K100008 · Cardo Medical, Inc. · LZO · May 18, 2010 · Orthopedic
Device Facts
Record ID
K100008
Device Name
CARDO MEDICAL CERAMIC HIP SYSTEM
Applicant
Cardo Medical, Inc.
Product Code
LZO · Orthopedic
Decision Date
May 18, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cardo Medical Ceramic Head components are for use in total hip arthroplasty as a result of: • Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative joint disease, osteoarthritis, and arthritis resulting from biologic or mechanical trauma to the hip, • Correction of functional deformity, • Avascular Necrosis, • Treatment of non-unions, femoral neck and trochanteric fractures of the proximal femur, • Difficult clinical management problems involving persistent pain and physical impairment where conventional arthodesis is not likely to achieve satisfactory results. These components are single use only and are intended for implantation with or without bone cement.
Device Story
Cardo Medical Ceramic Hip System is a line extension to a previously cleared hip system, adding 32mm and 36mm alumina ceramic femoral heads. The system is used in total hip arthroplasty; implanted by orthopedic surgeons in a clinical/hospital setting. The ceramic heads mate with existing 12/14 trunnion taper femoral stems, UHMWPE acetabular inserts, and titanium alloy bone screws. The device provides a bearing surface for hip joint replacement. Benefits include restoration of hip function and pain relief for patients with degenerative or traumatic hip conditions. The system is provided sterile for single use, with or without bone cement.
Clinical Evidence
No clinical data. Substantial equivalence supported by nonclinical bench testing, including stem fatigue, acetabular insert push-in/push-out/lever-out/rotational torque, range-of-motion (oscillation), and femoral head static burst, dynamic fatigue, and axial pull-off testing.
Technological Characteristics
Components: Alumina ceramic femoral heads, Ti-6Al-4V femoral stems/acetabular shells, UHMWPE acetabular inserts, Ti-6Al-4V bone screws. Interface: 12/14 trunnion taper. Sterilization: Sterile (SAL 10^-6). Single-use. Mechanical testing per FDA guidance for ceramic ball hip systems.
Indications for Use
Indicated for patients undergoing total hip arthroplasty due to rheumatoid arthritis, non-inflammatory degenerative joint disease, osteoarthritis, post-traumatic arthritis, functional deformity, avascular necrosis, non-unions, or proximal femoral fractures (femoral neck/trochanteric). Also indicated for management of persistent pain/impairment where arthrodesis is unsatisfactory.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
Accin™ Hip System (now Cardo Medical Hip System) (K073068, K094045)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "Cardo Medical Ceramic Head Premarket Notification Submission - 510(k)" on the left side of the image. On the right side of the image is the number K100008 written in cursive. The text on the left side of the image describes the document as a premarket notification submission for a Cardo Medical Ceramic Head. The number on the right side of the image is likely a reference or identification number for the submission.
MAY 1 8 2010
# Summary of Safety and Effectiveness
| Submitter: | Dina Weissman, J.D.<br>Cardo Medical, Inc., 10 Clifton Blvd., Suite B1, Clifton, NJ 07011<br>Tel: 973-777-8832, ext. 330<br>Fax: 973-777-8851<br>Email: dweissman@cardomedical.com |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | May 13, 2010 |
| Device: | Cardo Medical Ceramic Hip System |
| Classification: | 87 LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or<br>non-porous, uncemented, 21 CFR 888.3353 Class II |
| Predicate Device: | Accin™ Hip System (now Cardo Medical Hip System) - K073068, K094045<br>Stelkast 32mm Modular Ceramic Head - K033944 |
| Device Description: | The previously cleared Cardo Medical Hip System consists of<br>• commercially pure titanium plasma spray-coated titanium alloy (Ti-6Al-4V)<br>femoral stems,<br>• cobalt chrome (CoCr) femoral heads,<br>• commercially pure titanium plasma spray-coated titanium alloy (Ti-6Al-4V)<br>acetabular shells,<br>• ultra-high molecular weight polyethylene (UHMWPE) acetabular inserts, and<br>• titanium alloy (Ti-6Al-4V) bone screws.<br><br>The proposed devices are a line extension to the previously cleared system to<br>add 32mm and 36mm alumina ceramic femoral heads. |
| Intended Use: | The Cardo Medical Ceramic Head components are for use in total hip<br>arthroplasty as a result of:<br>• Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative<br>joint disease, osteoarthritis, and arthritis resulting from biologic or mechanical<br>trauma to the hip,<br>• Correction of functional deformity,<br>• Avascular Necrosis,<br>• Treatment of non-unions, femoral neck and trochanteric fractures of the<br>proximal femur,<br>• Difficult clinical management problems involving persistent pain and physical<br>impairment where conventional arthodesis is not likely to achieve satisfactory<br>results.<br><br>These components are single use only and are intended for implantation with or<br>without bone cement. |
## Comparison to Predicates:
The Cardo Medical Ceramic Heads are manufactured from alumina ceramic and are for use with the currently marketed femoral stem, acetabular inserts and bone screws. The proposed devices are similar in design and materials to the Cardo Medical Total Hip System (K073068 and K094045) and the Stelkast Ceramic Heads (K033944).
