1ML UNITRACT INSULIN SYRINGE 27G X 1/2 AND 29G X 1/2
K093400 · Unilife Medical Solutions Limited · FMF · Apr 1, 2010 · General Hospital
Device Facts
Record ID
K093400
Device Name
1ML UNITRACT INSULIN SYRINGE 27G X 1/2 AND 29G X 1/2
Applicant
Unilife Medical Solutions Limited
Product Code
FMF · General Hospital
Decision Date
Apr 1, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The syringe, as supplied, is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.
Device Story
1mL Unitract Insulin Syringe is a sterile, single-use, fixed-needle syringe for subcutaneous insulin delivery. Device features include automatic needle retraction and auto-disable mechanism to prevent reuse and sharps injuries. Operation involves manual filling and injection; upon completion of the injection stroke, the plunger engages a spring-based mechanism at the proximal end, triggering automatic needle retraction into the barrel. Simultaneously, the plunger locks into the outer collar, preventing further movement. Used by patients or clinicians for insulin administration. Benefits include reduced risk of needle-stick injuries and prevention of syringe reuse.
Clinical Evidence
Bench testing only. Performance testing verified design specifications, including peak activation forces for retraction, spring reaction forces, and auto-disable override forces. Chemical testing included extractable trace metals, non-volatile residue, and pH. Biocompatibility testing included cytotoxicity and pyrogens, with fluid path materials meeting USP Class VI requirements. Sterilization validated per DIN EN 550:1994.
Technological Characteristics
Fixed-needle piston syringe; 27G x 1/2" or 29G x 1/2" needle sizes. Materials: biocompatible plastics and stainless steel. Energy source: mechanical spring-based retraction. Sterilization: ethylene oxide (EO) per DIN EN 550:1994. Connectivity: none (standalone).
Indications for Use
Indicated for subcutaneous administration of U-100 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Unilife medical solutions. The word "UNILIFE" is in bold, sans-serif font, with the words "medical solutions" in a smaller font below. Above the logo is the number "2093400" written in cursive. To the right of the word "UNILIFE" is a square logo with a black and white design.
## SECTION 6
## Special 510(k) Summary
APR - 1 2010
This 510(k) summary has been prepared in compliance with 21 CFR 807.92.
| Applicant: | Unilife Medical Solutions Limited<br>633 Lowther Rd<br>Lewisberry, PA 17339<br>Tel: 717-938-9323<br>Fax: 717-938-9364 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Melissa L. DeHass, Senior Manager, Regulatory Affairs<br>Email: Melissa.dehass@unilife.com |
| Date Prepared: | 30 October 2009 |
| Trade Name: | 1mL Unitract® Insulin Syringe |
| Common Name: | Piston Syringe |
| Classification: | Syringe, Piston; 21 CFR 880.5860; FMF; Class II |
| Predicate Devices: | Unitract Insulin Syringe (K080877) |
| Device Description: | The 1mL Unitract® Insulin Syringe is a sterile, fixed needle, single<br>use, syringe, consisting of a barrel and plunger assembly. The<br>syringe is supplied in two needle sizes, being 27G x ½" and 29G x<br>½".<br><br>The devices function in a manner similar to standard syringes for<br>filling and mixing insulin and subcutaneous injection, but also<br>incorporates a sharps injury prevention (automatic needle retraction)<br>and reuse prevention (auto disable) feature.<br><br>The syringe, as supplied, is a device with a small barrel, plunger and<br>fixed needle, calibrated in units of insulin (U-100), to be used to<br>administer (infuse) insulin subcutaneously. It incorporates features<br>that include reuse prevention and sharps injury prevention.<br><br>The sharps injury prevention feature allows for automatic and full<br>retraction of the needle into the syringe barrel immediately after the<br>full dose has been expelled; where the rate of retraction may be<br>controlled by the user. The reuse prevention feature also engages<br>immediately after retraction, which prevents further movement of<br>the plunger in either direction. |
| Indications for Use: | The syringe, as supplied, is a device with a small barrel, plunger and<br>fixed needle, calibrated in units of insulin (U-100), to be used to<br>administer (infuse) insulin subcutaneously. It incorporates features<br>that include reuse prevention and sharps injury prevention. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for UNILIFE medical solutions. The word "UNILIFE" is written in a stylized font, with the letters connected to each other. Below the word "UNILIFE" is the phrase "medical solutions" in a smaller font. To the right of the word "UNILIFE" is a graphic that looks like a backwards "L" shape.
