PRIMUS STEAM STERILIZER PSS8-AA (16X16X26 MULTI FUNCTIONAL STERILIZER), LAB/LO STERILIZER,PRIMUS STEAM STERILIZER PSS8
K093333 · Primus Sterilizer Company, LLC · FLE · Apr 2, 2010 · General Hospital
Device Facts
Record ID
K093333
Device Name
PRIMUS STEAM STERILIZER PSS8-AA (16X16X26 MULTI FUNCTIONAL STERILIZER), LAB/LO STERILIZER,PRIMUS STEAM STERILIZER PSS8
Applicant
Primus Sterilizer Company, LLC
Product Code
FLE · General Hospital
Decision Date
Apr 2, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRIMUS PSS8 Steam Sterilizer Series are designed for use in the Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilizer Series provide efficient steam sterliization of non-porous and porous, heat and moisture stabile materials. The intended use for the sterilizers models listed above is to provide efficient steam sterilization of non-porous and porous, heat and moisture stabile materials, wrapped and unwrapped surgical instruments, hard goods, and linens.
Device Story
Steam sterilizer (autoclave) series for hospital/clinical use; processes non-porous and porous, heat/moisture-stable materials (instruments, hard goods, linens). Operates via vacuum, gravity, liquid, and test (Bowie-Dick) cycles. Features stainless steel chambers (316L internal surface) and optional carbon steel reinforced doors; available in horizontal or vertical sliding door configurations; some models pit-mounted for floor cart access. Controlled by validated software providing fail-safe monitoring, warnings, and signals for cycle conditions/malfunctions. Healthcare providers load materials, select factory-preset cycles, and monitor performance via chemical/biological indicators. Benefits include standardized, validated sterilization assurance (SAL 10^-6) for medical equipment.
Clinical Evidence
Bench testing only. Performance validated against ANSI/AAMI ST8:2001 consensus standards. Sterilization efficacy confirmed using biological indicators (BI) for each cycle type to achieve a Sterility Assurance Level (SAL) of 10^-6. Electrical safety verified per AAMI/ANSI/IEC 60601-1-2 and UL/CSA 61010 standards. 100% Factory Acceptance Testing performed on all units.
Technological Characteristics
Steam sterilizer with 316L stainless steel chambers and optional 304 stainless steel jackets. ASME Section VIII, Division 1 pressure vessel construction. Operates via saturated steam. Connectivity includes fail-safe control software. Electrical safety per IEC 60601-1-2, UL 61010A-1/2-041, and CAN/CSA C22.2 No. 1010. Sterilization validated per ANSI/AAMI ST8:2001.
Indications for Use
Indicated for steam sterilization of non-porous and porous, heat and moisture stable materials, including wrapped and unwrapped surgical instruments, hard goods, and linens in hospital operating suites, central sterile supply, and clinical laboratories.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
1093333
# APR - 2 2010
#### Section 5
## A 510(k) SUMMARY PERTAINING TO THE SAFETY AND EFFECTIVENESS OF PRIMUS PSS8 STEAM STERILIZERS
| Manufacturer: | PRIMUS Sterilizer Company, LLC.<br>117 South 25th Street<br>Omaha NE 68131<br>Phone: 402-344-4200<br>Fax: 402-344-4242 |
|----------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact: | Connie Mansfield, Mgr., Marketing Communications & Regulatory Compliance<br>Omaha NE 68131<br>Phone: 402-344-4200 Ext 1229 |
| Date Prepared: | March 12, 2009 |
### Introduction:
The PRIMUS PSS8 Steam Sterilizers (or Autoclaves) are Class II, Product Code FLE Medical Devices as defined by CFRS880.6880 and defined for use in Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilize Series provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The models contained within this submission request utilize the same technology, materials and updated software as predicate devices cleared under K991575 and K082817.
The proposed PRIMUS Steam Sterilizer chambers offered within this submission is equipped with the same options offered under the predicate device, in design and construction except for ASME approved (optional) carbon steel reinforced doors, the vessel size and (optional 304 stainless) vessel jackets configured with 316L internal surface stainless steel chambers: offered in either horizontal or vertical sliding door applications. The PSS8-J, K and M units may be pit-mounted, allowing the optional floor carts to roll directly in to the chamber. All sterilizer doors are designed to be efficient, reliable and inherently safe. Pass through (double-door) models are also available.
