← Product Code [GKZ](/productcode/GKZ) · K093394

# BC-3200 AUTO HEMATOLOGY ANALYZER,M-30D DILUENT, M-30R RINSE, M-30CFL LYSE, M-30E-Z CLEANSER, M-30PROBE CLEANSER, BC-3D C (K093394)

_Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · GKZ · Apr 8, 2010 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K093394

## Device Facts

- **Applicant:** Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Apr 8, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K093394 · Apr 8, 2010 | BC-3200 AUTO HEMATOLOGY ANALYZER,M-30D DILUENT, M-30R RINSE, M-30CFL LYSE, M-30E-Z CLEANSER, M-30PROBE CLEANSER, BC-3D C | Shenzhen Mindray Bio-Medical Electronics Co., Ltd. | Whole-blood samples from 121 donors | The sponsor conducted a Normal Population Study using 121 donor samples to establish and validate reference ranges for the hematology analyzer. | Reference Range Study; Donor samples; Clinical validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Normal Population Study; Observational study of donor samples | 121 donors; Sample Size: 121 | Not applicable for this study | Establishment of reference ranges for 13 hematology parameters |

## Indications for Use

The BC-3200 auto hematology analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter to be used in clinical laboratories for In Vitro Diagnostic purpose. The intended use of BC-3200 Auto Hematology Analyzer is to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.

## Device Story

The BC-3200 is an automated hematology analyzer for clinical laboratory use by trained professionals. It processes whole blood samples (closed vial or prediluted) using automated sampling, diluting, and mixing. It employs two measurement methods: the Coulter method for WBC, RBC, and PLT counting/sizing, and a colorimetric method for HGB concentration. The system generates 16 hematology parameters and three histograms. Results are compared against user-defined normal ranges; the device flags abnormal results (H/L) for further clinical review. This automated analysis assists clinicians in patient screening and identifying samples requiring manual differential or additional diagnostic studies.

## Clinical Evidence

No new clinical data was required for this submission. Analytical performance (accuracy, precision, linearity, carryover) was established under the previous clearance (k063407). The current submission focuses on software modifications to unmask four parameters (Mid%, Mid#, RDW, MPV) for QC reporting, supported by software validation and hazard analysis documentation.

## Technological Characteristics

Automated hematology analyzer; Coulter impedance principle for WBC, RBC, PLT; colorimetric method for HGB. Dimensions/form factor: benchtop. Connectivity: computer-controlled with data storage/review. Reagents: M-30 series (diluent, rinse, lyse, cleansers). Standards: IEC61010-1, IEC61010-2-101, ISO14971, EN 13640, EN 591, EN 375, EN 980, IEC 61326.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- COULTER® A⁶•T diff 2™ Analyzer (k990352)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k093394

B. Purpose for Submission:
To obtain clearance for unmasking of 4 previously shielded parameters of a cleared device (k063407)

C. Manufacturer and Instrument Name:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd
BC-3200 Auto Hematology Analyzer

D. Type of Test or Tests Performed:
WBC, Lymph #, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, MCV, MCH, MCHC, RDW, HCT, MPV, WBC Histogram, RBC Histogram, and PLT Histogram

E. System Descriptions:

1. Device Description:
The BC-3200 Auto Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter for In Vitro Diagnostic use in clinical laboratories. It consists of the analyzer, M-30 serial reagents (M-30D diluent, M-30CFL lyse, M-30R rinse, M-30E E-Z cleanser and M-30P probe cleanser), SC-CAL plus calibrators and BC-3D controls.
The BC-3200 Auto Hematology Analyzer is only to be used by trained medical professionals to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies. The analyzer provides analysis results of 16 parameter of human blood and three histograms.

2. Principles of Operation:
WBCs are counted and sized by the impedance method. This method is based on the measurement of changes in electrical resistance produced by a particle, which in this case is a blood cell suspended in a conductive diluent as it passes through an aperture of known dimensions. HGB is determined by the colorimetric method. RBCs and PLTs are counted by the impedance method also. In addition, for RBCs and PLTs, volumetric metering is used. An accurate cell count cannot be obtained unless the precise volume of diluted sample that passes through the aperture during the count cycle is known. The analyzer uses a volumetric metering unit to control the count cycle and to ensure that a precise volume of sample is analyzed for the measurement.

3. Modes of Operation:
Closed Vial Whole Blood Mode, Whole Blood Mode for venous blood, and Predilute Mode for capillary blood.

4. Specimen Identification:
Barcode or manual keyboard entry.

{1}

5. Specimen Sampling and Handling:
Samples are manually loaded into a sample compartment one at a time. The BC-3200 utilizes an automatic sampling and mixing device for sample processing. The Mindray calibrator is called SC-CAL PLUS.

6. Calibration:
The device has two calibration programs: manual calibration and auto calibration using commercial calibrators. The Mindray calibrator is called SC-CAL PLUS.

7. Quality Control:
The device has two QC programs: L-J Analysis and X-B Analysis. The Mindray three level control is called BC-3D)

8. Software:
All components of the device are controlled/monitored by software, which is responsible for the functionality, user interface, safety checks and performance accuracy.

FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
☐ Yes ☑ X or No

F. Regulatory Information:

1. Regulation section:
21 CFR §864.5220: Automated Differential Cell Counter

2. Classification:
Class II

3. Product code:
GKZ: Counter, Differential Cell

4. Panel:
81 Hematology

G. Intended Use:

1. Indication(s) for Use:
The BC-3200 auto hematology analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter to be used in clinical laboratories for In Vitro Diagnostic purpose.
The intended use of BC-3200 Auto Hematology Analyzer is to identify the normal patient, with all normal system-generated parameters, and to flag or identify patient results that require additional studies.

2. Special Conditions for Use Statement(s):
N/A

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:
COULTER® A⁶•T diff 2™ Analyzer, k990352

2. Comparison with Predicate Device:

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | A quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories | Same  |
|  Sample Types | Whole Blood Mode and Prediluted Mode | Same  |
|  Sample Processing | Utilizes an automatic sampling, diluting, and mixing device for sample processing | Same  |
|  Calibration | 2 calibration programs: manual calibration and auto calibration using commercial calibrators | Same  |
|  Aperture Alert | Minimize the possibility of reporting erroneous results caused by a partial or transient aperture clog or by other aperture disturbance | Same  |
|  Display | LCD | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Operating Modes | Closed Vial Whole Blood Mode | Closed Vial Whole Blood Mode and Open Vial Whole Blood Mode  |
|  Throughput | 1 minute/analysis | 60 seconds or less for each sample  |
|  Quality Control | 2 QC programs: L-J Analysis and X-B Analysis | 1 QC program: L-J Analysis  |
|  Recommended Controls | BC-3D: Low, Normal, & High | 4C PLUS: Abnormal Low, Normal, Abnormal  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   |  | High  |
|  Recommended Calibrator | CBC-CAL PLUS | S-CAL calibrator  |
|  Sample Volume Aspirated | 13 μL whole blood
20 μL prediluted blood | 18 μL whole blood
20 μL prediluted blood  |
|  Reagents Required | M-30D Diluent
M-30R Rinse
M-30CFL Lyse
M-30E E-Z Cleanser
M-30P Probe Cleanser | diff A^{C}-T Park or diff A^{C}-T Tain reagent park, both of which contain diluent and lytic reagent. A^{C}-T Rinse Shutdown Diluent  |
|  Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, Lymph%, Lymph#, Mid%, Mid#, Gran%, Gran#, RDW, MPV | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, Ly%, Ly#, MO%, MO#, GR%, GR#, RDW, MPV  |
|  WBC Operating Range | 0.0 - 299.9 (×10^{3}/μL) | 0.0 - 150 (×10^{3}/μL)  |
|  RBC Operating Range | 0.00 - 19.99 (×10^{6}/μL) | 0.00 - 8.00 (×10^{6}/μL)  |
|  HGB Operating Range | 0 - 29.9 (×g/dL) | 0.00 - 30.0 (×g/dL)  |
|  PLT Operating Range | 0 - 2999 (×10^{3}/μL) | 000 - 3000 (×10^{3}/μL)  |
|  MCV Operating Range | 0.0 - 249.9 fL | 50.0 - 130.0 fL  |

I. Special Control/Guidance Document Referenced (if applicable):

1. ISO 14971: Standards Title Medical device - Application of Risk Management to Medical Devices. Version 2001/A1:2003.
2. IEC 61010-1: Standards Title Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirement. Version 2001.
3. IEC 61010-2-101: Standards Title Safety requirements for electrical equipment for measurement, control, and laboratory use -Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment. Version 2002.
4. IEC 61326: Standards Title Electrical equipment for measurement, control and laboratory use – EMC requirements. Version 2002.
5. EN 13640: Standards Title Stability testing of in vitro diagnostic medical devices. Version 2002.
6. EN 13612: Standards Title Performance evaluation of in vitro diagnostic medical devices. Version 2002.

{4}

7. NCCLS EP5-A: Standards Title Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline. Version 1999.
8. NCCLS EP9-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (Second Edition). Version 2002.
9. NCCLS H20-A: Reference Leukocyte Differential (Proportional) and Evaluation of Instrumental Methods, Approved Standard. Version 1992.

## J. Performance Characteristics:

1. **Analytical Performance:**
a. **Accuracy:**
Established under k063407
b. **Precision/Reproducibility:**
Established under k063407
c. **Linearity:**
Established under k063407
d. **Carryover:**
Established under k063407
e. **Interfering Substances:**
The sponsor provided a list of commonly known interferences as supported by published references.

2. **Other Supportive Instrument Performance Data Not Covered Above:**
a. **Software Documentation:**
Modifications of the software that allow the sponsor to add the four parameters (i.e., Mid%, Mid#, RDW, MPV) to previously claimed 12 parameters for the BC-3D control and to add them at the QC screen have been sufficiently documented.

The sponsor has provided acceptable documentation demonstrating that the sponsor developed software for this device under an appropriate software development program; performed hazard analyses from both the patient's and user's standpoints, and addressed those hazards; and carried out an appropriate validation process. These procedures provide the foundation for assuring, to the extent possible, that the software will operate in a manner described in the specifications, and in no other way.

## K. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

## L. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K093394](https://fda.innolitics.com/device/K093394)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
