K100053 · Mattioli Engineering Corp. · ONF · Mar 10, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100053
Device Name
MATTIOLI PULSE TWO/THREE PLUS FAMILY
Applicant
Mattioli Engineering Corp.
Product Code
ONF · General, Plastic Surgery
Decision Date
Mar 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MATTIOLI PULSE TWO/THREE PLUS is an Intense Pulse Light (IPL) device family indicated for use in aesthetic applications (based on selective photothermolysis) , in the treatment of various benign pigmented lesions and hair removal and that produce different effects depending on the applicator that is used: SA APPLICATOR: Model SN: Wavelengths from 560 - 1200 nm are indicated for treatment of benign pigmente (epidermal and coutaneous) lesions, including hyperpigmentation, warts, lentigines, nevi, melasma, and cafe-au-lait. VA APPLICATOR Model VN: Wavelengths from 510 - 1200 nm are indicated for the treatment of benign vascular lessons including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider anglormas, polkiloderma of civatte; leg veins, facial veins and venous malformations. HR APPLICATORS indicated for the treatment of unwanted hair (i.e. hair removal). Model HN: Wavelengths from 650-1200 nm for skin types I-IV; Model HF: Wavelengths from 710 - 1200 nm for skin type V The equipment should only be used under medical supervision.
Device Story
Mattioli Pulse Two/Three Plus Family is an Intense Pulsed Light (IPL) system for aesthetic procedures. System consists of a console with electronic supply, capacitor bank, and computerized control system (CCS) with an 8.2" LCD touch screen. Handheld applicators contain a flash lamp, water-cooled reflector, and filtered light guide to select specific wavelengths. Operator selects parameters via touch screen to deliver light energy based on selective photothermolysis. HR applicators target hair bulbs for removal; SA applicators target pigmented lesions; VA applicators target vascular lesions. Used in clinical settings under medical supervision. Output parameters (pulse duration, fluence) are controlled by the CCS. Benefits include non-invasive treatment of skin conditions and hair removal.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
IPL generator; flash lamp source; water-cooled reflector; filtered light guide; 8.2" LCD touch screen interface; microprocessor-controlled; internal closed-loop water-air heat exchanger; desktop (Pulse Two Plus) or floor-standing (Pulse Three Plus) form factor.
Indications for Use
Indicated for aesthetic applications including hair removal (skin types I-V), treatment of benign pigmented lesions (e.g., warts, lentigines, nevi, melasma, cafe-au-lait), and benign vascular lesions (e.g., port wine stains, hemangiomas, telangiectasias, rosacea, venous malformations). For use under medical supervision.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Angelite Family of Intense Pulsed Light Systems (Advanced Technology Laser, Co., Ltd.)
Submission Summary (Full Text)
{1}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
.
・
. .
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
. .
。
.
510(k) Premarket Notification rev 1.0 page 1 of 3
K1000053
:
.
. .
. '
・. .
. .
·
. ・
. : i . . . . .
:・・
· · ·
: :
•
.
## 510(k) SUMMARY
| Submitter: | Mattioli Engineering Corporation<br>MAR 1 0 2010 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Gian Franco Bernabei |
| Date Summary Prepared: | december 11, 2009 |
| Device Trade Name: | Mattioli Pulse Two/Three Plus Family |
| Common Name: | IPL system generator |
| Classification Name: | Powered Light Based Non-Laser Surgical Instrument With<br>Thermal Effect<br>ONF<br>21 CFR 878.4810 |
| Equivalent Device: | Advanced Technology Laser, Co., Ltd.: Angelite Family of<br>Intense Pulsed Light Systems |
| Device Description: | The MATTIOLI PULSE TWO/THREE Plus Family is an<br>Intense Pulse Light (IPL) generator device producing on the<br>skin different effects depending on the applicator used by<br>the operator:<br><br>SR Applicator: its purpose is to stimulate the regenerative<br>process of the skin. It uses the principle of the selective<br>absorption of the different light wavelengths, without<br>damaging the skin structure.<br><br>HR Applicator: its purpose is to destroy thermally the hair<br>bulb, due to the absorption of a selected range of radiated<br>wavelength. The selective absorption of different<br>wavelengths is either used by HR to obtain the desired<br>effect without damaging the surrounding skin structures.<br><br>VR Applicator: Its purpose is to treat benign pigmented<br>(epidermal and cutaneous) lesions including warts scars<br>abnd striae. It is also intended for treatment of benign<br>(cutaneous) vascular lesions including port wine stains,<br>hemangiomas, facial, truncal and leg telangectasias, rosacea, |
{2}------------------------------------------------
melasmna, angiomas and spider angiomas, polkiloderma of civatte; leg veis, facial veins and venous malformations.
The two systems differ substantially only by the case: it is "desktop" style for MATTIOLI PULSE TWO PLUS device and a standard case, with wheels to allow easy movement on the floor, for PULSE THREE PLUS device.