PSS sterilizer units offered under this request for clearance are available in the following configurations. The Multifunction units also offer vacuum, gravity, liquids and test (VAC) configuration.
| PSS8-AA-M** | (16" x 16" x 26" Multi-Functional Sterilizer) | Single or Double Drains |
|---------------|-----------------------------------------------|-------------------------|
| PSS8-A-M** | (20" x 20" x 38" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-B-M** | (26" x 26" x 39" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-C-M** | (26" x 26" x 49" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-D-M** | (26" x 26" x 67" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-E-M** | (26" x 36" x 39" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-F-M** | (26" x 36" x 48" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-G-M** | (26" x 36" x 60" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-G.1-M*** | (32" x 36" x 48" Multi-Functional Sterilizer) | Single or Double Drains |
| PSS8-J-M** | (26" x 63" x 48" Multi-Functional Sterilizer) | Single or |
or Door OL Door OT Door DI Door or Door or Door Dr Double Door
Door
Door
Door
Section 5 Page 1 of 4
{1}------------------------------------------------
| PSS8-K-M** | (26" x 63" x 76" Multi-Functional Sterilizer) | Single or |
|-------------|-----------------------------------------------|-------------|
| | | Double Door |
| PSS8-M-M*** | (35" x 57" x 60" Multi-Function Sterilizer) | Single or |
| | | Double Door |
PRIMUS Steam Sterilizers models defined above are as described in Section 4 "Indications for Use Statement" and contained within this submission
The PRIMUS PSS8 series sterilizers are offered with factory pre-set sterilization cycles described within the Indication for Use Table (See attached Table og 3).
#### Safety
The PRIMUS PSS8 Steam Sterilizer Series have been validated against FDA recognized consensus standards for electrical safety:
- AAMI / ANSI / IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment -- Part 1-2: . Collateral Standard - General Requirements for Safety; Electromagnetic Compatibility - Requirements and Tests.
Additionally PRIMUS declares conformance to applicable industry and electrical codes as follows
- . UL 61010A-1, IEC 61010-1 Amendment 2, and the Part 2, Particular Requirements for Autoclaves Using Steam for the Treatment of Medical Materials & for Laboratory Process, IEC 61010-2-041, UL 61010A-2-041.
- CNL indicates the product was evaluated to the Canadian Standard for Laboratory Equipment, CAN/CSA-● C22.2 No. 1010 and the Part 2, Particular Requirements for Autoclaves Using Steam for the Treatment of Medical Materials and Laboratory Process, CAN/CSA-C22.2 No. 1010.2-041-96.
- PRIMUS PSS8 Pressure Vessels are designed, manufactured and tested in accordance with American . Society of Mechanical Engineers (ASME), Section VIII, Division 1 Unfired Pressure Vessels.
Validated software designed into the PRIMUS PSS8 series sterilizer provides for fail-safe controls that give appropriate warnings and signals when required conditions have not been met or if unit malfunctions. The technology designed into the PRIMUS PSS Steam Sterlizer Series provides for fail safe controls that provide end users with appropriate warnings and signals when required conditions have not been met or malfunctions occurs. Safety warnings and signals are challenged and verified on each unit, as a function of routine process control testing, during a 100% Factory Acceptance Test conducted at final inspection. Results are documented and maintained in the Device History Record (DHR).
#### Effectiveness:
PRIMUS has validated sterilizer performance for each load cycle type to ensure that the exposure time, provided in the directions for use statements, have been proven to assure the safety and effectiveness of the PRIMUS PSS8 series sterilizer. Internal validation protocols are documented and deemed to be comply to the ANSI/AAMI ST8:2001, Hospital steam sterilizers - (Sterility); a harmonized standard recognized FDA as applicable to the validation of steam sterilizers intended for use in hospitals and other health care facilities and that have a volume greater than 2 cubic feet (ft') (56 liter {L] ) with no exclusions taken. The recognized standard established the minimum construction requirements and performance requirements for hospital sterilizers, using saturated steam as the sterilizing agent.
Declarations of Conformance made by PRIMUS to recognized standards are based on a verification of product performance data, that was independently validated with BI indicators exposed during validation for each cycle type published within the Indications for Use statement. Data output from the validation report was reviewed and found to meet Sterility Assurance Level (SAL) of 10th when the sterilizer is used and maintained as directed.
