MODELS FORA V30 AND TD-4242 BLOOD GLUCOSE MONITORING SYSTEMS

K093635 · Taidoc Technology Corporation · NBW · Feb 18, 2010 · Clinical Chemistry

Device Facts

Record IDK093635
Device NameMODELS FORA V30 AND TD-4242 BLOOD GLUCOSE MONITORING SYSTEMS
ApplicantTaidoc Technology Corporation
Product CodeNBW · Clinical Chemistry
Decision DateFeb 18, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The FORA V30 Blood Glucose Monitoring System/TD-4242 Blood Glucose Monitoring System are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the FORA V30 Blood Glucose Monitoring System /TD-4242 Blood Glucose Monitoring System can be used only during steady-state blood glucose conditions. This meter has some speaking functions but is not intended for use by the visually impaired.

Device Story

System measures glucose in fresh capillary whole blood via electrochemical biosensor (glucose oxidase). Components include meter, test strips, control solution, lancing device, and lancets. Used by healthcare professionals or patients at home to monitor diabetes control programs. Device features speaking functions for audio output, though not intended for visually impaired users. Output provides quantitative glucose concentration to assist in diabetes management. Modifications from predicate include software updates, addition of voice chip/speaker, and physical appearance changes.

Clinical Evidence

No clinical data provided. Substantial equivalence established via design control activities, risk analysis, and verification/validation of device modifications.

Technological Characteristics

Glucose meter; electrochemical sensing principle; powered by two AAA alkaline batteries; includes voice chip and speaker for audio output; software-controlled display and unit settings (fixed mg/dL).

Indications for Use

Indicated for use by patients for the quantitative measurement of glucose in capillary whole blood as an aid in monitoring the effectiveness of diabetes control programs.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIVD Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER: k093635 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) Taidoc Technology Corporation, FORA G30 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4241 (k090187) 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: A. Modification to the software to: remove the one pre-set code from displaying on the screen to having no coding information displayed when the test strip is inserted and to set mg/dL as a preset setting so users can not change the unit of measurement. B. Addition of some speaking functions (including the addition of a voice chip and a speaker to the device) but not enough to support use by visually impaired users. C. The power source was changed from one CR2032 battery to two AAA alkaline batteries. D. The physical appearance of the meter has changed. E. The name of the device has changed. F. The labeling was modified to reflect the changes to the device. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...