ACUSON MODEL X300 DIAGNOSTIC ULTRASOUND SYSTEM

K093531 · Siemens Medi Cal Solutions, Inc. · IYN · Dec 1, 2009 · Radiology

Device Facts

Record IDK093531
Device NameACUSON MODEL X300 DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantSiemens Medi Cal Solutions, Inc.
Product CodeIYN · Radiology
Decision DateDec 1, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Device Story

ACUSON X300 is a multi-purpose diagnostic ultrasound system; utilizes various transducers (phased, convex, linear, TEE, intracardiac) to acquire ultrasound signals; transforms signals into diagnostic images and fluid flow data; used in clinical settings (radiology, cardiology, surgery) by trained healthcare professionals; provides anatomical measurements and analysis packages for clinical diagnosis; aids clinicians in visualizing internal structures and blood flow; benefits patients through non-invasive diagnostic assessment.

Clinical Evidence

Bench testing only. Modifications were verified and validated according to the company's design control process. No clinical data presented.

Technological Characteristics

Diagnostic ultrasound system; supports multiple transducer types (phased, convex, linear, TEE, intracardiac); complies with UL 60601-1, IEC 60601-2-37, AIUM/NEMA UD-2/UD-3, and ISO 10993-1 biocompatibility standards; includes proprietary software for image processing and analysis.

