K100141 · Siemens Medical Solutions USA, Inc. · MOS · Aug 27, 2010 · Radiology
Device Facts
Record ID
K100141
Device Name
14-CHANNEL EXTREMITY COIL FOR MAGNETOM ESSENZA
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 27, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of the new 14-Channel Extremity Coil is, in conjunction with THE 1.5T MAGNETOM ESSENZA Magnetic Resonance Scanner, the MR examination of the human knee, foot, ankle, hand and wrist. Used in the MAGNETOM ESSENZA, the 14-Channel Extremity Coil is indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal and oblique cross sectional images that display the internal structure and/or function of the knee, foot, ankle, hand and wrist. The images produced by the MAGNETOM ESSENZA with the, 14-Channel Extremity Coil reflects the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The intended use of the MAGNETOM ESSENZA is not affected in any way by the use of the new 14-Channel Extremity Coil.
Device Story
14-Channel Extremity Coil accessory for 1.5T MAGNETOM ESSENZA MRI scanner; receives radiofrequency signals from human knee, foot, ankle, hand, or wrist; transforms signals into transverse, sagittal, coronal, and oblique cross-sectional images; used in clinical radiology settings by trained physicians; output images display internal structure/function based on proton spatial distribution; aids physician diagnosis; provides high-resolution diagnostic imaging of extremities.
Clinical Evidence
Bench testing only. Performance parameters including Signal to Noise Ratio (SNR), image uniformity, and Specific Absorption Rate (SAR) were tested and compared to the predicate device, demonstrating equivalence.
Technological Characteristics
14-channel phased-array extremity coil; designed for 1.5T MAGNETOM ESSENZA MRI; RF sensing principle; materials identical to predicate; no new biocompatibility concerns; performance metrics include SNR, image uniformity, and SAR.
Indications for Use
Indicated for use with 1.5T MAGNETOM ESSENZA MR scanner for imaging human knee, foot, ankle, hand, and wrist to produce cross-sectional images for diagnostic purposes by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K100141
ge 1/3
**SIEMENS**
## Healthcare
AUG 2 7 2010
# Attachment B - 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
## I. General Information
| Establishment | Siemens Medical SolutionsUSA, Inc.<br>51 Valley Stream Parkway<br>Malvern, PA 19355 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 2240869 |
| Manufacturer | Siemens Mindit Magnetic Resonance Ltd.<br>Siemens MRI Center<br>Gaoxin C. Ave. 2nd<br>Hi-Tech Industrial Park,<br>ShenZhen 518057, PR. China |
| Registration Number | 3004754211 |
| Contact | Kim Rendon<br>Manager, Regulatory/Clinical Affairs<br>51 Valley Stream Parkway, MS G01<br>Malvern, PA 19355<br>Phone: (610) 717-8085<br>Fax: (610) 448-1787<br>E-mail address: kimberly.rendon@siemens.com |
| Device Name | Trade Name: 14-Channel Extremity Coil for<br>1.5T MAGNETOM ESSENZA<br>Classification Name: Coil, Magnetic Resonance Specialty<br>Device Class: Class II 21 CFR § 892.1000<br>Product Code: MOS<br>Classification Panel: Radiology |
| Performance Standards | None established under Section 514 the Food, Drug, and Cosmetic Act. |
K100141 - MAGNETOM Essenza 14-Channel Extremity Coil
{1}------------------------------------------------
K100141
Page 2/3
# SIEMENS
## Healthcare
## II. Safety and Effectiveness Information Supporting Substantial Equivalence.
#### Intended Use
The new 14-Channel Extremity Coil is indicated for use in conjunction with the 1.5T MAGNETOM ESSENZA, in the MR examination of the human knee, foot, ankle, hand and wrist.
Used in the MAGNETOM ESSENZA, the 14-Channel Extremity Coil is indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal and oblique cross sectional images that display the internal structure and/or function of the knee, foot, ankle, hand and wrist.
When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
The intended use of the MAGNETOM ESSENZA is not affected in any way by the use of the new 14-Channel Extremity Coil.
#### Device Description
The new 14-Channel Extremity Coil is indicated for use in conjunction with the 1.5T MAGNETOM ESSENZA, in the MR examination of the human knee, foot, ankle, hand and wrist.
Used in the MAGNETOM ESSENZA, the 14-Channel Extremity Coil is indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal and oblique cross sectional images that display the internal structure and/or function of the knee, foot, ankle, hand and wrist.
#### Substantial Equivalence
Siemens believes that, within the meaning of the Safe Medical Device Act of 1990, the 14-Channel Extremity Coil is substantially equivalent to the QED TxRx 15Ch Knee Coil 1.5T.
| Predicate Device Name | FDA Clearance<br>Number | FDA Clearance<br>Date |
|-------------------------------------|-------------------------|-----------------------|
| TxRx 15Ch Knee Coil 1.5T .<br>(QED) | K082636 | Sep. 25, 2008 |
#### General Safety and Effectiveness Concerns:
The following are the safety and performance parameters:
K100141 - MAGNETOM Essenza 14-Channel Extremity Coil
Page 57 of 60
{2}------------------------------------------------
K100141
page 3/3
# SIEMENS
## Healthcare
Safety
-Maximum Static Field -Rate of Change of Magnetic Field -Acoustic Noise Level
### Performance
-Geometric Distortion -Slice Profile, Thickness and Gap -High Contrast Spatial Resolution
Specified by the FDA Guidance document for MR, Diagnostic Devices are unaffected by the modifications described within this notification.
The following parameters were considered for the new 14-Channel Extremity Coil:
Safety -RF Power Deposition -Biocompatibility
### Performance -Signal to Noise Ratio -Image Uniformity
No new materials were used for the new 14-Channel Extremity Coil compared to the predicate device. Therefore no new biocompatibility tests were performed. Signal to Noise Ratio (SNR), image uniformity and SAR tests were performed for the new 14-Channel Extremity Coil and the results presented in this submission show that they are equivalent with the predicate devices.
Conclusion as to Substantial Equivalence
Laboratory testing was performed to support this claim of substantial equivalence and to show that the technological differences do not raise any new questions pertaining to effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a representation of human figures.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Judith Campbell Regulatory Technical Specialist Siemens Medical Solution USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355
AUG 2 7 2010
Re: K100141
Trade/Device Name: 14-Channel Extremity Coil for MAGNETOM ESSENZA Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: August 6, 2010 Received: August 9, 2010
Dear Ms. Campbell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device; subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, ' Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Donald J. Trump
Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K100141
Image /page/5/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The letters are black and appear to be slightly distressed, with some small imperfections and variations in the ink density. The background is white.
Attachment A - Indications for Use
510(k) Number (if known): K100141
Device Name: 14-Channel Extremity Coil for MAGNETOM ESSENZA
#### Indications for Use:
The intended use of the new 14-Channel Extremity Coil is, in conjunction with THE 1.5T MAGNETOM ESSENZA Magnetic Resonance Scanner, the MR examination of the human knee, foot, ankle, hand and wrist.
Used in the MAGNETOM ESSENZA, the 14-Channel Extremity Coil is indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal and oblique cross sectional images that display the internal structure and/or function of the knee, foot, ankle, hand and wrist. The images produced by the MAGNETOM ESSENZA with the, 14-Channel Extremity Coil reflects the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.
When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
The intended use of the MAGNETOM ESSENZA is not affected in any way by the use of the new 14-Channel Extremity Coil.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (0) O J V D
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K100141
K100141 - MAGNETOM Essenza 14-Channel Extremity Coil
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.