Both equipments consist of a console, containing the electronic supply system for capacitor bank charge , the computerized control system (CCS) , and one or two hand pieces in which are contained the light generators devices (flash lamp) and the necessary filtered light guide for the selection of the wavelengths range.
The computerized control system (CCS) monitors and controls either the equipment and interfaces with the operator, through an 8.2" colour back lighted LCD screen display and a touch screen panel, displaying all the selected parameters and indications on the status of the machine. The selection of each command and parameter simply occurs by touching the LCD screen.
Each applicator is composed basically by the flash-lamp, a water cooled reflector, a filtered light guide, a light guide cooler, a flash-counter and the pushbutton for the activation of flash emission.
The MATTIOLI PULSE TWO/THREE PLUS is an Intense Pulse Light (IPL) device family indicated for use in aesthetic and cosmetic applications (based on selective photothermolysis, in the treatment of acne, various benign pigmented lesions and hair removal and that produce different effects depending on the applicator that is used:
### HR APPLICATORS:
Intense Pulse Light Energy Wavelengths from 650 (HN type) or 710 (HF type ) - 960 nm are indicated for the treatment of unwanted hair (i.e., hair removal).
#### SR APPLICATOR:
Intense Pulse Light Energy Wavelengths from 560 - 960 nm are indicated for skin resurfacing procedure , for the treatment of inflammatory acne and decrease the appearance of dark spots.
#### VR APPLICATOR
Intended Use:
{3}------------------------------------------------
| | Intense Pulse Light Energy Wavelengths from 510 - 960 nm<br>are indicated for the treatment of benign pigmented<br>(epidermal and coetaneous) lesions including warts, scars<br>and striae. For the treatment of benign (cutaneous) vascular<br>lesions including port wine 'stains, hemangiomas, facial,<br>truncal and leg telangiectasias, rosacea, melasmna,<br>angiomas and spider anglormas, polkiloderma of civatte; leg<br>veins, facial veins and venous malformations. |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Rationale for Substantial<br>Equivalence: | The product specification, functionality, indication for use,<br>and treatment parameters of the Mattioli Pulse Two/Three<br>Plus Family are the same or very similar to the legally<br>marketed Angelite Family of Intense Pulsed Light Systems. |
| | Both systems have the same indication for use. |
| | Both systems uses a flash lamp, which emitted light is<br>delivered to the patient via a glass light guide. |
| | The Mattioli Pulse Two/Three Plus Family output<br>characteristics (including pulse duration and fluence) are<br>identical, or very similar, to those of the predicate device. |
| | Both systems are microprocessor controlled devices. |
| | Both systems utilize an internal closed loop water-air heath<br>exchanger circuit for optimal thermal control of system<br>temperature. |
The risks and benefits for the Mattioli Pulse Two/Three Plus Family are comparable to those for the predicate device. Therefore, the introduction of this IPL device should not raise new questions of Safety and Effectiveness .
Non-Clinical Performance Data: None
Clinical Performance Data:
None
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem of an eagle with outstretched wings, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Mattioli Engineering Corporation % Mr. Gian Franco Bernabei 8300 Greensboro Drive, Suite 800 McLean, Virginia 22102
MAR 1 0 2010
Re: K100053
Trade/Device Name: Mattioli Pulse Two/Three Plus Family
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and dermatology
Regulatory Class: Class II Product Code: ONF Dated: February 25, 2010 Received: March 01, 2010
Dear Mr. Bernabei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act ' or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Gian Franco Bernabei
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
y yours,
for
Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known): K100053
Device Name: Mattioli Pulse TWO/THREE Plus Family
Indications for Use:
The MATTIOLI PULSE TWO/THREE PLUS is an Intense Pulse Light (IPL) device family indicated for use in aesthetic applications (based on selective photothermolysis) , in the treatment of various benign pigmented lesions and hair removal and that produce different effects depending on the applicator that is used:
SA APPLICATOR:
Model SN: Wavelengths from 560 - 1200 nm are indicated for treatment of benign pigmente (epidermal and coutaneous) lesions, including hyperpigmentation, warts, lentigines, nevi, melasma, and cafe-au-lait.
#### VA APPLICATOR
Model VN: Wavelengths from 510 - 1200 nm are indicated for the treatment of benign vascular lessons including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider anglormas, polkiloderma of civatte; leg veins, facial veins and venous malformations.
HR APPLICATORS indicated for the treatment of unwanted hair (i.e. hair removal). Model HN: Wavelengths from 650-1200 nm for skin types I-IV; Model HF: Wavelengths from 710 - 1200 nm for skin type V
The equipment should only be used under medical supervision.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------------|------------|
| Division of Surgical, Orthopedic, and Restorative Devices | |
| Page | of _______ |
| 510(k) Number | K100053 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.