The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
Device Story
Ophthalmic surgical laser system; utilizes femtosecond laser pulses to perform anterior capsulotomy and phacofragmentation during cataract surgery. System inputs include eye surface localization via aiming beam; patient eye fixated via disposable contact lens and suction ring assembly. Device transforms inputs by scanning femtosecond pulses in 3D patterns (cylindrical for capsulotomy; intersecting ellipsoidal planes for phacofragmentation) to segment lens nucleus into quadrants via photodisruption. Operated by surgeons in clinical settings. Output is a pre-fragmented lens nucleus, reducing manual manipulation and ultrasound power requirements during subsequent phacoemulsification. Benefits include reduced mechanical effects and ultrasonic energy transmission to ocular tissue compared to traditional ultrasound-only phacoemulsification.
Clinical Evidence
Prospective clinical study (n=92) compared LenSx laser phacofragmentation followed by phacoemulsification (n=46) against standard ultrasound phacoemulsification control (n=46). Primary endpoints included manual manipulation and ultrasound power requirements. Results showed the laser group required significantly less manual manipulation and ultrasound power to achieve fragmentation. No significant adverse events reported; clinical outcomes consistent with advanced phacoemulsification techniques.
Technological Characteristics
Ophthalmic surgical laser; femtosecond pulse generation; 3D scanning delivery system; aiming beam for surface localization; disposable contact lens/suction ring interface. Operates via photodisruption (acoustic/cavitation effects). Software-controlled pulse patterns. Class II device (21 CFR 886.4390).
Indications for Use
Indicated for patients undergoing cataract surgery requiring anterior capsulotomy and/or laser phacofragmentation.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Alcon Series 20000 Legacy Phacoemulsifier (K952213)
Submission Summary (Full Text)
{1}------------------------------------------------
APR 2 3 2010
## 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
# SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED
| a. | Applicant: | LenSx Lasers, Inc.<br>33 Journey, Suite 175<br>Aliso Viejo, CA 92656<br>Tel: (949) 360-6010<br>Fax: (949) 360-6028 |
|----|------------|--------------------------------------------------------------------------------------------------------------------|
|----|------------|--------------------------------------------------------------------------------------------------------------------|
Judy Gordon, D.V.M. Contact Person: b. ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651 judy@clinregconsulting.com Tel: (949) 715-0609 Fax: (949) 715-0610
NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME
| a. Trade/Proprietary Name: | LenSx 550 Laser System |
|----------------------------|-------------------------------------|
| b. Common/Usual Name: | LenSx 550 Laser System |
| c. Classification Name: | Laser Instrument, Surgical, Powered |
| d. Classification Code(s): | 21 CFR 886.4390 ; 79 OOE |
#### PREDICATE DEVICES
| 510(K) # | TRADE NAME | MANUFACTURER |
|----------|----------------------------------------------|-----------------------|
| K082947 | LenSx 550 Laser System | LenSx Lasers, Inc. |
| K993154 | Dodick Laser Photolysis<br>System | ARC Laser Corporation |
| K952213 | Alcon Series 20000 Legacy<br>Phacoemulsifier | Alcon Laboratories |
{2}------------------------------------------------
### DEVICE DESCRIPTION
The LenSx 550 Laser is an ophthalmic surgical laser that has previously been cleared for use in anterior capsulotomy during cataract surgery (K082947). The LenSx 550 generates femtosecond laser pulses that are scanned in a three-dimensional pattern in the eye. Localization of these laser pulses is accomplished by an aiming beam that identifies the surfaces of the lens, and an off-the-shelf, disposable contact lens and suction ring assembly that fixates the eye relative to the delivery system. As described in K082947, anterior capsulotomy is performed by delivering a cylindrical pattern of laser pulses to intersect the anterior lens capsule.
The same surgical device, with only minor software changes, is now also being proposed for use in phacofragmentation during cataract surgery. As described in this premarket notification, laser phacofragmentation divides the lens into quadrants, in the same way that division of the lens into quadrants is performed with standard ultrasound phacofragmentation in the initial step of phacoemulsification.
