The ZZOMA Positional Sleeper is indicated for use and intended for professional use for the treatment of mild to moderate, predominantly positional obstructive sleep apnea (OSA) and to reduce or alleviate snoring. The ZZOMA Positional Sleeper is intended for use by professional healthcare personnel trained in its use. Caution: Federal law restricts the device to sale by or on the order of a physician or dentist.
Device Story
ZZOMA Positional Sleeper is a wearable, non-electronic, mechanical device designed to prevent supine sleeping. It consists of a semi-rigid synthetic foam wedge contained in a backpack-style harness with an adjustable Velcro elastic belt. The device is worn on the back; the wedge-shaped design physically keeps the user in a lateral (side) position during sleep. It is used in a clinical or home setting under the guidance of professional healthcare personnel. By preventing the supine position, the device reduces airway obstruction in patients with positional OSA, thereby alleviating snoring and improving sleep quality. It serves as an alternative to CPAP therapy for positional OSA patients.
Clinical Evidence
Clinical study compared ZZOMA Positional Sleeper to CPAP therapy in patients with positional OSA. Study findings: 1) Positional therapy is equivalent to CPAP in normalizing the apnea-hypopnea index (AHI) in positional OSA patients; 2) Decreased AHI by >30%; 3) Similar to CPAP regarding effects on sleep quality and nocturnal oxygenation; 4) Effective at maintaining patients in the non-supine position throughout the night.
Technological Characteristics
Mechanical positional therapy device. Materials: semi-rigid synthetic foam core, nylon outer cover, cotton-covered body contact surface with PVC dots for stability. Dimensions: 12 x 5.5 x 4 inches. Form factor: backpack-style harness with adjustable Velcro elastic belt. Non-electronic; no energy source. No software or connectivity.
Indications for Use
Indicated for patients with mild to moderate, predominantly positional obstructive sleep apnea (OSA) and for the reduction or alleviation of snoring. Intended for professional use by trained healthcare personnel.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
K100/60
# Exhibit 1 510(k) Summary
- 1 510(k) Summary
MAY 1 1 2010
Date of Summary Preparation: May 7, 2010
| 1.2 | Submitter: | Joseph Crocetti, DO<br>Sleep Specialists, LLC<br>P.O. Box 610<br>Abington, PA 19001<br>Phone: (610) 304-6408<br>E-mail: info@zzomasleep.com |
|-----|------------|---------------------------------------------------------------------------------------------------------------------------------------------|
|-----|------------|---------------------------------------------------------------------------------------------------------------------------------------------|
1.3 Trade Name: ZZOMA Positional Sleeper
Classification Name, Product Code, Class, Classification Reference: 1.4
| Classification Name | Common Name | Product Code | Class | 21CFR § |
|---------------------|-------------|--------------|-------|----------|
| | ZZOMA | MYB | II | 872.5570 |
| | Positional | | | |
| | Sleeper | | | |
1.5 Standards/Special Controls:
None
- 1.6 Indications for Use:
The ZZOMA Positional Sleeper is indicated for use and intended for professional use for the treatment of mild to moderate obstructive sleep apnea (OSA) and to reduce or alleviate snoring.
The ZZOMA Positional Sleeper is intended for use by professional healthcare personnel trained in its use.
#### 1.7 Device Description:
The ZZOMA Positional Sleeper is 12 x 5.5 x 4 inches in size and made of lightweight semi-rigid synthetic foam (Figures 1 and 2). It is contained in a backpack type material with an associated Velcro® elastic belt. The device is worn on the back, with the elastic belts brought around each side of the subject and secured anteriorly with the adjustable straps. The particular size and wedge-shaped design on both sides of this device keeps the subject comfortably positioned on their side, and prevents him/her from assuming the supine position. The ZZOMA has a firm inner core made of foam and the outer part of the device is covered in nylon, and the part that touches the subject's body is cotton covered with a coating of PVC dots that help keep the ZZOMA in place while you sleep.
Substantially Equivalent Commercially Available Devices: · 1.8
The ZZOMA Positional Sleeper device is substantially equivalent to the predicate device described herein with respect to indications for use:
· Sleep Devices, Inc. Sona Pillow - K040161
Aeolus International Corporation, Sniff Position Pillows/Popitz Pillows - K023010
The predicate devices (Sona Pillow and Sniff Position Biollows/Popitz Pillows) are commercially available and marketed Class II devices.
