ELECSYS TESTOSTERONE II IMMUNOASSAY
K093421 · Roche Diagnostics Corp. · CDZ · Apr 23, 2010 · Clinical Chemistry
Device Facts
| Record ID | K093421 |
| Device Name | ELECSYS TESTOSTERONE II IMMUNOASSAY |
| Applicant | Roche Diagnostics Corp. |
| Product Code | CDZ · Clinical Chemistry |
| Decision Date | Apr 23, 2010 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1680 |
| Device Class | Class 1 |
| Attributes | Pediatric |
Intended Use
Immunoassay for the in vitro quantitative determination of testosterone in human serum and plasma. The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys and cobas e immunoassay analyzers. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and androgenital syndromes.
Device Story
The Elecsys Testosterone II Immunoassay is an in vitro diagnostic test for measuring testosterone levels in human serum and plasma. It utilizes a competitive immunoassay principle with streptavidin-coated microparticles and electrochemiluminescence (ECLIA) detection. The assay is performed on automated Elecsys and cobas e immunoassay analyzers. The system uses a 2-point calibration and a master curve provided via reagent bar code to determine results. Healthcare providers use the quantitative output to assess androgen status in patients presenting with symptoms of hypogonadism, puberty disorders, or virilization. Results are intended to be interpreted in conjunction with clinical history and other diagnostic findings. The device benefits patients by providing standardized, accurate hormone quantification to guide clinical management of endocrine disorders.
Clinical Evidence
No clinical trials performed. Analytical performance validated via bench testing: precision (CLSI EP5-A2), linearity (CLSI EP6-A), LoD/LoQ (CLSI EP17-A), and interference studies. Method comparison study (n=387) against predicate showed Deming regression Y=0.989X-2.87 (R=0.992). Pediatric method comparison (n=123) showed Y=1.06X-2.79 (R=0.93). Reference ranges established for healthy adults and pediatric populations.
Technological Characteristics
Competitive immunoassay using streptavidin-coated microparticles and electrochemiluminescence (ECLIA). Reagents include R1, R2, and M (microparticles). Standardized to ID-GC/MS. Compatible with Elecsys and cobas e series analyzers. Measuring range 2.5–1500 ng/dL. Connectivity via automated analyzer platform.
Indications for Use
Indicated for the quantitative determination of testosterone in human serum and plasma to aid in the diagnosis and treatment of androgen-related disorders, including primary/secondary hypogonadism, delayed/precocious puberty, and male impotence; and in females, hirsutism and virilization associated with tumors, polycystic ovaries, and adrenogenital syndromes.
Regulatory Classification
Identification
A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
Predicate Devices
- Elecsys Testosterone Assay (k964889)
Related Devices
- K211685 — Elecsys Testosterone II · Roche Diagnostics · May 6, 2022
- K120009 — TESTOSTERONE TEST SYSTEM; CALIBRATOR; AND QUALITY CONTROL MATERIAL (ASSAYED AND UNASSAYED) · Abbott Laboratories · Sep 11, 2012
- K964889 — ELECSYS TESTOSTERONE ASSAY · Boehringer Mannheim Corp. · Mar 18, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
093421
Elecsys® Testosterone II Immunoassay
:
Confidential
| 510(k) Summary | APR 2 8 2010 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. |
| Submitter<br>Name, Address,<br>Contact | Roche Diagnostics<br>9115 Hague Road<br>Indianapolis, IN 46250<br><br>Contact person: Sarah Baumann<br>Phone: (317) 521-3952<br>Fax: (317) 521-2324<br>Email: sarah.baumann@roche.com<br><br>Secondary contact: Stephanie Greeman<br>Phone: (317) 521-2458<br>Fax: (317) 521-2324<br>Email: stephanie.greeman@roche.com<br><br>Date Prepared: October 23, 2009 |
| Device Name | Proprietary name: Elecsys® Testosterone II Immunoassay<br>Common name: Testosterone II Assay<br>Classification name: Testosterone Test System |
| Device<br>Description | The Elecsys Testosterone II immunoassay is based on a competitive test<br>principle with streptavidin-coated microparticles and<br>electrochemiluminescence detection. Results are determined using a<br>calibration curve that is generated specifically on each instrument by a 2-point<br>calibration and a master curve provided with the reagent bar code.<br>The Elecsys Testosterone II reagent kit consists of a Reagent Pack (R1, R2,<br>and M[Streptavidin-coated microparticles]). |
| Substantial<br>Equivalence | The Elecsys Testosterone II Test System is substantially equivalent to other<br>devices legally marketed in the United States. We claim equivalency to the<br>currently marketed Elecsys Testosterone Assay (K964889). |
.