To perform phacofragmentation, the LenSx system delivers a series of laser pulses to form two intersecting ellipsoidal planes. Photodisruption of the lens tissue in this pattern, which appears as a cross from a surgical (top) view, segments the nucleus into four sections before the eye is physically entered by any instruments or devices.
Following entry into the eye and then into lens, the fragmented nuclear quadrants created by the LenSx 550 are further fragmented and removed using a standard phacoemulsification probe.
#### STATEMENT OF INTENDED USE
The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON
The LenSx 550 Laser System proposed for use in phacoframentation is identical with respect to all technological characteristics to the previously cleared LenSx 550 for use in anterior capsulotomy (K082947). Substantial equivalence is also drawn to previously cleared phacofragmentation systems.
The LenSx 550 Laser System delivers femtosecond laser pulses to produce a pattern of photodisruption in the lens nucleus. Highly localized, laser mediated acoustic and cavitation effects associated with photodisruption precisely fragment the lens nucleus.
Predicate devices include laser and ultrasound phacoemulsification devices cleared for phacofragmentation, such as the Alcon Series 20000 Legacy (Alcon Laboratories, Inc. K952213) and the Dodick Photolysis system (ARC Laser Corp., K993154). Ultrasound
{3}------------------------------------------------
phacofragmentation devices utilize a piezoelectric ultrasonic source that generates similar acoustic and cavitation effects that, along with mechanical effects from the probe, result in phacofragmentation.
Laser devices cleared for phacofragmentation, including the Dodick Photolysis System, use high energy nanosecond laser pulses delivered via a probe to create the mechanical, cavitation and acoustic effects that mediate phacofragmentation.
Since the femtosecond pulses are delivered optically through the cornea, the LenSx Laser performs phacofragmentation without the mechanical effects and ultrasonic shockwaves/energy transmitted to the tissue by handheld phacofragmentation probes.
# BRIEF SUMMARY OF PRECLINICAL AND CLINICAL PERFORMANCE TEST RESULTS
The performance data supporting substantial equivalence of the LenSx 550 femtosecond laser system to the predicate devices are summarized as follows:
- Pre-clinical studies applying the LenSx phacofragmentation pattern in plastic . substrates demonstrated a high degree of accuracy and reproducibility for laser pulse placement.
- LenSx phacofragmentation performed in ex vivo porcine eyes further confirmed . the accuracy and reproducibility of the localized laser tissue effects. This testing also demonstrated that the ultrasound power required to complete the phacofragmentation procedure was significantly reduced in comparison to procedures performed using standard phacofragmentation and phacoemulsification techniques with the Alcon Series 20000 Legacy, the most common such device.
- A prospective clinical study was performed outside the U.S. to evaluate clinical . performance, with a total of 92 subjects randomly assigned to one of the following treatment groups:
- 1. A study group of 46 eyes underwent initial laser phacofragmentation using the LenSx 550 Laser, followed by phacoemulsification using the Alcon Series 20000 Legacy.
- 2. A control group of 46 eyes underwent phacofragmentation and phacoemulsification with the Alcon Series 20000 Legacy (predicate device), using a surgical technique previously demonstrated to significantly reduce ultrasound power requirements.
Compared to the control group where only standard ultrasonic phacoemulsification was used, the study group (using laser phacofragmentation) required significantly less manual manipulation and ultrasound power to achieve the desired phacofragmentation effect. Clinical results were excellent in both groups, with no significant adverse events. Results in both groups were consistent with those of advanced phacoemulsification techniques reported in previous studies (Can, 2004).
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is stylized with thick, black lines, giving it a bold and recognizable appearance.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
APR 2 3 2010
LenSx Lasers, Incorporated c/o Judy F. Gordon, D.V.M. ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, California 92651
Re: K094052
Trade/Device Name: LenSx 550 Laser System, Model 550 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE, HQC Dated: March 23, 2010 Received: March 25, 2010
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Edith m.h
Malvina B. Eydelman, M.Ø Director Division of Ophthalmic, Neurological, and Ear, Nose, Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
Device Name(s): LenSx 550 Laser System
Indications for Use:
The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _ 1 _ of
Daryl Kaufman M.A.
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices .
510(k) Number K0-94052
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.