FIXCET SPINAL FACET SCREW SYSTEM

K100154 · X-Spine Systems, Inc. · MRW · May 21, 2010 · OR

Device Facts

Record IDK100154
Device NameFIXCET SPINAL FACET SCREW SYSTEM
ApplicantX-Spine Systems, Inc.
Product CodeMRW · OR
Decision DateMay 21, 2010
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The X-spine Systems, Inc. Fixcet Spinal Facet Screw system is intended for posterior fixation to the lumbar spine (L1 to S1 inclusive). The system is intended for bilateral, transfacet fixation of the facet joint in order to provide stability for fusion. The system.is indicated for posterior surgical treatment of any or all of the following at the L1 to S1 (inclusive) spinal levels: - Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc . as confirmed by patient history and radiographic studies), - . Degenerative disease of the facets with pain and/or instability on plain flexion and extension lateral radiographs where there is movement of the vertebral bodies relative to each other of more than 4mm, - . Trauma (i.e., fractures and/or dislocations), - . Spondylolisthesis, - . Spondylolysis, - . Pseudoarthrosis and/or failed previous fusions.

Device Story

Fixcet Spinal Facet Screw System provides bilateral, transfacet fixation for lumbar spine fusion; enhances spinal stability. System consists of 4.5mm diameter self-tapping titanium alloy (Ti6Al4V) bone screws; available in single-threaded (25-45mm) and dual-threaded (30-45mm) configurations. Used by surgeons in clinical/OR settings to transfix facet articular processes. Implants provided non-sterile; require steam autoclave sterilization by healthcare professional prior to use. Mechanical performance supports spinal fusion; provides structural stability to vertebral bodies.

Clinical Evidence

Bench testing only. No clinical data provided. Biomechanical testing included static and fatigue compression bending (ASTM F2193), torsional yield strength, insertion/removal torque, pullout force, and self-tapping performance (ASTM F543).

Technological Characteristics

Materials: Titanium alloy (Ti6Al4V) per ASTM F136. Dimensions: 4.5mm diameter; lengths 25-45mm. Configuration: Single or dual-threaded bone screws. Sterilization: Steam autoclave (user-performed). Mechanical testing: ASTM F2193, ASTM F543.

Indications for Use

Indicated for patients requiring posterior lumbar (L1-S1) stabilization for degenerative disc disease, facet joint degeneration with instability, trauma (fractures/dislocations), spondylolisthesis, spondylolysis, or pseudoarthrosis/failed fusion.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 100154 # MAY 2 1 2010 # 510(k) SUMMARY # ADMINISTRATIVE INFORMATION | Manufacturer Name: | X-spine Systems, Inc.<br>452 Alexandersville Rd.<br>Miamisburg, OH 45342 | |-------------------------|--------------------------------------------------------------------------| | | Telephone (937) 847-8400<br>FAX (937) 847-8410 | | Official Contact: | David Kirschman, M.D.<br>Chief Medical Officer | | Date Prepared: | January 15, 2010 (revised April 21, 2010) | | DEVICE NAME | | | Trade/Proprietary Name: | Fixcet™ Spinal Facet Screw System | | Classification Name: | System, Facet Screw Spinal Device | | Common Name: | Facet Screw | # ESTABLISHMENT REGISTRATION NUMBER X-spine Systems, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903. ## DEVICE CLASSIFICATION FDA has classified facet screw spinal devices as Unclassified. The product code for System, Facet Screw Spinal Device is MRW. These device classifications are reviewed by the Orthopedic Devices Branch. ### INTENDED USE The X-spine Systems, Inc. Fixcet Spinal Facet Screw system is intended for posterior fixation to the lumbar spine (L1 to S1 inclusive). The system is intended for bilateral, transfacet fixation of the facet joint in order to provide stability for fusion. The system.is indicated for posterior surgical treatment of any or all of the following at the L1 to S1 (inclusive) spinal levels: page 1 of 3 {2}------------------------------------------------ - Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc . as confirmed by patient history and radiographic studies), - . Degenerative disease of the facets with pain and/or instability on plain flexion and extension lateral radiographs where there is movement of the vertebral bodies relative to each other of more than 4mm, - . Trauma (i.e., fractures and/or dislocations), - . Spondylolisthesis, - . Spondylolysis, - . Pseudoarthrosis and/or failed previous fusions. # DEVICE DESCRIPTION The X-spine Fixcet Spinal Facet Screw System is designed to provide bilateral, transfacet fixation of the spinal facet joint in the lumbar spine. This system consists of titanium alloy bone screws designed to transfix the facet articular process in the spine to enhance spinal fusion and stability. The self-tapping 4.5mm diameter facet screws are available in two configurations: (1) single-threaded screws offered in lengths of 25-45mm (in 5 mm increments); and (2) dual-threaded screws offered in lengths of 30-45mm (in 5 mm increments). The screws are composed of medical grade Titanium alloy (Ti6Al4V) that complies with ASTM F-136. The implant components are provided clean and non-sterilized by a healthcare professional using Steam Autoclave in accordance to the instructions for use provided by X-spine Systems Inc., as well as the instructions provided by the manufacturer of the Autoclave. ### EQUIVALENCE TO MARKETED PRODUCT X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the Fixcet Spinal Facet Screw system is substantially equivalent to the Depuy Spine Discovery System (K012773) and the SpineFrontier Chameleon System (K071420) based on a comparison of the following: - . FDA Product Code - . Intended Uses - Surgical Approach ● - . Levels of Fixation - Implant Materials . - � Product Dimensions - . Mechanical Performance Page 2 of 3 {3}------------------------------------------------ ## PERFORMANCE DATA The implant components were tested using the following standards: ASTM F2193 – Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System - Static Compression Bending . - Fatigue Compression Bending . ASTM F543 - Standard Specification and Test Methods for Metallic Medical Bone Screws - . Torsional Yield Strength - . Insertion Torque - Removal Torque . - . Pullout Force - � Self-Tapping Performance In conclusion, biomechanical testing results indicate that the Fixcet Spinal Facet Screw System is substantially equivalent to predicate device performance and is capable of performing in accordance with its intended use. Page 3 of 3 {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings and a flowing design below, possibly representing water or movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 # MAY 2 1 2010 X-Spine Systems, Inc. % David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342 Re: K100154 Trade/Device Name: Fixcet™ Spinal Facet Screw System Regulation Number: Unclassified Product Code: MRW Dated: April 21, 2010 Received: May 19, 2010 Dear Dr. Kirschman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - David Kirschman, M.D. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Soubare buchanD Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K100154 Device Name: Fixcet™ Spinal Facet Screw System Indications for Use: The X-spine Systems, Inc. Fixcet Spinal Facet Screw System is intended for posterior fixation to the lumbar spine (L1 to S1 inclusive). The system is intended for bilateral, transfacet fixation of the facet joint in order to provide stability for fusion. The system is indicated for posterior surgical treatment of any or all of the following at the L1 to S1 (inclusive) spinal levels: - . Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), - Degenerative disease of the facets with pain and/or instability on plain flexion and extension lateral . radiographs where there is movement of the vertebral bodies relative to each other of more than 4mm, - . Trauma (i.e., fractures and/or dislocations), - . Spondylolisthesis, - . Spondylolysis, - . Pseudoarthrosis and/or failed previous fusions. Prescription Use AND/OR × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number_ K100154 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...