific (Shenzhen) Co., Ltd<br>Floor 1-3, Cybio Electronic Building, Langshan 2nd Street, North area of High-tech Park,<br>Nanshan District, Shenzhen, Guangdong, 518057, P. R. China<br>Contact Person:<br>Ms. Xiaoli Shi, International Regulatory / Mr. David Zhang, CTO<br>TEL:86-13418601356   FAX:(86-755)86026251<br>Email:stepbystep0326@hotmail.com |
| Correspondent:      | Ms. Diana Hong / Mr. Lee Fu<br>Shanghai Mid-Link Business Consulting Co., Ltd<br>Suite 5D, No.19, Lane 999,Zhongshan Road (S-2),Shanghai, 200030, China<br>TEL: +86-21-22815850   FAX: +86-21-22815850                                                                                                                                                                   |
| Proposed Device     | FuStar™ Steerable Introducers                                                                                                                                                                                                                                                                                                                                            |
| Classification:     | Class II<br>Product Code: DQY<br>Classification Name: catheter, percutaneous<br>Regulation Number: 21 CFR 870.1250<br>Review Panel: Cardiovascular                                                                                                                                                                                                                       |
| Predicate Device:   | Predicate Device 1:<br>510(k) Number: K072313<br>Trade Name: AMPLATZER Torqvue Delivery System<br>Manufacturer: AGA Medical Corporation<br><br>Predicate Device 2:<br>510(k) Number: K061119<br>Trade Name: Enpath Medical Steerable Sheath                                                                                                                              |

{1}------------------------------------------------

Premarket Notification Section 510(k) Submission FuStar Steerable Introducers Exhibit #5 510(k) Summary Ref No .: A2009-018-075

Image /page/1/Picture/1 description: The image shows the logo for "SesrCare Lifetech Scientific". The word "SesrCare" is written in a bold, sans-serif font, with the letters connected. The letters are black. The words "Lifetech Scientific" are written in a smaller, thinner font below "SesrCare".

Intended Use: FuStarTM Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

Device Description: The FuStarTM Steerable Introducers are designed to perform as a guiding sheath and/or introducer sheath, which provides a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature. Its low-profile design and atraumatic tip reduce the potential risk for trauma during percutaneous punctures while facilitating passage of a wide variety of catheters and devices.

> The proposed introducer can give physicians good control during operation, the curl options are available at 30mm (Small) and 50mm (Large). In each option, there are 2 connector types which are luer lock connector and silicone valve connector, and several specifications combination. Also, the visible knob with scale of controller facilitates the physicians to estimate the deflecting angle of the sheath while intra-operation.

> The products are available in different effective length (550mm, 700mm and 900mm) and diameter (French size, 5F, 6F, 7F, 8F, 9F and 10F). The deflectable angle of sheath is from 0 degree to 180 degree.

Testing Conclusion: Performance testing was conducted to validate and verify that the proposed device met all design specifications including:

> Performance Test: Primary Dimension Test; Exterior Surface Condition X-Ray Visible Testing; Bending Test, Cycle (Fatigue) Test: Conical Fitting Test: Hub Test; Connection Strength Test: Corrosion Resistance test, Leakage and Aspiration Test,

Biocompatibility Test: Cytotoxicity, Acute Systemic Toxicity, Intracutaneous Reactivity Test, Hypersensitivity Test, Hemolysis Test, Complement System Test. Thrombosis Test Sterilization: Sterliization Validation, Package Integrity Test, Pyrogen Test and Endotoxicity Test.

- SE Conclusion: FuStar™ Steerable Introducers, can be claimed to be Substantially Equivalent (SE) to the predicate device, Predicate Device AMPLATZER Torqvue Delivery System K072313 and Enpath Medical Steerable Sheath, K061119.
2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Lifetech Scientific (Shenzhen) Co., Ltd C/O Ms. Diana Hong Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D. No.19, Lane 999, Zhongshan Road (S-2), Shanghai, 200030, China

FEB 1 4 201

K100184 Trade/Device Name: FuStar™ Steerable Introducers Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: January 18, 2011 Received: January 19, 2011

Dear Ms. Hong:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

1:00 PM IST STATE

{3}------------------------------------------------

Page 2 - Ms. Diana Hong

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?rgblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Dunn P. Viener

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Premarket Notification Section 510(k) Submission FuStar Steerable Introducers Section II Indication for Use Statement · Ref No.: A2009-018-075

Image /page/4/Picture/1 description: The image shows the logo for "SesrCare Lifetech Scientific". The word "SesrCare" is in a bold, sans-serif font, with the "TM" symbol in superscript next to it. Below "SesrCare" is the text "Lifetech Scientific" in a smaller, lighter font. The logo appears to be for a company involved in life sciences or scientific research.

Section II Indication for Use Statement

510(k) Number: K106184

Device Name: FuStar

TM

Steerable Introducers

Indications for Use:

FuStar™ Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

humph R. Vebus

sion Sign-Off) (Division Sign-Oll)
Division of Cardiovascular Devices

510(k) Number k100184

II-1

---

**Source:** [https://fda.innolitics.com/device/K100184](https://fda.innolitics.com/device/K100184)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
