ACUMED WRIST ARTHRODESIS PLATE SYSTEM

K100123 · Acumed, LLC · HRS · Oct 6, 2010 · Orthopedic

Device Facts

Record IDK100123
Device NameACUMED WRIST ARTHRODESIS PLATE SYSTEM
ApplicantAcumed, LLC
Product CodeHRS · Orthopedic
Decision DateOct 6, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Acumed Wrist Arthrodesis Plate System is intended for wrist arthrodesis and fractures of other small bones. Specific indications include post-traumatic arthritis of the joints of the wrist, rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis r frame wrist, severe unremitting wrist pain related to motion, brachial plexus nerve palsies, turnor resection, and spastic deformities.

Device Story

Acumed Wrist Arthrodesis Plate System consists of pre-contoured, anatomic metallic plates and 2.3 mm/3.5 mm cortical locking and non-locking screws. Used by orthopedic surgeons in clinical settings for bone fixation and wrist fusion. Device provides structural support to stabilize bones during healing or fusion. Benefits include restoration of joint function and pain relief. No software or electronic components.

Clinical Evidence

Bench testing only. Engineering analysis compared plate strength based on cross-sectional area and material properties. Mechanical testing compared distal screw pull-out strength between subject and predicate devices.

Technological Characteristics

Titanium alloy (Ti 6Al 4V) per ASTM F136. Pre-contoured anatomic plates with multiple holes for 2.3 mm and 3.5 mm cortical locking and non-locking screws. Mechanical fixation device.

Indications for Use

Indicated for patients requiring wrist arthrodesis or fixation of small bone fractures, including those with post-traumatic arthritis, rheumatoid deformities, complex carpal instability, post-septic arthritis, severe motion-related wrist pain, brachial plexus nerve palsies, tumor resection, or spastic deformities.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Innovative Solustions 15100123 # OCT 6 2010 | 510(k) Summary | | |--------------------|---------------------------------------| | Contact: | Lino Tsai | | | Acumed, LLC | | | 5885 NW Cornelius Pass Rd. | | | Hillsboro, OR 97124-9432 | | | (503) 627-9957 x 1370 | | | FAX: (503) 686-7102 | | Device Trade Name: | Acumed Wrist Arthrodesis Plate System | | Manufacturer: | Acumed, LLC | | | 5885 NW Cornelius Pass Rd. | | | Hillsboro, OR 97124-9432 | | Common Name: | Plate, Fixation, Bone | | Classification: | 21 CFR 888.3030 | | Class: | II | | Product Code: | HRS | | Date: | September 24, 2010 | ### Indications for Use: The Acumed Wrist Arthrodesis Plate System is intended for wrist arthrodesis and fractures of other small bones. Specific indications include post-traumatic arthritis of the joints of the wrist, rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis r frame wrist, severe unremitting wrist pain related to motion, brachial plexus nerve palsies, turnor resection, and spastic deformities. # Device Description: Device Book prist Arthrodesis Plates are pre-contoured with an anatomic design with multiple rric Acaned Which will be a combination of distal and proximal holes which utilize 2.3 mm and 3.5 mm cortical locking and non-locking screws. {1}------------------------------------------------ Innovative Solustions ### Technological Characteristics: The plates and screws are made of titanium alloy, Ti 6Ai 4V, per ASTM F136. The plates are supplied in multiple wrist positions. The predicate devices share these dimensional and material characteristics. There are no technological characteristics that raise new issues of safety or effectiveness. An engineering analysis of plate strength was conducted to compare the Acumed Wrist Arthrodesis Plate to the predicate device. Calculations based on cross-sect6ional area and plate material showed that the Acumed Wrist Arthrodesis Plate is able to withstand more load than the predicate device before plastic deformation. Mechanical testing of distal screw pull-out strength was conducted comparing the Acumed Wrist Arthrodesis Plate to the predicate device. The results showed that the Acumed Wrist Arthrodesis Plate required equivalent force and significantly more energy to cause distal screw pull-out than the predicate device. ## Predicate Device(s): Synthes LCP Wrist Fusion Plate - K042355 Synthes Small Titanium Wrist Fusion Plate - K023879 Synthes Straight Wrist Fusion Plate - K011458 Synthes Wrist Fusion Plate - K000558 DVO Wrist Fusion Plate - K052754 KMI Wrist Fusion System - K991873 Based upon the similarities of the Acumed Wrist Arthrodesis Plate System and the predicate devices studied, the safety and effectiveness of the Acumed Wrist Arthrodesis Plate System is substantially equivalent to the predicate devices referenced. 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Acumed LLC % Ms. Lino Tsai Regulatory Specialist 5885 Northwest Cornelius Pass Road Hillsboro, Oregon 97124-9432 OCT 6 2010 Re: K100123 Trade/Device Name: Acumed Wrist Arthrodesis Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: August 26, 2010 Received: August 27, 2010 Dear Ms. Tsai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ # Page 2 - Ms. Lino Tsai device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours, ely yours, for N. Malkerson Mark N. Melke Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "ACUMED" in all caps with a registered trademark symbol. The font is blocky and sans-serif. The letters are black and the background is white. Innovative Solutions Indications for Use 510(k) Number (if known): k100123 Device Name: Acumed Wrist Arthrodesis Plate System Indications for Use: The Acumed Wrist Arthrodesis Plate System is intended for wrist arthrodesis and fractures of other small bones. Specific indications include post-traumatic arthritis of the wrist, · rheumatoid wrist deformities requiring restoration, complex carpal instability, post-septic arthritis of the wrist, severe unremitting wrist pain related to motion, brachial plexus nerve palsies, tumor resection, and spastic deformities. Prescription Use _ × (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Soutte for mxn (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K100123 Acumed LLC Page 42 1 of 6
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%