AF 620 PH.I.S.I.O., AF 640 PH.I.S.I.O. ARTERIAL FILTERS
K093986 · Sorin Group Italia S.R.L. · DTM · Jan 22, 2010 · Cardiovascular
Device Facts
Record ID
K093986
Device Name
AF 620 PH.I.S.I.O., AF 640 PH.I.S.I.O. ARTERIAL FILTERS
Applicant
Sorin Group Italia S.R.L.
Product Code
DTM · Cardiovascular
Decision Date
Jan 22, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AF 620 Ph.I.S.I.O. with 20 micron screen with phosphoryIcholine coating and the AF 640 Ph.I.S.I.O. with 40 micron screen with phosphory choline coating are recommended for use in the arterial line of the extracorporeal nicuti during any procedure that requires cardiopulmonary bypass. The filters, are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
Device Story
Sterile, disposable arterial line blood filters for cardiopulmonary bypass circuits; flow rates up to 6.0 L/min. AF 620 (20 micron) and AF 640 (40 micron) models. Input: arterial line perfusate containing gaseous emboli and particulate debris. Operation: physical filtration via double-pleated polyester screen; phosphorylcholine coating for hemocompatibility. Output: filtered blood returned to patient. Used in clinical settings (e.g., OR) by perfusionists/surgeons. Benefits: reduction of emboli and debris reaching patient during bypass. Modifications from predicate include ultrasonic welding (replacing polyurethane potting), reduced housing size, double-pleated screen, and updated phosphorylcholine formulation.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993), sterility, ETO residuals, hemolysis, acute systemic toxicity, mutagenicity, cytotoxicity, irritation, and sensitization. In vitro performance testing (structural integrity, pressure integrity, pressure drop, filtration efficiency, cell depletion) conducted on aged devices (up to 3 years) compared to unmodified predicates. All results met established specifications.
Indicated for use in the arterial line of the extracorporeal circuit during any procedure requiring cardiopulmonary bypass to trap and remove gaseous emboli and particulate debris. Contraindicated for use longer than 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”
Predicate Devices
D731 MICRO Ph.I.S.I.O. Pediatric Arterial Filter
D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filter
Submission Summary (Full Text)
{1}------------------------------------------------
6093986
AF-620 Ph.I.S.I.O. and AF 640 Ph.I.S.I.O. Arterial Filters = 3 Sorin Group Italia S.r.I.
Special 510(k) December 23, 2009
### 510(k) SUMMARY
SUBMITTER:
Sorin Group Italia 86, Via Statale 12 Nord 41037 Mirandola (MO) Italy JAN 22 2010
CONTACT PERSON:
Luigi Vecchi Phone: 39 0535 29811 Fax: 39 0535 25229
December 23, 2009
DATE PREPARED:
Sorin AF 620-640 Ph.I.S.I.O.
DEVICE TRADE NAME:
COMMON NAME:
Arterial Filter
CLASSIFICATION NAME:
Cardiopulmonary Bypass Arterial Line Blood Filter
UNMODIFIED DEVICES:
D731/D733 MICRO Ph.I.S.I.O. Pediatric Arterial Filters
#### DEVICE DESCRIPTION:
The AF 620 Ph.1.S.I.O. and 640 Ph.I.S.I.O are sterile, non-pyrogenic disposable filter for use in the arterial line of the cardiopulmonary bypass circuit with flow rate not exceeding 6.0 liters/minute. The AF 620 Ph.I.S.I.O. and 640 Ph.I.S.I.O are Arterial Filters with 20 and 40 micron filters screen, respectively, designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris, greater than the pore size, from the arterial line perfusate. The AF 620 Ph.I.S.I.O. and 640 Ph.I.S.I.O are a modified version of the currently marketed D731 and D733 MICRO Ph.I.S.I.O. The modifications consist of: a different port orientation of the blood outlet port in order to improve the ease of use, ergonomics and fluid dynamics properties; change from polyurethane potting to ultrasonic welding for improved overall biocompatibility; the size of the filter housing has been reduced thus the filter net is double pleated rather than single pleated: the pore size of the filter screen for the AF 620 Ph.I.S.I.O. has been reduced from 27 micron to 20 micron with respect to the D731 MICRO Ph.I.S.I.O. for improved filtration efficiency; a and different formulation of phosphorylcholine monomer has been used to improve wettability. As a consequence of these modifications, the labeling has been updated.
The modified device has unchanged intended use, operating principles, manufacturing, control mechanisms, sterilization process and fundamental scientific technology.
{2}------------------------------------------------
The manufacturing process in regards to the coating is also unchanged with respect to the unmodified device.
