HALO 360 & SIZING BALLOON, MODEL 3441C

K093855 · Barrx Medical, Incorporated · GEI · Jan 15, 2010 · General, Plastic Surgery

Device Facts

Record IDK093855
Device NameHALO 360 & SIZING BALLOON, MODEL 3441C
ApplicantBarrx Medical, Incorporated
Product CodeGEI · General, Plastic Surgery
Decision DateJan 15, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HALO360+ Sizing Balloon model 3441C is intended to be used for the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. The HALO660+ Sizing Balloon model 3441C is indicated for use for the coagulation of bleeding and nonbleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

Device Story

The HALO360+ Sizing Balloon is a disposable, single-use catheter used to assess esophageal lumen size to facilitate selection of an appropriate ablation catheter. It consists of a sizing balloon, catheter shaft with markings, and an electrical connector. It is used in conjunction with HALO360 or HALORLEX energy generators. The device is inserted into the esophagus; the balloon is inflated to measure the lumen diameter. The sizing data helps clinicians select the correct size for the subsequent HALO360+ Coagulation Catheter. The device is intended for use by physicians in a clinical setting. The modification allows for measurement of diameters greater than 33.7 mm and identification of balloon migration into the stomach.

Clinical Evidence

Bench testing only. Performance data demonstrated that the device met specifications for sizing accuracy, structural strength, material compatibility, and sterility, confirming it is as safe and effective as the predicate devices.

Technological Characteristics

Consists of a sizing balloon, catheter shaft with markings, and electrical connector. Used with HALO360 or HALORLEX energy generators. Disposable, single-use. Modified balloon material. Measures esophageal lumen diameter. No software or complex algorithms described.

Indications for Use

Indicated for coagulation of bleeding and non-bleeding sites in the GI tract, including esophagus. Conditions include esophageal ulcers, Mallory-Weiss tears, arteriovenous malformations, angiomata, Barrett's esophagus, Dieulafoy lesions, and angiodysplasia.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### 510(k) SUMMARY 5093855 JAN 1 5 2010 ### BARRX Medical's [MODIFIED DEVICE] . ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared BARRX Medical Inc. 540 Oakmead Pkwv Sunnyvale, CA 94085 Phone: 408-328-7302 Facsimile: 408-328-7395 Contact Person: Viorica Filimon Date Prepared: December 20, 2009 # Name of Device and Name/Address of Sponsor HALO360+ Sizing Balloon BARRX Medical Inc. 540 Oakmead Pkwy Sunnyvale, CA 94085 ### Common or Usual Name Sizing Balloon ## Classification Name Product code: GEI CFR Section: 878.4400 Electrosurgical, cutting & coagulation & accessories Device Class: II Classification panel: General & Plastic Surgery #### Predicate Devices HALO360 Sizing Balloon model 3441B (K051168) HALO360+ Ablation Catheter model 32041-XX (K083711, K071543) # Purpose of the Special 510(k) notice. The HALO360+ Sizing Balloon model 3441C is a modification to HALO360 Sizing Balloon model 3441B and HALO360+ Ablation Catheter model 32041-XX the predicate devices. #### Intended Use 510(k) HALO360+ Sizing Balloon Model 3441C Page 93 {2}------------------------------------------------ The HALO360+ Sizing Balloon model 3441C is intended to be used for the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. The HALO660+ Sizing Balloon model 3441C is indicated for use for the coagulation of bleeding and nonbleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia. ## Technological Characteristics The HALO360+ Sizing Balloon model 3441C is used in conjunction with either the HALO360 Energy Generator models 1100C-115B (or 1100C-230B), or HALORLEX Energy Generator model 1190A-115A (or 1190A-230A) for assessing the size of the esophageal lumen, and facilitate the selection of the disposable single-use HALO360+ Coagulation Catheter. The HALO360+ Sizing Balloon model 3441C, like the predicate device HALO360 Sizing Balloon model 3441B is comprised of a sizing balloon, a catheter shaft with markings, and an electrical connector. The HALO360+ Sizing Balloon model 3441B is substantially equivalent in design, performance, and mode of operation with the already cleared predicate devices: HALO360 Sizing Balloon model 3441B and HALO360+ Ablation Catheter model 32041-XX. ### Performance Data The HALO360+ Sizing Balloon model 3441C met the same specifications requirements as the HALO360 Sizing Balloon and HALO360+ Ablation Catheter for the following characteristics: - 0 Sizing accuracy - Structural strength . - Material compatibility - Sterility . ## Substantial Equivalence HALO360+ Sizing Balloon model 3441C has the same intended use and indications, principles of operation, and technological characteristics as HALO360 Sizing Balloon model 3441B and HALO360+ Ablation Catheter model 32041-XX. The minor differences in the HALO360+ Sizing Balloon model 3441C are: - . Changes in the balloon material - Increase the sizing balloon capability to measure esophageal diameter . higher than 33.7 mm and as result identify when the balloon migrates in the stomach. These changes do not raise any new questions of safety or effectiveness. Performance data demonstrates that the HALO360+ Sizing Balloon is as safe and effective as HALO360 Sizing Balloon model 3441B and HALO360+ Ablation catheter model 32041-XX. Thus, the HALO360+ Sizing Balloon model 3441B is substantially equivalent to its predicate devices. {3}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 BARRX Medical, Inc. % Ms. Viorica Filimon Vice President of Quality/Regulatory Affairs 540 Oakmead Parkway Sunnyvale, California 94085 JAN 1 5 2010 Re: K093855 Trade/Device Name: HALO360 + Sizing Balloon Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: December 15, 2009 Received: December 16, 2009 Dear Ms. Filimon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Viorica Filimon If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Singular Sincerely yours, Mark N. Melkerson Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known):___ Device Name: HALO360 + Sizing Balloon Indications for Use: HALO360 + Sizing Balloon model 3441C is indicated for the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia. Prescription Use X (Per 21 C.F.R. 801.109) Subpart C) AND/OR Over-The-Counter Use (Per 21 C.F.R. 807 # (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) FOR M. MELKERSON (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K093855 510(k) HALO360+ Sizing Balloon Model 3441C Page 99
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...