BE micro/Trea intended use is collecting and recording physiological data to be used in neurology (EEG), polysomnography (PSG) and sleep disorder studies. The BE micro is intended for pediatric to adult population , and can be used either home or hospital environments The BE micro/Trea is only to be used under the direction or supervision of a physician, or other trained health care professional.
Device Story
BE micro/Trea is a portable, multi-channel recording system for EEG and PSG studies. It captures and digitally amplifies physiological signals from patient sensors; data is stored on internal NAND flash memory or transmitted to a PC via USB. Designed for wearable use, allowing patient mobility during attended or unattended recordings. Clinicians use the system to collect data for neurological and sleep disorder diagnosis. Output includes raw physiological waveforms and derived metrics like Heart Rate and Pulse Transit Time (PTT). The device is operated by clinicians or patients (under supervision) in hospital or home environments. Healthcare providers review the recorded data to assist in clinical decision-making and diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was verified through software validation and verification (V&V) testing and electrical safety/electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2 standards. Results confirmed the device functions as intended.
Technological Characteristics
Portable, wearable multi-channel recorder. Battery or USB powered. 16-bit A/D resolution. Sampling rate up to 2048 samples/s. Internal NAND flash memory storage. Connectivity via USB bridge. Includes impedance check. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-26, CSA C22.2 No 601.1-M90, and UL 60601-1.
Indications for Use
Indicated for pediatric to adult patients requiring collection and recording of physiological data for neurology (EEG), polysomnography (PSG), and sleep disorder studies. Use restricted to physician or trained healthcare professional supervision in home or hospital settings.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
# K093728 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92.
#### 807 92(a)(1)
## Submitter Information
EBNeuro, S.p.A. Via Pietro Fanfani 97/A Florence, Italy 50127
Contact Person: Jamie Austin, RAC Regulatory Consultant, Anson Group Phone: (317) 569-9500 x 118 Facsimile: (317) 569-9520
Date: December 2, 2009
#### 807.92(a)(2)
Trade Name: BE micro/Trea
Regulation Number: 21 CFR 882.1400
Regulation Name(s): Electroencephalograph
Regulatory Class: Class II
Product Code: OLV
Additional Product Codes: GWQ, GWL, MNR, and DQA
# 807.92(a)(3)
# Predicate Device(s)
K# K010460 K052066 K061996
Name Track-it Netlink Traveler Sandman Pocket Manufacturer
Lifelines Ltd Bio-logic System Corp. EBNeuro S.p.A.
Additional substantial equivalence information is provided below.
APR 2 9 2010
{2}------------------------------------------------
#### 807.92 (a)(4)
# Device Description
The BE micro is manufactured by EBNeuro and intended for use in collecting and recording physiological data to be used in diagnosis in EEG (electroencephalography) and PSG (sleep disorder studies) clinical investigation.
The BE micro is a portable multi-channel recording system that capture and digitally amplifies neurological (EEG) and/or Polysomnography (PSG) physiological activity and records the acquired data onto an internal NAND flash memory chip and/or sends those data directly to a PC via the USB port. The BE micro system is a small, lightweight recording system that is comfortable for the patient and simple for the clinician to set up.
The Amplifier can be worn by the patient allowing in such way the patient to freely move or sleep while recordings are being made, allowing a more comfortable and efficient test.
#### 807.92(a)(5)
#### Intended Use(s)
BE micro/Trea intended use is collecting and recording physiological data to be used in neurology (EEG), polysomnography (PSG) and sleep disorder studies. The BE micro is intended for pediatric to adult population , and can be used either home or hospital environments
The BE micro/Trea is only to be used under the direction or supervision of a physician, or other trained health care professional.
#### 807.92(a)(6)
# Technological Characteristics
The BE micro/Trea system utilizes the same technological characteristics as the predicate devices. A comparison of these characteristics can be summarized as follows:
- . Design
- o BE micro/Trea and predicate(s) are designed for the same target populations and environments of use.
- o BE micro/Trea and predicate(s) utilize same type of operator interface.
- o BE mirco/Trea and predicate(s) have same type of configuration.
- BE micro/Trea and predicate(s) have same/similar sampling rates, data o storing, data compression, A/D resolution, and maximum number of channels.
- . Material
- o BE mirco/Trea and predicate use same microprocessor.
- Energy source .
- o BE micro/Trea and predicate(s) utilize same type of energy source(s).
{3}------------------------------------------------
## Summary of Substantial Equivalence
EBNeuro believes that the BE micro/Trea system is substantially equivalent to the Track-it (K010460), Netlink Traveler (K052066) and Sandman Pocket (K061996). Please see table below for detailed comparison of the specifications, characteristics, and performance/safety testing of the BE micro/Trea to the respective predicate devices.
#### 807.92(b)(1)
# Summary of Non-Clinical Performance Data
The following non-clinical performance data were gathered:
- . Data to verify and validate the measurement algorithm and software functionality.
- o RESULT: PASS
- . Data to evaluate electrical safety and electromagnetic compatibility (per IEC 60601-1, IEC 60601-1-2, and other standards).