#### INDICATION FOR USE:
The AF 620 Ph.I.S.I.O. with 20 micron screen with phosphoryIcholine coating and the AF 640 Ph.I.S.I.O. with 40 micron screen with phosphoryIcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
### TECHNOLOGICAL CHARACTERISTICS:
The AF 620 Ph.I.S.I.O. and AF 640 Ph.I.S.I.O. arterial filters have control mechanisms and operating principles when compared to the unmodified device. The pore size of the filter screen for the AF 620 Ph.I.S.I.O. has been reduced from 27 micron to 20 micron with respect to the D731 MICRO Ph.I.S.I.O. for improved filtration efficiency. The AF 640 Ph.I.S.I.O pore size (40 micron) is unchanged with respect to the unmodified device D733 MICRO Ph.I.S.I.O.
The size of the filter housing has been reduced and the polyester net of the filter screen is now double pleated rather than the single pleated filter of the unmodified devices.
The AF 620 Ph.I.S.I.O.,and AF 640 Ph.I.S.I.O. utilize the same filtering medium. The proposed AF 620/640 Ph.I.S.I.O. arterial filters present a different formulation of the same phosphorylcholine monomer. Such different formulation has been used in other products already cleared in US. No new materials are used as a result of these changes.
The current port orientation is parallel to the main axis so that the blood outlet flow is now parallel to the housing centerline
No change to the intended use has been made as a result of these modifications. There are no differences in packaging type and material between AF 620/640 Ph.I.S.I.O. arterial filters and the D731/D733 unmodified devices.
The AF 620/640 Ph.I.S.I.O. Arterial Filters are ethylene oxide sterilized and have a nonpyrogenic fluid path. Thev are for single use only.
{3}------------------------------------------------
# NON CLINICAL TEST RESULTS:
A complete battery of tests was carried out for the AF 640 Ph.I.S.I.O. Arterial Filter in accordance with the requirements of ISO 10993 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on raw materials.
Testing was performed on devices accelerated aged to an equivalent of three years real time aging. Sterility, ETO residuals, Hemolysis, Acute Systemic Toxicity, Mutagencial Cytotoxicity, Irritation, Sensitization and Hemocompatibility were conducted.
The results of the testing met established specifications.
## IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for Industry, dated November 29, 2000 necessary to demonstrates both the substantial equivalence with the unmodified devices and also to comply with safety and effectiveness requirements. The devices were aged up to 3 years (+ 1 additional year of aging in order to test a truly worst case) and tested for operating blood volume, structural integrity test, pressure integrity test, pressure drop, filter flow rate capacity, in vitro hemolysis/cell depletion, filtration efficiency, leaching of the coating and in handling characteristics. For comparative purposes all tests, when applinable, were performed on sterilized aged devices comparing the AF 620/640 Ph.I.S.I.O.a. herial filters vs. the unmodified devices operated at same max blood flow. The results of these tests met established specifications.
## CONCLUSIONS:
The AF 620 Ph.I.S.I.O. and AF 640 Ph.I.S.I.O. arterial filters perform in a manner substantially equivalent to the unmodified devices, D731 MICRO Ph.I.S.I.O. and D733 MICRO Ph.I.S.I.O. with respect to the functional parameters in common with the unmodified devices. Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.
In conclusion, test results of this study suggest the AF 620/640 Ph.J.S.I.O. arterial filters are equivalent to the D731/D733 MICRO Ph.I.S.I.O. arterial filters with respect to device function.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of an eagle or other bird with outstretched wings, rendered in a simple, graphic style.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
JAN 2 2 2010
Sorin Group Italia s.r.1. c/o Mr. Barry Sall Principal Consultant Parexel International Consulting 195 West Street Waltham, WA 02451
Re: K093986
AF 620 and AF640 Ph.I.S.I.O. Arterial Filters Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary Bypass Arterial Line Blood Filter Regulatory Class: Class II Product Code: DTM Dated: December 23, 2009 Received: December 24, 2009
Dear Mr. Sall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CIFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
AF 620 Ph.I.S.I.O. and AF 640 Ph.I.S.I.O. Arterial Filters Sorin Group Italia S.r.I.
Special 510(k) December 23, 2009
で
1093986 510(k) Number (if known):
:
Device Name: Sorin AF 620-640 Ph.I.S.I.O. Indication for Use:
The AF 620 Ph.I.S.I.O. with 20 micron screen with phosphoryIcholine coating and the AF 640 Ph.I.S.I.O. with 40 micron screen with phosphory choline coating are recommended for use in the arterial line of the extracorporeal nicuti during any procedure that requires cardiopulmonary bypass. The filters, are used to trap and remove gaseous emboli as well as particulate debris that may be introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
Prescription Use X (Part 21CFR 801 Subpart D)
Over-the- Counter Use AND/OR (21 CFR 807 Subpart C)
(Please do Not WRITE BELOW THIS LINE -- Continue on another PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | |
| 510(k) Number | K093986 |
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.