K093185 · Diros Technology, Inc. · GXD · Jan 22, 2010 · Neurology
Device Facts
Record ID
K093185
Device Name
OWL UNIVERSAL RF SYSTEM URF-3AP (ML)
Applicant
Diros Technology, Inc.
Product Code
GXD · Neurology
Decision Date
Jan 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DIROS OWL™ URF-3AP(ML) LESION GENERATOR is intended for the following: 1. Lesioning nerve tissue for functional neurosurgical procedures such as thalamotomies, pallidotomies, tractomies, and myelotomies; or 2. radiofrequency heat lesion procedures for the relief of pain. The MLA-4 Multi-Lesion adaptor is not intended for use in brain surgery.
Device Story
Diros OWL URF-3AP(ML) is a radiofrequency (RF) lesion generator used by qualified medical personnel to create heat ablative lesions in nerve tissue. Input: RF energy (up to 50W at 481kHz) delivered via probes; system monitors probe temperature, power, duration, and impedance. Operation: Controlled heating of tissue surrounding electrode tip; thermosensor feedback regulates temperature; impedance monitoring ensures safety. Output: RF energy for ablation; low-frequency stimulus pulses (monopolar mode) for nerve localization. The MLA-4 adaptor enables up to four simultaneous lesions (excluding brain surgery). Clinical use: Treatment of movement disorders (e.g., Parkinson's tremor) or pain relief (e.g., facet joint nerves). Healthcare providers use real-time display data to guide lesion creation, directly impacting surgical precision and patient outcomes through targeted nerve destruction.
Clinical Evidence
Bench testing only. Performance validated through risk analysis (ISO 14971) and electrical/mechanical safety testing per IEC 60601-1, UL 60601-1, CAN/CSA C22.2 No. 601.1-M90, IEC 60601-1-2, FCC 15 Subpart B, and IEC 60601-2-2. Software performance testing conducted to verify functional requirements.
Technological Characteristics
RF generator (481kHz, 50W max); monopolar/bipolar capability; thermosensor-based temperature control; impedance monitoring; MLA-4 multi-lesion adaptor. Standards: IEC 60601-1, UL 60601-1, CAN/CSA C22.2 No. 601.1-M90, IEC 60601-1-2, FCC 15 Subpart B, IEC 60601-2-2. Connectivity: Standalone system with probe inputs.
Indications for Use
Indicated for patients requiring functional neurosurgical procedures (thalamotomies, pallidotomies, tractomies, myelotomies) or radiofrequency heat lesioning for pain relief. MLA-4 adaptor contraindicated for brain surgery.
Regulatory Classification
Identification
A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.
{0}------------------------------------------------
Koa93185
Image /page/0/Picture/1 description: The image shows a logo with the letters "DTI" in a stylized font. The "D" is large and textured, the "T" is represented by a caduceus symbol, and the "I" is a simple vertical line with a textured pattern. Above the caduceus, there are some smaller letters that are difficult to read due to the image quality.
#### DIROS TECHNOLOGY INC. 510(K) SUBMITTAL OWL™ RF LESION GENERATOR MODEL URF-3AP WITH MULTI-LESION ADAPTOR (MLA-4)
## 510(k) Summary
#### Submitter:
DIROS TECHNOLOGY INC 232 Hood Road Markham, ON L3R3K8, Canada
JAN 2 2 2010
#### Contact:
Mr. George Darmos Telephone: 905-415-3440 Fax: 905-415-0667 gpdarmos@dirostech.com
Date: September 28, 2009
Trade Name: Diros OWL™ URF-3AP(ML)
Common Name: OWL Radiofrequency System
Classification Name: Generator, Lesion, Radiofrequency
Regulatory Class: II
Product Code: GXD
Regulation Number: 882.4400
REGISTRATION NO: 8043398
OWNER/OPERATOR NO: 9001301
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the letters "DTI" stacked vertically. The "D" and "I" are large and stylized, with the "D" on top and the "I" on the bottom. Between the "D" and "I" is a smaller symbol that appears to be a caduceus, a symbol often associated with medicine. The logo has a textured or patterned fill, giving it a slightly three-dimensional appearance.
### DIROS TECHNOLOGY INC. 510(K) SUBMITTAL OWL™ RF LESION GENERATOR MODEL URF-3AP WITH MULTI-LESION ADAPTOR (MLA-4)
## Predicate Devices:
We are making the claim that the DIROS OWL URF-3AP(ML) which is the URF-3AP with the multiple lesion adaptor accessory MLA-4 is substantially equivalent to the predicated devices listed in the chart below.
| LEGALLY<br>MARKETED<br>PREDICATE DEVICE | MANUFACTURE<br>NAME | REGULATORY<br>CLASS AND<br>PRODUCT<br>CODE | 510(K)<br>REGISTRATION<br>NUMBER |
|-----------------------------------------|------------------------------|--------------------------------------------|----------------------------------|
| OWL URF-3AP | DIROS<br>TECHNOLOGY<br>INC | Class II/GXD | K062758 |
| PMG-115 | BAYLIS<br>MEDICAL<br>COMPANY | Class II/GXD | K072478 |
The rationale of declaring the DIROS OWLIM URF-3AP(ML) is substantially equivalent to the above 2 predicate devices is based on the following:
- > Same Indications for use: All systems provide treatments by making heat ablative lesions in central or peripheral nerve tissue for the treatment of movement disorders or for the relief of pain. All systems are using the same fundamental scientific technology.