,
:
.
{1}------------------------------------------------
The following table compares the Elecsys Testosterone II Immunoassay with Substantial Equivalencethe predicate device. Comparison
| Table 1. Testosterone Immunoassay Comparison | | |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) |
| Intended Use | Immunoassay for the in vitro<br>quantitative determination of<br>testosterone in human serum and<br>plasma. The electrochemiluminescence<br>immunoassay "ECLIA" is intended for<br>use on Elecsys and cobas e<br>immunoassay analyzers. | Immunoassay for the in vitro<br>quantitative determination of<br>testosterone in human serum<br>and plasma. The<br>electrochemiluminescence<br>immunoassay "ECLIA" is<br>intended for use on the<br>Boehringer Mannheim Elecsys<br>1010 and 2010 immunoassay<br>analyzers. |
| Indications for<br>Use | Measurements of testosterone are used<br>in the diagnosis and treatment of<br>disorders involving the male sex<br>hormones (androgens), including<br>primary and secondary hypogonadism,<br>delayed or precocious puberty,<br>impotence in males and, in females<br>hirsutism (excessive hair) and<br>virilization (masculinization) due to<br>tumors, polycystic ovaries, and<br>adrenogenital syndromes. | Same |
| Assay Protocol | Competition principle | Same |
| Detection Protocol | Electrochemiluminescence<br>immunoassay (ECLIA) | Same |
| Traceability/<br>Standardization | ID-GC/MS (Isotope Dilution Gas-<br>Chromatography/Mass Spectrometry) | Same |
| Sample Type | Human serum and plasma | Same |
| Table 1. Testosterone Immunoassay Comparison, continued | | |
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) |
| Instrument<br>Platform | Elecsys 2010<br>(Request for CLIA categorization<br>has been made to add the<br>MODULAR ANALYTICS E170,<br>cobas e 411, and cobas e 601<br>analyzers according to the<br>Replacement Reagent and<br>Instrument Policy). | Elecsys 2010<br>(Elecsys 1010, MODULAR<br>ANALYTICS E170, cobas e 411,<br>and cobas e 601 analyzers added<br>subsequent to clearance) |
| Measuring Range | 2.5 – 1500 ng/dL<br>(0.087 – 52.0 nmol/L) | 2.0 – 1500 ng/dL<br>(0.069 – 52.0 nmol/L) |
| Calibrator | Testosterone II CalSet II<br>Calibrators 1 and 2<br>The stability, value assignment and<br>matrix is identical to Testosterone<br>CalSet II (cleared on K003411). | Same |
| Calibration<br>Interval | Once per reagent lot and<br>• After 1 month (28 days) when<br>using the same reagent lot<br>• After 7 days (when using the<br>same reagent kit on the analyzer)<br>• As required: e.g. quality control<br>findings outside the specified<br>limits | Same |
| Controls | PreciControl Universal 1 and 2<br>(cleared on K090541) | Same |
| Reagent Stability | • Unopened at 2-8°C – up to the<br>expiration date | • Same |
| | • After opening at 2-8°C – 12 weeks | • After opening at 2-8°C – 8 weeks |
| | • Onboard the analyzer – 8 weeks | • Same |
| Table 2. Testosterone Immunoassay Performance Comparison | | |
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) |
| Precision | Elecsys 2010:<br><br>Intermediate Precision (Total)<br>18.5% CV @ 4.5 ng/dL<br>8.4% CV @ 9.5 ng/dL<br>3.2% CV @ 69.1 ng/dL<br>2.8% CV @ 216 ng/dL<br>2.8% CV @ 867 ng/dL<br>3.4% CV @ 1300 ng/dL<br>2.4% CV @ 1450 ng/dL<br><br>Repeatability (Within-Run)<br>10.2% CV @ 4.5 ng/dL<br>4.7% CV @ 9.5 ng/dL<br>2.1% CV @ 69.1 ng/dL<br>1.9% CV @ 216 ng/dL<br>2.6% CV @ 867 ng/dL<br>1.2% CV @ 1300 ng/dL<br>1.5% CV @ 1450 ng/dL | Elecsys 2010:<br><br>Intermediate Precision (Total)<br>7.4% CV @ 24 ng/dL<br>2.6% CV @ 195 ng/dL<br>2.2% CV @ 275 ng/dL<br>1.6% CV @ 620 ng/dL<br>1.7% CV @ 701 ng/dL<br><br>Repeatability (Within-Run)<br>4.6% CV @ 24 ng/dL<br>1.7% CV @ 195 ng/dL<br>1.4% CV @ 275 ng/dL<br>0.9% CV @ 620 ng/dL<br>1.1% CV @ 701 ng/dL |
| LoQ<br>(Functional<br>Sensitivity) | 12.0 ng/dL | Same |
| LoB<br>(Limit of Blank) | 1.2 ng/dL | N/A |
| LoD<br>(Limit of<br>Detection) | 2.5 ng/dL | 2.0 ng/dL<br>(Lower Detection Limit, LDL) |
| Limitations | The assay is unaffected by:<br><br>• Bilirubin: < 30 mg/dL<br>• Hemoglobin: < 600 mg/dL<br>• Intralipid: < 1000 mg/dL<br>• Biotin: < 30 ng/mL | The assay is unaffected by:<br><br>• Bilirubin: < 25 mg/dL<br>• Hemoglobin: < 1 g/dL (1000 mg/dL)<br>• Intralipid: < 1500 mg/dL<br>• Biotin: < 30 ng/mL |