Indications for Use

Indicated for diagnostic ultrasound imaging or fluid flow analysis in general radiology, fetal, abdominal, intraoperative, pediatric, small parts, neonatal/adult cephalic, cardiac, transesophageal, pelvic, transcranial, OB/GYN, urology, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093531 ACUSON X300 ™ Diagnostic Ultrasound System 510(k) Submission ### 510(k) Summary | Sponsor: | Siemens Medical Solutions, Inc.,<br>Ultrasound Division<br>1230 Shorebird Way<br>Mountain View, California 94043 | | | | | | | | | | | DEC - 1 2009 | |------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------|--|--|--|--|--|--|--|--|--|--------------| | Contact Person: | Shelly Pearce<br>Telephone: (650) 694-5988<br>Fax: (650) 694-5580 | | | | | | | | | | | | | Submission Date: | October 06, 2009 | | | | | | | | | | | | | Device Name: | ACUSON X300™ Diagnostic Ultrasound System | | | | | | | | | | | | | Common Name: | Diagnostic Ultrasound System with Accessories | | | | | | | | | | | | | Classification: | | | | | | | | | | | | | | Regulatory Class: | II | | | | | | | | | | | | | Review Category: | Tier II | | | | | | | | | | | | | Classification Panel: | Radiology | | | | | | | | | | | | | Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN | | | | | | | | | | | | Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN | |------------------------------------------|---------------|---------------------| | Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO | | Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX | | Diagnostic Intravascular Catheter | FR # 870.1200 | Product Code 74-DQ | ### A. Legally Marketed Predicate Devices The Siemens Acuson X300 ultrasound system is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to our current product, the Siemens Acuson X300 ultrasound system (K090276, K071036 and K061946) and the following: - K061980, Siemens Acuson X500™ Ultrasound System ◆ - K081121, ACUSON X150 Ultrasound System ● - · K090276, ACUSON X300™ Diagnostic Ultrasound System, Premium Edition ### B. Device Description: The Acuson X300 has been designed to meet the following product safety standards: - UL 60601-1, Safety Requirements for Medical Equipment 트 - IEC 60601-2-37 Diagnostic Ultrasound Safety Standards 제 - I CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment - 로 AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment - 트 AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound - 93/42/EEC Medical Devices Directive 트 - Safety and EMC Requirements for Medical Equipment - « EN/IEC 60601-1 발 {1}------------------------------------------------ - 트 EN/IEC 60601-1-1 - M EN/IEC 60601-1-2 - 내 EN/IEC 60601-1-4 - I EN/IEC 60601-1-6 - 피 EN/IEC 60601-2-18 - EN/IEC 60601-2-25 I - IEC 1157 Declaration of Acoustic Power - ISO 10993-1 Biocompatibility ### C. Intended Use The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. ### D. Substantial Equivalence The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics. ### E. Performance Data The Acuson X300 modifications are verified and validated according to the company's design control process. ### X300 510{k} Submission {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, facing to the right. Public Health Service DEC - 1 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Siemens Medical Solutions USA, Inc., Ultrasound Group % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313 Re: K093531 Trade/Device Name: ACUSON X300™ Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, and DOO Dated: November 13, 2009 Received: November 16, 2009 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the ACUSON X300™ Diagnostic Ultrasound System, as described in your premarket notification: Transducer Model Number P4-2 Phased Sector Array CH5-2 Convex Array VF10-5 Linear Array L9-5 Linear Array EC9-4 Convex Array {3}------------------------------------------------ EV9-4 Convex Array VF13-5 Linear Array P8-4 Phased Sector Array BE9-4 Convex Array CW2 Continuous Wave Doppler CW5-Continuous Wave Doppler AcuNav 8F Intracardiac AcuNav 10F Intracardiac V5Ms TEE 4V1c Phased Sector Array VF13-5SP Linear Array C8-5 Tightly Curved Array 8L3 Linear "Regel" 10V4 Phased Array C7F2 Curved Array Mechanical 3D/4D EV9F4 Curved Array Mechanical 3D/4D L13F5 3D/4D Mechanical Wobbler Linear VF8-3 Linear Array P5-1 Phased Array C6-2 Convex Array SoundStar 10F P9-4 Phased Array P4-2 Litho Phased Sector Array BP9-4 Convex Array If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please {4}------------------------------------------------ note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242. Sincerely yours, hozu h. Why 5 Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {5}------------------------------------------------ Indications for Use Statement 63531 510(k) Number (if known): Device Name: ACUSON X300™ Diagnostic Ultrasound System Indications For Use: The Siemens Acuson X300 ultrasound imaging system is intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign- Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number X300 510(k) Submission {6}------------------------------------------------ #### ACUSON X300™ Diagnostic Ultrasound System SIO(k) Submission #### Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: Intended Use: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | |---------------------------|-------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------| | Other<br>(Track1<br>Only) | Specific<br>(Tracks11& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Fetal | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-operative<br>(Note 6) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-operative<br>(Neuro) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Fetal | Laparoscopic | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Pediatric | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Imaging<br>& Other | SmallOrgan<br>(Note 1) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Neonatal Cephalic | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-rectal | P | P | P | P | | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-vaginal | P | P | P | P | | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-urethral | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-esoph.<br>(non-Card.) | P | P | P | P | | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Musculo-skel.<br>(Convent.) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Musculo-skel.