NEUTRASAL
K093642 · Invado Pharmaceuticals · LFD · Dec 11, 2009 · DE
Device Facts
| Record ID | K093642 |
| Device Name | NEUTRASAL |
| Applicant | Invado Pharmaceuticals |
| Product Code | LFD · DE |
| Decision Date | Dec 11, 2009 |
| Decision | SESE |
| Submission Type | Special |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
(OTC): Neutrasal® is indicated for the dryness of the mouth (hyposalivation, xerostomia) (OTC): Neutrasal® is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atropine or other anticholinergic agents that suppress salivary secretion. (OTC): Neutrasal® may be used as part of an oral hygiene program for patients with dry mouth. Neutrasal® also provides intensive hygiene of the oral cavity. (Rx): Neutrasal® is also indicated as an adjunct to standard oral care in relieving the discomfort associated with oral mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain. (Rx): Neutrasal® may be used for relief of dryness of the oral mucosa when hyposalivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy, infection or dysfunction of the salivary glands; fever; emotional factors such as fear or anxiety; obstruction of the salivary ducts; Sjögren's syndrome
Device Story
Neutrasal is a powder formulation containing calcium chloride, sodium phosphate, sodium bicarbonate, and silica gel. When dissolved in tap water, it creates a supersaturated calcium and phosphate oral rinse. Used by patients in home or clinical settings to restore oral ionic/pH balance, maintain moisture, and loosen mucus. Effervescent formulation enhances mucosal penetration. Provides relief from dryness, fissuring, and painful tongue conditions; facilitates chewing and speaking; improves denture adherence. Acts as an adjunct to standard oral care for mucositis pain relief. No complex electronics or software involved.
Clinical Evidence
Bench testing only. Performance testing included assessment of physical properties, pH, dissolution, activity of the supersaturated state, and content testing of active ingredients. Biocompatibility is claimed to be identical to the predicate device.
Technological Characteristics
Powder formulation consisting of calcium chloride, sodium phosphate, sodium bicarbonate, and silica gel (anti-caking agent). Dissolves in water to form a supersaturated calcium and phosphate electrolyte solution. Non-sterile. No software or electronic components.
Indications for Use
Indicated for patients with dry mouth (hyposalivation, xerostomia) caused by medications (antihistamines, atropine, anticholinergics), surgery, radiotherapy, chemotherapy, infection, salivary gland dysfunction, fever, emotional factors, salivary duct obstruction, or Sjögren's syndrome. Also indicated as an adjunct for oral mucositis discomfort.
Predicate Devices
Submission Summary (Full Text)
{1}------------------------------------------------
#### 5. 510(k) Summary K093642
| A. From: | Edward Kobus | (Owner) | DEC 11 2009 |
|-------------|-------------------------|---------|-------------|
| | 25 Ravenna Drive | | |
| | Pomona, NY 10970 | | |
| | Invado Pharmaceuticals | | |
| | 914-715-6232 | | |
| | ekobus@invadopharma.com | | |
| Trade Name: | Neutrasal® | | |
Powder for Supersaturated Calcium Phosphate Rinse Common Name:
Classification Names: Saliva, Artificial
# B. Regulatory Information:
Unclassified Device Class:
LFD Product Code:
- K030802, K991938 C. Predicates:
Caphosol ®, Artificial Saliva
Product Code: LFD
# K091718
Neutrasal
Product Code" LFD
{2}------------------------------------------------
# D. Device Description:
Neutrasal® is a powder that when dissolved in normal tap water becomes a supersaturated calcium and phosphate rinse. Neutrasal® was designed in part to replace the normal ionic and pH balance in the oral cavity. Neutrasal® also maintains moistness in the oral cavity. It relieves diffuse dryness and fissuring of the oral mucosa, as well as painful tongue conditions due to hyposalivation.
Neutrasal® facilitates chewing and speaking; loosens tough mucus; prevents mucous membranes from sticking together; and improves adherence of dentures.
Additionally, Neutrasal® can be used as an adjunct to standard oral care for relieving the discomfort associated with oral mucositis that may be caused by high dose chemotherapy or radiation. Relief of dryness of the oral mucosa in these conditions is associated with an amelioration of pain.
Neutrasal® is a comprised of calcium chloride, sodium phosphate and sodium bicarbonate plus inactive ingredients. When Neutrasal® is dissolved in normal tap water, the water becomes supersaturated with both calcium and phosphate ions.
Silica gel, an inactive ingredient in Neutrasal, is an amorphous silica that is recommended for free-flow and anti-caking for hydrogscopic food powders, industrial powders and other applications (pharmaceuticals) where humidity must be kept to a minimum.