Substantial Equivalence:
The 1mL Unitract® Insulin Syringe has identical indications for use to the cited predicate # K080877 which is to infuse drugs; whilst incorporating a sharps injury prevention feature. The device design and technology is identical to K080877. The label changes to the 1mL Unitract® Insulin Syringe do not raise new questions of safety and effectiveness based on its similarities to the cited predicates in terms of design and intended use.
The ImL Unitract® Insulin Syringe has identical needle retraction mechanisms to the cited predicate # K080877. The syringe is filled; and at the end of the injection stroke once the full dose has been expelled, the plunger engages the needle to allow retraction into the syringe barrel. The auto disable feature of the 1mL Unitract® Insulin Syringe is integrated within the plunger design, where after full retraction, the plunger engages with the collar outer to prevent movement of the plunger in either direction. This feature requires no secondary action by the user to disable the syringe.
The characteristics of the 1mL Unitract® Insulin Syringe spring based mechanism are identical to the technology of the cited predicate # K080877. The 1mL Unitract® Insulin Syringe spring mechanism is located on the proximal end of the syringe to allow retraction into the barrel.
The design of the 1mL Unitract® Insulin Syringe integrates the technological design characteristics of the cited predicate, and having identical operating procedures for aspiration and injection, has been demonstrated to have no significant differences.
The syringe components are manufactured from biocompatible materials and stainless steel.
Mechanical performance tests were conducted to verify the device meets design specifications based on the requirements of the FDA Consensus Standard and the voluntary standards.
Additional performance testing was conducted on the safety features relating to peak activation forces of the retraction mechanism, spring reaction forces and forces to override the auto disable feature.
Chemical testing included Reducing (Oxidisable) Matter, Extractable Trace Metals, Non-Volatile Residue, pH and chemical residues (EO Sterilisation).
Biocompatibility testing of the finished devices included Cytotoxicity and Pyrogens; with all materials in the fluid path meeting USP Class VI requirements.
Sterilisation processes are validated to European Standard DIN EN 550:1994, Sterilisation of medical devices - Validation and routine control of ethylene oxide sterilisation to a sterility assurance level (SAL) of 10°.
Manufacturing process control requires functional, chemical and biological tests be performed on each production lot number.
The 1mL Unitract® Insulin Syringe with 27G x 1/2" or 29G x 1/2" needle submitted in this 510(k) is determined as substantially equivalent to the cited predicate devices based on performance testing, intended use and design principles of operation and technology, used for the subcutaneous infusion of insulin. Any
Characteristics:
Technological
Non Clinical Performance Data:
Conclusion:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Unilife medical solutions. The word "UNILIFE" is in bold, sans-serif font. Below the word "UNILIFE" is the phrase "medical solutions" in a smaller font. To the right of the word "UNILIFE" is a graphic of a curved line.
differences cited do not raise any new questions of safety and effectiveness.
[INTENTIONALLY LEFT BLANK BELOW LINE]
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Melissa L. DeHass Senior Manager, Regulatory Affairs Unilife Medical Solutions Limited 633 Lowther Road Lewisberry, Pennsylvania 17339
APR - 1 2010
Re: K093400
Trade/Device Name: 1mL Unitract® Insulin Syringe Regulation Number: 21CFR 880.8560 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: March 3, 2010 Received: March 4, 2010
Dear Ms. DeHass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address > http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Th for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "UNILIFE medical solutions". The logo is in black and white and features a stylized font. To the right of the text is a graphic that resembles a leaf.
## SECTION 5
## Indications for Use
510(k) Number (if known):
K093400
1mL Unitract® Insulin Syringe Device Name:
Indications for Use:
The syringe, as supplied, is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use ______________________________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number. K093400
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.