← Product Code [DZL](/productcode/DZL) · K100182

# SALVIN DENTAL SPECIALTIES FIXATION SCREW (K100182)

_Salvin Dental Specialties, Inc. · DZL · May 10, 2010 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K100182

## Device Facts

- **Applicant:** Salvin Dental Specialties, Inc.
- **Product Code:** [DZL](/productcode/DZL.md)
- **Decision Date:** May 10, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4880
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Salvin Bone Tack System is designed to stabilize a barrier membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial, or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.

## Device Story

The Salvin Bone Tack System consists of 3mm and 5mm titanium (Ti-6Al-4V) bone tacks and associated instrumentation. The system is used by clinicians to stabilize barrier membranes onto cortical bone in craniofacial, maxillofacial, or mandibular regions. The device is provided non-sterile. It functions as a mechanical fixation system to hold membranes in place during bone healing or grafting procedures. The clinician manually applies the tacks using the provided instruments. The system benefits patients by ensuring membrane stability, which is critical for successful bone regeneration.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing, including engineering review and inspection of actual parts.

## Technological Characteristics

Materials: Titanium (Ti-6Al-4V). Components: 3mm and 5mm bone tacks and associated instrumentation. Sterilization: Provided non-sterile. Energy source: None (manual).

## Regulatory Identification

An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.

## Predicate Devices

- IMTEC Bone Tac ([K973180](/device/K973180.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K100182

Image /page/0/Picture/1 description: The image shows the logo for "SALVIN Dental Specialties". The word "SALVIN" is in large, bold, black letters, with a registered trademark symbol to the right of the "N". Below "SALVIN" is the phrase "Dental Specialties" in a smaller, sans-serif font, set against a black background.

MAY 1 0 2010

## Salvin Dental Specialties, Inc 3450 Latrobe Drive · Charlotte, NC 28211 · Phone 704-442-5400 · Fax 704-442-5424

## 510(k) Summary

A 510(k) Owner Salvin Dental Specialties, Inc 3450 Latrobe Drive Charlotte, NC 28211

Contact

Robert Salvin CEO Salvin Dental Specialties, Inc. 3450 Latrobe Drive Charlotte, NC 28211 (704) 442-5400 (704) 442-5424 bobsalvin@salvin.com

Preparation Date

B

C

December 22, 2009

Trade Name Salvin Dental Specialties, Inc. Bone Tack System

Membrane Fixation System

Screw, Fixation, Intraosseous

K973180 – IMTEC Bone Tac

Common Name

Classification Name

Predicate Device(s)

D Device Description The Salvin Bone Tack System consists of 3mm and 5mm Titanium (Ti-6Al-4V) bone tacks and associated instrumentation.

The devices are provided non-sterile.

(21 CFR 872.4880, Product code DZL)

E Intended Use The Salvin Bone Tack System is designed to stabilize a barrier membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial, or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.

{1}------------------------------------------------

| F | Technological<br>Characteristics | As was established in this submission, the subject device<br>is substantially equivalent to other devices cleared by the<br>agency for commercial distribution in the United States.                                                                                                                                                                                   |
|---|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|   |                                  | Engineering drawings, labeling, and detailed inspectional<br>analysis have demonstrated that the subject device is<br>substantially equivalent, if not identical, to its predicate<br>devices in terms of design, materials of composition,<br>indications for use, and such other characteristics as may<br>be associated with the manufacture of any medical device. |
| G | Non-Clinical Testing             | Analysis based on engineering review and inspection of<br>actual parts                                                                                                                                                                                                                                                                                                 |
| H | Clinical Testing                 | Not applicable to this device                                                                                                                                                                                                                                                                                                                                          |
| I | Conclusions                      | Based on the 510(k) Summary and the information<br>provided herein, we conclude that the Salvin Bone Tack<br>System is substantially equivalent to the existing legally<br>marketed devices under the Federal Food, Drug and<br>Cosmetic Act.                                                                                                                          |

1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -

..

. .

. . . . .

.

。

:

:

.

 t 1

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 0 2010

Mr. Robert Salvin Chief Executive Office Salvin Dental Specialties, Incorporated 3450 Latrobe Drive Charlotte, North Carolina 28211

Re: K100182

Trade/Device Name: Salvin Dental Specialties Fixation Screw Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw or Wire Regulatory Class: II Product Code: DZL Dated: April 2, 2010 Received: April 30, 2010

Dear Mr. Salvin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Salvin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucml 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Susan Runyon

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): 1100182

Device Name: Salvin Dental Specialties Fixation Screw

Indications for Use:

The Salvin Bone Tack System is designed to stabilize a barrier membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial, or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kei Mulaylian MSR
(Division Sign-Off)

Divisigont Anasthasiology GegarahHospital Intection Gootrol, Degran Dayiges

Page of

510(k) Number: K100182

---

**Source:** [https://fda.innolitics.com/device/K100182](https://fda.innolitics.com/device/K100182)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