<br>(Superfic) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra -vascular | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Other<br>(Specify) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Cardiac Adult | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11,12 | | Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-vascular<br>(Cardiac) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-esophageal<br>(Cardiac) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-Cardiac | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Other (Specify) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11,12 | | | Other (Specify) | P | P | P | P | P | P | P | P | BMDC | Note<br>2,3,4,5,7,8,9,11 | N = new indication; P = proviously cleared by K090276 Nole 1 For example: breast, testes, thyroid, penis, prostate, etc. Note 2 Ensemble tissue harmonic imaging Note 3 Note 3 SD Imaging Note 4 PWM BioScanner andramara B&W SieScape panoramic imaging Power SieScape panoramic imaging Note 4 Note 5 Note 6 For example: abdominal, vascular Note 7 Contrast agent imagi Note 7 Note 8 Contrast agent imaging SieClear multi-view spatial compounding Note 9 Tissue Equalization Technology Note 10 Note 11 Intracardiac Imaging Dynamic TCE Note 12 AHP > (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aomr Mz Wz (Division Si roductive, Abdominal and Page 38 of 1531 510(k) Number {7}------------------------------------------------ #### ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission #### Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: Intended Use: P4-2 Phased Sector Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | |-----------------------------|-------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------| | Other<br>( Track 1<br>Only) | Specific<br>(Tracks11& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | P | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | P | P | P | P | P | P | P | | BMDC | Note<br>2,3,5,6,7,8,9,11 | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Fetal | Pediatric | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Imaging<br>& Other | SmallOrgan<br>(Note 1) | | | | | | | | | | | | | Neonatal Cephalic | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | | | | | | | | | | | | Musculo-skel.<br>(Superfic) | | | | | | | | | | | | | Intra-vascular | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac Adult | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Cardiac | Cardiac Pediatric | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | Trans-esophageal<br>(Cardlac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | Peripheral | Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | Vessel | Other (Specify) | | | | | | | | | | | Note 1 For example: breast, testes, thyroid, penis, prostate, etc. Note 2 Note 3 Ensemble lissue hamonic imaging JD imaging B&W SieScape panoramic imaging Note 4 Power SieScape panoramic imaging For example: abdominal, vascular Note 5 Note 6 Note 6 For exemple: abdominal, vas Note 7 Contrast agent imaging Nole 7 Contrast agent imaging SieClear multl-view spatial compounding Tissue Equalization Technology Note 8 Note 9 Note 9 Tissue Equalization Technology Note 10. Additional scans Note 10 Intracardiac Imaging Note 11 Dynamic TCE > (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aora M. Whay (Division Sign-Division of Reproductive, Abdominal, and Radiological D Page 39 of 1531 510(k) N {8}------------------------------------------------ ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: Intended Use: CH5-2 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | | |----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|--| | Other<br>( Track1<br>Only) | Specific<br>(Tracks1& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Abdominal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Fatal | Pediatric | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Imaging<br>& Other | SmallOrgan<br>(Note 1) | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | P | P | P | | P | P | | BMDC | Note<br>2.3.4.5.7.8.9,11 | | | | Musculo-skel.<br>(Superfic) | | | | | | | | | | | | | | Intra-vascular | | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | | Cardlac Adult | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Cardlac | Cardiac Pedlatric | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | Peripheral | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Vessel | Other (Specify) | | | | | | | | | | | | N = new indication; P = prevlously cleared by K090276 Note 1 Note 2 For example: breast, testes, thyrold, penls, prostate, atc. Ensemble tissue harmonic imaging Note 3 Note 4 Note 5 2008. Imaging manoramic imaging BBW Sging Siessape panoramic limaging B&Wer SiesScape panoramic limaging For example: abodominal, vascularing Contrast agent imaging Note 6 Note 7 Note 8 SieClear multi-view spatial compounding Note 9 Tissue Equalization Technology Intracardiac imaging Note 10 Note 11 Dynamic TCE > (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concence of CDRH, Office of Device Evaluation (ODE) > Concence of CDRH, Office of Device Evaluation (ODE) . --- (Division Sign-Off) productive. Abdominal. Division of Re and Radiological Dev 510(k) Numbe Page 40 of 1531 {9}------------------------------------------------ ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: r Intended Use: ASHOON XSAC Di ACUSON X300 Dlagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | Mode of Operation | | | | | | | | | | | |-----------------------------|-------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------| | Other<br>(Track 1<br>Only) | Specific<br>(Tracks1& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | SmallOrgan<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Musculo-skel.<br>(Superfic) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-vascular | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Note 12 | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11,12 | | | Other (Specify) | | | | | | | | | | | | | N = new indication: P = previously cleared by K090278 | | | | | | | | | | | For example: breast, testes, thyroid, penis, prostate, etc. Ensemble: breast, testes, thyroid, penis, prostate, etc. Ensemble tissue harmonic imaging Note 1 Note 2 Note 3 Note 4 Rule 4 DAW Seascapes panoramic image Note 6 Penna-Olga 1x6 U 3D imaging B&W SieScape panoramic imaging Power SleScape panoramic imaging For example: abdominal, wascular For example: abdominal, vascular Contrast agent imaging Note S Note B Note 7 Note 8 Note 9 SleClear multi-view spatial compounding Tissue Equalization Technology Intracardiac imaging Note 10 Note 11 Note 12 Dynamic TCE AHP (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Prescription Use (Per 21 CFR 801.109) Ammal m. Why (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological De 510(k) Number # Page 41 of 1531 {10}------------------------------------------------ ACUSON X300m Diagnostic Ultrasound System 510(k) Submission Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: Intended Use: L.9-5 Linear Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | |-----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------| | Other<br>(Track 1<br>Only) | Specific<br>(Tracks11& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | SmallOrgan<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Musculo-skel.