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) |
| Limitations, continued | • In patients receiving therapy with high biotin doses (i.e. > 5 mg/day), no sample should be taken until at least 8 hours after the last biotin administration. | • In patients receiving therapy with high biotin doses (i.e. > 5 mg/day), no sample should be taken until at least 8 hours after the last biotin administration. |
| | • No interference was observed from rheumatoid factors up to a concentration of 1000 IU/mL. | |
| | • In vitro tests were performed with 18 commonly used pharmaceuticals. No interference with the assay was found. | • In vitro tests were performed with 16 commonly used pharmaceuticals. No interference with the assay was found. |
| | • Three additional drugs were tested: heparin clexane, dexamethasone, and Nandrolone. A strong interaction with Nandrolone was found. Do not use samples from patients under Nandrolone treatment. | |
| | • In isolated cases, elevated testosterone levels can be seen in samples from female patients with end stage renal disease (ESRD). | |
{2}------------------------------------------------
Substantial Equivalence-Comparison (continued)
The following table compares the Elecsys Testosterone II Immunoassay with the predicate device.
{3}------------------------------------------------
The following table compares the performance of the Elecsys Testosterone II Substantial Equivalence-Immunoassay with the predicate device. Comparison (continued)
{4}------------------------------------------------
The following table compares the performance of the Elecsys Testosterone II Substantial Equivalence-Immunoassay with the predicate device. Comparison (continued)
{5}------------------------------------------------
:
The following table compares the performance of the Elecsys Testosterone II Substantial Equivalence-Immunoassay with the predicate device. Comparison (continued)
| Table 2. Testosterone Immunoassay Performance Comparison, continued | | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) |
| Limitations, continued | In rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur. These effects are minimized by suitable test design. For diagnostic purposes, the results should always be assessed in conjunction with the patient's medical history, clinical examination and other findings. Implausible elevated values in female samples should be verified by an extraction method or validated LC-MS/MS tandem method. | In rare cases interference due to extremely high titers of antibodies to streptavidin can occur. The risk of interference from potential immunological interactions between test components and rare sera have been minimized by the inclusion of suitable additives. For diagnostic purposes, the results should always be assessed in conjunction with the patient's medical history, clinical examination and other findings. |
{6}------------------------------------------------
The following table compares the performance of the Elecsys Testosterone II Substantial Immunoassay with the predicate device. Equivalence-Comparison (continued)
| Table 2. Testosterone Immunoassay Performance Comparison, continued | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|----------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Method<br>Comparison | Four method comparison studies were<br>performed to demonstrate the accuracy<br>of the Elecsys Testosterone II assay.<br><br>(1) Isotope dilution/liquid<br>chromatographic-tandem mass<br>spectrometry (ID/LC-MS/MS) was<br>validated against the reference method,<br>isotope-dilution/gas chromatography-<br>mass spectrometry (ID-GC/MS).<br><br>A total of 52 serum samples with<br>testosterone values ranging from 8 to<br>1383 ng/dL were measured.<br><br>The following table summarizes the<br>results. | The Elecsys Testosterone assay was<br>compared to the Coat-A-Count®<br>Total Testosterone Assay.<br><br>A total of 71 clinical samples with<br>testosterone values ranging from<br>20 – 1269 ng/dL were tested in<br>singlicate.<br><br>The following table summarizes the<br>results. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Deming Regression n 52 Range 8-1383 ng/dL Slope 1.024 y-int -0.060 Correlation<br>Coefficient r = 0.996 | | | | | | | | | | | | | Passing/Bablok Least Squares n 71 71 Range 20-1269 ng/dL 20-1269 ng/dL Slope 1.02 0.956 y-int -0.108 0.049 tau/r 0.963 0.963 | | | | | | | | | | | | | | | | | | |
