CLEVER CHOICE AUTO-CODE PRO BLOOD GLUCOSE MONITORING SYSTEM MODEL TD-4267
Applicant
Taidoc Technology Corporation
Product Code
CGA · Clinical Chemistry
Decision Date
Apr 1, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger and the following alternative sites: palm, forearm, upper arm, calf and thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus. The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System cannot be used on neonates. The alternative site testing in above systems can be used only during steady-state blood glucose conditions.
Device Story
System measures glucose concentration in fresh capillary whole blood; utilizes electrochemical meter and dry reagent biosensor test strips. Amperometric principle: electrical current generated by glucose-reagent reaction is proportional to glucose concentration. Used by healthcare professionals and patients at home for diabetes management. Provides quantitative glucose readings to assist in monitoring diabetes control programs. Not for diagnostic or screening purposes.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence supported by design control activities, risk analysis, and verification/validation of software and hardware modifications.
Technological Characteristics
Electrochemical glucose monitoring; USB data interface; physical meter enclosure with integrated display and navigation buttons (M, S, up/down). Software modified to remove coding display. Sterilization: N/A (meter).
Indications for Use
Indicated for quantitative blood glucose measurement in patients with diabetes mellitus. Suitable for use by healthcare professionals and patients at home. Samples: fresh capillary whole blood from finger, palm, forearm, upper arm, calf, or thigh. Contraindications: Not for use on neonates; alternative site testing only during steady-state glucose conditions; not for diagnosis or screening of diabetes.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
FORA G30 blood glucose monitoring system (K090187)
Related Devices
K093005 — FORA G72 BLODO GLUCOSE MONITORING SYSTEM/TD-4264 BLOOD GLUCOSE MONITORING SYSTEM, MODEL TD-4264 · Taidoc Technology Corporation · Feb 5, 2010
K094005 — FORA G31A/TD-4256A BLOOD GLUCOSE MONITORING SYSTEM AND FORA G31B/TD-4256B BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · May 5, 2010
K072854 — DIACHEX BASIC BLOOD GLUCOSE MONITORING SYSTEM, DIACHEX SUPERB BLOOD GLUCOSE MONITORING SYSTEM · Tyson Bioresearch, Inc. · Jun 30, 2008
K060467 — PRODIGY BLOOD GLUCOSE TEST SYSTEM · Diagnostic Devices, Inc. · Jul 13, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# 510 (k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
ко93506 The Assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
| 1. Submitter Information | |
|--------------------------|-------------------------------|
| Company name | TaiDoc Technology Corporation |
| Contact person | Teling Hsu |
| Address | 6F, No. 127, Wugong 2nd Rd, |
| | Wugu Township, Taipei County, |
| | 24888, Taiwan |
| Phone | (+886-2) 6625-8188 #1176 |
| FAX | (+886-2) 6625-0288 |
| E-mail | teling.hsu@taidoc.com.tw |
| Date Prepared | November 6th, 2009 |
2. Name of Device Proprietary Names
Common Name
Product Code
Classification Name
CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System Blood Glucose Test System Class II devices (21 CFR Section 862.1345) NBW / CGA
3. Predicate Device Trade/Proprietary Name: Common/Usual Name: Manufacturer 510 (k) Number
FORA G30 Blood Glucose Monitoring System Blood Glucose Test System TaiDoc Technology Corporation K090187
#### 4. Device Description
The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System is designed to quantitatively measure the concentration of glucose in fresh capillary whole blood. The test principle of system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in whole blood and control solutions.
{1}------------------------------------------------
#### 5. Intended Use
The CLEVER CHO!CE Auto-Code Pro Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger and the following alternative sites: palm, forearm, upper arm, calf and thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus.
The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System cannot be used on neonates.
The alternative site testing in above systems can be used only during steady-state blood glucose conditions.
# 6. Comparison to Predicate Device
The intended use, test principle, and operating technology of CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System are all the same as the predicate device. The only difference between the predicate and proposed devices is in engineering modification.
This system uses amperometry for blood glucose measurement. The test is based on the measurement of electrical current generated by the reaction of glucose with reagents of the test strip.
•
# 7. Performance Studies
CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System has the same performance characteristics as the predicate device.
The studies of software verification and validation testing, method comparison, and meter reliability test demonstrated that the performance of systems meets the intended use.
## 8. Conclusion
CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System demonstrates satisfactory performance and is suitable for the intended use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and appears to be a simple, official representation of the department.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
TaiDoc Technology Corporation c/o Teling Hsu Regulatory Affairs Specialist 6F, No. 127, Wugong 2nd Road, Wugu Township Taipei County, China (Taiwan) 248
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
APR 0 1 2010
Re: k093506
Trade/Device Name: CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: February 26, 2010 Received: March 5, 2010
Dear Teling Hsu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Signature
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): k093506
Device Name:
CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System
Indications for Use:
The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood samples from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus.
The CLEVER CHOICE Auto-Code Pro Blood Glucose Monitoring System cannot be used on neonates.
The alternative site testing in above systems can be used only during steady-state blood glucose conditions.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Benson
Division Sign-off Office of in Vitro Diagnostic Devices Evaluation and Safety 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.