- o RESULT: PASS
# Non-Clinical Performance Data Conclusions
Software V&V testing demonstrated that the software functions as intended. Electrical safety and electromagnetic compatibility testing demonstrated that, like the referenced predicate device(s), the BE micro/Trea complies with IEC 60601-1, IEC 60601-1-2, and other safety/EMC standards (per table below).
{4}------------------------------------------------
| Product Characteristic | BEmicro (Trea)<br>Submission device | Track-it<br>(Predicate Device) | Netlink Traveler<br>(Predicate Device) | Sandman Pocket<br>(Predicate device) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | EBNeuro S.p.A. | Lifelines Ltd | Bio-logic System Corp. | EBNeuro S.p.A. |
| 510(k) number | TBA | K010460 | K052066 | K061996 |
| Device class | Class II | Class II | Class II | Class II |
| Intended use | BE micro intended use is<br>collecting and recording<br>physiological data to be used<br>in neurology (EEG),<br>polysomnography (PSG) and<br>sleep disorder studies. The<br>BE micro is intended for<br>pediatric to adult population,<br>and can be used either home<br>or hospital environments | The lifelines Track it<br>Recorder is a 36 channel<br>ambulatory<br>electroencephalograph<br>that is designed for use in<br>a variety of monitoring<br>application to record<br>physiological data for<br>EGG and Sleep studies | The Bio-logic Netlink Traveler is<br>indicated for use in the recording and<br>analysis of EEG tests. Typical routine<br>EEG tests are 20-30 minutes in<br>duration, but the Netlink Traveler can<br>also be used for longer tests, including<br>continuous long-term EEG monitoring<br>with patient video. Similarly,<br>the Netlink Traveler is indicated for use<br>in the recording and analysis of human<br>physiological data necessary for the<br>diagnosis of Sleep-related disorders. It<br>is intended to record and present this<br>data in a form that can improve the<br>speed of diagnosis and assist in<br>potential treatment decisions. In<br>general, EEG and Sleep testing is<br>indicated for use whenever it is<br>necessary to measure and record a<br>patient's electrophysiological activity,<br>including the electrical activity of the<br>brain, by attaching multiple electrodes<br>at various locations on the body. | Intended for use in<br>collecting and recording<br>physiological data to be<br>used in<br>polysomnography and<br>sleep disorder studies.<br>For use in either home<br>or hospital environments<br>with a pediatric through<br>adult patient population. |
| Target population | Pediatric through adult | Patients of all ages, from<br>children to adults. | Patients of all ages, from<br>children to adults, including<br>geriatric patients. | Pediatric through adult<br>(including all pediatric<br>subpopulations) |
| Environment of use | Hospital and home | Hospital and home | Hospital and home | Hospital and home |
| Product Characteristic | BEmicro (Trea)<br>Submission device | Track-it<br>(Predicate Device) | Netlink Traveler<br>(Predicate Device) | Sandman Pocket<br>(Predicate device) |
| Recording modality | Attended and Unattended | Attended and Unattended | Attended and Unattended | Attended and Unattended |
| Prescription status | Available only on the order<br>of a physician. | Available only on the order<br>of a physician. | Available only on the order of a<br>physician. | Available only on the<br>order of a physician. |
| Power | Battery powered or USB<br>powered | Battery powered | Battery Powered | Battery powered or USB<br>powered |
| Communication<br>Interfaces | Physiological signals are<br>sent from the patient<br>sensors to the device<br>through the sensor cables.<br>In the recorder the data are<br>stored in flash memory in<br>both attended and<br>unattended studies. During<br>attended studies, the data is<br>also transmitted to a<br>computer in real-time via a<br>USB bridge. After<br>unattended studies, data<br>can be downloaded from<br>the recorder using the<br>USB bridge. | Physiological signals are<br>sent from the patient<br>sensors to the device<br>through the sensor cables.<br>In the recorder the data are<br>stored in flash memory in<br>both attended and<br>unattended studies. During<br>attended studies, the data<br>is also transmitted to a<br>computer wireless<br>(Bluetooth) interface.<br>After unattended studies,<br>data can be downloaded<br>from the recorder using<br>the Bluetooth connection | Physiological signals are sent<br>from the patient sensors to the<br>yoke through the sensor<br>cables.<br>The yoke sends the data to the<br>recorder where the data is<br>stored in flash memory in both<br>attended and unattended<br>studies. During attended<br>studies, the data is also<br>transmitted to a computer in<br>real-time via a Ethernet LAN<br>connection. After unattended<br>studies, data can be<br>downloaded from the recorder<br>using the LAN connection | Physiological signals are<br>sent from the patient<br>sensors to the yoke<br>through the sensor<br>cables.<br>The yoke sends the data<br>to the recorder where the<br>data is stored in flash<br>memory in both attended<br>and unattended studies.<br>During attended studies,<br>the data is also<br>transmitted to a<br>computer in real-time<br>via a USB cable. After<br>unattended studies, data<br>can be downloaded from<br>the recorder using a<br>USB cable. |
