← Product Code [KDI](/productcode/KDI) · K093964

# VENOFER PUMP (K093964)

_Renal Solutions, Inc. · KDI · Jul 9, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K093964

## Device Facts

- **Applicant:** Renal Solutions, Inc.
- **Product Code:** [KDI](/productcode/KDI.md)
- **Decision Date:** Jul 9, 2010
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Venofer® Pump Module is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.

## Device Story

The Venofer Pump Module is an optional accessory for the Fresenius 2008T Hemodialysis Machine. It is designed to administer Venofer (iron sucrose) to patients with chronic kidney disease undergoing hemodialysis. The module consists of a control panel, vial holder, fluid detector, and a peristaltic pump. It fits into the module compartment of the 2008T machine and cannot operate independently. The device delivers the drug into the extracorporeal blood set, not directly into the patient. It is operated by healthcare professionals in a clinical setting. By automating the delivery of iron sucrose, the device provides consistent dosing, assisting in the management of iron deficiency anemia, which is a common complication in dialysis patients. The healthcare provider monitors the delivery via the machine's interface, ensuring the drug is administered according to the physician's prescription.

## Clinical Evidence

No clinical data. The submission relied on non-clinical testing, including verification, safety, performance, and software testing of the Venofer Pump module as part of the Fresenius 2008T Hemodialysis System.

## Technological Characteristics

The device is an electromechanical accessory module featuring a peristaltic pump, fluid detector, vial holder, and control panel. It integrates into the Fresenius 2008T Hemodialysis Machine module compartment. It operates as a non-standalone system, relying on the host machine for power and control. The design leverages the existing Heparin Pump technology.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Heparin Pump (component of Allient 1500 Sorbent Hemodialysis System) ([K070739](/device/K070739.md))
- Allient 1500 Sorbent Hemodialysis System
- Fresenius 2008T Hemodialysis System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text '_K093964' on the top line and 'pg. 1 of 2' on the second line. The text is handwritten in a casual style. The text is black against a white background.

#### Venofer Pump 510(k) Premarket Notification

#### Section 5: 510(k) Summary

# 510(k) Summary

### JUL - 9 2010

| Official Contact      | David J. Vanella<br>Senior Vice President, Quality Systems<br>Renal Solutions, Inc.<br>770 Commonwealth Drive<br>Warrendale, Pa 15086<br>Phone: (724) 720-2840<br>FAX: (724) 772-6925 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date                  | July 8, 2010                                                                                                                                                                          |
| Classification Name   | Dialyzer, High Permeability With Or Without Sealed<br>Dialysate System                                                                                                                |
| Regulation Number     | 876.5860                                                                                                                                                                              |
| Product Code          | KDI                                                                                                                                                                                   |
| Common/Usual Name     | Hemodialysis System                                                                                                                                                                   |
| Proprietary Name      | Fresenius Model 2008T Hemodialysis Machine                                                                                                                                            |
| Predicate Devices     | Allient 1500 Sorbent Hemodialysis System<br>Fresenius 2008T Hemodialysis System                                                                                                       |
| Reason for submission | New Device                                                                                                                                                                            |

## Substantial Equivalence

The addition of the Venofer Pump Module on the Fresenius Model 2008T Hemodialysis Machine leverages the company's Heparin Pump design previously used in the Allient 1500 Sorbent Hemodialysis System. The proposed predicate device for the Venofer Pump Module is the Heparin Pump, a component of the Allient 1500 Sorbent Hemodialysis System that was cleared by the FDA in 2007 under K070739.

The Venofer Pump Module device is substantially equivalent to the predicate device in terms of the following:

- Intended use .
- Environment of use .
- . Operating principle
- Technology ●

Like the Heparin Pump, the Venofer Pump Module is not a stand alone device. The Venofer Pump Module cannot operate unless it is connected to the 2008T Hemodialysis Machine. In addition, similar to the Heparin Pump, the Venofer Pump delivers the contents of the vial into the extracorporeal blood set (not directly into

{1}------------------------------------------------

## Venofer Pump

._K093964
pg. 2 of 2

510(k) Premarket Notification

#### Section 5: 510(k) Summary

the patient). Therefore, similar to the Heparin Pump, the Venofer Pump is considered an accessory to the Fresenius Model 2008T Hemodialysis Machine and thus would also carry the same classification as KDI, 876.5860.

## Indication for Use

Fresenius Model 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.

### Intended Use/General Safety and Effectiveness

The Venofer® Pump Module is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.

### Non-Clinical Testing

The non-clinical testing submitted, referenced, and relied on in this 510(k) premarket notification included verification, safety, performance and software testing of the Venofer Pump module as part of the Fresenius 2008T Hemodialysis System. The conclusions drawn from this testing demonstrates that the Fresenius 2008T Hemodialysis System is as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices to which it was compared.

### Device Description

The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.

The Venofer Pump Module is intended to provide ease and consistent delivery of Venofer, an iron sucrose supplement, indicated for the treatment of iron deficiency anemia for Hemodialysis dependent-chronic kidney disease patients. Anemia commonly occurs in patients with chronic kidney disease undergoing Hemodialysis and effective anemia management is recognized as an important factor in improving the outcomes of these patients. Iron deficiency is a frequent contributing factor which complicates the treatment of anemia in chronic kidney disease patients. Iron sucrose is commonly delivered to improve iron status in patients who have chronic kidney disease. Venofer is an injectable iron preparation drug approved for treatment of iron deficiency and is the most commonly used drug to treat iron deficiency anemia in dialysis patients. Renal Solutions has developed the Venofer Pump Module that provides the means to deliver Venofer when undergoing Hemodialysis on a Fresenius 2008T Dialysis System.

The Venofer Pump Module fits into the module compartment of existing Fresenius 2008T Machines and consists of a control panel, vial holder, fluid detector, and a peristaltic pump.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its body, representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-Gt Silver Spring, MD 20993-0002

Mr. David J. Vanella Senior Vice President, Quality Systems Renal Solutions®, Inc. 770 Commonwealth Drive, Suite 101 WARRENDALE PA 15086

JUL - 9 2010

Re: K093964

Trade/Device Name: Fresenius 2008T Hemodialysis Machine with Venofer Pump Module Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system

Regulatory Class: II Product Code: KDI Dated: July 7, 2010

Received: July 8, 2010

Dear Mr. Vanella:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Venofer® Pump

#### 510(k) Premarket Notification

#### Section 4: Indications for Use Statement

## Indications for Use Statement

510(k) Number (if known): N/A

Device Name: Fresenius 2008T Hemodialysis Machine with Venofer Pump Module

Indications for Use:

Same as K080964

Fresenius 2008T is indicated for acute and chronic dialysis therapy.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Page 1

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal, and
Radiological Devices
510(k) Number K093964

---

**Source:** [https://fda.innolitics.com/device/K093964](https://fda.innolitics.com/device/K093964)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