{7}------------------------------------------------
The following table compares the performance of the Elecsys Testosterone II Substantial Equivalence-Immunoassay with the predicate device. Comparison (continued)
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay<br>(K964889) | | | | | | | | | | | | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Method<br>Comparison<br>continued | (2) The Elecsys Testosterone II assay was<br>compared to the ID-GC/MS reference method.<br><br>A total of 55 serum samples with testosterone<br>values ranging from 7.6 to 1383 ng/dL were<br>measured.<br><br>The following table summarizes the results.<br>Deming Regression n 55 Range 7.6-1383 ng/dL Slope 1.032 y-int -0.055 Correlation<br>Coefficient r = 0.999 | | | | | | | | | | | | | |
| | (3) The Elecsys Testosterone II assay was<br>compared to the validated ID/LC-MS/MS<br>method.<br><br>A total of 142 female serum samples with<br>testosterone values ranging from<br>3 - 517 ng/dL were measured.<br><br>The following table summarizes the results.<br>Deming Regression n 142 Range 3-517 ng/dL Slope 0.928 y-int 0.047 Correlation<br>Coefficient r = 0.959 | | | | | | | | | | | | | |
{8}------------------------------------------------
The following table compares the performance of the Elecsys Testosterone II Substantial Equivalence-Immunoassay with the predicate device. Comparison (continued)
| Table 2. Testosterone Immunoassay Performance Comparison, continued | | | | | | | | | | | | | | |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Feature | Elecsys Testosterone II Assay | Predicate Device<br>Elecsys Testosterone Assay (K964889) | | | | | | | | | | | | |
| Method Comparison continued | (4) The Elecsys Testosterone II assay was compared to the predicate device, the Elecsys Testosterone assay (K964889).<br><br>A total of 239 male and 148 female serum samples with testosterone values ranging from 6.3 – 1400 ng/dL and 2.5 – 926 ng/dL, respectively, were measured in singlicate.<br><br>The following tables summarize the results. | | | | | | | | | | | | | |
| | <b>Male samples</b> | | | | | | | | | | | | | |
| | Deming Regression n 239 Range 6.3-1400 ng/dL Slope 0.971 y-int 0.085 Correlation Coefficient r = 0.985 | | | | | | | | | | | | | |
| | <b>Female samples</b> | | | | | | | | | | | | | |
| | Deming Regression n 148 Range 2.5-926 ng/dL Slope 0.984 y-int -7.34 Correlation Coefficient r = 0.972 | | | | | | | | | | | | | |
| | <b>Male and Female Samples Combined</b> | | | | | | | | | | | | | |
| | Deming Regression n 387 Range 2.5-1400 ng/dL Slope 0.989 y-int -2.87 Correlation Coefficient r = 0.992 | | | | | | | | | | | | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, represented by a series of stylized lines.
Roche Diagnostics Corporation c/o Sarah Baumann 9115 Hague Road P.O.Box 50410 Indianapolis, IN 46250
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring. MD 20993-0002
APR 2 3 2010
Re: k093421
Trade Name: Elecsys® Testosterone II Assay Regulation Number: 21 CFR §862.1680 Regulation Name: Testosterone test system Regulatory Class: Class I, reserved Product Codes: CDZ Dated: April 19, 2010 Received: April 20, 2010
Dear Ms. Baumann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
CA
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### Indication for Use
510(k) Number (if known): k093421
8 Device Name: Elecsys Testosterone II Assay
Indication For Use:
Immunoassay for the in vitro quantitative determination of testosterone in human serum and plasma. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers.
Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and androgenital syndromes.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation' and Safety (OIVD)
Carol Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K09342/