{1}------------------------------------------------
#### 1.9 Substantial Equivalence Comparison:
·
l
| ZZOMA Positional Sleeper | Predicate Sleep Devices,<br>Inc, Sona Pillow –<br>K040161 | Equivalency |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| The ZZOMA Positional Sleeper is<br>indicated for use and intended for<br>non-professional, over-the-counter<br>use and for professional use for the<br>treatment of mild to moderate<br>obstructive sleep apnea (OSA) and to<br>reduce or alleviate snoring. | May stop or decrease snoring<br>May be used to treat mild<br>obstructive sleep apnea<br>May improve the quality of sleep | Same - Treats<br>mild obstructive<br>sleep apnea<br>Same –<br>regarding<br>effects on<br>snoring<br>Different – does |
| | | not disrupt<br>sleep quality<br>Different – |
| | | treats both mild<br>and moderate<br>sleep apnea |
# INDICATIONS FOR USE
# METHODOLOGY
| ZZOMA Positional Sleeper | Predicate | Equivalency |
|--------------------------|--------------------------------|-------------------------------------------------------------------------------|
| Positions on the back | Positions on the head and neck | Same - both<br>devices prevent<br>subjects from<br>sleeping on<br>their backs |
| | | Different -<br>Pillow positions<br>head, Zzoma<br>positions back |
## PERFORMANCE
| ltem | ZZOMA Positional<br>Sleeper | Predicate | Equivalency |
|-------------|-----------------------------------------|-----------------------------------------|-------------|
| Storage | Room temperature | Room temperature | Same |
| Positioning | Keep subject off back while<br>sleeping | Keep subject off back<br>while sleeping | Same |
A clinical study was conducted to examine the non-inferiority of the Zzoma Positional Sleeper and CPAP therapy on the apnea-hypopnea index after one night of therapy in patients with positional obstructive sleep apnea.
{2}------------------------------------------------
i A secondary objective was to evaluate the ability of the Zzoma Positional Sleeper to maintain the patient in the lateral position was to evaluate the ability of in objective is to evaluate the initial effects of the in the lateral position during sloop. Thistier observer, and compare it to initial effects seen with CPAP therapy.
There were 4 major findings in this study: 1) in patients with positional OSA, positional therapy is There were 4 major in this study. 7/11 pations with in addition to decreasing the AHI by > equivalent to CPAP therapy at nomalizing the Arr o - regards to effects on sleep quality and nocturnal 30%, 2) positional meraly is siniliar to Or X. " trerapy in 70garane AHI in patients with positional oxygenation, of there is milliffer the at maintaining patients in the non-supine position throughout the night.
# 1.10 Indications and Contraindications:
Relative indications and contraindications for the ZZOMA Positional Sleeper and positional commercially relative indications and contrained uses (treating moderate positional sleep apena) are similar.
1.11 Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Sleep Specialists, LLC concludes that the new information provided in the pro manect nosffective and substantially equivalent to the predicate devices as described herein.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding a stylized symbol. The symbol is a representation of a human figure embracing or protecting another person, conveying a sense of care and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
# MAY 1 1 2010
Sleep Specialists, LLC c/o Mr. Howard Mann 1 Congressional Drive, Apt C Greenville, Delaware 19807
Re: K100160
Trade/Device Name: ZZOMA Positional Sleeper Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: MYB Dated: March 29, 2010 Received: April 6, 2010
Dear Mr. Mann:
-
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. Howard Mann
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Rua. Hien, M.D.
Malvina B. Eydelman, M Director Division of Ophthalmic, Neuroloical, and Ear, Nose, Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### EXHIBIT B Indications for Use Statement
Indications for Use .
### 510(k) K100160:
Device Name: ZZOMA Positional Sleeper
Indications for Use:
The ZZOMA Positional Sleeper is indicated for use and intended for professional use for the treatment of mild to moderate, predominantly positional obstructive sleep apnea (OSA) and to reduce or alleviate snoring.
The ZZOMA Positional Sleeper is intended for use by professional healthcare personnel trained in its use.
Caution: Federal law restricts the device to sale by or on the order of a physician or dentist.
AND/OR Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anatolet
Page 1 of
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number k100160
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.