INFINITY ACUTE CARE SYSTEM WORKSTATION CRITICAL CARE
K093633 · Draegerwerk AG & CO Kgaa · CBK · Mar 30, 2010 · Anesthesiology
Device Facts
Record ID
K093633
Device Name
INFINITY ACUTE CARE SYSTEM WORKSTATION CRITICAL CARE
Applicant
Draegerwerk AG & CO Kgaa
Product Code
CBK · Anesthesiology
Decision Date
Mar 30, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The Infinity Acute Care System Workstations Critical Care consist of monitoring and control displays and additional therapy units. They are intended to be used as integrated, networked, and configurable workstations to provide critical care specific therapy. The Infinity Acute Care System Workstations Critical Care are intended to be used by qualified and trained medical personnel. The Infinity C Series Medical Cockpits, consisting of the C500 and the C700, are monitoring and control displays for the Infinity Acute Care System (IACS). Medical Cockpits are intended to be used to monitor waveforms, parameter information, and alarms as well as to control settings. The Infinity Series Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare professionals. The Evita V500 ventilation unit of the Infinity Acute Care System is intended for the ventilation of adult, pediatric and neonatal patients. Evita V500 offers mandatory ventilation modes and ventilation modes for spontaneous breathing support and airway monitoring. The Evita V500 ventilation unit is used with Infinity C Series Dräger Medical Cockpits. The Evita V500 ventilation unit is intended for use in different medical care areas. Evita V500 is intended for stationary use in hospitals and medical rooms or for patient transportation within the hospital.
Device Story
Infinity Acute Care System (IACS) integrates Infinity Medical Cockpit C500/C700 and Evita Infinity V500 ventilator; optional GS500 gas supply unit. System provides critical care ventilation and patient monitoring. Ventilator inputs include patient airway pressure/flow; microprocessor-controlled system executes various volume-controlled (VC), pressure-controlled (PC), and spontaneous (Spn) breathing modes. Cockpit displays waveforms, parameters, and alarms; allows clinician control of ventilation settings. Used in hospitals/medical rooms or during in-hospital transport by trained healthcare professionals. Output enables real-time patient status assessment and therapy adjustment; supports clinical decision-making for respiratory management; benefits patients through precise, synchronized, and automated ventilation support.
Clinical Evidence
Bench testing only. Performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-12, IEC 60601-1-6, IEC 60601-1-8, EN ISO 14971, and EN ISO 17664. No clinical data provided.
Technological Characteristics
Microprocessor-controlled ventilator; modular system with Medical Cockpit display and ventilation unit. Connectivity: Networked/integrated workstation. Standards: IEC 60601-1, 60601-1-2, 60601-2-12, 60601-1-6, 60601-1-8, ISO 14971, ISO 17664. Features include flow triggering, inspiratory termination, and automatic apnea ventilation.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring mechanical ventilation and monitoring in hospital settings. Used by qualified medical personnel for critical care therapy, including mandatory ventilation, spontaneous breathing support, and airway monitoring.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Related Devices
K041775 — GE DATEX-OHMEDA ENGSTROM CARESTATION · Datex-Ohmeda, Inc. · Oct 26, 2004
K131252 — PURITAN BENNETT 980 SERIES VENTILATOR SYSTEM · Covidien · Feb 24, 2014
K050597 — GE DATEX-OHMEDA ENGSTROM CARESTATION · Datex-Ohmeda, Inc. · Mar 25, 2005
K162738 — Puritan Bennett 980 Ventilator System · Covidien · Oct 27, 2017
K222024 — Evita V800, Evita V600 · Draegerwerk AG & CO Kgaa · May 16, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
**Drägermed**
A Dräger and Siemens Company
# 510(k) Summary according to 21 CFR 807.92
Appendix 14.1
### Applicants Name and Address:
Draeger Medical AG & Co. KG Moislinger Allee 53-55 23542 Luebeck Germany
Establishment Registration Number: 9611500
Contact Person:
Ulrich Schröder Director of Regulatory & Clinical Affairs
Tel. No.: + 49 (451) 882-3648 Fax No.: + 49 (451) 882-3018
#### Applicants US Contact Person
Ms. Joyce Kilroy VP Processes, Quality & Regulatory Affairs
Tel. No.: (215) 660-2626 Fax No.: (267) 885-9989
#### Date submission was prepared:
March 2010
#### Device Name:
Trade Name: Common Name: Infinity Acute Care System Workstation Critical Care Intensive Care Ventilator
MAR 3 0 2010
、「
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for Dräger medical. The logo is in black and white, with the word "Dräger" in a bold, sans-serif font. The word "medical" is in a thinner, sans-serif font. Below the logo, there is a tagline that reads "A Dräger and Siemens Company."
#### General description of the device
The Infinity Acute Care System Workstation Critical Care provides critical-care specific functionality. It is made up of the Infinity Medical Cockpit C500 and the ventilation unit Evita Infinity V500. The optional gas supply unit GS500 can be added. The ventilation unit provides critical-care specific ventilation and monitoring data of ventilation parameters. The Infinity Medical Cockpit is the control and display unit witch runs a ventilation application. The patient monitoring data are displayed on the Infinity Medical Cockpit, which is also used to control ventilation settings including alarms. The optional gas supply unit provides compressed air to the ventilation unit in case of central gas failure or in house transport.