#### Operator Information
PRIMUS provides information in the User's Manual that is intended to ensure safe and effective use of the sterilizer. Additional information concerning recommended practices for end users in monitoring sterilizer performance can be found in the ANSI/AAMI ST79:2006 Comprehensive guide to steam sterility assurance in the health
{2}------------------------------------------------
care facilities or other applicable standards to assure safe and effective use of the steam sterilization processes for application.
Based on the testing and comparison to the consensus standard, PRIMUS concludes that each of the chamber sizes included under this submission performs as intended and raises no new safety or effectiveness issues when used as directed.
Connie L. Mangold
Mgr., Marketing Communications and Regulation Compliance PRIMUS Sterilizer Company, LLC
{3}------------------------------------------------
| Cycle # | Cycle Type | Sterilize Temp | Sterilize Time (min) | Dry Time (min) | No. of Pre-vacs | 16x16x26 | 20x20x38 | 26x26x39 | 26x26x49 | 26x26x67 | 26x36x39 | 26x36x48 | 26x36x60 | 32x36x48 | 26x63x48 | 26x63x76 | 35x57x60 | |
|-----------------------------|------------|--------------------|----------------------|----------------|-----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|----------------|-------------------------------|
| 2 | Vacuum | 132°C<br>(270°F) | 4 | 1 | 3 | N/A | 1 | 1 | 1 | N/A | 1 | 1 | N/A | 1 | N/A | N/A | N/A | Porous Single |
| 3 | Vacuum | 132°C<br>(270°F) | 4 | 30 | 3 | N/A | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | Instrument<br>each tray |
| 4 | Vacuum | 132°C<br>(270°F) | "4 | 30 | 3 | N/A | 2 | 4 | 6 | N/A | 6 | 12 | N/A | 15 | N/A | N/A | N/A | 9x6, 12 lbs. |
| 6 | Gravity | 121.1°C<br>(250°F) | 30 | 30 | 0 | N/A | 2 | 2 | 4 | 6 | 12 | 16 | 16 | 16 | 30 | 60 | 42 | maximum size:<br>pounds |
| 7 | Liquids | 121.1°C<br>(250°F) | 30 | 15* | 0 | 57 | 111 | 149 | 189 | N/A | 207 | 255 | N/A | 284 | N/A | N/A | N/A | Plate glass<br>ml or smaller. |
| 8 | Test (VAC) | 132°C<br>(270°F) | 3-1/2 | 2 | 3 | N/A | 136 | 181 | 228 | 313 | 252 | 309 | 388 | 390 | 543 | 860 | 615 | |
| | | | | | | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack | |
| | | | | | | Shelving | Loading Eq. | Shelving | Loading Eq. | Shelving | Loading Eq. | Shelving | Loading Eq. | Shelving | Loading Eq. | Shelving | Loading Eq. | |
| Maximum Load Configurations | | | | | | | | | | | | | | | | | | |
** Dry time in LIQUIDS cycle is liquid cool time.
WARNING
THE INTENDED USE OF A LIQUID CYCLE IS NOT FOR STERILIZING MATERIALS FOR DIRECT PATIENT CONTACT
WARNING
PRIMUS RECOMMENDS THE VALIDATED FACTORY PRESET CYCLES. CHANGES TO THE CYCLE PARAMETERS,
PER ANSI/AAM ST79:2006, ARE NOT RECOMMENDED. CONTACT PRIMUS FOR FURTHER INFORMATION.