- > Similar key design technical characteristics- Pain relief is achieved by the controlled heating of the area surrounding the electrode tip. A thermosensor in the probe is used to control tissue temperature. Impedance values are displayed to guard against unsafe conditions. The PMG-115 includes an accessory that provides up to four simultaneous lesions.
- Same/similar components for treatment and measurement. >
- > Similar size, power source, and performance
## Description:
The Diros OWL™ radiofrequency generator URF-3AP(ML) is used by qualified medical personnel to make heat ablative lesions in central or peripheral nerve tissue for the treatment of movement disorders or for the relief of pain. The lesions are ablative in order to be therapeutic; i.e. the destruction of a small portion of the thalamus within the brain interferes with the motor pathway causing the tremor of Parkinson's disease, thereby relieving the tremor; or the destruction of the facet joint nerves in the lumbar vertebrae to block pain transmission by these nerves and thereby relieve certain types of low back pain.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo or emblem with the letters 'DTI' prominently displayed. The 'D' is a large, blocky letter on the left, and the 'I' is a vertical line on the right. In the center, between the 'D' and the 'I', there is a symbol that resembles a stylized caduceus, which is often associated with medicine or commerce. The overall design appears to be somewhat textured or patterned, giving it a slightly rough or vintage look.
#### DIROS TECHNOLOGY INC. 510(K) SUBMITTAL OWL™ RF LESION GENERATOR MODEL URF-3AP WITH MULTI-LESION ADAPTOR (MLA-4)
The URF-3AP(ML) supplies up to 50 Watts of Radio Frequency energy at 481kHz under power or temperature control while continuously monitoring and displaying actual power delivered, measured probe temperature, time of power duration, and measured impedance. This 50W RF source can be directed to one of four channels, one at a time, to create simultaneously up to four lesions.
When used with monopolar probes, the system also delivers low-frequency stimulus pulses in either voltage or current controlled modes.
## Indications for Use:
The DIROS OWL™ URF-3AP(ML) SYSTEM is intended for the following:
- 1. Lesioning nerve tissue for functional neurosurgical procedures such as thalamotomies, pallidotomies, tractomies, and myelotomies; or
- 2. radiofrequency heat lesion procedures for the relief of pain
The MLA-4 Multi-Lesion adaptor is not intended for use in brain surgery.
## Summary of Performance Testing:
A risk analysis identifying potential hazards and documenting mitigations of the hazards has been developed and applied as part of the DIROS OWL TMURF-3AP(ML) product development cycle. The risk analysis is based on EN 1441/ISO14971 - Risk Analysis for Medical Devices.
Testing was performed to validate the functional performance of the DIROS system. In particular, specific performance testing of the software was performed to show that the performance was met.
The DIROS OWL™ URF-3AP(ML) has been subjected to performance testing to applicable safety, electrical, mechanical, EMC standards. The DIROS OWL URF-3AP(ML) system has been evaluated and has passed all mechanical and electrical safety according to CSA International. Standards that were investigated are: IEC 60601-1, UL 60601-1 and CAN/CSA C22.2No.601.1-M90 certified, IEC 60601-1-2 and FCC 15 Subpart B. The URF-3AP(ML) has also been evaluated and satisfies the requirements to IEC 60601-2-2 Medical electrical equipment, Part 2: Particular requirements for the safety of high frequency surgical equipment.
## Conclusion:
As stated above, DIROS TECHNOLOGY INC.'s conclusion is that the DIROS OWL™ URF-3 AP(ML) is safe and effective and complies with the appropriate medical standards and is substantially equivalent to the above identified predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of an eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Diros Technology, Inc. c/o Mr. George Darmos President 232 Hood Rd. Markham, Ontario Canada L3R 3K8
JAN 2 2 2010
Re: K093185
Trade/Device Name: DIROS OWL™ URF-3AP(ML) Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency Lesion Generator Regulatory Class: II Product Code: GXD Dated: December 2, 2009 Received: December 7, 2009
Dear Mr. Darmos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CcntersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K093185
Device Name; DIROS OWL™ URF-3AP(ML)
#### Indications for Use:
The DIROS OWL™ URF-3AP(ML) LESION GENERATOR is intended for the following:
- Lesioning nerve tissue for funetional netwosurgical procedures such as thalamotorinies, .............................................................................................................................................................................. pallidotomies, tractomies, and myelotomics; or
- 2. radiofrequency heat lesion procedures for the relief of pain
The MLA-4 Multi-Lesion adaptor is not intended for use in brain surgery.
Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Exaluation (ODE
*(Division Sign-Off)*
Division of Ophthalmic, Neurological an Nose and Throat Devices
510(k) Number K093185
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.