### Femoral Stems
There has been no change to the existing femoral stems.
Page 1 of 2
{1}------------------------------------------------
Cardo Medical Ceramic Head Premarket Notification Submission - 510(k)
Image /page/1/Picture/1 description: The image shows a handwritten alphanumeric string. The string is "K100008". The characters are written in black ink on a white background.
### Femoral Heads
The proposed alumina ceramic femoral heads have the same 12/14 trunnion taper as the previously cleared cobalt chrome femoral heads, offered in sizes 22 through 36 mm OD in a variety of offsets (K073068 and K094045) and are equivalent in size, design, and sterilization. They are equivalent in material to the Stelkast ceramic femoral heads.
#### Acetabular Shells
There has been no change to the existing acetabular shells. They are designed to mate with the acetabular inserts and have holes for bone screws, if necessary.
#### Acetabular Inserts
There has been no change to the existing UHMWPE acetabular inserts, offered in sizes 22 through 36mm. They are designed to mate with the previously cleared acetabular shells.
#### Bone Screws
There has been no change to the existing titanium alloy bone screws cleared in K073068, which are in a 6.5mm diameter, in lengths from 15mm to 60mm.
All Cardo medical implant components are provided as sterile, single use only to a sterility assurance level (SAL) of 10-6.
### Summary
Cardo Medical, Inc. has determined that any differences in the proposed device will not impact the safety or effectiveness of the hip system for its intended use. Testing has shown that the proposed device meets the requirements of the FDA Guidance document "Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems" and the guidance on total hip arthroplasty. The proposed device is substantially equivalent to the predicate device.
### Synopsis of Test Methods and Results:
#### Nonclinical Performance
All mechanical testing required by the ĝuidance document for ceramic heads is included in this submission including technical reports for stem fatique mechanical testing; the push-in, push-out, leverout and rotational torque forces for acetabular inserts; the Range-of-Motion (oscillation) of the proposed system; and static burst testing, dynamic fatigue testing and Axial Pull-Off testing of the femoral heads. Results indicate that the Cardo Medical ceramic femoral heads and inserts is equivalent to devices currently legally marketed, is compatible with the previously cleared 12/14 femoral heads and capable of withstanding in vivo loading.
#### Clinical Performance
Clinical data and conculsions are not needed for this device.
Page 2 of 2
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cardo Medical Inc. % Ms. Dina Weissman, J.D. Director, Quality Assurance, Regulatory Affairs and Government 10 Clifton Boulevard, Suite B1 Clifton, New Jersey 07011
MAY 18 2010
Re: K100008
Trade/Device Name: Cardo Medical Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. Regulatory Class: II Product Code: LZO Dated: April 30, 2010 Received: May 3, 2010
Dear Ms. Weissman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Dina Weissman, J.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Harbarbriclino
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Form
100008 510(k) Number (if known):
Device Name: Cardo Medical Hip System
## Indications for Use:
The Cardo Medical Hip System components are for use in total hip arthroplasty as a result of:
- . Hip arthritis caused by rheumatoid disease, non-inflammatory degenerative joint disease, osteoarthritis, and arthritis resulting from biologic or mechanical trauma to the hip
- . Correction of functional deformity
- Avascular Necrosis
- . Treatment of non-unions, femoral neck and trochanteric fractures of the proximal femur
- . Difficult clinical management problems involving persistent pain and physical impairment where conventional arthodesis is not likely to achieve satisfactory results
These components are single use only and are intended for implantation with or without bone cement.
Prescription Use × (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Suita for mxn
(Devision Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K100008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.