The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi- arthroplasty is determined by the surgeon to be the preferred method of treatment. The cemented primary humeral stem, long/revision stem, fracture stems and all Equinoxe glenoids are intended for cemented fixation. The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon. The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. Humeral Heads are intended for use in cemented and press-fit applications. The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head. The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
Device Story
Equinoxe Shoulder System components include modular humeral stems, heads, and glenoids for total/hemi-arthroplasty and fracture repair. Subject devices are Cage Glenoids (UHMWPE surface on grit-blasted Ti-6Al-4V cage/pegs) and XL Keeled/Pegged Glenoids. Cage Glenoids include posterior augment version to preserve bone by limiting eccentric reaming. XL versions interface with 47-53mm humeral heads. Modifications to existing Equinoxe glenoids include size addition, reduced UHMWPE thickness (5mm to 4mm), radial machining pattern, and reduced edge radius. Used in OR by orthopedic surgeons for joint replacement. Provides stable articulating surface to restore joint function and mobility; reduces pain associated with degenerative disease or fracture.
Clinical Evidence
No clinical data. Evidence consists of bench testing including finite element analyses, mechanical tests (edge loading, displacement, wear, thermal expansion, disassembly force), engineering analyses, biocompatibility studies, and simulated surgical implantations (sawbones and cadaver).
Indicated for skeletally mature patients with glenohumeral degenerative diseases (osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic) or fractures (proximal humerus, displaced tuberosities, 3/4-part fractures). Also indicated for failed previous reconstructions, congenital abnormalities, and cases requiring revision or restoration of mobility. Reverse system indicated for irreparable rotator cuff deficiency or failed arthroplasty with superior humeral head migration.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
| K093430 | | | |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|--------|
| | | r-1/4 | |
| I. Submitter Information<br>Company: | Exactech, Inc<br>2320 NW 66th Court<br>Gainesville, FL 32653<br>800-392-8711 | SEP | 2 2010 |
| Date: | August 31, 2010 | | |
| Contact Person:<br>Graham L. Cuthbert Regulatory Affairs Specialist II | | | |
| II. Device Information<br>Proprietary Name: | Exactech Equinoxe® Cage Glenoid™<br>Exactech Equinoxe® XL Keeled & XL Pegged Glenoid | | |
| Common Name: | Glenoid Component | | |
| Classification Name: | Shoulder joint metal/polymer non-constrained cemented<br>prosthesis (21 CFR 888.3650, Class II, Product Code KWT)<br>Prosthesis, Shoulder, Semi-constrained, metal/polymer<br>cemented (21 CFR 888.3660, Class II, Product Code KWS) | | |
Of
### Legally Marketed Devices to Which Substantial Equivalence Is Claimed III. Exactech Equinoxe Cage Glenoid
- Depuy Anchor Peg Glenoid #K981487 .
- Tornier Affiniti Glenoid #K081707 .
- Smith and Nephew Cofield Glenoid: #K955767 and #K070565. .
- Biomet Modular Hybrid Glenoid: #K060694 ●
## Exactech Equinoxe XL Keeled and XL Pegged Glenoid
- Exactech Equinoxe Glenoids: #K042021 .
- Tornier Acqualis Glenoids: #K994393 .
- t Depuy Global Glenoids: #K981487
#### Device Description IV.
The Equinoxe Shoulder System comprises both cemented and press-fit semi-constrained glenohumeral prostheses for use in hemi-shoulder and total-shoulder joint replacement procedures and a cemented semi-constrained glenohumeral fracture prosthesis for use in fractures of the proximal humerus. The system includes various sizes and types of modulat humeral stems, humeral heads, glenoids, replicator plates, and screws for use in primary, revision, and fracture applications.
## Exactech Equinoxe Cage Glenoid
The Equinoxe Caged Glenoids are designed to interface and articulate with the Equinoxe total-shoulder system. They are composed of an Ultra High Molecular Weight Polyethylene (UHMWPE) articulating surface modularly connected to a grit blasted titanium (Ti-6Al-4V) bone cage and three grit blasted titanium (Ti-6Al-4V) pegs. A posterior augment version of
{1}------------------------------------------------
# K093430
p. 2/4
this cage glenoid is provided to preserve bone by limiting the amount of eccentric glenoid reaming necessary to correct the patient's glenoid version. All components are supplied sterile.
## Exactech Equinoxe XL Keeled an XL Pegged Glenoid
The Equinoxe XL Keeled and XL Pegged Glenoids are designed to interface and articulate with the 47, 50, & 53 mm humeral head from the Equinoxe total-shoulder system. The subject devices are a modification of the previously cleared Equinoxe cemented keeled and pegged glenoids (#K042021). All components are supplied sterile.
Additionally, this "Traditional" 510(k) proposes several modifications to our previously cleared Equinoxe cemented keeled and pegged glenoids. Specifically: 1) an addition of an XL keeled and pegged glenoid size, 2) a decrease in minimum UHMWPE thickness from 5mm to 4mm, 3) a modification in machining pattern on the backside surface (change from horizontal grooves to radial grooves that follow the profile of the device), and 4) a decrease in edge radius around each fixation surface from 0.047 inches to 0.031 inches.
#### V. Indications Use
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi- arthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems and all . Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be used ● with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented applications . or in revision cascs when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit applications. .
Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:
| P | L/R | F | Indications |
|---|-----|---|------------------------------------------------------------------------------------------------------------------------|
| √ | √ | | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems |
| √ | √ | | congenital abnormalities in the skeletally mature |
| √ | | | primary and secondary necrosis of the humeral head. |
| √ | | √ | humeral head fracture with displacement of the tuberosities |
| √ | √ | | pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable |
| √ | √ | | revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved) |
| | | √ | displaced three-part and four-part upper humeral fractures |
| | √ | | spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases) |
{2}------------------------------------------------
K093430
| √ | | revision of failed previous reconstructions when distal anchorage is required |
|---|---|-------------------------------------------------------------------------------|
| √ | √ | to restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
#### VI. Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
## Exactech Equinoxe Cage Glenoid
- Intended Use. Exactech Equinoxe Cage Glenoids and predicate devices are intended . for use in Total Shoulder Arthroplasty. The Equinoxe cage glenoids and posterior augment cage glenoids are intended for cemented fixation.
- Materials. Exactech Equinoxe Cage Glenoids and predicate devices are composed of t equivalent materials conforming to recognized industry standards for permanent implants.
- Dimensions. Exactech Equinoxe Cage Glenoids and predicate devices are available in . equivalent size ranges.
- Sterilization processes. Exactech Equinoxe Cage Glenoids and predicate devices are . sterilized using equivalent sterilization processes conforming to recognized industry standards.
- Performance specifications. Exactech Equinoxe Cage Glenoids and predicate devices . conform to recognized performance standards for total shoulder joint replacement devices.
## Exactech Equinoxe XL Keeled and XL Pegged Glenoid
- Intended Use. Exactech Equinoxe XL Keeled, XL Pegged Glenoid and predicate . devices are intended for use in Total Shoulder Arthroplasty. The XL keeled and pegged glenoids are intended for cemented fixation.
- Materials. Exactech Equinoxe XL Keeled, XL Pegged Glenoid and predicate devices . composed of equivalent materials conforming to recognized industry standards for permanent implants.
{3}------------------------------------------------
K093430
p. 4/4
- Dimensions. Exactech Equinoxe XL Keeled, XL Pegged Glenoid and predicate devices . are available in equivalent size ranges.
- Sterilization processes. Exactech Equinoxe XL Keeled, XL Pegged Glenoid and . predicate devices are sterilized using equivalent sterilization processes conforming to recognized industry standards.
- Performance specifications. Exactech Equinoxe XL Keeled, XL Pegged Glenoid and . predicate devices conform to recognized performance standards for total shoulder joint replacement devices.
### Summary of Non-Clinical Performance Data
Literature reviews, finite element analyses, mechanical tests (edge loading and displacement studies, wear, thermal expansion and disassembly force), engineering analyses, biocompatibility studies and simulated surgical (sawbones and cadaver) implantations were conducted to demonstrate the safety and effectiveness of the Exactech Equinoxe Cage Glenoids, the Exactech Equinoxe XL Keeled and XL Pegged Glenoids in support of the claim of substantial equivalence to the predicates listed above.
### Substantial Equivalence Conclusion
Results from these tests and analyses provided within this 510(k) demonstrate that the Exactech Equinoxe Cage Glenoids, XL Keeled glenoids, XL Pegged Glenoids and the modified cemented keeled and pegged glenoids are substantially equivalent to the identified predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Exactech, Inc. % Mr. Graham L. Cuthbert Regulatory Affairs Specialist II 2320 NW 66th Court Gainesville, Florida 32653
2 2010 SEP
Re: K093430
Trade/Device Name: Exactech Equinoxe XL Keeled, XL Pegged and Cage Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: KWS. KWS. KWT Dated: August 19, 2010 Received: August 24, 2010
Dear Mr. Cuthbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Graham L. Cuthbert
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Pavare Buehukb
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K093430
p-1/2
## Exactech®, Inc. Indications for Use
510(k) Number:
Device Name:
Exactech Equinoxe® XL Keeled, XL Pegged and Cage GlenoidTM
## INDICATIONS FOR USE:
The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi- arthroplasty is determined by the surgeon to be the preferred method of treatment.
- The cemented primary humeral stem, long/revision stem, fracture stems and all . Equinoxe glenoids are intended for cemented fixation.
- The press-fit humeral stems are intended for press-fit applications but may be used . with bone cement at the discretion the surgeon.
- The reverse humeral components are intended to be used in cemented applications . or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.
- Humeral Heads are intended for use in cemented and press-fit applications. .
Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:
| P | L/R | F | Indications |
|---|-----|---|---------------------------------------------------------------------------------------------------------------------------|
| √ | √ | | rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic<br>degenerative problems |
| √ | √ | | congenital abnormalities in the skeletally mature |
| √ | | | primary and secondary necrosis of the humeral head. |
| √ | | √ | humeral head fracture with displacement of the tuberosities |
| √ | √ | | pathologies where arthrodesis or resectional arthroplasty of the humeral<br>head are not acceptable |
| √ | √ | | revisions of humeral prostheses when other treatments or devices have<br>failed (where adequate fixation can be achieved) |
| | | √ | displaced three-part and four-part upper humeral fractures |
| √ | √ | | spiral and other fractures of the mid-humerus (in combination with<br>glenohumeral degenerative diseases) |
| √ | √ | | revision of failed previous reconstructions when distal anchorage is<br>required |
| √ | √ | √ | to restore mobility from previous procedures (e.g. previous fusion) |
The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative discases of the glenohumeral joint and a grossly deficient, irreparable
{7}------------------------------------------------
K093430
P.2/2
rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.
| Prescription Use | <b>X</b> |
|-------------------|----------|
| (Per CFR 801.109) | |
or
| Over-the-counter Use | |
|----------------------|--|
|----------------------|--|
Concurrence of CDRH, Office of Device Evaluation (ODE)
K093430
(Division Sign-Off) Valtine Prechus forger
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.