NOTE
STERILIZE TEMP PARAMETER IS PROGRAMMED TO ENSURE STERILIZE TEMP MEETS ANSI/AAM ST8
REQUIREMENTS FOR EACH FACTORY VALIDATED CYCLE.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract image of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR - 2 2010
Ms. Connie Mansfield Manager, Marketing Communications Primus Sterilizer Company, LLC 117 South 250 Street Omaha, Nebraska 68131
Re: K093333
Trade/Device Name: PRIMUS Steam Sterilizers Regulation Number: 21CFR 880.6880 Regulation Name: Steam Sterilization Regulatory Class: II Product Code: FLE Dated: March 26, 2010 Received: March 26, 2010
Dear Ms. Mansfield:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Mansfield
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Section 4
#### Indications for Use Statement
Applicant: PRIMUS Sterilizer Company, LLC
510(k) Number: K0933333
Device Name: PRIMUS Steam Sterilizers
Indication for Use:
The PRIMUS PSS8 Steam Sterilizer Series are designed for use in the Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilizer Series provide efficient steam sterliization of non-porous and porous, heat and moisture stabile materials. The PRIMUS PSSS Steam Sterilizer Series are available in the following configurations:
| PSS8-AA-M** | (16" x 16" x 26" Multi-Functional Sterilizer) | Single or<br>Double Door |
|---------------|-----------------------------------------------|--------------------------|
| PSS8-A-M** | (20" x 20" x 38" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-B-M** | (26" x 26" x 39" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-C-M** | (26" x 26" x 49" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-D-M** | (26" x 26" x 67" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-E-M** | (26" x 36" x 39" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-F-M** | (26" x 36" x 48" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-G-M** | (26" x 36" x 60" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-G.1-M*** | (32" x 36" x 48" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-J-M** | (26" x 63" x 48" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-K-M** | (26" x 63" x 76" Multi-Functional Sterilizer) | Single or<br>Double Door |
| PSS8-M-M*** | (35" x 57" x 60" Multi-Function Sterilizer) | Single or<br>Double Door |
Sterilizers are available in multi-functional models which feature vacuum, gravity, liquids and test (VAC) cycles.
· PRIMUS Sterilizer Company, LLC recommends that a suitable chemical indicator or biological challenge test be used on a periodic basis to test effectiveness of the sterilizer in accordance with health care facilities docurented plan for monitoring SAL 10d performance.
The intended use for the sterilizers models listed above is to provide efficient steam sterilization of non-porous and porous, heat and moisture stabile materials, wrapped and unwrapped surgical instruments, hard goods, and linens. The cycles to be cleared for each of the models listed above are found in the table below:
{7}------------------------------------------------
| Load Type | Cycle # | Cycle Type | Sterilize Temp | Sterilize Time (min) | Dry Time (min) | No. of Pre-vacs | 16x16x26 | 20x20x38 | 26x26x39 | 26x26x49 | 26x26x67 | 26x36x39 | 26x36x48 | 26x36x60 | 32x36x48 | 26x63x48 | 26x63x76 | 35x57x60 | |
|-----------------------------------------------------------------|---------|------------|--------------------|----------------------|----------------|-----------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|----|
| Unwrapped Nonporous Single Instrument | 2 | Vacuum | 132°C<br>(270°F) | 4 | 1 | 3 | N/A | 1 | 1 | 1 | N/A | 1 | 1 | N/A | 1 | N/A | N/A | N/A | |
| | | | | | | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Shelving | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
| Double wrapped instrument trays 16 pounds each tray | 3 | Vacuum | 132°C<br>(270°F) | 4 | 30 | 3 | N/A | 2 | 4 | 6 | 8 | 6 | 12 | 14 | 15 | 30 | 60 | 42 | |
| | | | | | | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Shelving | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
| Textile packs 9x9x6, 12 lbs. | 4 | Vacuum | 132°C<br>(270°F) | 4 | 30 | 3 | N/A | 4 | 6 | 12 | 12 | 14 | 12 | 16 | 18 | 16 | 30 | 60 | 42 |
| | | | | | | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Shelving | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
| Fabric packs maximum size:<br>12x12x20, 12 pounds | 6 | Gravity | 121.1°C<br>(250°F) | 30 | 30 | 0 | N/A | 2 | 2 | 4 | 6 | 9 | 12 | 14 | 16 | 20 | 50 | 40 | |
| | | | | | | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Shelving | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
| Vented borosilicate glass containers, 1000 ml or smaller, 4"x7" | 7 | Liquids | 121.1°C<br>(250°F) | 30 | 15* | 0 | N/A | 57 | 111 | 149 | 189 | 207 | 255 | N/A | 284 | N/A | N/A | N/A | |
| | | | | | | | N/A | 136 | 181 | 228 | 313 | 252 | 309 | 388 | 390 | 543 | 860 | 615 | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
| Bowie-Dick Test | 8 | Test (VAC) | 132°C<br>(270°F) | 3-1/2 | | 2 | | | | | | | | | | | | | |
| | | | | | | | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | 1 Test Pack | |
| | | | | | | | | | | | | | | | | | | Shelving | |
| | | | | | | | | | | | | | | | | | | Loading Eq. | |
.
.
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.