| Operator Interface | LCD Display | LCD Display | LCD Display | LCD Display |
| Microprocessor | TMS320UC5409 | Unknown | Unknown | TMS320UC5402 |
| Product Characteristic | BEmicro (Trea)<br>Submission device | Track-it<br>(Predicate Device) | Netlink Traveler<br>(Predicate Device) | Sandman Pocket<br>(Predicate device) |
| A/D Resolution | 16 bit | 16 bits | 18 bit | 16 bit |
| Data Storing | On internal NAND flash<br>chip | PCMCIA type II Compact<br>flash card | Compact flash memory | On internal NAND flash<br>chip |
| Data compression | No | No | Yes | No |
| Configuration | Wearable | Wearable | Wearable | Wearable |
| Sensors | Commercially available<br>sensors only | Commercially available<br>sensors only | Commercially available<br>sensors only | Commercially available<br>sensors only |
| Maximum number of<br>channels | 21 | 36 | 36 | 22 |
| Derived channel | Pulse Transit Time (PTT)<br>- A calculation of the time<br>between the occurrence of<br>the R-wave on the EKG<br>and 50% ascending slope<br>on the plethysmogram.<br>Heart rate - Derived from<br>the ECG channel | Heart rate - Derived from<br>the ECG channel | Heart rate - Derived from the<br>ECG channel | Pulse Transit Time<br>(PTT) - A calculation of<br>the time between the<br>occurrence of the R-<br>wave on the EKG and<br>50% ascending slope on<br>the plethysmogram.<br>Heart rate - Derived<br>from the ECG channel |
| Oximetry (option) | Yes | Yes | Yes | Yes |
| Sampling rate | Up to 2048 samples/s | 1-256 sample/s selectable | 256, 512 samples/s | up to 2048 samples/s |
| Product Characteristic | BEmicro (Trea)<br>Submission device | Track-it<br>(Predicate Device) | Netlink Traveler<br>(Predicate Device) | Sandman Pocket<br>(Predicate device) |
| Passbands | 0.5 – 128 Hz | 0.16 – 70 Hz | 0.1 – 100 Hz | 0.1 – 256Hz<br>0.1 – 64 Hz<br>DC – 10 Hz<br>(depending on channel type) |
| CMRR | >100 dB | >100 dB | >100 dB | >100 dB |
| Impedance check | Yes | Yes | Yes | Yes |
| Safety | Device complies with the<br>International Standard<br>IEC 60601-1<br>IEC 60601-1-4<br>IEC 60601-1-26<br>CSA C22.2 No 601.1-M90<br>UL STD No 60601-1 | Unknown | Device complies with the<br>International Standard<br>IEC 601-1<br>UL 2601<br>UL 60601<br>CSA C22.2 No 601-1 | Device complies with<br>the International<br>Standard<br>IEC 60601-1<br>IEC 60601-1-4<br>IEC 60601-1-26<br>CSA C22.2 No 601-1-<br>M90<br>UL 2601 |
.
.
،
{5}------------------------------------------------
:
ﺗ
·
:
.
{6}------------------------------------------------
. :
{7}------------------------------------------------
:
、
・
:
.
.
{8}------------------------------------------------
| Product Characteristic | BEmicro (Trea)<br>Submission device | Track-it<br>(Predicate Device) | Netlink Traveler<br>(Predicate Device) | Sandman Pocket<br>(Predicate device) |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electromagnetic<br>Compatibility (EMC) | Device complies with the<br>International Standard<br>IEC 60601-1-2, including<br>the following standards:<br>IEC 61000-3-2<br>IEC 61000-3-3<br>IEC 61000-4-2<br>IEC 61000-4-3,<br>IEC 61000-4-4,<br>IEC 61000-4-5,<br>IEC 61000-4-6,<br>IEC 61000-4-8<br>IEC 61000-4-11<br>EN55011 class B<br>EN55014-1 | Unknown | Unknown | Device complies with<br>the International<br>Standard<br>IEC 60601-1-2,<br>including the following<br>standards:<br>IEC 61000-3-2<br>IEC 61000-3-3<br>IEC 61000-4-2<br>IEC 61000-4-3,<br>IEC 61000-4-4,<br>IEC 61000-4-5,<br>IEC 61000-4-6,<br>IEC 61000-4-8<br>IEC 61000-4-11<br>EN55011 class B<br>EN55014-1 |
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
EB Neuro S.P.A. c/o Ms. Jamie L. Austin, RAC The Anson Group. LLC 11460 North Meridian Street, Suite 150 Carmel, IN 46032
Re: K093728
APR 2 9 2010
Trade/Device Name: BE Micro Trea Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLV, GWQ, GWL, MNR, and DQA Dated: March 22, 2010 Received: March 24, 2010
Dear Ms. Austin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
Page 2 - Ms. Jamie L. Austin, RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including; but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Runk Thumpton
Malvina B. Eydelman Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use
510(k) Number (if known): K o93728
Device Name:______________ (Trea -__OEM version )
Indications For Use:
BE micro/Trea intended use is collecting and recording physiological data to be used in neurology (EEG), polysomnography (PSG) and sleep disorder studies. The BE micro is intended for pediatric to adult population, and can be used either home or hospital environments
The BE micro/Trea is only to be used under the direction or supervision of a physician, or other trained health care professional.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.