{3}------------------------------------------------
| Product | Method of<br>Use | Applications<br>per day | Type of<br>Product | Presentation |
|------------|--------------------------------|-------------------------------------------------------|-------------------------|--------------|
| Neutrasal® | Dissolve<br>Powder in<br>Water | 2-10 as<br>directed by<br>physician and<br>indication | Electrolyte<br>Solution | Non-Sterile |
## E. Indications for Use:
(OTC): Neutrasal® is indicated for the dryness of the mouth (hyposalivation, xerostomia)
(OTC): Neutrasal® is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atropine or other anticholinergic agents that suppress salivary secretion.
(OTC): Neutrasal® may be used as part of an oral hygiene program for patients with dry mouth. Neutrasal® also provides intensive hygiene of the oral cavity.
(Rx): Neutrasal® is also indicated as an adjunct to standard oral care in relieving the discomfort associated with oral mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.
(Rx): Neutrasal® may be used for relief of dryness of the oral mucosa when hyposalivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy, infection or dysfunction of the salivary glands; fever; emotional factors such as fear or anxiety; obstruction of the salivary ducts; Sjögren's syndrome
### F. Manufacturing
Neutrasal® will be manufactured according to product specifications and under the guidelines of current Good Manufacturing Practices (cGMP). Risk analysis has been performed to identify possible failure mode during manufacturing. Manufacturing controls have been developed
{4}------------------------------------------------
and implement to address the identified risks factors based on the criticality of failure mode. All established GMPs will assure that device manufactured meets all established specifications prior to release and is safe and effective for its intended use.
# G. Summary of Non-Clinical Testing: Performance
The Performance Testing of Neutrasal® included the assessment of the physical properties of Neutrasal® and its ability to achieve its intended use. Bench testing was completed by Scientech Laboratories, Inc., an FDA-approved analytical laboratory testing service as well internal testing done. The following bench tests were performed:
- . pH
- Dissolution Tests .
- Activity Testing of Supersaturated State .
- Content Testing of Active Ingredients .
### H. Biocompatibility testing
Neutrasal® is identical to the 510(k)-approved product predicate product Neutrasal.
# I. GRAS Listing of Ingredients
The EAFUS list of substances contains ingredients added directly to food or medications that FDA has either approved as food additives or listed or affirmed as GRAS (generally recognized as safe). All ingredients contained in Neutrasal® are EAFUS approved food additives labeled as GRAS. Only USP-approved chemicals, with validation, will be used during the manufacturing of Neutrasal®.
{5}------------------------------------------------
#### Other Information J.
Effervescence has proved it utility as an oral drug delivery system that offers advantages to other forms of drug delivery. Neutrasal® offers more portability because it is more easily transported than liquid medication because no water is added until the product is ready to use. Additionally, Neutrasal® offers improved penetration in to the oral mucosa because of the effervescent formulation enhancement of sodium bicarbonate.
# K. Substantial Equivalence Conclusion
The indications of use, technological properties, performance testing for Neutrasal® (Powder for supersaturated calcium phosphate rinse) are identical to Neutrasal (supersaturated calcium phosphate rinse). They both share the same labeling and intended uses.
{6}------------------------------------------------
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
020 1 1 2009
Mr. Edward D. Kobus President Invado Pharmaceuticals 25 Ravenna Drive Pomona, New York 10970
Re: K093642
Trade/Device Name: Neutrasal® Regulation Number: Unclassified Regulation Name: None Regulatory Class: None Product Code: LFD Dated: November 19, 2009 Received: November 24, 2009
Dear Mr. Kobus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2- Mr. Kobus
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reporta Problem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
fir
Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### 4. Indications for Use
K093642 510(k) Number (if known):
Device Name:
Neutrasal® Supersaturated Calcium Phosphate Oral Rinse
Indications for Use:
(OTC): Neutrasal® is indicated for the dryness of the mouth (hyposalivation, xerostomia)
(OTC): Neutrasal® is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atropine or other anticholinergic agents that suppress salivary secretion.
(OTC): Neutrasal® may be used as part of an oral hygiene program for patients with dry mouth. Neutrasal® also provides intensive hygiene of the oral cavity.
(Rx): Neutrasal® is also indicated as an adjunct to standard oral care in treating the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.
(Rx): Neutrasal® may be used for relief of dryness of the oral mucosa when hyposalivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy, infection or dysfunction of the salivary glands; fever; emotional factors such as fear or anxiety; obstruction of the salivary ducts; Sjögren's syndrome;
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
RoBetz DDS for Dr. KP. Medley (Acting)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control D
510(k) Number: K093642