← Product Code [LIT](/productcode/LIT) · K093965

# ULTRAVERSE 014 AND 018 BALLOON DILATATION CATHETERS (K093965)

_C.R. Bard, Inc. · LIT · Mar 17, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K093965

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [LIT](/productcode/LIT.md)
- **Decision Date:** Mar 17, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters, are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popliteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

## Device Story

Over-the-wire PTA balloon dilatation catheter; features semi-compliant, low-profile balloon with two radiopaque marker bands for placement visualization; compatible with 0.014" and 0.018" guidewires. Used by physicians in clinical settings for peripheral artery angioplasty. Device is inserted into target vasculature; balloon is inflated to dilate stenotic vessels; marker bands assist in positioning. Benefits include restoration of blood flow in peripheral arteries. No software or electronic components.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through in vitro testing including balloon burst strength, fatigue, inflation/deflation, trackability, crossing profile, and mechanical integrity of bonds and hubs.

## Technological Characteristics

Over-the-wire catheter; semi-compliant balloon; radiopaque marker bands; compatible with 0.014" and 0.018" guidewires. Mechanical device; no energy source or software. Sterilization method not specified.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Ultraverse® Small Vessel PTA Balloon Dilatation Catheter ([K012913](/device/K012913.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters

510(k) Summary

# MAR 1 7 2010

## 21 CFR 807.92

As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the information upon which substantial equivalence determination is based as follows:

1. Submitter Information:

Applicant: Bard Peripheral Vascular, Inc.

1625 West Third Street

Tempe, Arizona 85281

Phone: 480-303-2662

Fax: 480-449-2546

Contact: Candace Wade, Regulatory Affairs Associate

Date: December 22, 2009

2. Subject Device:

| Device Trade Name:    | Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters |
|-----------------------|----------------------------------------------------------|
| Common or Usual Name: | Catheter, Angioplasty, Peripheral, Transluminal/         |
|                       | Catheter, Percutaneous                                   |
| Classification:       | Class II                                                 |
| Classification Panel: | Cardiovascular                                           |

#### 3. Predicate Device:

| Device Trade Name:      | Ultraverse® Small Vessel PTA Balloon Dilatation Catheter |
|-------------------------|----------------------------------------------------------|
| Cleared 510(K) Number:  | K012913                                                  |
| Date of Cleared 510(K): | 09/28/01                                                 |

#### 4. Summary of Change:

The modifications from the predicate device, the Ultraverse® Small Vessel PTA Balloon Dilatation Catheter, to the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters, were to the balloon design, catheter design, and performance characteristics.

{1}------------------------------------------------

## 5. Device Description:

The subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are small vessel balloon catheters consisting of an over the wire catheter with a balloon fixed at the distal tip. The semi-compliant, low profile balloon has two radiopaque marker bands to delineate the working length of the balloon and aid in balloon placement. The coaxial, over the wire catheter is compatible with 0.014" and 0.018" guidewires.

14093969

## 6. Indications for Use of Device:

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters, are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popliteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

#### 7. Technological Comparison to Predicate Device:

The technological characteristics of the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters, are substantially equivalent to those of the predicate device, the Ultraverse Small Vessel PTA Balloon Dilatation Catheter, in terms of intended use, indications for use, fundamental scientific technology, target population, operating principle, sterility assurance level and method of sterilization.

#### 8. Performance Testing Summary:

To demonstrate substantial equivalence of the subject device, the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters to the predicate device, the technological characteristics and performance criterion were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed:

- Catheter Length .
- Guidewire Inner Lumen Diameter .
- Shaft Outer Diameter .
- Balloon Outer Diameter at Operating Pressure ●
- Balloon Length .
- Marker Band Spacing/Alignment .
- . Tip Length
- . Tip Taper
- Guidewire Compatibility .

{2}------------------------------------------------

- Distensibility .
- Balloon Fatigue .
- . Balloon Inflation and Deflation
- Balloon Burst Strength .
- Balloon to Catheter Shaft Bond Strength .
- Joint Strength of Hub to Shaft .
- Hub Torsion Testing .
- Hub Stress Testing .
- Simulated Use: Introducer Sheath Compatibility .
- Catheter Shaft Leaks .
- Catheter Elongation .
- Crossing Profile .
- Simulated Use: Trackability 4
- Marker Band Visibility .
- . Compatibility with Standard Luer Locks
- Rewrapping with Stylet ●
- . Package Integrity
- . Dye Penetration of Packaging
- . Pouch Seal Strength

The results from these tests demonstrate that the technological characteristics and performance criteria of the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use.

### 9. Conclusions:

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are substantially equivalent to the legally marketed predicate device, the Ultraverse® Small Vessel PTA Balloon Dilatation Catheter.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three curved lines, with its arms outstretched. The logo is simple and uses a monochromatic color scheme.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

MAR ] 7 2010

Bard Peripheral Vascular, Inc. c/o Candace Wade Regulatory Affairs Associate 1625 West Third Street Tempe, AZ 85281

Re: K093965

Trade/Device Name: Ultraverse 014 and 018 PTA Balloon Dilatation Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: LIT, DOY Dated: February 5, 2010 Received: February 12, 2010

Dear Ms. Wade:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

https://www.fda.gov/MedicalDevices/ResourcesforYouIndustry/default.htm

Sincerely yours,

er R. V. Mmer

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 73

#### Indications for Use

510(k) Number (if known):

Device Name:

Indications for Use:

KO93965

Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popiteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

Prescription Use × (Part21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rma R. V. Lines

(Division Sign-Off) (Division Sign Silvascular Devices

510(k) Number K 093965

TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.

BARD

---

**Source:** [https://fda.innolitics.com/device/K093965](https://fda.innolitics.com/device/K093965)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
