K093167 · Diagnostica Stago, Inc. · JPA · Dec 22, 2009 · Hematology
Device Facts
Record ID
K093167
Device Name
STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER
Applicant
Diagnostica Stago, Inc.
Product Code
JPA · Hematology
Decision Date
Dec 22, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.5425
Device Class
Class 2
Intended Use
The STA Compact® Automated Multi-Parametric Analyzer is a fully automatic clinical instrument designed to perform tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Device Story
STA Compact Automated Multi-Parametric Analyzer; performs clinical laboratory analysis. Modifications include: new hardware/mechanical components for reliability; updated software for user enhancements; modified data post-processing; addition of Hook Effect Detection function. Device operates in clinical laboratory settings; used by laboratory technicians/personnel. Input: clinical samples; Output: analytical results. Hook Effect Detection improves result reliability by identifying potential assay interference. Healthcare providers use output for clinical decision-making.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures (21 CFR 820.30). Risk analysis conducted to assess impact of modifications; results demonstrated predetermined acceptance criteria were met.
Technological Characteristics
Fully automatic multi-parametric analyzer. Performs chronometric and photometric assays. Features barcoding for reagent management. Software includes data post-processing for Hook Effect Detection. Hardware updated for reliability. No specific materials or connectivity standards provided.
Indications for Use
Indicated for use on human plasma samples to aid in the diagnosis of coagulation abnormalities or to monitor anticoagulant therapy.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
STA Compact® Automated Multi-Parametric Analyzer (K961579)
Related Devices
K093001 — STA-R EVOLUTION EXPERT SERIES HEMOSTASIS SYSTEM AUTOMATED MULTI-PARAMETRIC ANALYZER · Diagnostica Stago, Incorporated · Dec 18, 2009
K151867 — STA R MAX · Diagnostica Stago S.A.S · Aug 7, 2015
K961579 — STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER · American Bioproducts Co. · Jul 10, 1996
K962857 — COAG-A-MATE MTX · Organon Teknika Corp. · Jan 22, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
### CONFIDENTIAL
Image /page/0/Picture/1 description: The image shows the word "Stago" in a serif font. Above the word is a logo consisting of two abstract shapes, one resembling a leaf and the other a crescent moon. The logo is positioned directly above the "t" in "Stago", creating a visual connection between the text and the symbol.
### Attachment - 6
# DEC 2 2 2 2009
### 510 (k) Summary
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Umberto V. Parrotta, Jr. Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
Phone: (973) 631-1200, x-2044 Facsimile: (973) 695-0095
Contact Person: Umberto V. Parrotta, Jr.
Date Prepared: October 05, 2009.
#### Name of Modified Device and Name/Address of Sponsor:
STA Compact® Automated Multi-Parametric Analyzer or STA Compact®.
Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
#### Common or Usual Name:
- IVD Coagulation Device/Instrument. .
- Automated and Semi-Automated Hematology Device. .
- Multi-Parametric Analyzer .
#### Classification Name:
System, Multipurpose for In Vitro Coagulation Studies
#### Predicate Device:
Diagnostica Stago's STA Compact® Automated Multi-Parametric Analyzer (K961579).
#### Purpose of the Special 510(k) Notice:
The STA Compact® is a modification to the company's own legally marketed device, STA Compact® (K961579).
The modification of the device involves replacement of components for reliability of operation or obsolescence reasons, change in the device application software to provide operator/user use enhancements and modification of the data post-processing to increase results reliability. The software modification essentially consists of
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "Stago" in a serif font. Above the word is a logo consisting of two curved shapes, one above the other, that resemble stylized water droplets or crescent moons. The logo is positioned directly above the "t" in "Stago", creating a visual connection between the symbol and the text.
### Attachment - 6
addition of a quality improvement function which provides Hook Effect Detection when the vWF (Von Willebrand factor) test is performed on the sample plasma.