#### Infinity Medical Cockpit
The user interface of the Infinity Acute Care System Workstation Critical Care is the Infinity Medical Cockpit, a standardized display and control unit for the connected monitoring and therapy units.
The Infinity Medical Cockpit is a standard platform, using common hardware, software, and user interface components to facilitate ease of use for clinicians.
#### Evita Infinity V500
The ventilation unit of the Infinity Acute Care System is a microprocessor-controlled ventilator. The Evita Infinity V500 provides overpressure ventilation and adjustable oxygen concentration with pressure- and volume-controlled automatic and spontaneous breathing modes:
· volume-controlled (VC):
- · VC-SIMV (Synchronized Intermittent Mandatory Ventilation)
- o VC-AC (Assisted Controlled)
- o VC-CMV (Continuous Mandatory Ventilation)
- o VC-MMV (Mandatory Minute Volume Ventilation)
- · pressure-controlled (PC):
- o PC-SIMV (Synchronized Intermittent Mandatory Ventilation)
- o PC-SIMV+ (Pressure Control-biphasic positive airway pressure, spontaneous breathing under continuous positive airway pressure with 2 different pressure levels)
- o PC-AC (Assisted Controlled)
- o PC-CMV (Continuous Mandatory Ventilation)
- o PC-APRV (Option) (Airway Pressure Release Ventilation)
- o PC-PSV (Pressure Support Ventilation)
- · Spontaneous (Spn):
- o Spn-CPAP/VS (Continuous Positive Airway Pressure / Ventilation Support)
- o Spn-CPAP/PS (Continuous Positive Airway Pressure / Pressure Support)
- o Spn-PPS (Option) (Proportional Pressure Support)
Additionally the ventilation unit features special modes to complement the ventilation modes. If breathing of a spontaneously breathing patient stops, Apnea Ventilation switches over automatically to volume-controlled mandatory ventilation. Automatic Tube Compensation ATC reduces the breathing effort attributable to the tube. By switching on the flow trigger, the mandatory strokes are synchronised with the patient's spontaneous breathing attempts. Inspiratoric termination determines the duration of inspiration in PS, VS and PPS mode. It defines at which relation (in percent) of max. insp. flow the inspiration ends and the expiration starts. Auto release determines the duration of pressure release using in PC-
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for Dräger Medical. The logo is in black and white and features the company name in a bold, sans-serif font. The word "Dräger" is larger than the word "medical". The text "A Dräger and Siemens Company" is below the logo.
## List of performance testing:
Performance was tested in compliance with following standards:
| IEC 60601-1 :2006 | Medical electrical equipment - Part 1:<br>General requirements for basic safety and<br>essential performance |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 :2007 | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and<br>essential performance - Collateral standard:<br>Electromagnetic compatibility -<br>Requirements and tests |
| IEC 60601-2-12:2001 | Medical electrical equipment - Part 2-12:<br>Particular Requirements for the safety of<br>lung ventilators - critical care ventilators |
| IEC 60601-1-6 :2004 | Medical electrical equipment - Part 1-6:<br>General requirements for basic safety and<br>essential performance - Collateral standard<br>- Usability |
| IEC 60601-1-8:2006 | Medical electrical equipment - Part 1-8:<br>General requirements for basic safety and<br>essential performance - Collateral<br>standard: General requirements, tests and<br>guidance for alarm systems in medical<br>electrical equipment and medical electrical<br>systems |
| EN ISO 14971:2003 | Medical devices - Application of risk<br>management to medical devices |
| EN ISO 17664:2004 | Sterilization of medical devices -<br>information to be provided by the<br>manufacturer for the reprocessing of<br>resterilizable medical devices |
| FDA Guidance for Ventilators: 1995 | |
.
March 2010
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a bird-like form. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Dräger Medical AG & Company KG C/O Ms. Joyce Kilroy Draeger Medical System, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969
MAR 3 0 2010
Re: K093633
Trade/Device Name: Infinity Acute Care System Workstation Critical Care Regulation Number: 21CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: March 16, 2010 Received: March 18, 2010
Dear Ms. Kilroy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Kilroy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
for
Anthony D. Watson. B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
## 510(k) Number (if known): K (P) 36- 33
Device Name: Infinity Acute Care System Workstation Critical Care
Indications for Use:
The Infinity Acute Care System Workstations Critical Care consist of monitoring and control displays and additional therapy units. They are intended to be used as integrated, networked, and configurable workstations to provide critical care specific therapy. The Infinity Acute Care System Workstations Critical Care are intended to be used by qualified and trained medical personnel.
The Infinity C Series Medical Cockpits, consisting of the C500 and the C700, are monitoring and control displays for the Infinity Acute Care System (IACS). Medical Cockpits are intended to be used to monitor waveforms, parameter information, and alarms as well as to control settings. The Infinity Series Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare professionals.
The Evita V500 ventilation unit of the Infinity Acute Care System is intended for the ventilation of adult, pediatric and neonatal patients. Evita V500 offers mandatory ventilation modes and ventilation modes for spontaneous breathing support and airway monitoring. The Evita V500 ventilation unit is used with Infinity C Series Dräger Medical Cockpits. The Evita V500 ventilation unit is intended for use in different medical care areas.
Evita V500 is intended for stationary use in hospitals and medical rooms or for patient transportation within the hospital.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
(
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K093633
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.