<br>(Superfic) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Intra-vascular | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Note 12 | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | Intra-Cardaic | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11,12 | | | Other (Specify) | | | | | | | | | | | N = new indication; P = previously cleared by K090276 For example: breast, testes, thyrold, penis, prostate, etc. Note 1 Note 2 Ensemble tissue harmonic Imaging Note 3 Note 4 3D Imaging B&W SieScape 3D imaging B&W SieScape panoramic Imaging Power SieScape panoramic Imaging Note 4 Note 5 Power SieScape panoramic Imaging For example: abdominal, vascular Nota 6 Note 7 Contrast agent imaging SieClear multi-view spatlal compounding Note 8 Note 9 Note 10 Tissua Equalization Technology Intracardiac imaging Note 11 Dynamic TCE Note 12 AHP > (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Aopai Th Way (Division Si Division of Reproductive, Abdomina and Radiological Dev Page 42 of 1531 510(k) Number {11}------------------------------------------------ #### ACUSON X300™ Diagnostic Ultrasound System SIO(k) Submission Dlagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: Intended Use: EC9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | |-----------------------------|-------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------| | Other<br>( Track 1<br>Only) | Specific<br>(Tracks1& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | | | | | | | | | | | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric<br>SmallOrgan<br>(Note 1) | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Neonatal Cephalic | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-vaginal | P | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | | | | | | | | | | | | Musculo-skel.<br>(Superfic) | | | | | | | | | | | | | Intra-vascular | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | Intra-Cardiac | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | Note 1 For example: breast, testes, thyroid, penis, prostate, etc. - Note 2 Ensemble tissue harmonic imaging Note 3 3D imaging - Note 4 - B&W SieScape panoramic imaging Note 5 - Power SieScape panoramic imaging Note f - Note 7 Contrast agent imaging Note 8 - SieClear multi-view spatial compounding Tissue Equalization Technology - Note 9 Note 10 Intracardiac imaging - Nole 11 Dynamic TCE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Lori M. Whay (Divi Sign-Off Division of Reproductive, Abdominal, and Radiological De 510(k) Number # Page 43 of 1531 {12}------------------------------------------------ ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: ,' , Intended Use: EV9-4 Convex Array Transducer for use with: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | |-----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------| | Other<br>(Track1<br>Only) | Specific<br>(Tracks11& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Abdominal | | | | | | | | | | | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | | | | | | SmallOrgan<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-vaginal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | | | | | | | | | | | | Musculo-skel.<br>(Superfic) | | | | | | | | | | | | | Intra-vascular | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac Adult | | | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | Intra-Cardiac | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | Peripheral | Peripheral vessel | | | | | | | | | | | | Vessel | Other (Specify) | | | | | | | | | | | N = new indication; P = previously cleared by K090276 Note 1 For example: breast, testes, thyroid, penis, prostate, etc. Ensemble tissue harmonic imaging Note 2 Note 3 3D imaging B&W SieScape panoramic imaging Nole 4 Note 5 Power SieScape panoramic Imaging For example: abdominal, vascular Nota 6 Note 7 Note B Contrast agent imaging SieClear multi-view spatial compounding Tissue Equalization Technology Note 9 Note 10 Intracardiac imaging Note 11 Dynamic TCE > (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) > Concurrence Prescription Use (Per 21 CFR 801.109) Tony M. Wieg . . (Division Reproductive, Abdominal, grid 510(k) N # Page 44 of 1531 {13}------------------------------------------------ ### ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission ### Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: VF13-5 Linear Array Transducer for use with: Intended Use: ACUSON X300 Diagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation | | | | | | | | | | |-----------------------------|-------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------|--| | Other<br>( Track 1<br>Only) | Specific<br>(Tracks1i& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Abdominal | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Intra-operative<br>(Note 6) | | | | | | | | | - | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | SmallOrgan<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note<br>2.3,4.5,7,8,9,11 | | | | Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note<br>2.3,4,5,7,8,9,11 | | | | Adult Cephalic | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | | Musculo-skel.<br>(Convent.) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Musculo-skel.<br>(Superfic) | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Intra-vascular | | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Note 12 | | | | Cardiac Pediatric | | | | | | | | | | | | | | Intra-vascular.<br>(Cardiac) | | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | Other (Specify) | | | | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | P | P | P | | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11,12 | | | | Other (Specify) | | | | | | | | | | | | N = new indication; P = previously cleared by K090276 Note 1 For example: breast, testes, thyrold, penis, prostate, etc. Ensemble tissue harmonic imaging Note 2 Note 3 3D imaging B&W SieScape panoramic imaging Note 4 Note 5 Note 6 Power SieScape panoramic imaging For example: abdominal, vascular Note 7 Contrast agent imaging Note 8 SleClear multi-view spatial compounding Tissue Equalization Technology Note 9 Intracardiac Imaging Note 10 Note 11 Dynamic TCE AHP Note 12 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evalualion (ODE) Prescription Use (Per 21 CFFR 801, 109) Aorall M. Whing Abdominal oductive and Ra Page 45 of 1531 510(k) Numbe {14}------------------------------------------------ #### ACUSON X300™ Diagnostic Ultrasound System 510(k) Submission Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): Device Name: P8-4 Phased Sector Array Transducer for use with: Intended Use: ACUSON X300 Dlagnostic Ultrasound Systems Diagnostic imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | | | | | |-----------------------------|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------|--| | Other<br>( Track 1<br>Only) | Specific<br>(Tracks11& 3) | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | | | Fetal | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Abdominal | | P | P | P | P | P | P | | BMDC | Note<br>2,3,5,6,7,8,9,11 | | | | Intra-operative<br>(Note 6) | | | | | | | | | | | | | | Intra-operative<br>(Neuro) | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Fetal | Pediatric | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Imaging<br>& Other | SmallOrgan<br>(Note 1) | | | | | | | | | | | | | | Neonatal Cephalic | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Adult Cephalic | | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | | Trans-esoph.<br>(non-Card.) | | | | | | | | | | | | | | Musculo-skel.<br>(Convent) | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Musculo-skel.<br>(Superfic) | | | | | | | | | | | | | | Intra-vascular | | | | | | | | | | | | | | Other<br>(Specify) | | | | | | | | | | | | | | Cardiac Adult | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | Cardiac | Cardiac Pediatric | | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,7,8,9,11 | | | | Intra-vascular<br>(Cardiac) | | | | | | | | | | | | | | Trans-esophageal<br>(Cardiac) | | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | | | Other (Specify) | | | | | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...