#### Indication/Intended Use:
The STA Compact® Automated Multi-Parametric Analyzer is a fully automatic clinical instrument designed to perform tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
#### Principles of Operation:
The STA Compact Automated Multi-Parametric Analyzer is designed as a fully automatic system. Samples and test reagents are loaded into the instrument where sample handling, reagent delivery, analysis, and reporting of results are performed automatically. A central processing unit controls instrument functions such as, management of patient results, quality control, support for instrument maintenance, and work load optimization.
The instrument utilizes Diagnostica Stago reagents in addition to open adaptation of other currently available reagents. Barcoding of test reagents, calibrators, and controls facilitate their use on the system and permits reagent management simple. Manual entry of reagent information enables the use of non-barcoded reagents. The instrument performs multiple test methodologies in random access as selected by the user. These include clotting time or clot-based tests (i.e. chronometric) measurements and photometric assays (at specific wavelengths) on plasma samples.
The STA Compact® is the modified version of the Company's STA Compact® (K961579), legally marketed Predicate Device. The modification of the device involves replacement of components for reliability of operation or obsolescence reasons, change in the device application software to provide operator/user use enhancements and modification of the data post-processing to increase results reliability.
Principles of Operation for the aforementioned analyses are the same between the subject submission device and the Predicate Device.
#### Substantial Equivalence:
The STA Compact® and its Predicate Device, STA Compact® (K961579) have the same Indications for Use, Technology, Principles of Operation and comparable Performances. The previously mentioned characteristics have been slightly modified to improve performance, without changing the principle of operation or the technological characteristics. The modifications consist of new hardware and mechanical components to increase the operation reliability or to replace obsolete components. The software is also updated to include operator/user use enhancements and modification of the data post-processing is done to increase results reliability. The software modification, subject of this Special 510(k) essentially consists of addition of a function, Hook Effect Detection when the vWF test is performed on the sample
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "Stago" in a serif font. Above the word is a logo that appears to be two abstract shapes, one above the other. The top shape is dark and curved, while the bottom shape is a crescent shape with some small white circles on it.
## Attachment - 6
plasma. This function is operational during vWF (Von Willebrand factor) testing. This function identifies the hook effect (showing abnormally low results for high concentrations of the vWF factor) which subsequently allows the instrument to perform appropriate re-dilutions to obtain reliable results.
As evidenced by Risk Assessment and Validation Studies (including all the changes mentioned above), no new questions were raised regarding the Safety, Effectiveness, Performance, Indications for Use, Technology and the Principles of Operation. Therefore, the STA Compact® is Substantially Equivalent to the STA Compact® Predicate Device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
DEC 2 2 2 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Diagnostica Stago, Inc. c/o Umberto V. Parrotta Director of Regulatory Affairs and Quality Assurance 5 Century Drive Parsippany, New Jersey 07054
Re: K093167
Trade/Device Name: STA Compact® Automated Multi-Parametric Analyzer Regulation Number: 21 CFR 864.7290 Regulation Name: Multipurpose System for In Vitro Coagulation Studies Regulatory Class: Class II Product Code: JPA Dated: December 4, 2009 Received: December 8, 2009
Dear Mr. Parrotta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
Page 2 - Mr. Parrotta
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ia m chan
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Stago" in a serif font, with the "S" capitalized and the rest of the letters in lowercase. Above the word "Stago" is a graphic that resembles a stylized leaf or abstract shape. The graphic is dark and solid, with some texture or detail visible within its form.
### Attachment - 2A
### Indications for Use
510(k) Number (if known): K093167
Device Name: STA Compact® Automated Multi-Parametric Analyzer
Indications for Use:
The STA Compact® Automated Multi-Parametric Analyzer is a fully automatic clinical instrument designed to perform tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Serik R. S
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K09316 7
DSI-STA-CompactB6-Attachment-02Av1-IndicationsforUseStatementForm.doc
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.